1 · Light, colour and the spectrophotometer
Most chemistry tests end with a coloured product. The instrument does not "see" colour like we do. It shines light of one chosen wavelength through the solution and measures how much light gets through. More analyte → more colour → less light passes.
Light basics
- Light behaves as a wave and as particles (photons).
- Wavelength (in nanometres, nm) = distance between two wave crests. 📌 Shorter wavelength = more energy.
- Order from short to long: gamma → X-ray → ultraviolet (UV) → visible (about 400–700 nm) → infrared (IR) → microwaves → radio.
- The colour you see is the light a solution transmits (lets through). The photometer is set to the wavelength the solution absorbs most.
- When an atom or molecule absorbs a photon, an electron jumps to a higher energy level. It is unstable, falls back, and may give off light of a characteristic wavelength (a "line spectrum" that identifies the substance).
The light path (memorise the order)
| Part | Job | Key exam facts |
|---|---|---|
| Lamp (light source) | Bright, steady light | Tungsten = visible and near-IR (continuous). Deuterium (or hydrogen) = UV. Mercury vapour = sharp lines (used to check wavelength). ⚠️ After changing a lamp → recalibrate (light hits the monochromator at a new angle). |
| Monochromator | Picks a narrow band of wavelengths | Glass / interference filters → simple photometers. Prisms and diffraction gratings → spectrophotometers. |
| Bandpass | Width of the band passed (measured at half of peak transmittance) | Centred on the setting: 10-nm bandpass at 540 nm = 535–545 nm. |
| Cuvet | Holds the solution; fixed light path (b) | Quartz for UV, glass/plastic for visible. Scratches = errors. |
| Detector | Changes light into an electric signal | See Section 2. |
Beer's law — the heart of photometry
Cu = (0.300 ÷ 0.400) × 100 = 75 mg/dL. Check: lower absorbance than the standard → answer must be below 100.
Blanks and two-wavelength reading
- Reagent blank (reagents, no sample): zeroes the instrument and removes the colour of the reagents only. It does not correct lipemia, hemolysis or sample chromogens.
- Bichromatic analysis: every sample is read at two wavelengths (peak and base). The difference removes background colour (bilirubin, hemoglobin) — each sample blanks itself.
Checking the spectrophotometer
| Check | Question it answers | Tool |
|---|---|---|
| Wavelength accuracy | Is "540 nm" really 540 nm? | Holmium oxide (UV/visible) or didymium (visible/near-IR) glass; nickel sulfate solution; mercury or deuterium emission lines |
| Absorbance (photometric) accuracy | Is A = 0.500 really 0.500? | Solutions / filters of known absorbance |
| Linearity | Does A rise in a straight line with concentration? | A series of increasing concentrations |
| Stray light | Is light of other wavelengths leaking to the detector? | Cut-off filters / solutions |
Reflectance photometry (dry slides)
On dry-slide analyzers the reaction happens in thin reagent layers on a slide. Light passes through the layers, bounces off a white reflective layer, and the unabsorbed, reflected light returns to the detector. Reflectance is not linear with concentration, so the computer uses an algorithm to convert it.
Transmittance, the zero settings and when Beer's law fails
- Setting up a single-beam instrument: block the light path completely → set 0 %T; put in the reagent blank → set 100 %T (A = 0); then read calibrators and patients.
- Double-beam instrument: a chopper and mirrors send the beam alternately through the sample and a reference cuvet onto one detector. This cancels drift of the lamp and changes in detector sensitivity.
- Beer's law holds only when: the light is truly monochromatic, the solvent hardly absorbs, the concentration stays inside the linear range, no other substance absorbs at that wavelength, and the analyte does not react with the solvent or other solutes. A wide bandpass (broad light) usually gives a lower absorbance than expected.
- Standard (calibration) curve: plot A of several calibrators against their concentration. Use at least two calibrators, and never use a single calibration factor above the linear part of the curve.
- Absorbed energy can also change how bonds vibrate or rotate (the basis of infrared absorption), not only move electrons.
Setting up a method and a run
- Why λmax? Highest sensitivity, and a small wavelength error changes the reading least. Avoid the steep side of a peak.
- Bandpass vs a narrow peak: if the bandpass is wider than the peak (e.g., NADH at 340 nm), the shoulders are included and the peak reads LOWER. Two instruments, same NADH solution → the one reading HIGHER has the narrower bandpass.
- Wavelength check filters: holmium oxide for narrow-bandpass instruments; didymium for instruments with a bandwidth of about 8 nm or more.
- UV region is below about 380 nm (340 nm = NADH); visible about 380–750 nm.
Source: SUCCESS! in Clinical Laboratory Science, 4th ed., Ch.1 §I A–B (pp. 2–3), §I S.3b (p. 14); review questions Q1–13 (pp. 95–96) and rationales (pp. 145–147). · Tietz 7th ed., Ch 16 Optical Techniques, pp. 203–204 (gap check v-next) · Tietz 7th ed., Ch 16 Optical Techniques, pp. 201, 206, 208 (update 25 Oct 2026)
2 · Detectors — turning light into a signal
The detector turns light energy into an electrical signal. More light reaching it = bigger signal. The readout then shows the result (digital display, printout, computer).
| Detector | How it works | Remember |
|---|---|---|
| Photomultiplier tube (PMT) | Light knocks electrons off a sensitive surface; 10–15 dynode stages multiply them | 🔥 Very sensitive, very fast, amplifies up to ~106. Must be shielded from room light. Fast enough to read chopped (pulsed) light. |
| Photodiode array | 256–2048 photodiodes in a line, each reading its own wavelength | Whole UV–visible spectrum at once (resolution about 1–2 nm). Also an HPLC detector. |
| Barrier-layer cell / photocell | Iron plate + selenium layer; makes current directly | Simple, cheap, no amplification. "Iron + selenium" = NOT a PMT. |
| Phototube | Single light-sensitive cathode in a tube | Simpler than a PMT. |
Source: SUCCESS! Ch.1 §I B.5–6 (p. 3); review questions Q3–5 (p. 95); rationales Q3–5 (pp. 145–146).
