1 · The quality system and the three phases of testing
Simple idea: a laboratory result is only useful if the right patient, the right specimen, the right analysis and the right report all come together. Every step can fail, so the laboratory controls every step.
| Term | Meaning | Example |
|---|---|---|
| Quality control (QC) | Statistical checks of the analytical process with control materials | Two controls per run plotted on a Levey–Jennings chart |
| Quality assurance / assessment (QA) | Monitoring of the whole process (pre, analytical, post) | Turnaround time, specimen rejection rate, critical-value calls |
| Quality management system (QMS) | Organisation, documents, training, audits and improvement for the whole laboratory | ISO 15189, CAP checklists, CLIA rules |
| External QC = proficiency testing (PT / EQA) | Unknown samples from an outside provider, graded against other labs | CAP surveys three times a year |
Quality improvement tools
- PDCA — Plan, Do, Check, Act: the basic improvement cycle.
- Six Sigma (DMAIC) — Define, Measure, Analyze, Improve, Control; reduces variation and defects.
- Lean — removes waste (waiting, extra movement, rework) → shorter turnaround time.
- Root cause analysis (RCA) — after a serious error, ask "why?" until the system cause is found.
- First step in any problem (for example a turnaround-time complaint): collect data (measure) before changing anything.
Total quality management (TQM) and ISO standards
- TQM = a management approach that improves every factor that affects testing, not only the analyzer. Its five parts (the "5 Q" model): quality planning, the quality laboratory process, quality control, quality assessment and quality improvement.
- ISO 9000 / ISO 9001 = general international quality-management standards for any organisation. ISO 15189 = the version written for medical laboratories (quality and competence). ISO 22870 = point-of-care testing.
- ISO meaning of quality: the result is fit for its intended use (it meets the stated requirement).
Tietz 7th ed., Ch 3 Governance, Risk and Quality Management, p. 60.e1–60.e16; Ch 5, p. 89; SUCCESS! Ch.13 §II; BOC Ch.8 Quality Assessment. · Tietz 7th ed., Ch 3 Governance, Risk and Quality Management, p. 60.e1, 60.e5–60.e6 (gap check v-next)
2 · The preanalytical phase and specimen errors
Patient identification
- Use two patient identifiers (for example full name + date of birth or medical record number). A room or bed number is never an identifier.
- Label tubes at the bedside, after collection, in front of the patient. Never pre-label.
Patient preparation and collection
| Factor | Effect |
|---|---|
| Recent meal | Triglycerides ↑↑, glucose ↑, phosphate ↑ (ALP may rise after fat). Cholesterol hardly changes. |
| Prolonged tourniquet (> 1 min) | Hemoconcentration: proteins, protein-bound substances (calcium, lipids, enzymes) ↑; K ↑ |
| Fist pumping | Potassium ↑ (can be > 1 mmol/L), phosphate and lactate ↑ |
| Drawing above an IV line | Dilution: analytes in the fluid ↑ (glucose, Na, Cl), others ↓ (K, CO2, Ca, proteins) |
| Wrong order of draw (EDTA carryover) | K ↑, Ca ↓, Mg ↓, ALP ↓ |
Tube additives
| Top | Additive | Chemistry use / warning |
|---|---|---|
| Red / gold (SST) | None or clot activator ± gel | Most chemistry; gel may absorb some drugs |
| Green / light green | Lithium or sodium heparin (± gel) | Plasma chemistry, blood gases; not for lithium if lithium heparin |
| Lavender | K2/K3 EDTA | HbA1c, CBC; not for K, Ca, Mg, ALP |
| Gray | Sodium fluoride + potassium oxalate | Glucose, lactate, alcohol; fluoride inhibits glycolysis (enolase) and some enzymes |
| Light blue | Sodium citrate | Coagulation |
Handling and delay
- Unseparated blood keeps metabolising: glucose falls about 10 mg/dL per hour (about 5–7%) at room temperature; potassium, phosphate, LD and AST leak out of cells; CO2 is lost. Separate serum within about 2 hours.
- Refrigerating unseparated blood stops the Na/K pump → potassium rises even without visible hemolysis.
- Bilirubin is destroyed by light — protect from light.
- Stable after delay: urea, creatinine, uric acid.
Hemolysis, icterus and lipemia (HIL)
| Problem | Main effects | What to do |
|---|---|---|
| Hemolysis (commonest preanalytical error) | K, LD, AST, Mg, phosphate, iron ↑; spectral interference | In vitro: recollect (or report unaffected tests with comment). In vivo: report with comment. |
| Icterus | Bilirubin absorbs light and reacts in peroxidase methods (creatinine, cholesterol, uric acid) | Check method limits; dilution or other method |
| Lipemia | Turbidity (light scatter); volume displacement (pseudohyponatremia by indirect ISE); non-homogeneous sample | Ultracentrifugation (CLSI recommended); never for lipid tests |
Blood gas specimens
- Heparinized syringe, no clots, no air bubbles, mixed gently.
- Air bubble: pO2 ↑, pCO2 ↓, pH ↑. Delay (cell metabolism): pO2 ↓, pCO2 ↑, pH ↓.
- CLSI: plastic syringe at room temperature, analyse within 30 min; ice slurry only for glass syringes when delay is expected. Excess liquid heparin dilutes the sample.
Biological variation and patient preparation
| Factor | What happens | Exam example |
|---|---|---|
| Circadian / diurnal (24-hour rhythm, linked to sleep–wake) | Some analytes peak at a fixed time of day | Cortisol highest in the early morning, lowest late evening; potassium lower in the afternoon. Draw at the time used for the reference interval (usually 7–9 am). |
| Seasonal (circannual) and monthly cycles | Repeating changes over the year or the menstrual cycle | 25-OH vitamin D higher in summer; total T3 about 20% lower in summer; reproductive hormones follow the menstrual cycle |
| Posture | Standing up moves plasma water into tissues → large molecules and cells concentrate | Supine → upright: proteins, cells, protein-bound substances ↑; total calcium ↑ 5–10%, free calcium unchanged. Rest at least 15 min before drawing. |
| Exercise | Muscle releases contents; mostly transient (back to baseline in about 3 h) | K ↑ (even fist clenching), CK, LD, ALT, troponin ↑, WBC ↑; regular heavy training keeps CK/LD high for longer |
| Stress (physical or mental) | Stress hormones are released | Cortisol ↑ (and total cholesterol ↑); other pituitary hormones and aldosterone may change |
Venipuncture technique — key facts
- Identification: inpatient → compare the ID band with the requisition (or scan it). No band → do not draw until the patient's nurse or physician identifies the patient; record that person's name. Outpatient → patient states two identifiers (e.g., name + date of birth).
- Site: median cubital vein is preferred (large, near the surface); cephalic next; basilic last (close to nerve and brachial artery). Avoid an arm with an IV line, hematoma, fistula or on the side of a mastectomy.
- Needle gauge: the bigger the number, the smaller the bore. Adult usual = 20–21 G; small or fragile veins / children = 23 G, often with a butterfly set.
- Needle enters bevel up at a low angle (about 15–30°). Release the tourniquet as soon as blood flows (within 1 min).
- Draw the smallest volume needed — repeated draws cause iatrogenic anemia (about 7 g/L hemoglobin lost per 100 mL drawn).
- Invert additive tubes gently (CLSI: citrate 3–4 times, serum tubes 5, heparin/EDTA/fluoride 8). Never shake.
- Activate the needle safety device and discard into a sharps container; never recap. Label at the bedside: patient name, ID number, date and time, collector's initials.
How the anticoagulants work
| Additive | Mechanism and amount | Exam point |
|---|---|---|
| EDTA (lavender) | Chelates (binds) Ca²⁺ and Mg²⁺; about 1.5 mg per mL blood (1–2 g/L) | Also inhibits ALP and CK (cofactor removal); too much EDTA (short draw) shrinks RBCs |
| Heparin (green) | Speeds up antithrombin → thrombin is neutralised; about 0.2 mg (≈18 units) per mL blood | Ammonium heparin → false high ammonia; lithium heparin → false lithium |
| Sodium citrate 3.2% (light blue) | Binds calcium; 9 parts blood : 1 part citrate | Fill to the mark (±10%). Underfilled tube → too much citrate → falsely long PT/aPTT |
| Gel separator (SST/PST) | Thixotropic gel moves between cells and serum/plasma during centrifugation | Not for blood bank; can absorb some drugs |
Transport, storage and chain of custody
- Ice and speed: ammonia (analyse within about 30 min) and lactate, pyruvate, ACTH, renin, gastrin → send on ice. Cells keep making ammonia and lactate at room temperature.
- Keep the cap on during centrifugation: prevents loss of CO₂ (pH ↑, ionized Ca changes) and of volatiles (ethanol ↓), reduces evaporation and aerosols.
- If a whole-blood sample cannot be spun within 2 h, keep it at room temperature, not in the fridge.
- Separated serum/plasma: refrigerate (2–8 °C) for short storage (about 2–3 days for many analytes); freeze at −20 °C or colder for long storage. Avoid frost-free freezers (temperature swings).
- Chain of custody (forensic drug or alcohol testing): sealed container, signatures and a document that follows the specimen through every hand-over, so it can be defended in court.
Tietz 7th ed., Ch 4 Specimen Collection and Processing, p. 64–70; Ch 5 Preanalytical Variation, p. 81–94 and p. 116–119; Harr §5.11; BOC Ch.3 Quality Assessment. · Tietz 7th ed., Ch 5 Preanalytical Variation, p. 81–82, 87–88 (gap check v-next) · Tietz 7th ed., Ch 4 Specimen Collection and Processing, p. 61, 63–66 (Table 4.2) (gap check v-next) · Tietz 7th ed., Ch 4, p. 65 (Table 4.2), 69–70; Ch 5, p. 120 (gap check v-next) · Tietz 7th ed., Ch 4, p. 72, 76–77 (gap check v-next)
3 · Quality control: materials, Levey–Jennings charts and Westgard rules
Simple idea: a control is a sample with a known expected value. If the control result is where it should be, the analyzer is probably working and patient results can be reported.
Control materials
- At least two levels (normal and abnormal), near medical decision points; matrix like patient samples; stable (one lot for about a year); go through the same pretreatment as samples.
- Calibrator = sets the signal–concentration relationship. Control = checks it. Never calibrate with a control.
- New control lot: target from the lab's own results over at least about 10 days; keep the established long-term SD. Package-insert and peer-group SDs are too wide.
- CLIA minimum for quantitative tests: two levels every day of testing (or an approved IQCP). Many labs run controls every shift.
Levey–Jennings chart
Westgard multirules
| Rule | Meaning | Error type |
|---|---|---|
| 12s | One result beyond ±2 SD | WARNING only (5% false alarms) |
| 13s | One result beyond ±3 SD | Random (or large systematic) |
| 22s | Two consecutive results (or both levels in one run) beyond 2 SD on the same side | Systematic |
| R4s | Range between two results in a run (or two consecutive) > 4 SD | Random |
| 41s | Four consecutive results beyond 1 SD on the same side (across runs and levels) | Systematic |
| 10x | Ten consecutive results on the same side of the mean | Systematic (high false-alert rate) |
Shift, trend, random error
| Pattern | Looks like | Typical causes |
|---|---|---|
| Shift | Sudden jump; then 6+ results on one side of the mean | New reagent or calibrator lot, wrongly made calibrator, recalibration, new lamp, temperature or wavelength change |
| Trend | 6+ results moving steadily in one direction | Reagent or calibrator deterioration, evaporation, weakening lamp, dirty optics, slow temperature drift |
| Random error | Scatter in both directions, 13s or R4s | Bubbles, clots, pipetting error, unstable power, poor mixing, operator technique |
Other QC charts
- CUSUM — sums the deviations; earliest warning of small shifts and trends.
- Youden (Tonks–Youden) plot — level 1 against level 2 for many labs; points far along the diagonal = systematic error.
- Moving average of patient results — patient-based QC between control runs.
More about control materials
- Assayed control = the manufacturer lists expected values for each method. Unassayed control = no values given; the lab sets its own mean and SD (cheaper). Even with assayed controls, the lab should establish its own target.
- Lyophilized (freeze-dried) controls: reconstitute exactly (same volumetric pipette or device, ideally the same person) to avoid vial-to-vial variation. Respect stability after reconstitution or opening.
- Qualitative tests: run a negative and a positive control; graded tests (e.g., urine dipstick) also need a graded (weakly positive) level.
- On a Levey–Jennings chart, loss of precision looks like wider scatter of points around a normal mean.
Tietz 7th ed., Ch 6 Quality Control of the Analytical Examination Process, p. 131–142 (Table 6.3); Harr §5.4; SUCCESS! Ch.13 §II. · Tietz 7th ed., Ch 6 Quality Control of the Analytical Examination Process, p. 135–136 (gap check v-next)
4 · QC failure, troubleshooting, reagent lots, peer comparison and sigma
- All parameters fail together (e.g., blood gas analyzer) → common cause → recalibrate first.
- Reagent carryover: errors follow a particular test (e.g., triglyceride after lipase). Sample carryover: a low sample reads high after a very high sample.
- Probe obstruction: all tests from one probe read very low or undetectable.
New reagent lots
QC materials may react differently with a new reagent lot (noncommutability). So run patient samples on the old and new lots; accept the lot if patient results agree, and then update the QC target if needed (keep the SD).
Peer comparison and PT
- Same difference at both control levels = constant bias (often calibrator value).
- Difference growing with concentration = proportional bias.
Sigma metric
Point-of-care QC
Electronic QC checks only the reader (electronics, optics). It does not check the reagents in cartridges or strips — liquid QC is still needed (new lots, shipments, set intervals or an IQCP).
Tietz 7th ed., Ch 6, p. 133–158 (Fig. 6.10, Fig. 6.13); Ch 30 Point-of-Care Testing, p. 340–341; Harr §5.4 and §5.11.
5 · Method evaluation and reference intervals
| Property | Meaning | How it is checked |
|---|---|---|
| Precision (random error) | Closeness of replicate results to each other | Replicates within run and between days → SD, CV |
| Trueness / bias (systematic error) | Closeness of the mean to the true value | Reference material, comparison of methods, PT |
| Accuracy | Closeness of one result to the true value (bias + imprecision) | — |
| Recovery | % of added analyte that is measured | Detects proportional error |
| Interference | Effect of another substance | Detects constant error |
| Linearity / AMR | Range measured directly without dilution | Linearity material, dilutions |
| Analytical sensitivity | Smallest amount detected | Limit of detection |
| Analytical specificity | Measures only the intended analyte | Interference studies |
| Reliability | Keeps accuracy and precision over time | QC history |
Comparison of methods (regression y = a + bx)
- Intercept a ≠ 0 → constant systematic error.
- Slope b ≠ 1 → proportional systematic error (slope 0.90 = −10%).
- Sy/x (standard error of the estimate) → random error around the line.
- r (correlation coefficient) → association only; depends on the range; not agreement.
- Paired t-test → difference of means (bias); F-test → difference in precision (variances).
- Bias (difference) plot: y − x against x; points below zero = negative bias; widening = proportional.
Target diagram
Tight cluster away from the centre = precise, not accurate. Scatter around the centre = accurate on average, imprecise. Tight cluster in the centre = precise and accurate.
Reference intervals
- Establish: at least 120 healthy reference subjects per partition; central 95% (2.5th–97.5th percentile, nonparametric).
- Verify (transfer) a manufacturer's interval: 20 subjects; accept if ≤ 2 are outside. If more fail, test 20 more; then consider a full study.
- Reference interval ≠ decision limit (e.g., HbA1c ≥ 6.5% for diabetes).
Choosing a new instrument
Tietz 7th ed., Ch 2 Statistical Methodologies, p. 17, 30–35; Ch 9 Reference Intervals, p. 178, 186; Harr §5.4; BOC Ch.8. · Book only: SUCCESS! 4th ed., Ch 13 §II.D, p. 997; Concise Review 2nd ed., Ch 11 §VII, p. 382–383 (gap check v-next)
6 · The postanalytical phase
- Delta check: compares a result with the patient's previous result (usually within 24–72 h). A failure means: wrong patient / mislabel, IV contamination, interference, analytical error — or a real change. Look at the pattern before deciding.
- Critical (panic) values: life-threatening results → phone promptly to the responsible caregiver, read-back, document who, date and time (TJC National Patient Safety Goal).
- Autoverification: software releases results that pass validated rules; holds results with flags (QC, HIL, delta, critical).
- Corrected reports: if a wrong result was released, correct it, notify the clinician, keep the original traceable.
- Turnaround time: from order (or collection) to report — a key quality indicator.
- Report contents (CLIA): patient ID, testing laboratory name and address, test, result and units, reference interval, specimen condition comments.
- The anion gap is a handy internal check: a gap near zero or negative suggests an electrolyte error (often sodium).
Tietz 7th ed., Ch 6, p. 130 and 149–150 (Table 6.7); Ch 3, p. 60.e4; Harr §5.4 and §5.11; SUCCESS! Ch.13.
7 · Regulation, accreditation and laboratory materials (US-based items)
| Body / law | Role |
|---|---|
| CLIA '88 | US law covering every lab that tests human specimens for patient care |
| CMS | Administers CLIA: certificates, inspections, PT approval, sanctions |
| FDA | Clears tests and categorizes complexity (waived, moderate, high) |
| CDC | Scientific support for CLIA |
| CAP, TJC, COLA | Accrediting organizations with deemed status (meet or exceed CLIA) |
| CLSI | Writes consensus standards (e.g., order of draw, reference intervals) |
| HIPAA | Privacy of patient health information |
| ASCP BOC | Certifies individuals (certification ≠ accreditation ≠ licensure) |
Key CLIA requirements (nonwaived testing)
- QC: at least two control levels per day of testing for quantitative tests, or an approved IQCP (risk assessment → QC plan → quality assessment).
- Proficiency testing: three events per year, usually five challenges each; test like patients; no referral and no inter-lab communication before submission; review ungraded results too. No PT available → alternative assessment twice a year.
- Competency: six elements (direct observation, monitoring reporting, record review, maintenance observation, blind or PT samples, problem solving); at 6 and 12 months in the first year, then annually.
- Calibration verification / AMR: AMR = range measured directly without dilution.
- Procedure manual: approved written procedures; package inserts may be referenced but are not procedures.
- Waived tests: Certificate of Waiver, follow manufacturer's instructions. A modified test becomes high complexity.
Laboratory materials and equipment
- Class A volumetric glassware for standards; check reusable glassware for detergent residue.
- Class S weights check analytical balances; Class M weights calibrate weights.
- TD blow-out pipettes have two etched rings at the top; TC pipettes are rinsed out.
- CLRW (clinical laboratory reagent water) for quantitative tests, standards and buffers.
- