C(ASCPi) CHEMISTRY · LABORATORY OPERATIONS · MODULE 17

Instrumentation & Analytical Principles

Spectrophotometry & Beer's law · Detectors · Atomic absorption · Nephelometry, turbidimetry, fluorescence & luminescence · Chromatography (TLC, GLC, HPLC) & mass spectrometry · Potentiometry, amperometry, coulometry & ion-selective electrodes · Osmometry · Electrophoresis & hemoglobin electrophoresis · Automation, centrifuges, maintenance & point-of-care testing · Immunoassays & their interferences

Simple English · Question sources: BOC · Harr · SUCCESS! · CLSR (adapted) + original questions · Answers checked against Tietz Textbook of Laboratory Medicine, 7th ed. · Works offline · questions · Version 1.5

Independent exam-preparation material. Not affiliated with or endorsed by ASCP or the ASCP Board of Certification.

📘 Study progress

Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 very high-yield · 📌 must memorize · ⚠️ exam trap · 🧠 memory trick · 🔬 lab connection · 💡 understand this.

START HERE

1 · Light, colour and the spectrophotometer

Most chemistry tests end with a coloured product. The instrument does not "see" colour like we do. It shines light of one chosen wavelength through the solution and measures how much light gets through. More analyte → more colour → less light passes.

Light basics

  • Light behaves as a wave and as particles (photons).
  • Wavelength (in nanometres, nm) = distance between two wave crests. 📌 Shorter wavelength = more energy.
  • Order from short to long: gamma → X-ray → ultraviolet (UV) → visible (about 400–700 nm) → infrared (IR) → microwaves → radio.
  • The colour you see is the light a solution transmits (lets through). The photometer is set to the wavelength the solution absorbs most.
  • When an atom or molecule absorbs a photon, an electron jumps to a higher energy level. It is unstable, falls back, and may give off light of a characteristic wavelength (a "line spectrum" that identifies the substance).

The light path (memorise the order)

Power supply → LAMP → entrance slit → MONOCHROMATOR → exit slit → CUVET (sample) → DETECTOR → READOUT
PartJobKey exam facts
Lamp (light source)Bright, steady lightTungsten = visible and near-IR (continuous). Deuterium (or hydrogen) = UV. Mercury vapour = sharp lines (used to check wavelength). ⚠️ After changing a lamp → recalibrate (light hits the monochromator at a new angle).
MonochromatorPicks a narrow band of wavelengthsGlass / interference filters → simple photometers. Prisms and diffraction gratings → spectrophotometers.
BandpassWidth of the band passed (measured at half of peak transmittance)Centred on the setting: 10-nm bandpass at 540 nm = 535–545 nm.
CuvetHolds the solution; fixed light path (b)Quartz for UV, glass/plastic for visible. Scratches = errors.
DetectorChanges light into an electric signalSee Section 2.

Beer's law — the heart of photometry

A = a × b × c (A = absorbance, a = absorptivity, b = light path in cm, c = concentration) Same cuvet + same wavelength → A is proportional to c → Cu = (Au ÷ As) × Cs Absorbance ↔ transmittance: A = 2 − log %T 100 %T → A = 0 10 %T → A = 1.0 1 %T → A = 2.0
🔬 WORKED EXAMPLE 1 — A 100 mg/dL standard reads A = 0.400; the patient reads A = 0.300.
Cu = (0.300 ÷ 0.400) × 100 = 75 mg/dL. Check: lower absorbance than the standard → answer must be below 100.
🔬 WORKED EXAMPLE 2 — A solution transmits 25 % of the light. A = 2 − log 25 = 2 − 1.398 = 0.602.

Blanks and two-wavelength reading

  • Reagent blank (reagents, no sample): zeroes the instrument and removes the colour of the reagents only. It does not correct lipemia, hemolysis or sample chromogens.
  • Bichromatic analysis: every sample is read at two wavelengths (peak and base). The difference removes background colour (bilirubin, hemoglobin) — each sample blanks itself.

Checking the spectrophotometer

CheckQuestion it answersTool
Wavelength accuracyIs "540 nm" really 540 nm?Holmium oxide (UV/visible) or didymium (visible/near-IR) glass; nickel sulfate solution; mercury or deuterium emission lines
Absorbance (photometric) accuracyIs A = 0.500 really 0.500?Solutions / filters of known absorbance
LinearityDoes A rise in a straight line with concentration?A series of increasing concentrations
Stray lightIs light of other wavelengths leaking to the detector?Cut-off filters / solutions

Reflectance photometry (dry slides)

On dry-slide analyzers the reaction happens in thin reagent layers on a slide. Light passes through the layers, bounces off a white reflective layer, and the unabsorbed, reflected light returns to the detector. Reflectance is not linear with concentration, so the computer uses an algorithm to convert it.

🔥 VERY HIGH-YIELD: Tungsten = visible · Deuterium = UV · Monochromator = picks the wavelength · Holmium oxide = wavelength check · Reagent blank = reagent colour only · A = 2 − log %T · Cu = (Au/As) × Cs.

Transmittance, the zero settings and when Beer's law fails

%T = (I ÷ I₀) × 100 (I = light leaving the sample, I₀ = light entering it) More analyte → I falls → %T falls (log curve) → A rises (straight line)
  • Setting up a single-beam instrument: block the light path completely → set 0 %T; put in the reagent blank → set 100 %T (A = 0); then read calibrators and patients.
  • Double-beam instrument: a chopper and mirrors send the beam alternately through the sample and a reference cuvet onto one detector. This cancels drift of the lamp and changes in detector sensitivity.
  • Beer's law holds only when: the light is truly monochromatic, the solvent hardly absorbs, the concentration stays inside the linear range, no other substance absorbs at that wavelength, and the analyte does not react with the solvent or other solutes. A wide bandpass (broad light) usually gives a lower absorbance than expected.
  • Standard (calibration) curve: plot A of several calibrators against their concentration. Use at least two calibrators, and never use a single calibration factor above the linear part of the curve.
  • Absorbed energy can also change how bonds vibrate or rotate (the basis of infrared absorption), not only move electrons.
🔬 LAB CONNECTION — Patient absorbance above the highest linear calibrator: dilute the sample (for example 1 + 1 = 1:2), re-assay, and multiply by 2. Example: diluted result 210 mg/dL × 2 = 420 mg/dL. Common error: forgetting the dilution factor, or extrapolating the curve instead of diluting.
⚠️ EXAM TRAP — %T is NOT proportional to concentration; absorbance is. A plot of %T against concentration curves downwards.

Setting up a method and a run

New method: scan the coloured product → choose the wavelength of MAXIMUM absorbance (λmax) → read the standards → plot the curve → check linearity Every run: set the wavelength FIRST → set zero absorbance with the blank (at that wavelength) → standards → controls → patients
  • Why λmax? Highest sensitivity, and a small wavelength error changes the reading least. Avoid the steep side of a peak.
  • Bandpass vs a narrow peak: if the bandpass is wider than the peak (e.g., NADH at 340 nm), the shoulders are included and the peak reads LOWER. Two instruments, same NADH solution → the one reading HIGHER has the narrower bandpass.
  • Wavelength check filters: holmium oxide for narrow-bandpass instruments; didymium for instruments with a bandwidth of about 8 nm or more.
  • UV region is below about 380 nm (340 nm = NADH); visible about 380–750 nm.
⚠️ EXAM TRAP — 'Wavelength of greatest %T' is the WORST choice: maximum transmittance means minimum absorbance.

Source: SUCCESS! in Clinical Laboratory Science, 4th ed., Ch.1 §I A–B (pp. 2–3), §I S.3b (p. 14); review questions Q1–13 (pp. 95–96) and rationales (pp. 145–147). · Tietz 7th ed., Ch 16 Optical Techniques, pp. 203–204 (gap check v-next) · Tietz 7th ed., Ch 16 Optical Techniques, pp. 201, 206, 208 (update 25 Oct 2026)

INSTRUMENT PARTS

2 · Detectors — turning light into a signal

The detector turns light energy into an electrical signal. More light reaching it = bigger signal. The readout then shows the result (digital display, printout, computer).

DetectorHow it worksRemember
Photomultiplier tube (PMT)Light knocks electrons off a sensitive surface; 10–15 dynode stages multiply them🔥 Very sensitive, very fast, amplifies up to ~106. Must be shielded from room light. Fast enough to read chopped (pulsed) light.
Photodiode array256–2048 photodiodes in a line, each reading its own wavelengthWhole UV–visible spectrum at once (resolution about 1–2 nm). Also an HPLC detector.
Barrier-layer cell / photocellIron plate + selenium layer; makes current directlySimple, cheap, no amplification. "Iron + selenium" = NOT a PMT.
PhototubeSingle light-sensitive cathode in a tubeSimpler than a PMT.
⚠️ EXAM TRAP: "A PMT cannot be used with a chopper" — FALSE. Its fast response is exactly why it works with a chopper (for example in atomic absorption).
🧠 MEMORY TRICK: PMT = Powerful (amplifies) · Must hide from light · Turbo-fast.

Source: SUCCESS! Ch.1 §I B.5–6 (p. 3); review questions Q3–5 (p. 95); rationales Q3–5 (pp. 145–146).

METALS

3 · Atomic absorption spectrophotometry (AAS)

Simple idea: free metal atoms in their ground state absorb light of their own special wavelength. Send that exact wavelength through the sample and see how much disappears.

Hollow-cathode lamp (cathode made of the metal being measured) ↓ light of that metal's lines CHOPPER (makes the light pulsed) ↓ FLAME (atomizer + "cuvet": turns the sample into free ground-state atoms) ↓ atoms absorb part of the light Monochromator → Detector (PMT, tuned to the pulsed signal) → Readout
  • Hollow-cathode lamp: anode + cylindrical cathode made of the analyte metal + inert gas (argon, neon or helium). The ionized gas hits the cathode, excites the metal atoms, and they emit that metal's lines.
  • Flame: breaks compounds into atoms and keeps most of them at ground state (about 99%) — it does not excite them.
  • What is measured: the lamp beam after it passes through the flame (with and without sample). The decrease = absorbed light, related to concentration.
  • Chopper + tuned amplifier: only the pulsed lamp light is amplified, so steady light from the flame and light emitted by excited analyte atoms is ignored. They do not fix flame-temperature changes or lamp ageing.
  • Flameless AAS (graphite furnace / carbon rod, tantalum or platinum): sample vaporized in a heated chamber → more sensitive than flame.
  • Used for metals: lead, zinc, copper, aluminium, magnesium, calcium, lithium.
⚠️ EXAM TRAP: "What does the AAS actually MEASURE?" → the intensity of the lamp beam after the flame — not "the light absorbed" (that is calculated).
🧠 MEMORY TRICK: Lamp = Light, Flame = Free atoms, Chopper = Code (pulses) the detector listens for.

Flame emission, AAS interferences and ICP-MS

Flame emission photometryAtomic absorption
Flame's jobExcites the atoms → they EMIT their own colourOnly frees the atoms → they stay at ground state and ABSORB lamp light
SignalEmitted light ∝ concentrationFall in lamp light ∝ concentration
Flame coloursNa yellow · K violet · Li red · Mg blue—
StatusOnce used for Na, K, Li; now mostly replaced by ISEsAbout 100× more sensitive than flame emission and very specific
  • Chemical interference: phosphate binds calcium in the flame → falsely low Ca. Fix: add lanthanum (or strontium), which takes the phosphate.
  • Ionization interference: an easily ionized element (such as K) changes how much of the analyte becomes ions. Fix: add a large amount of an easily ionized element to every sample.
  • Background absorption (mainly flameless AAS): corrected by the Zeeman method — a strong magnetic field splits the absorption line, so total and background absorbance can be read separately at the same wavelength.
  • ICP-MS (inductively coupled plasma mass spectrometry) is replacing AAS for trace metals: it measures many metals at once. AAS stays cheaper and simpler.
🧠 MEMORY TRICK — "La La for Ca": add LAnthanum when phosphate steals calcium.

Source: SUCCESS! Ch.1 §I C (pp. 3–4); review questions Q21–23 (pp. 97–98); rationales (pp. 148–150). · Tietz 7th ed., Ch 16 Optical Techniques, pp. 209–211 (gap check v-next)

LIGHT METHODS

4 · Light scatter, fluorescence and luminescence

Nephelometry vs turbidimetry

NephelometryTurbidimetry
MeasuresLight SCATTERED by particles, read at an angle (forward scatter for large particles)Light BLOCKED by particles = decrease in transmitted light, read straight through
InstrumentNephelometerOrdinary spectrophotometer
Main problemsBackground scatter: scratched cuvets, particles in reagents. Avoid UV light (can cause fluorescence → falsely high).Particle size must be similar in samples and standards; aggregation/settling → keep timing constant. Temperature is not critical (room temperature).

Fluorescence and the luminescence family

TypeWhat excites the electrons?Key points
FluorescenceAbsorbed light (often UV)Emits longer wavelength (lower energy) almost immediately. Detector at 90° to the lamp. About 1000× more sensitive than absorbance and more specific (two wavelengths chosen). Problems: quenching, self-absorption, pH/temperature/solvent changes; needs very pure reagents.
PhosphorescenceAbsorbed lightLike fluorescence but with a longer delay (> about 10-4 s) before emission.
ChemiluminescenceEnergy of a chemical reaction (no light needed)Extremely sensitive (attomole–zeptomole). Labels: acridinium ester, isoluminol. Read with a luminometer.
BioluminescenceAn enzyme-catalysed oxidationLuciferase + luciferin (firefly). A special type of chemiluminescence; more sensitive than direct fluorescence.
FLUOROMETER: lamp → primary filter (excitation λ) → cuvet → [90°] → secondary filter (emission λ) → detector Fluorometer = filters Spectrofluorometer = gratings / prisms

Fluorescence polarization (FPIA) — preview of Section 9

Plane-polarized light excites a fluorescent label. A label bound to antibody is big and turns slowly → its light stays polarized. A free label is small and spins fast → light becomes depolarized. More patient analyte → more free label → less polarization (inverse relationship). No separation step → homogeneous.

Counting radioactivity

Scintillation counter: a fluor turns radiation into flashes of light → photomultiplier tube. Gamma → solid sodium iodide crystal with thallium. Beta → liquid scintillation. Alpha particles are not measured in clinical laboratories. (Radioimmunoassay is now rare.)

🧠 MEMORY TRICK: "Nephele" = cloud → light scattered by the cloud. Turbidity = the cloud blocks the light.
FLUORO = light lights it · CHEMI = chemistry lights it · BIO = an enzyme lights it · PHOSPHO = lights it later.

Limits of fluorescence and how labs beat them

  • Stokes shift = gap between the peak excitation wavelength and the peak emission wavelength (energy lost before light is given off). A larger shift makes it easier to keep excitation light away from the detector.
  • Inner filter effect: at high concentration the sample itself absorbs the exciting or emitted light → the signal stops rising in a straight line. Fluorescence is linear only in dilute solutions.
  • Background: Rayleigh and Raman scatter of the exciting light, and natural fluorescence of serum (proteins, bilirubin, drugs), add a false signal.
  • Time-resolved fluorescence: labels made from lanthanides (europium, samarium) glow for microseconds, while background fluorescence fades within nanoseconds. A pulsed lamp plus a delayed reading removes the background → much better signal-to-noise.
🔥 VERY HIGH-YIELD — Europium label = time-resolved fluorescence = background removed by waiting before reading.

Source: SUCCESS! Ch.1 §I D–F (pp. 4–5), §I U.2c (p. 17); review questions Q12, Q14–20 (pp. 96–97); rationales (pp. 147–148). · Tietz 7th ed., Ch 16 Optical Techniques, pp. 212, 217–218 (gap check v-next)

SEPARATION

5 · Chromatography and mass spectrometry

Simple idea: a mobile phase (gas or liquid) carries the sample over a stationary phase. Each substance travels at its own speed because of differences in solubility, volatility, charge or size — so they separate.

Thin-layer (TLC)Gas–liquid (GLC / GC)High-performance liquid (HPLC)
Stationary phaseSilica gel on a glass plate or plastic sheet (planar)Liquid coated on inert packing (e.g. Celite) or a capillary columnOrganic material bonded to silica in a column (polar or nonpolar)
Mobile phaseLiquid solvent rising by capillary actionInert gas: helium, nitrogen, hydrogen, argonLiquid pumped through: water, methanol, ethanol, acetonitrile, isopropanol
Separates bySolubility in solvent vs attraction to silicaVolatility and interaction with the columnDistribution between the two phases
Identify / measureRf value + colour sprays, UV, fluorescence, heatRetention time (identity); peak area/height (amount)Retention time; peak size
Typical useUrine drug screeningVolatile substances, alcohols, drugs (with MS)TDM, vitamins, hormones

TLC details

  • Steps: extract → concentrate → spot on the plate → develop → detect (sprays, UV, fluorescence, heat) → compare Rf with standards.
  • Solvent too polar → solute runs with the solvent front. Solvent not polar enough → solute stays at the origin.
  • Rf changes with chamber saturation, temperature, humidity and solvent composition.
🔬 WORKED EXAMPLE — Rf = distance origin → spot centre ÷ distance origin → solvent front. 48 mm ÷ 141 mm = 0.34. (Never use "spot to front" in the formula.)

GLC details

  • Parts: carrier gas + flow control, heated injector (about 50 °C hotter than the column, so the sample vaporizes), column (packed or capillary) in an oven, detector, computer.
  • More volatile solute, or less interaction with the column → elutes faster. Carrier-gas flow rate must be steady for reproducible retention times.
  • Detectors: flame ionization, thermal conductivity, electron capture, mass spectrometer (best identification).

HPLC details

  • Parts: solvent reservoir(s), pump(s), loop injector, pre-column/guard column, column (usually room temperature), detector (UV/visible, photodiode array, fluorometer, electrochemical, MS).
  • Isocratic elution = constant solvent strength. Gradient elution = solvent strength increases during the run (better resolution).
  • Normal phase = polar stationary + nonpolar mobile. Reversed phase = nonpolar stationary + polar mobile.
  • Other column methods: ion exchange separates by the sign and size of charge; size exclusion by molecular size.

Mass spectrometry (MS)

Sample → ION SOURCE (electron/chemical ionization; electrospray with HPLC) → fragments → ANALYZER (e.g. quadrupole) separates ions by MASS-TO-CHARGE ratio (m/z) [inside a VACUUM] → DETECTOR (electron multiplier) → COMPUTER compares the spectrum with a library
  • The mass spectrum is unique for a compound (identification); the number of ions ∝ amount (quantification).
  • The vacuum stops ions colliding with other molecules.
  • Coupled systems: GC-MS and HPLC-MS; tandem MS (GC/MS/MS, HPLC/MS/MS) adds a second fragmentation step for more selectivity.
  • Uses: drugs and metabolites, steroids, amino-acid composition of proteins, proteomics.
⚠️ EXAM TRAP: In GLC the mobile phase is a GAS. "Liquid pumped through the column" = HPLC. Celite = GLC support, not an electrophoresis medium.
🧠 MEMORY TRICK: TLC = plate & Rf · GC = gas & retention time · HPLC = pump & liquid · MS = m/z fingerprint.

Five ways liquid chromatography separates

TypeSeparates byRemember
Adsorption (liquid–solid)Sticking to and leaving the surface of a solid support (often silica)Solutes that like the mobile phase move fastest
Partition (liquid–liquid)Sharing between a liquid stationary layer and the mobile phase (polarity)Basis of normal- and reversed-phase HPLC; most clinical LC
Ion exchangeBinding to fixed charges on a resin (cation or anion exchanger)Mixed cation + anion resins deionize water; cation-exchange HPLC quantifies Hb A2, Hb F and Hb variants
Size exclusion (gel filtration)Size — small molecules enter the pores and are delayedLargest molecules come out FIRST
AffinityA biological binding partner (antibody, lectin, boronate)Most specific

Gas chromatography also comes in two kinds: gas–solid (solid adsorbent) and gas–liquid (liquid coated on a solid; the usual clinical type).

⚠️ EXAM TRAP — In size exclusion the big molecule is NOT slowed down: it cannot enter the pores, so it takes the short route and elutes first.

Mass spectrometry — ion sources and analyzers

PartOptionsKey point
Ion sourceElectron ionization (EI) · chemical ionization (CI) · electrospray (ESI) · APCI · MALDI · ICPEI breaks molecules into many fragments → library "fingerprint" (GC-MS). CI, ESI and MALDI are soft (little fragmentation). ESI = the most used clinical source (with HPLC).
AnalyzerQuadrupole · time-of-flight (TOF) · ion traps (incl. Orbitrap) · magnetic sectorQuadrupole scans m/z one value after another; TOF takes the whole spectrum at once (lighter ions arrive first); magnetic sector is rare (large, costly).
Tandem (MS/MS)Triple quadrupole: MS1 → collision cell (gas breaks the selected ion) → MS2Selected (multiple) reaction monitoring follows one parent-ion → product-ion pair → very selective quantification.
🔬 LAB CONNECTION — MALDI-TOF identifies bacteria and yeasts from colonies; LC-MS/MS is used for steroids, vitamin D, immunosuppressant drugs and newborn screening.

Chromatography troubleshooting and modes

  • All chromatography works because compounds share themselves differently between a mobile and a stationary phase. Modes: adsorption, partition, ion exchange, size exclusion (gel filtration), affinity.
  • Size exclusion: large molecules cannot enter the pores and leave the column FIRST; small ones take the long path.
  • HPLC pressure rising + abnormal peaks → something is blocked (frit, guard column, tubing). Detector or injection faults change the peaks, not the pressure.
  • GC: analytes must be volatile (or derivatized). TLC Rf = spot distance ÷ solvent-front distance (0–1); Rf 0.5 = halfway.

Source: SUCCESS! Ch.1 §I G–K (pp. 5–8); review questions Q48–58 (pp. 100–101); rationales (pp. 154–155). · Tietz 7th ed., Ch 19 Chromatography, pp. 242, 254–257; Ch 77 Hemoglobin and Hemoglobinopathies, p. 1066 (gap check v-next) · Tietz 7th ed., Ch 20 Mass Spectrometry, pp. 275–288 (gap check v-next) · Tietz 7th ed., Ch 19 Chromatography, pp. 241, 245, 247, 254–255, 268 (update 25 Oct 2026)

ELECTRODES

6 · Electrochemistry — potentiometry, amperometry, coulometry, polarography

MethodWhat is measuredClinical examples
PotentiometryVoltage (potential) between an indicator and a reference electrode, at (near) zero currentpH, PCO2, ion-selective electrodes for Na+, K+, Cl−, Ca2+, NH4+
AmperometryCurrent from oxidation or reduction at an electrode held at a fixed potentialPO2 (Clark) electrode, amperometric glucose electrode
CoulometryQuantity of electricity (current × time) — Faraday's lawChloride titrator (sweat chloride)
Polarography / anodic stripping voltammetryCurrent while the voltage is changed (polarogram)Lead and other heavy metals (ASV)

Potentiometry basics

  • A half-cell (electrode) = metal conductor in an electrolyte. Two half-cells joined by a salt bridge make a cell; the potential difference is measured.
  • Standard hydrogen electrode = 0.000 V (the universal standard).
  • Reference electrodes (constant potential): calomel (mercury + mercurous chloride + saturated KCl) and silver/silver chloride.
  • Indicator electrode: its potential changes with the analyte.
  • Electrons flow anode → cathode: oxidation at the anode, reduction at the cathode.

Blood gas electrodes

ElectrodeTypeHow it worksCalibration
pHPotentiometric glassThin pH-sensitive glass tip filled with HCl/AgCl and an internal Ag/AgCl wire. The glass must be hydrated (H+ exchanges with metal ions in the glass). Store in the manufacturer's recommended solution.Two standard buffers, at constant 37 °C
PCO2 (Severinghaus)Potentiometric — "a pH electrode in disguise"pH electrode in bicarbonate solution behind a membrane permeable to CO2 gas (not to H+). CO2 in → carbonic acid → pH falls. pH is inversely related to log PCO2.Calibrating gases; barometric pressure needed
PO2 (Clark)AmperometricPlatinum cathode + Ag/AgCl anode + polarizing voltage, gas-permeable (polypropylene) membrane. O2 reduced at the cathode → current directly proportional to PO2.Calibrating gases; barometric pressure needed
🔥 MEASURED: pH, PCO2, PO2. CALCULATED: bicarbonate, base excess (Henderson–Hasselbalch).

Ion-selective electrodes (ISE)

  • Potassium: liquid ion-exchange membrane with valinomycin as the neutral carrier.
  • Ammonium: nonactin + monactin.
  • Sodium: special glass (silicon dioxide, sodium oxide, aluminium oxide). Errors from protein build-up on the membrane.

Amperometric glucose electrode

Glucose + O2 --glucose OXIDASE--> gluconic acid + H2O2 H2O2 oxidized at a PLATINUM electrode (positive potential) → current ∝ glucose (no hexokinase!)

Chloride coulometer

A constant current flows from a silver anode (to a second electrode), generating Ag+ ions, which precipitate chloride as AgCl. When all chloride is used up, free Ag+ appears and is reduced at a separate (amperometric) indicator electrode, so the current suddenly rises → generation stops. The time of current flow is proportional to chloride. Principle = coulometry + amperometry. Still used for sweat chloride.

Anodic stripping voltammetry (ASV)

Step 1: a negative potential plates trace metal ions onto the electrode (preconcentration). Step 2: the metal is stripped off while the current is recorded; the pattern identifies and measures the metal. Based on polarography, not potentiometry. Classic for lead in blood.

🧠 MEMORY TRICKS: "AN OX, RED CAT" (Anode = Oxidation, Reduction = Cathode) · K-VAL (potassium = valinomycin) · CHLORIDE = COULOMETRY · Oxygen = Amperes.

Osmometry — counting dissolved particles

Osmolality = the number of dissolved particles per kilogram of water. It does not matter whether the particle is big or small, charged or neutral — each particle counts once.

Add solute particles → osmotic pressure ↑ · vapor pressure ↓ · boiling point ↑ · freezing point ↓ (these four changes are the COLLIGATIVE properties)
Osmometer typeWhat it readsKey point
Freezing point depression (routine)Sample is supercooled, freezing is started by agitation, a thermistor reads the plateau temperatureDetects volatile solutes (ethanol, methanol); not affected by room temperature. Used by almost all clinical labs.
Vapor pressure (historical)Fall in vapor pressure (dew point)Misses volatile solutes such as methanol and ethanol — they escape into the vapor.
Osmolality (mOsm/kg) = freezing point depression (°C) ÷ 1.86 × 1000 Example: −0.53 °C → 0.53 ÷ 1.86 × 1000 ≈ 285 mOsm/kg
🔬 LAB CONNECTION: Calculated osmolality ≈ 2 × Na + glucose/18 + BUN/2.8 (mg/dL units). Osmolal gap = measured − calculated. A raised gap warns of an extra solute, most often a toxic alcohol (methanol, ethylene glycol, isopropanol, ethanol). Comparing measured and calculated values also helps to confirm pseudohyponatremia.
⚠️ EXAM TRAP: Suspected methanol poisoning + vapor pressure osmometer = falsely "normal" osmolality. Only a freezing point osmometer shows the gap.
🧠 MEMORY TRICK: Colligative = COUNT the particles, never weigh them. NaCl gives ~2 particles, glucose gives 1.

Source: Tietz Textbook of Laboratory Medicine, 7th ed., Ch 37 Electrolytes and Blood Gases, pp. 415.e9–415.e11.

Nernst equation and direct vs indirect ISE

E = E° + (0.0592 ÷ n) × log (activity) at 25 °C (0.0615 at 37 °C) n = charge of the ion: Na⁺, K⁺ → n = 1 · Ca²⁺ → n = 2

🔬 Worked example: a 10-fold rise in K⁺ activity changes the potential by about 59 mV at 25 °C (≈ 61.5 mV at 37 °C). For Ca²⁺ the same 10-fold rise gives only about 30 mV. Common error: forgetting to divide by n for divalent ions.

Selectivity: every ISE responds a little to other ions; the selectivity coefficient tells how much. A small coefficient = a good, selective electrode.

Direct ISEIndirect ISE
SampleUndiluted (blood gas analyzers, POCT, whole blood)Diluted first (large chemistry analyzers)
MeasuresActivity in plasma WATER (molality)Concentration in total plasma volume
Very high lipids or proteinNot affectedFalsely low Na⁺ (pseudohyponatremia) — less water per litre, but the dilution assumes normal water
🔥 VERY HIGH-YIELD — Plasma is normally about 93% water; with severe lipemia or paraproteinemia it can drop to about 80%, so indirect ISE (and flame photometry) can read up to ~20% low. Direct ISE gives the true value.
⚠️ EXAM TRAP — Same patient, Na⁺ 128 on the main analyzer and 140 on the blood gas analyzer, triglycerides very high → not an instrument fault: indirect vs direct ISE.

Troubleshooting pH meters and ISEs

  • An ISE senses the ACTIVITY of the free ion and is SELECTIVE, not perfectly specific; the membrane (special glass, valinomycin, ion exchanger) decides which ion it 'sees'.
  • Noisy, drifting readings with EVERY calibrator → look at the REFERENCE electrode: a liquid junction blocked by KCl crystals or a low filling solution. Dissolve the crystals, refill with fresh electrolyte, recalibrate.
  • Slow, sluggish response only with serum (fast with aqueous calibrators) → protein film on the ion-selective membrane.

Source: SUCCESS! Ch.1 §I L–O (pp. 8–11); review questions Q24–36 (pp. 98–99); rationales (pp. 149–152). · Tietz 7th ed., Ch 17 Electrochemistry and Chemical Sensors, pp. 224.e2–224.e3, 224.e8 (gap check v-next) · Tietz 7th ed., Ch 17 Electrochemistry and Chemical Sensors, pp. 224.e1–224.e4 (update 25 Oct 2026)

SEPARATION BY CHARGE

7 · Electrophoresis (including hemoglobin electrophoresis)

Simple idea: charged molecules move in an electric field. Their speed depends on net charge, size and shape, the buffer, the support medium, temperature and the strength of the field.

Buffer pH 8.6 (alkaline) → proteins are ABOVE their isoelectric point → NEGATIVE charge → move to the ANODE (+) Albumin fastest (nearest the anode) … gamma-globulins slowest (nearest the cathode; may drift back by endosmosis) Isoelectric point (pI) = pH where net charge = 0 → the protein does NOT move
TopicKey facts
Support mediaAgarose gel, cellulose acetate (5–7 fractions; routine), starch gel and polyacrylamide (smaller pores, ≈20 fractions). Celite is NOT an electrophoresis medium (GLC).
StainsSerum proteins: Ponceau S, amido black, Coomassie brilliant blue · CSF/urine proteins: silver (≈30× more sensitive than Coomassie; no preconcentration needed) · Lipoproteins: fat red 7B, oil red O · LD isoenzymes: nitrotetrazolium blue
QuantificationScan the stained bands with a densitometer
TroubleshootingHoles ("doughnut") in a band = protein too concentrated (dilute and repeat) · Very slow migration = voltage too low · Sample precipitates = pH too high/low or too much heat
High-resolutionAgarose, higher voltage, cooling, more concentrated buffer → up to 12 zones
Isoelectric focusing (IEF)A type of zone electrophoresis. Ampholytes create a pH gradient; each protein moves until gel pH = its pI, then stops. Separates proteins with very close pIs.
Capillary electrophoresisBased on electro-osmotic flow toward the cathode; cations move fastest.

Hemoglobin electrophoresis

  • Hemoglobin = 4 globin chains + 4 heme groups + 4 iron atoms. Hb A1 = α2β2, Hb A2 = α2δ2 (< about 4%), Hb F = α2γ2.
  • Hb S = valine replaces glutamic acid at position 6 of the beta chain. Hb C = lysine replaces glutamic acid at the same position.
  • Hb F resists alkali denaturation (alkali test). Normal Hb F after 1 year of age is below about 1%.
  • Solubility test (sodium hydrosulfite + concentrated phosphate buffer): turbid = Hb S. Screening only.
SystemWhat happens
Cellulose acetate, pH 8.6 (screening)Toward the anode: A1 > F > S = G = D > A2 = C = E = O. A2 and C (also S and D) cannot be told apart here.
Citrate agar, pH 6.2 (confirmation)A different pattern: A1, A2, D, E and G run together as one group, so C separates from A2 and S separates from D/G.
⚠️ BOOK NOTE (needs verification, not taught): the book's outline gives an exact migration direction for the pH 6.2 agar system that looks reversed compared with common teaching. Only learn here that the acid system separates C from A2 (and S from D). See the track's "Needs verification" list.
🧠 MEMORY TRICK: Alkaline order "A Fast Swimmer Arrives Clean" (A1, F, S, A2/C). Doughnut band = too much "dough" (protein).

What else changes the speed of migration

  • Ionic strength of the buffer: higher → thicker ion cloud around each protein → slower migration but sharper bands, and more current and heat. Lower → faster but broader bands.
  • Heat (Joule effect): current through the gel makes heat → water evaporates → resistance falls → current rises further. A constant-current power supply limits this; too much heat denatures proteins.
  • Wick flow: evaporation pulls buffer into the support from both ends, which can shift proteins. The cover/enclosed chamber reduces evaporation.
  • Electroendosmosis is small on starch, purified agarose and polyacrylamide (few surface charges) and large on media with many fixed negative charges.
🧠 MEMORY TRICK — "Strong buffer = Slow but Sharp."

Hemoglobin electrophoresis — reading alkaline + acid together

  • Hb A has a pI of about 6.8: negative at alkaline pH (moves to the anode) and positive at acid pH (moves to the cathode).
  • Modern screening uses agarose with barbital buffer, pH ≈ 9.2 (older books: cellulose acetate pH 8.6). Stains: amido black or Ponceau S.
  • Acid electrophoresis uses citrate agarose, pH ≈ 6.0–6.4, run only when the alkaline gel shows an abnormal band. Its band order is F – A – S – C, with F and C at opposite ends; D, G, E, O, A2 run with A.
Alkaline bandAcid bandPresumptive Hb
S positionS positionHb S
S positionA positionHb D or G
C positionC positionHb C
C positionA positionHb E (or A2)
C positionbetween A and SHb O-Arab
🔥 VERY HIGH-YIELD — Alkaline gel cannot split S from D/G or C from E/O/A2; the acid gel splits them. Results are presumptive — confirm with HPLC, IEF or DNA tests.
⚠️ EXAM TRAP — Buffer too WEAK (low ionic strength): current reading LOW, bands run FASTER and FARTHER but are blurred. Fix = fresh buffer of correct strength; turning down the voltage does not restore resolution. The buffer carries the current and sets the pH (and so the charge of each protein).

Source: SUCCESS! Ch.1 §I P–Q (pp. 11–13); review questions Q37–47, Q63–68 (pp. 99–102); rationales (pp. 152–154, 156–157). · Tietz 7th ed., Ch 18 Electrophoresis, pp. 226–229 (gap check v-next) · Tietz 7th ed., Ch 77 Hemoglobin and Hemoglobinopathies, pp. 1055–1057, 1067–1068 (gap check v-next) · Tietz 7th ed., Ch 18 Electrophoresis, pp. 226, 229 (update 25 Oct 2026)

LAB OPERATIONS

8 · Automation, instrument maintenance and point-of-care testing

Automation words

TermMeaning
Discrete analysisEach sample–reagent mixture in its own vessel (may be single-test or multi-test)
Centrifugal analysisCentrifugal force moves samples and reagents into cuvets for simultaneous analysis
Random accessAny test or panel in any order; STAT samples can be added ahead
Batch analysisSamples processed as a group
ThroughputMaximum number of tests per hour
Turnaround timeTime to generate one result
Dead volumeSpecimen in the cup/tube that cannot be aspirated
Carry-overContamination of a sample by the previously aspirated sample
Reflex testingA first result decides whether more tests are added or cancelled
Stand-alone / modular workcell / multiple platform / integrated modular systemOne instrument of one discipline / ≥2 instruments of one discipline with one controller / one instrument doing tests from ≥2 disciplines / ≥2 analytical modules sharing one sample-and-reagent system
Total laboratory automationSamples received, centrifuged, sorted and loaded by conveyor without an operator (high-volume labs)

How analyzers work

  • Specimen handling: level-sensing probes, bar-code readers (patient, sample and reagent identification).
  • Sampling errors (falsely low results): short sample, air bubble at the bottom of the cup, fibrin clot in the probe.
  • Reagents: dry (lyophilized powder or tablets to reconstitute; dry multilayer slides) or liquid.
  • Testing: cuvets of quartz glass (permanent) or plastic (disposable); common reaction temperatures 37 °C and 30 °C.
  • Kinetic (rate) assay = concentration from the change in absorbance over time. Endpoint assay = incubate a fixed time, read once, compare with calibrators.
  • Data management: QC review and delta checks (compare with the patient's previous result) before release; results and QC archived (with the laboratory information system).

Keeping instruments reliable

ActivityExamples
Function verificationMonitor temperature, electronic parameters, calibration, QC data — and correct when needed
Performance verificationConfirm the instrument performs to specification
Preventive maintenanceScheduled cleaning, replacing worn parts, adjustments (per shift, daily, weekly, monthly, yearly) for all equipment — balances, refrigerators, centrifuges, water baths, pipettes, analyzers — always documented. Purpose: less downtime, fewer costly repairs, longer equipment life.

Point-of-care testing (POCT)

  • Testing at or near the patient, outside the main laboratory, for immediate decisions (intensive care, emergency, coronary care, neonatal ICU).
  • Every POC instrument must be evaluated: results must correlate with the main laboratory analyzer; linearity, control ranges and reference ranges established.
  • Training of every operator — documented. Quality control, statistics and proficiency testing — documented.
⚠️ EXAM TRAP: Calibration, temperature and electronic checks = function verification, not "preventive maintenance". Dead volume ≠ carry-over.

Centrifuges

RCF (× g) = 1.118 × 10⁻⁵ × r (cm, centre of rotor to tube bottom) × rpm² Example: r = 10 cm, 3000 rpm → about 1000 × g
  • RCF, not rpm, describes the separating force — rpm is squared, so a small speed error changes the force a lot.
  • Speed check (tachometer): within about 5% of the rated speed, at each speed used.
  • Timer check against a stopwatch: within about 10%.
  • Refrigerated centrifuge: temperature checked (about monthly), within about 2 °C of the setting; ordinary centrifuges warm up during a run.
  • Balance opposite loads (difference about 1% or the manufacturer's limit). Use only tubes rated for the RCF.
  • Safety: keep tubes stoppered until after spinning (aerosols); do not free clots with wooden sticks (hemolysis); after a breakage clean the chamber and cushions as a biohazard.
🔥 VERY HIGH-YIELD: RCF formula, speed ±5%, timer ±10%, balanced and capped.

Source: Tietz Textbook of Laboratory Medicine, 7th ed., Ch 15 Principles of Basic Techniques and Laboratory Safety, pp. 199.e10–199.e11.

More automation terms and specimen-handling checks

  • Continuous-flow analyzer (the first automated analyzer): samples are pulled one after another through the same tubing with liquid reagents. Shared tubing → carry-over risk. Today most analyzers are discrete.
  • Automated stand-alone: a single-discipline instrument with extra built-in functions such as auto-repeat and auto-dilution (the analyzer dilutes high samples itself and calculates the final result).
  • Liquid-level sensing works by pressure or capacitance; clot detection stops a short aspiration or a blocked probe. "Clot errors" in a sample with no visible clot → think of a very viscous sample (hyperviscosity, e.g. paraprotein).
  • Pneumatic tube system: fast transport, but sudden acceleration can hemolyse samples; the system must be validated and carriers well packed.
  • Automation pros: more tests per hour, less variation, fewer manual errors, smaller sample and reagent volumes. Cons: high start-up cost, risk that the product is discontinued, and skilled staff needed.

POCT devices — what can go wrong

  • Glucose meters read strips by reflectance or, now mostly, electrochemically (enzyme + mediator such as ferrocene → current).
  • Glucose oxidase strips: very specific for glucose, but high PO₂ (patient on oxygen) → falsely low glucose.
  • Glucose dehydrogenase strips: not affected by oxygen; older types read maltose (and other sugars) as glucose → falsely high (newer engineered enzymes fixed this).
  • Hematocrit extremes still disturb many strips.
  • Immunoassay POCT (pregnancy, cardiac, infection tests) uses solid-phase lateral-flow (immunochromatography) or flow-through cassettes, read by eye or by reflectance/fluorescence readers.
⚠️ EXAM TRAP — ICU patient on high-flow oxygen with a low meter glucose but normal laboratory glucose → suspect a glucose-oxidase strip affected by oxygen.

Centrifuges, centrifugal analyzers and sampling faults

  • RCF = 1.118 × 10⁻⁵ × r × rpm², r = RADIUS in cm. Worked example: rotor 60 cm across, 3000 rpm → r = 30 cm → 1.118 × 10⁻⁵ × 30 × 9,000,000 ≈ 3000 × g. Common errors: using the diameter (≈ 6000 × g) or not squaring rpm.
  • Speed is checked with a tachometer, timer with a stopwatch — dial settings are not proof.
  • Centrifugal analyzer: spinning throws sample and reagent from the inner wells into the outer cuvets, mixing them so all reactions start together; the cuvets are read as they pass the light beam.
  • Wrong results grouped by one probe (e.g., every photometric test very low or 'below detection' while electrolytes from a second probe look normal) → clot/fibrin in that probe. Stop, clean, rerun QC and the samples.
  • POCT: main advantage = short turnaround time and small sample volume; usually higher cost per test and many non-laboratory operators to train.

Source: SUCCESS! Ch.1 §I R–T (pp. 13–15); review questions Q52, Q59–62 (pp. 101–102); rationales (pp. 155–156). · Tietz 7th ed., Ch 29 Automation in the Clinical Laboratory, pp. 324.e3, 324.e6, 324.e10–324.e11, 324.e15 (gap check v-next) · Tietz 7th ed., Ch 30 Point-of-Care Testing, pp. 329, 331 (gap check v-next) · Tietz 7th ed., Ch 15 p. 199.e10; Ch 29 pp. 324.e6, 324.e10–e11; Ch 30 pp. 327, 331 (update 25 Oct 2026)

ANTIGEN + ANTIBODY

9 · Immunochemical techniques (immunoassays)

Simple idea: an antibody binds the analyte (the antigen) very specifically. A label on the antigen or the antibody gives a measurable signal. This lets us measure tiny amounts: hormones, drugs, vitamins (B12, folate), tumor markers, CK-MB mass. Electrolytes are not measured this way (ion-selective electrodes).

  • Polyclonal antiserum = antibodies from many cell clones (a mixture). Monoclonal = from one clone (uniform) → higher specificity.
  • All immunoassays (and competitive protein binding) use a binder with limited binding sites. HPLC does not — it separates physically.

Competitive binding

Patient antigen (unlabelled, varies) + labelled antigen (fixed) → compete for → limited antibody sites (fixed) More patient antigen → LESS labelled antigen bound → more labelled antigen left FREE Measure the bound OR the free label and relate it to the patient concentration (via calibrators)
HeterogeneousHomogeneous
Bound and free label must be physically separated (wash): RIA, ELISA, IRMANo separation needed — the signal differs between bound and free label: EMIT, FPIA, LOCI

Labels: radioactive (I-125, old RIA), enzyme (alkaline phosphatase, glucose-6-phosphate dehydrogenase), fluorophore (fluorescein), chemiluminescent (acridinium ester, isoluminol), ruthenium (electrochemiluminescence).

The named techniques

TechniqueHow it worksSignal vs analyte
EMIT — enzyme-multiplied immunoassay techniqueEnzyme is coupled to the antigen (drug). Patient drug and enzyme-labelled drug compete for antibody. Antibody binding blocks the enzyme. Only free enzyme-labelled drug reacts with substrate + coenzyme (absorbance change, e.g. at 340 nm).Enzyme activity rises with drug concentration
FPIA — fluorescence polarization immunoassayPolarized excitation; bound label = polarized emission, free label = depolarized. Used for drugs, hormones, fetal lung maturity.Polarization falls as concentration rises (inverse)
Chemiluminescent immunoassayIsoluminol or acridinium ester label; with H2O2 and a catalyst isoluminol is oxidized and emits light (425 nm)Signal proportional to analyte (format dependent)
LOCI — luminescent oxygen channeling immunoassayAntigen links two antibody-coated beads: a sensitizer bead (photosensitive dye) and an acceptor bead (chemiluminescent precursor + fluorophore). Light → singlet oxygen → diffuses to the nearby acceptor → chemiluminescence → fluorophore emits longer-wavelength light. Homogeneous.Signal proportional to analyte
Electrochemiluminescence (ECLIA)Ruthenium label; complexes on magnetic beads are captured at an electrode; after washing, the electrode triggers light read by a PMT. Sandwich or competitive.Format dependent
⚠️ EXAM TRAPS: In EMIT the enzyme reacts with its SUBSTRATE, not with the patient's drug · In EMIT the enzyme labels the ANTIGEN, not the antibody · FPIA is homogeneous and INVERSE · LOCI starts with LIGHT (radiant energy).
🧠 MEMORY TRICK: HOMOgeneous = "one pot, no wash" (EMIT, FPIA, LOCI). HETERO = "wash it" (RIA, ELISA, IRMA).

Immunoassay interferences

InterferenceMechanismEffectHow to detect / fix
High-dose hook effectHuge antigen excess saturates capture AND labelled antibodies in a one-step sandwich assayFalsely LOWRe-assay on dilution (result rises). Two-step assays with a wash are not prone.
Heterophile / HAMA (human anti-animal antibodies)Patient antibodies bridge the animal capture and detection antibodiesUsually falsely HIGH in sandwich assays (false low possible)Non-linear dilution, other platform disagrees, heterophile-blocking tube. Reagents contain non-immune animal serum/IgG.
Biotin (high-dose supplements)Free biotin blocks streptavidin binding sitesSandwich → falsely LOW; competitive → falsely HIGHBiotin-free method, wash-out period (≥ about 8 h), streptavidin beads, dilution
⚠️ EXAM TRAP: Biotin + thyroid tests can look like hyperthyroidism (TSH falsely low by sandwich, FT4 falsely high by competitive).
🧠 MEMORY TRICK: HOOK pulls the result DOWN · HAMA builds a FAKE sandwich · Biotin: Sandwich Sinks, Competitive Climbs.

Source: Tietz Textbook of Laboratory Medicine, 7th ed., Ch 26 Immunochemical Techniques, pp. 311, 319–320; Ch 5 Preanalytical Variation, p. 107.

Noncompetitive (sandwich, immunometric) immunoassay

Capture antibody on solid phase (EXCESS) + patient antigen → wash → add labelled second antibody (EXCESS) against a DIFFERENT epitope → wash → bound label measured: MORE antigen → MORE signal (direct relationship)
CompetitiveNoncompetitive (sandwich)
AntibodyLimited amountExcess (two antibodies)
What is labelledAntigenAntibody (e.g. IRMA, most ELISAs)
Signal vs analyteInverseDirect
Best forSmall molecules with one epitope (drugs, T4, steroids)Large molecules with ≥ 2 epitopes (TSH, hCG, troponin, ferritin)
Detection limit set byAntibody affinityHow well the label can be detected → most sensitive (especially with chemiluminescent labels)

One-step (simultaneous) sandwich assays are quicker but can show the hook effect; two-step (sequential) assays wash before adding the labelled antibody and are not prone to it.

🧠 MEMORY TRICK — Sandwich = Same direction (more antigen, more signal). Competition = Contrary (more antigen, less signal).
🧠 MEMORY TRICK — Label sensitivity (Tietz label table): enzyme-amplified chemiluminescence ≈ 1 zmol · electrochemiluminescence ≈ 20 · ¹²⁵I ≈ 1000 · enzyme read by colour ≈ 2,000,000. So luminescence > radioactivity > colour. Enzyme labels replaced isotopes partly because the product can be read on an ordinary photometer.

Source: SUCCESS! Ch.1 §I U (pp. 15–18); review questions Q16, Q69–76 (pp. 97, 102–103); rationales (pp. 147, 157–159). · Tietz 7th ed., Ch 26 Immunochemical Techniques, pp. 310–312 (gap check v-next) · Tietz 7th ed., Ch 26 Immunochemical Techniques, pp. 312–313 (update 25 Oct 2026)

⚠️ PUT IT TOGETHER

10 · Choosing the best answer — instrument "look-alikes"

If the question says…Think…Not…
Continuous visible + near-IR lightTungsten lampMercury vapour (lines), deuterium (UV)
Selects the wavelengthMonochromatorDetector
Iron plate + seleniumBarrier-layer cellPhotomultiplier tube
Whole spectrum at oncePhotodiode arrayPMT
Checks wavelengthHolmium oxide / didymiumLinearity standards
Light scattered at an angleNephelometryTurbidimetry (blocked light)
Detector at 90°Fluorometer (keeps lamp light out)"Prevents loss of emitted light"
Chemical energy makes lightChemiluminescence (enzyme → bioluminescence)Fluorescence
Ground-state atoms, hollow-cathode lampAAS (flame = atomizer, not exciter)Flame emission
Mobile phase is a gasGLCHPLC (liquid pumped)
Separates by charge sign and sizeIon exchangeSize exclusion
Identifies by mass-to-chargeMass spectrometryRetention time
Mercury + mercurous chloride + KClCalomel reference electrodeAg/AgCl
ValinomycinPotassium ISESodium (glass)
Silver ions generated, time ∝ chlorideCoulometry + amperometryPotentiometry
Current ∝ oxygenAmperometric PO2 electrodepH glass electrode
Plate then strip; leadAnodic stripping voltammetry (polarography)Potentiometry
pH gradient, ampholytesIsoelectric focusingZone electrophoresis with one pH
Electro-osmotic flowCapillary electrophoresisIontophoresis
Doughnut bandProtein too concentratedDenatured protein
Each test in its own vesselDiscrete analysisContinuous flow
No separation of bound/freeHomogeneous: EMIT, FPIA, LOCIRIA, ELISA, IRMA
🔥 TOP 10 TRAPS
1. Reagent blank ≠ sample blank (no correction for lipemia/hemolysis).
2. Bandpass is split in half around the setting.
3. Rf uses origin→spot ÷ origin→front.
4. PMT works WITH a chopper.
5. AAS measures the lamp beam after the flame.
6. pH calibration = two buffers at 37 °C (gases need barometric pressure).
7. Measured blood gases: pH, PCO2, PO2 only.
8. Amperometric glucose = glucose oxidase, not hexokinase.
9. Silver stain is MORE sensitive than Coomassie.
10. FPIA polarization is inversely related to concentration; EMIT activity rises with concentration.

Summary of SUCCESS! Ch.1 §I (pp. 2–18) and review questions Q1–76 (pp. 95–103).