3 · Atomic absorption spectrophotometry (AAS)
Simple idea: free metal atoms in their ground state absorb light of their own special wavelength. Send that exact wavelength through the sample and see how much disappears.
- Hollow-cathode lamp: anode + cylindrical cathode made of the analyte metal + inert gas (argon, neon or helium). The ionized gas hits the cathode, excites the metal atoms, and they emit that metal's lines.
- Flame: breaks compounds into atoms and keeps most of them at ground state (about 99%) — it does not excite them.
- What is measured: the lamp beam after it passes through the flame (with and without sample). The decrease = absorbed light, related to concentration.
- Chopper + tuned amplifier: only the pulsed lamp light is amplified, so steady light from the flame and light emitted by excited analyte atoms is ignored. They do not fix flame-temperature changes or lamp ageing.
- Flameless AAS (graphite furnace / carbon rod, tantalum or platinum): sample vaporized in a heated chamber → more sensitive than flame.
- Used for metals: lead, zinc, copper, aluminium, magnesium, calcium, lithium.
Flame emission, AAS interferences and ICP-MS
| Flame emission photometry | Atomic absorption | |
|---|---|---|
| Flame's job | Excites the atoms → they EMIT their own colour | Only frees the atoms → they stay at ground state and ABSORB lamp light |
| Signal | Emitted light ∝ concentration | Fall in lamp light ∝ concentration |
| Flame colours | Na yellow · K violet · Li red · Mg blue | — |
| Status | Once used for Na, K, Li; now mostly replaced by ISEs | About 100× more sensitive than flame emission and very specific |
- Chemical interference: phosphate binds calcium in the flame → falsely low Ca. Fix: add lanthanum (or strontium), which takes the phosphate.
- Ionization interference: an easily ionized element (such as K) changes how much of the analyte becomes ions. Fix: add a large amount of an easily ionized element to every sample.
- Background absorption (mainly flameless AAS): corrected by the Zeeman method — a strong magnetic field splits the absorption line, so total and background absorbance can be read separately at the same wavelength.
- ICP-MS (inductively coupled plasma mass spectrometry) is replacing AAS for trace metals: it measures many metals at once. AAS stays cheaper and simpler.
Source: SUCCESS! Ch.1 §I C (pp. 3–4); review questions Q21–23 (pp. 97–98); rationales (pp. 148–150). · Tietz 7th ed., Ch 16 Optical Techniques, pp. 209–211 (gap check v-next)
4 · Light scatter, fluorescence and luminescence
Nephelometry vs turbidimetry
| Nephelometry | Turbidimetry | |
|---|---|---|
| Measures | Light SCATTERED by particles, read at an angle (forward scatter for large particles) | Light BLOCKED by particles = decrease in transmitted light, read straight through |
| Instrument | Nephelometer | Ordinary spectrophotometer |
| Main problems | Background scatter: scratched cuvets, particles in reagents. Avoid UV light (can cause fluorescence → falsely high). | Particle size must be similar in samples and standards; aggregation/settling → keep timing constant. Temperature is not critical (room temperature). |
Fluorescence and the luminescence family
| Type | What excites the electrons? | Key points |
|---|---|---|
| Fluorescence | Absorbed light (often UV) | Emits longer wavelength (lower energy) almost immediately. Detector at 90° to the lamp. About 1000× more sensitive than absorbance and more specific (two wavelengths chosen). Problems: quenching, self-absorption, pH/temperature/solvent changes; needs very pure reagents. |
| Phosphorescence | Absorbed light | Like fluorescence but with a longer delay (> about 10-4 s) before emission. |
| Chemiluminescence | Energy of a chemical reaction (no light needed) | Extremely sensitive (attomole–zeptomole). Labels: acridinium ester, isoluminol. Read with a luminometer. |
| Bioluminescence | An enzyme-catalysed oxidation | Luciferase + luciferin (firefly). A special type of chemiluminescence; more sensitive than direct fluorescence. |
Fluorescence polarization (FPIA) — preview of Section 9
Plane-polarized light excites a fluorescent label. A label bound to antibody is big and turns slowly → its light stays polarized. A free label is small and spins fast → light becomes depolarized. More patient analyte → more free label → less polarization (inverse relationship). No separation step → homogeneous.
Counting radioactivity
Scintillation counter: a fluor turns radiation into flashes of light → photomultiplier tube. Gamma → solid sodium iodide crystal with thallium. Beta → liquid scintillation. Alpha particles are not measured in clinical laboratories. (Radioimmunoassay is now rare.)
FLUORO = light lights it · CHEMI = chemistry lights it · BIO = an enzyme lights it · PHOSPHO = lights it later.
Limits of fluorescence and how labs beat them
- Stokes shift = gap between the peak excitation wavelength and the peak emission wavelength (energy lost before light is given off). A larger shift makes it easier to keep excitation light away from the detector.
- Inner filter effect: at high concentration the sample itself absorbs the exciting or emitted light → the signal stops rising in a straight line. Fluorescence is linear only in dilute solutions.
- Background: Rayleigh and Raman scatter of the exciting light, and natural fluorescence of serum (proteins, bilirubin, drugs), add a false signal.
- Time-resolved fluorescence: labels made from lanthanides (europium, samarium) glow for microseconds, while background fluorescence fades within nanoseconds. A pulsed lamp plus a delayed reading removes the background → much better signal-to-noise.
Source: SUCCESS! Ch.1 §I D–F (pp. 4–5), §I U.2c (p. 17); review questions Q12, Q14–20 (pp. 96–97); rationales (pp. 147–148). · Tietz 7th ed., Ch 16 Optical Techniques, pp. 212, 217–218 (gap check v-next)
5 · Chromatography and mass spectrometry
Simple idea: a mobile phase (gas or liquid) carries the sample over a stationary phase. Each substance travels at its own speed because of differences in solubility, volatility, charge or size — so they separate.
| Thin-layer (TLC) | Gas–liquid (GLC / GC) | High-performance liquid (HPLC) | |
|---|---|---|---|
| Stationary phase | Silica gel on a glass plate or plastic sheet (planar) | Liquid coated on inert packing (e.g. Celite) or a capillary column | Organic material bonded to silica in a column (polar or nonpolar) |
| Mobile phase | Liquid solvent rising by capillary action | Inert gas: helium, nitrogen, hydrogen, argon | Liquid pumped through: water, methanol, ethanol, acetonitrile, isopropanol |
| Separates by | Solubility in solvent vs attraction to silica | Volatility and interaction with the column | Distribution between the two phases |
| Identify / measure | Rf value + colour sprays, UV, fluorescence, heat | Retention time (identity); peak area/height (amount) | Retention time; peak size |
| Typical use | Urine drug screening | Volatile substances, alcohols, drugs (with MS) | TDM, vitamins, hormones |
TLC details
- Steps: extract → concentrate → spot on the plate → develop → detect (sprays, UV, fluorescence, heat) → compare Rf with standards.
- Solvent too polar → solute runs with the solvent front. Solvent not polar enough → solute stays at the origin.
- Rf changes with chamber saturation, temperature, humidity and solvent composition.
GLC details
- Parts: carrier gas + flow control, heated injector (about 50 °C hotter than the column, so the sample vaporizes), column (packed or capillary) in an oven, detector, computer.
- More volatile solute, or less interaction with the column → elutes faster. Carrier-gas flow rate must be steady for reproducible retention times.
- Detectors: flame ionization, thermal conductivity, electron capture, mass spectrometer (best identification).
HPLC details
- Parts: solvent reservoir(s), pump(s), loop injector, pre-column/guard column, column (usually room temperature), detector (UV/visible, photodiode array, fluorometer, electrochemical, MS).
- Isocratic elution = constant solvent strength. Gradient elution = solvent strength increases during the run (better resolution).
- Normal phase = polar stationary + nonpolar mobile. Reversed phase = nonpolar stationary + polar mobile.
- Other column methods: ion exchange separates by the sign and size of charge; size exclusion by molecular size.
Mass spectrometry (MS)
- The mass spectrum is unique for a compound (identification); the number of ions ∝ amount (quantification).
- The vacuum stops ions colliding with other molecules.
- Coupled systems: GC-MS and HPLC-MS; tandem MS (GC/MS/MS, HPLC/MS/MS) adds a second fragmentation step for more selectivity.
- Uses: drugs and metabolites, steroids, amino-acid composition of proteins, proteomics.
Five ways liquid chromatography separates
| Type | Separates by | Remember |
|---|---|---|
| Adsorption (liquid–solid) | Sticking to and leaving the surface of a solid support (often silica) | Solutes that like the mobile phase move fastest |
| Partition (liquid–liquid) | Sharing between a liquid stationary layer and the mobile phase (polarity) | Basis of normal- and reversed-phase HPLC; most clinical LC |
| Ion exchange | Binding to fixed charges on a resin (cation or anion exchanger) | Mixed cation + anion resins deionize water; cation-exchange HPLC quantifies Hb A2, Hb F and Hb variants |
| Size exclusion (gel filtration) | Size — small molecules enter the pores and are delayed | Largest molecules come out FIRST |
| Affinity | A biological binding partner (antibody, lectin, boronate) | Most specific |
Gas chromatography also comes in two kinds: gas–solid (solid adsorbent) and gas–liquid (liquid coated on a solid; the usual clinical type).
Mass spectrometry — ion sources and analyzers
| Part | Options | Key point |
|---|---|---|
| Ion source | Electron ionization (EI) · chemical ionization (CI) · electrospray (ESI) · APCI · MALDI · ICP | EI breaks molecules into many fragments → library "fingerprint" (GC-MS). CI, ESI and MALDI are soft (little fragmentation). ESI = the most used clinical source (with HPLC). |
| Analyzer | Quadrupole · time-of-flight (TOF) · ion traps (incl. Orbitrap) · magnetic sector | Quadrupole scans m/z one value after another; TOF takes the whole spectrum at once (lighter ions arrive first); magnetic sector is rare (large, costly). |
| Tandem (MS/MS) | Triple quadrupole: MS1 → collision cell (gas breaks the selected ion) → MS2 | Selected (multiple) reaction monitoring follows one parent-ion → product-ion pair → very selective quantification. |
Chromatography troubleshooting and modes
- All chromatography works because compounds share themselves differently between a mobile and a stationary phase. Modes: adsorption, partition, ion exchange, size exclusion (gel filtration), affinity.
- Size exclusion: large molecules cannot enter the pores and leave the column FIRST; small ones take the long path.
- HPLC pressure rising + abnormal peaks → something is blocked (frit, guard column, tubing). Detector or injection faults change the peaks, not the pressure.
- GC: analytes must be volatile (or derivatized). TLC Rf = spot distance ÷ solvent-front distance (0–1); Rf 0.5 = halfway.
Source: SUCCESS! Ch.1 §I G–K (pp. 5–8); review questions Q48–58 (pp. 100–101); rationales (pp. 154–155). · Tietz 7th ed., Ch 19 Chromatography, pp. 242, 254–257; Ch 77 Hemoglobin and Hemoglobinopathies, p. 1066 (gap check v-next) · Tietz 7th ed., Ch 20 Mass Spectrometry, pp. 275–288 (gap check v-next) · Tietz 7th ed., Ch 19 Chromatography, pp. 241, 245, 247, 254–255, 268 (update 25 Oct 2026)
6 · Electrochemistry — potentiometry, amperometry, coulometry, polarography
| Method | What is measured | Clinical examples |
|---|---|---|
| Potentiometry | Voltage (potential) between an indicator and a reference electrode, at (near) zero current | pH, PCO2, ion-selective electrodes for Na+, K+, Cl−, Ca2+, NH4+ |
| Amperometry | Current from oxidation or reduction at an electrode held at a fixed potential | PO2 (Clark) electrode, amperometric glucose electrode |
| Coulometry | Quantity of electricity (current × time) — Faraday's law | Chloride titrator (sweat chloride) |
| Polarography / anodic stripping voltammetry | Current while the voltage is changed (polarogram) | Lead and other heavy metals (ASV) |
Potentiometry basics
- A half-cell (electrode) = metal conductor in an electrolyte. Two half-cells joined by a salt bridge make a cell; the potential difference is measured.
- Standard hydrogen electrode = 0.000 V (the universal standard).
- Reference electrodes (constant potential): calomel (mercury + mercurous chloride + saturated KCl) and silver/silver chloride.
- Indicator electrode: its potential changes with the analyte.
- Electrons flow anode → cathode: oxidation at the anode, reduction at the cathode.
Blood gas electrodes
| Electrode | Type | How it works | Calibration |
|---|---|---|---|
| pH | Potentiometric glass | Thin pH-sensitive glass tip filled with HCl/AgCl and an internal Ag/AgCl wire. The glass must be hydrated (H+ exchanges with metal ions in the glass). Store in the manufacturer's recommended solution. | Two standard buffers, at constant 37 °C |
| PCO2 (Severinghaus) | Potentiometric — "a pH electrode in disguise" | pH electrode in bicarbonate solution behind a membrane permeable to CO2 gas (not to H+). CO2 in → carbonic acid → pH falls. pH is inversely related to log PCO2. | Calibrating gases; barometric pressure needed |
| PO2 (Clark) | Amperometric | Platinum cathode + Ag/AgCl anode + polarizing voltage, gas-permeable (polypropylene) membrane. O2 reduced at the cathode → current directly proportional to PO2. | Calibrating gases; barometric pressure needed |
Ion-selective electrodes (ISE)
- Potassium: liquid ion-exchange membrane with valinomycin as the neutral carrier.
- Ammonium: nonactin + monactin.
- Sodium: special glass (silicon dioxide, sodium oxide, aluminium oxide). Errors from protein build-up on the membrane.
Amperometric glucose electrode
Chloride coulometer
A constant current flows from a silver anode (to a second electrode), generating Ag+ ions, which precipitate chloride as AgCl. When all chloride is used up, free Ag+ appears and is reduced at a separate (amperometric) indicator electrode, so the current suddenly rises → generation stops. The time of current flow is proportional to chloride. Principle = coulometry + amperometry. Still used for sweat chloride.
Anodic stripping voltammetry (ASV)
Step 1: a negative potential plates trace metal ions onto the electrode (preconcentration). Step 2: the metal is stripped off while the current is recorded; the pattern identifies and measures the metal. Based on polarography, not potentiometry. Classic for lead in blood.
Osmometry — counting dissolved particles
Osmolality = the number of dissolved particles per kilogram of water. It does not matter whether the particle is big or small, charged or neutral — each particle counts once.
| Osmometer type | What it reads | Key point |
|---|---|---|
| Freezing point depression (routine) | Sample is supercooled, freezing is started by agitation, a thermistor reads the plateau temperature | Detects volatile solutes (ethanol, methanol); not affected by room temperature. Used by almost all clinical labs. |
| Vapor pressure (historical) | Fall in vapor pressure (dew point) | Misses volatile solutes such as methanol and ethanol — they escape into the vapor. |
Source: Tietz Textbook of Laboratory Medicine, 7th ed., Ch 37 Electrolytes and Blood Gases, pp. 415.e9–415.e11.
Nernst equation and direct vs indirect ISE
🔬 Worked example: a 10-fold rise in K⁺ activity changes the potential by about 59 mV at 25 °C (≈ 61.5 mV at 37 °C). For Ca²⁺ the same 10-fold rise gives only about 30 mV. Common error: forgetting to divide by n for divalent ions.
Selectivity: every ISE responds a little to other ions; the selectivity coefficient tells how much. A small coefficient = a good, selective electrode.
| Direct ISE | Indirect ISE | |
|---|---|---|
| Sample | Undiluted (blood gas analyzers, POCT, whole blood) | Diluted first (large chemistry analyzers) |
| Measures | Activity in plasma WATER (molality) | Concentration in total plasma volume |
| Very high lipids or protein | Not affected | Falsely low Na⁺ (pseudohyponatremia) — less water per litre, but the dilution assumes normal water |
Troubleshooting pH meters and ISEs
- An ISE senses the ACTIVITY of the free ion and is SELECTIVE, not perfectly specific; the membrane (special glass, valinomycin, ion exchanger) decides which ion it 'sees'.
- Noisy, drifting readings with EVERY calibrator → look at the REFERENCE electrode: a liquid junction blocked by KCl crystals or a low filling solution. Dissolve the crystals, refill with fresh electrolyte, recalibrate.
- Slow, sluggish response only with serum (fast with aqueous calibrators) → protein film on the ion-selective membrane.
Source: SUCCESS! Ch.1 §I L–O (pp. 8–11); review questions Q24–36 (pp. 98–99); rationales (pp. 149–152). · Tietz 7th ed., Ch 17 Electrochemistry and Chemical Sensors, pp. 224.e2–224.e3, 224.e8 (gap check v-next) · Tietz 7th ed., Ch 17 Electrochemistry and Chemical Sensors, pp. 224.e1–224.e4 (update 25 Oct 2026)
7 · Electrophoresis (including hemoglobin electrophoresis)
Simple idea: charged molecules move in an electric field. Their speed depends on net charge, size and shape, the buffer, the support medium, temperature and the strength of the field.
| Topic | Key facts |
|---|---|
| Support media | Agarose gel, cellulose acetate (5–7 fractions; routine), starch gel and polyacrylamide (smaller pores, ≈20 fractions). Celite is NOT an electrophoresis medium (GLC). |
| Stains | Serum proteins: Ponceau S, amido black, Coomassie brilliant blue · CSF/urine proteins: silver (≈30× more sensitive than Coomassie; no preconcentration needed) · Lipoproteins: fat red 7B, oil red O · LD isoenzymes: nitrotetrazolium blue |
| Quantification | Scan the stained bands with a densitometer |
| Troubleshooting | Holes ("doughnut") in a band = protein too concentrated (dilute and repeat) · Very slow migration = voltage too low · Sample precipitates = pH too high/low or too much heat |
| High-resolution | Agarose, higher voltage, cooling, more concentrated buffer → up to 12 zones |
| Isoelectric focusing (IEF) | A type of zone electrophoresis. Ampholytes create a pH gradient; each protein moves until gel pH = its pI, then stops. Separates proteins with very close pIs. |
| Capillary electrophoresis | Based on electro-osmotic flow toward the cathode; cations move fastest. |
Hemoglobin electrophoresis
- Hemoglobin = 4 globin chains + 4 heme groups + 4 iron atoms. Hb A1 = α2β2, Hb A2 = α2δ2 (< about 4%), Hb F = α2γ2.
- Hb S = valine replaces glutamic acid at position 6 of the beta chain. Hb C = lysine replaces glutamic acid at the same position.
- Hb F resists alkali denaturation (alkali test). Normal Hb F after 1 year of age is below about 1%.
- Solubility test (sodium hydrosulfite + concentrated phosphate buffer): turbid = Hb S. Screening only.
| System | What happens |
|---|---|
| Cellulose acetate, pH 8.6 (screening) | Toward the anode: A1 > F > S = G = D > A2 = C = E = O. A2 and C (also S and D) cannot be told apart here. |
| Citrate agar, pH 6.2 (confirmation) | A different pattern: A1, A2, D, E and G run together as one group, so C separates from A2 and S separates from D/G. |
What else changes the speed of migration
- Ionic strength of the buffer: higher → thicker ion cloud around each protein → slower migration but sharper bands, and more current and heat. Lower → faster but broader bands.
- Heat (Joule effect): current through the gel makes heat → water evaporates → resistance falls → current rises further. A constant-current power supply limits this; too much heat denatures proteins.
- Wick flow: evaporation pulls buffer into the support from both ends, which can shift proteins. The cover/enclosed chamber reduces evaporation.
- Electroendosmosis is small on starch, purified agarose and polyacrylamide (few surface charges) and large on media with many fixed negative charges.
Hemoglobin electrophoresis — reading alkaline + acid together
- Hb A has a pI of about 6.8: negative at alkaline pH (moves to the anode) and positive at acid pH (moves to the cathode).
- Modern screening uses agarose with barbital buffer, pH ≈ 9.2 (older books: cellulose acetate pH 8.6). Stains: amido black or Ponceau S.
- Acid electrophoresis uses citrate agarose, pH ≈ 6.0–6.4, run only when the alkaline gel shows an abnormal band. Its band order is F – A – S – C, with F and C at opposite ends; D, G, E, O, A2 run with A.
| Alkaline band | Acid band | Presumptive Hb |
|---|---|---|
| S position | S position | Hb S |
| S position | A position | Hb D or G |
| C position | C position | Hb C |
| C position | A position | Hb E (or A2) |
| C position | between A and S | Hb O-Arab |
Source: SUCCESS! Ch.1 §I P–Q (pp. 11–13); review questions Q37–47, Q63–68 (pp. 99–102); rationales (pp. 152–154, 156–157). · Tietz 7th ed., Ch 18 Electrophoresis, pp. 226–229 (gap check v-next) · Tietz 7th ed., Ch 77 Hemoglobin and Hemoglobinopathies, pp. 1055–1057, 1067–1068 (gap check v-next) · Tietz 7th ed., Ch 18 Electrophoresis, pp. 226, 229 (update 25 Oct 2026)
8 · Automation, instrument maintenance and point-of-care testing
Automation words
| Term | Meaning |
|---|---|
| Discrete analysis | Each sample–reagent mixture in its own vessel (may be single-test or multi-test) |
| Centrifugal analysis | Centrifugal force moves samples and reagents into cuvets for simultaneous analysis |
| Random access | Any test or panel in any order; STAT samples can be added ahead |
| Batch analysis | Samples processed as a group |
| Throughput | Maximum number of tests per hour |
| Turnaround time | Time to generate one result |
| Dead volume | Specimen in the cup/tube that cannot be aspirated |
| Carry-over | Contamination of a sample by the previously aspirated sample |
| Reflex testing | A first result decides whether more tests are added or cancelled |
| Stand-alone / modular workcell / multiple platform / integrated modular system | One instrument of one discipline / ≥2 instruments of one discipline with one controller / one instrument doing tests from ≥2 disciplines / ≥2 analytical modules sharing one sample-and-reagent system |
| Total laboratory automation | Samples received, centrifuged, sorted and loaded by conveyor without an operator (high-volume labs) |
How analyzers work
- Specimen handling: level-sensing probes, bar-code readers (patient, sample and reagent identification).
- Sampling errors (falsely low results): short sample, air bubble at the bottom of the cup, fibrin clot in the probe.
- Reagents: dry (lyophilized powder or tablets to reconstitute; dry multilayer slides) or liquid.
- Testing: cuvets of quartz glass (permanent) or plastic (disposable); common reaction temperatures 37 °C and 30 °C.
- Kinetic (rate) assay = concentration from the change in absorbance over time. Endpoint assay = incubate a fixed time, read once, compare with calibrators.
- Data management: QC review and delta checks (compare with the patient's previous result) before release; results and QC archived (with the laboratory information system).
Keeping instruments reliable
| Activity | Examples |
|---|---|
| Function verification | Monitor temperature, electronic parameters, calibration, QC data — and correct when needed |
| Performance verification | Confirm the instrument performs to specification |
| Preventive maintenance | Scheduled cleaning, replacing worn parts, adjustments (per shift, daily, weekly, monthly, yearly) for all equipment — balances, refrigerators, centrifuges, water baths, pipettes, analyzers — always documented. Purpose: less downtime, fewer costly repairs, longer equipment life. |
Point-of-care testing (POCT)
- Testing at or near the patient, outside the main laboratory, for immediate decisions (intensive care, emergency, coronary care, neonatal ICU).
- Every POC instrument must be evaluated: results must correlate with the main laboratory analyzer; linearity, control ranges and reference ranges established.
- Training of every operator — documented. Quality control, statistics and proficiency testing — documented.
Centrifuges
- RCF, not rpm, describes the separating force — rpm is squared, so a small speed error changes the force a lot.
- Speed check (tachometer): within about 5% of the rated speed, at each speed used.
- Timer check against a stopwatch: within about 10%.
- Refrigerated centrifuge: temperature checked (about monthly), within about 2 °C of the setting; ordinary centrifuges warm up during a run.
- Balance opposite loads (difference about 1% or the manufacturer's limit). Use only tubes rated for the RCF.
- Safety: keep tubes stoppered until after spinning (aerosols); do not free clots with wooden sticks (hemolysis); after a breakage clean the chamber and cushions as a biohazard.
Source: Tietz Textbook of Laboratory Medicine, 7th ed., Ch 15 Principles of Basic Techniques and Laboratory Safety, pp. 199.e10–199.e11.
More automation terms and specimen-handling checks
- Continuous-flow analyzer (the first automated analyzer): samples are pulled one after another through the same tubing with liquid reagents. Shared tubing → carry-over risk. Today most analyzers are discrete.
- Automated stand-alone: a single-discipline instrument with extra built-in functions such as auto-repeat and auto-dilution (the analyzer dilutes high samples itself and calculates the final result).
- Liquid-level sensing works by pressure or capacitance; clot detection stops a short aspiration or a blocked probe. "Clot errors" in a sample with no visible clot → think of a very viscous sample (hyperviscosity, e.g. paraprotein).
- Pneumatic tube system: fast transport, but sudden acceleration can hemolyse samples; the system must be validated and carriers well packed.
- Automation pros: more tests per hour, less variation, fewer manual errors, smaller sample and reagent volumes. Cons: high start-up cost, risk that the product is discontinued, and skilled staff needed.
POCT devices — what can go wrong
- Glucose meters read strips by reflectance or, now mostly, electrochemically (enzyme + mediator such as ferrocene → current).
- Glucose oxidase strips: very specific for glucose, but high PO₂ (patient on oxygen) → falsely low glucose.
- Glucose dehydrogenase strips: not affected by oxygen; older types read maltose (and other sugars) as glucose → falsely high (newer engineered enzymes fixed this).
- Hematocrit extremes still disturb many strips.
- Immunoassay POCT (pregnancy, cardiac, infection tests) uses solid-phase lateral-flow (immunochromatography) or flow-through cassettes, read by eye or by reflectance/fluorescence readers.
Centrifuges, centrifugal analyzers and sampling faults
- RCF = 1.118 × 10⁻⁵ × r × rpm², r = RADIUS in cm. Worked example: rotor 60 cm across, 3000 rpm → r = 30 cm → 1.118 × 10⁻⁵ × 30 × 9,000,000 ≈ 3000 × g. Common errors: using the diameter (≈ 6000 × g) or not squaring rpm.
- Speed is checked with a tachometer, timer with a stopwatch — dial settings are not proof.
- Centrifugal analyzer: spinning throws sample and reagent from the inner wells into the outer cuvets, mixing them so all reactions start together; the cuvets are read as they pass the light beam.
- Wrong results grouped by one probe (e.g., every photometric test very low or 'below detection' while electrolytes from a second probe look normal) → clot/fibrin in that probe. Stop, clean, rerun QC and the samples.
- POCT: main advantage = short turnaround time and small sample volume; usually higher cost per test and many non-laboratory operators to train.
Source: SUCCESS! Ch.1 §I R–T (pp. 13–15); review questions Q52, Q59–62 (pp. 101–102); rationales (pp. 155–156). · Tietz 7th ed., Ch 29 Automation in the Clinical Laboratory, pp. 324.e3, 324.e6, 324.e10–324.e11, 324.e15 (gap check v-next) · Tietz 7th ed., Ch 30 Point-of-Care Testing, pp. 329, 331 (gap check v-next) · Tietz 7th ed., Ch 15 p. 199.e10; Ch 29 pp. 324.e6, 324.e10–e11; Ch 30 pp. 327, 331 (update 25 Oct 2026)
9 · Immunochemical techniques (immunoassays)
Simple idea: an antibody binds the analyte (the antigen) very specifically. A label on the antigen or the antibody gives a measurable signal. This lets us measure tiny amounts: hormones, drugs, vitamins (B12, folate), tumor markers, CK-MB mass. Electrolytes are not measured this way (ion-selective electrodes).
- Polyclonal antiserum = antibodies from many cell clones (a mixture). Monoclonal = from one clone (uniform) → higher specificity.
- All immunoassays (and competitive protein binding) use a binder with limited binding sites. HPLC does not — it separates physically.
Competitive binding
| Heterogeneous | Homogeneous |
|---|---|
| Bound and free label must be physically separated (wash): RIA, ELISA, IRMA | No separation needed — the signal differs between bound and free label: EMIT, FPIA, LOCI |
Labels: radioactive (I-125, old RIA), enzyme (alkaline phosphatase, glucose-6-phosphate dehydrogenase), fluorophore (fluorescein), chemiluminescent (acridinium ester, isoluminol), ruthenium (electrochemiluminescence).
The named techniques
| Technique | How it works | Signal vs analyte |
|---|---|---|
| EMIT — enzyme-multiplied immunoassay technique | Enzyme is coupled to the antigen (drug). Patient drug and enzyme-labelled drug compete for antibody. Antibody binding blocks the enzyme. Only free enzyme-labelled drug reacts with substrate + coenzyme (absorbance change, e.g. at 340 nm). | Enzyme activity rises with drug concentration |
| FPIA — fluorescence polarization immunoassay | Polarized excitation; bound label = polarized emission, free label = depolarized. Used for drugs, hormones, fetal lung maturity. | Polarization falls as concentration rises (inverse) |
| Chemiluminescent immunoassay | Isoluminol or acridinium ester label; with H2O2 and a catalyst isoluminol is oxidized and emits light (425 nm) | Signal proportional to analyte (format dependent) |
| LOCI — luminescent oxygen channeling immunoassay | Antigen links two antibody-coated beads: a sensitizer bead (photosensitive dye) and an acceptor bead (chemiluminescent precursor + fluorophore). Light → singlet oxygen → diffuses to the nearby acceptor → chemiluminescence → fluorophore emits longer-wavelength light. Homogeneous. | Signal proportional to analyte |
| Electrochemiluminescence (ECLIA) | Ruthenium label; complexes on magnetic beads are captured at an electrode; after washing, the electrode triggers light read by a PMT. Sandwich or competitive. | Format dependent |
Immunoassay interferences
| Interference | Mechanism | Effect | How to detect / fix |
|---|---|---|---|
| High-dose hook effect | Huge antigen excess saturates capture AND labelled antibodies in a one-step sandwich assay | Falsely LOW | Re-assay on dilution (result rises). Two-step assays with a wash are not prone. |
| Heterophile / HAMA (human anti-animal antibodies) | Patient antibodies bridge the animal capture and detection antibodies | Usually falsely HIGH in sandwich assays (false low possible) | Non-linear dilution, other platform disagrees, heterophile-blocking tube. Reagents contain non-immune animal serum/IgG. |
| Biotin (high-dose supplements) | Free biotin blocks streptavidin binding sites | Sandwich → falsely LOW; competitive → falsely HIGH | Biotin-free method, wash-out period (≥ about 8 h), streptavidin beads, dilution |
Source: Tietz Textbook of Laboratory Medicine, 7th ed., Ch 26 Immunochemical Techniques, pp. 311, 319–320; Ch 5 Preanalytical Variation, p. 107.
Noncompetitive (sandwich, immunometric) immunoassay
| Competitive | Noncompetitive (sandwich) | |
|---|---|---|
| Antibody | Limited amount | Excess (two antibodies) |
| What is labelled | Antigen | Antibody (e.g. IRMA, most ELISAs) |
| Signal vs analyte | Inverse | Direct |
| Best for | Small molecules with one epitope (drugs, T4, steroids) | Large molecules with ≥ 2 epitopes (TSH, hCG, troponin, ferritin) |
| Detection limit set by | Antibody affinity | How well the label can be detected → most sensitive (especially with chemiluminescent labels) |
One-step (simultaneous) sandwich assays are quicker but can show the hook effect; two-step (sequential) assays wash before adding the labelled antibody and are not prone to it.
Source: SUCCESS! Ch.1 §I U (pp. 15–18); review questions Q16, Q69–76 (pp. 97, 102–103); rationales (pp. 147, 157–159). · Tietz 7th ed., Ch 26 Immunochemical Techniques, pp. 310–312 (gap check v-next) · Tietz 7th ed., Ch 26 Immunochemical Techniques, pp. 312–313 (update 25 Oct 2026)
10 · Choosing the best answer — instrument "look-alikes"
| If the question says… | Think… | Not… |
|---|---|---|
| Continuous visible + near-IR light | Tungsten lamp | Mercury vapour (lines), deuterium (UV) |
| Selects the wavelength | Monochromator | Detector |
| Iron plate + selenium | Barrier-layer cell | Photomultiplier tube |
| Whole spectrum at once | Photodiode array | PMT |
| Checks wavelength | Holmium oxide / didymium | Linearity standards |
| Light scattered at an angle | Nephelometry | Turbidimetry (blocked light) |
| Detector at 90° | Fluorometer (keeps lamp light out) | "Prevents loss of emitted light" |
| Chemical energy makes light | Chemiluminescence (enzyme → bioluminescence) | Fluorescence |
| Ground-state atoms, hollow-cathode lamp | AAS (flame = atomizer, not exciter) | Flame emission |
| Mobile phase is a gas | GLC | HPLC (liquid pumped) |
| Separates by charge sign and size | Ion exchange | Size exclusion |
| Identifies by mass-to-charge | Mass spectrometry | Retention time |
| Mercury + mercurous chloride + KCl | Calomel reference electrode | Ag/AgCl |
| Valinomycin | Potassium ISE | Sodium (glass) |
| Silver ions generated, time ∝ chloride | Coulometry + amperometry | Potentiometry |
| Current ∝ oxygen | Amperometric PO2 electrode | pH glass electrode |
| Plate then strip; lead | Anodic stripping voltammetry (polarography) | Potentiometry |
| pH gradient, ampholytes | Isoelectric focusing | Zone electrophoresis with one pH |
| Electro-osmotic flow | Capillary electrophoresis | Iontophoresis |
| Doughnut band | Protein too concentrated | Denatured protein |
| Each test in its own vessel | Discrete analysis | Continuous flow |
| No separation of bound/free | Homogeneous: EMIT, FPIA, LOCI | RIA, ELISA, IRMA |
1. Reagent blank ≠ sample blank (no correction for lipemia/hemolysis).
2. Bandpass is split in half around the setting.
3. Rf uses origin→spot ÷ origin→front.
4. PMT works WITH a chopper.
5. AAS measures the lamp beam after the flame.
6. pH calibration = two buffers at 37 °C (gases need barometric pressure).
7. Measured blood gases: pH, PCO2, PO2 only.
8. Amperometric glucose = glucose oxidase, not hexokinase.
9. Silver stain is MORE sensitive than Coomassie.
10. FPIA polarization is inversely related to concentration; EMIT activity rises with concentration.
Summary of SUCCESS! Ch.1 §I (pp. 2–18) and review questions Q1–76 (pp. 95–103).
⚡ Last-Minute Review — Instrumentation & Analytical Principles
1. Spectrophotometry
- Lamp → entrance slit → monochromator → exit slit → cuvet → detector → readout.
- Tungsten = visible/near-IR (continuous) · deuterium = UV · mercury vapour = lines. New lamp → recalibrate.
- Filters = photometers; prisms/gratings = spectrophotometers. Bandpass 10 nm at 540 → 535–545 nm.
- A = abc · A = 2 − log %T (10 %T → 1.0) · Cu = (Au/As) × Cs.
- Reagent blank = reagent colour only. Bichromatic = two wavelengths, self-blanking.
- Holmium oxide / didymium = wavelength accuracy.
2. Detectors
- PMT: sensitive, fast, amplifies ~106, works with a chopper, shield from light.
- Photodiode array: a line of diodes → whole spectrum at once. Iron + selenium = barrier-layer cell.
3. Atomic absorption
- Hollow-cathode lamp (cathode = the metal) → chopper → flame makes ground-state atoms → detector reads the beam after the flame.
- Chopper + tuned amplifier reject flame light and analyte emission. Graphite furnace = more sensitive.
4. Scatter & luminescence
- Nephelometry = scattered light (angle; avoid scratches, particles, UV). Turbidimetry = blocked light (particle size, settling; temperature not critical).
- Fluorescence: detector at 90°; ≈1000× more sensitive + more specific; quenching; very pure reagents.
- Phosphorescence = delayed. Chemiluminescence = chemical energy. Bioluminescence = enzyme (luciferase).
- Scintillation counter = beta/gamma (gamma: NaI(Tl) crystal; beta: liquid).
5. Chromatography & MS
- TLC: silica plate, Rf = origin→spot ÷ origin→front; urine drug screen; solvent too polar → runs with the front.
- GLC: gas mobile phase; volatility; retention time = identity, peak area = amount; Celite support.
- HPLC: pumped liquid; isocratic vs gradient; reversed phase = nonpolar stationary/polar mobile.
- Ion exchange = sign and size of charge. MS = mass-to-charge, vacuum, unique spectrum; tandem MS.
6. Electrochemistry
- Potentiometry = voltage (pH, PCO2, ISE). Standard H2 electrode = 0 V. Reference: calomel, Ag/AgCl.
- Anode = oxidation; cathode = reduction.
- pH: hydrated glass, store in manufacturer's medium, 2 buffers at 37 °C. PCO2 = pH electrode in bicarbonate behind a CO2 membrane (log). PO2 = amperometric, current ∝ O2.
- Measured: pH, PCO2, PO2; bicarbonate is calculated.
- K+ = valinomycin; NH4+ = nonactin + monactin; Na+ = glass (protein build-up error).
- Amperometric glucose = glucose oxidase + H2O2 + platinum. Chloride = coulometry + amperometry (sweat). ASV = plate & strip, lead.
7. Electrophoresis
- pH 8.6 → proteins negative → anode; pI = zero charge = no movement.
- Stains: Ponceau S / amido black / Coomassie (proteins), silver (CSF), fat red 7B / oil red O (lipoproteins), nitrotetrazolium blue (LD). Densitometer quantifies.
- Doughnut = too concentrated · slow = low voltage · precipitate = pH or heat.
- IEF = ampholytes → pH gradient. Capillary = electro-osmotic flow.
- Hb A1 α2β2 · A2 α2δ2 · F α2γ2 · S = β6 Glu→Val · C = β6 Glu→Lys. F = alkali-resistant; solubility test = S.
- Cellulose acetate pH 8.6: A1 > F > S > A2/C. Citrate agar pH 6.2 separates C from A2.
8. Automation & POCT
- Discrete = own vessel · random access = STATs anytime · throughput = tests/hour · turnaround = time per result · dead volume · carry-over · reflex testing · delta check.
- Short sample / air bubble / fibrin clot → falsely low. Kinetic = ΔA/time; endpoint = one reading.
- Function verification = temperature, electronics, calibration. Preventive maintenance = scheduled, documented, ALL equipment.
- POCT: correlate with the main lab, documented training, QC and proficiency testing.
9. Immunoassays
- Monoclonal = one clone = specific. Competitive: more analyte → less label bound.
- Heterogeneous (wash): RIA, ELISA, IRMA. Homogeneous: EMIT, FPIA, LOCI.
- EMIT: enzyme on the drug; free label active; activity ↑ with drug. FPIA: polarization ↓ with drug. LOCI: light → singlet oxygen. ECLIA: ruthenium + magnetic beads + electrode.
- Electrolytes are NOT immunoassays (ISE).
Gap topics: osmometry, centrifuges, immunoassay interferences
- Freezing point osmometer = routine; vapor pressure misses volatile alcohols. mOsm/kg = ΔT ÷ 1.86 × 1000 (−0.53 °C ≈ 285).
- Osmolal gap = measured − calculated (2 Na + glucose/18 + BUN/2.8) → think toxic alcohol.
- RCF = 1.118 × 10⁻⁵ × r × rpm² · speed ±5% · timer ±10% · balance loads · spin capped.
- Hook = falsely low → dilute. HAMA = false sandwich (high) → blocking tube. Biotin: sandwich low, competitive high.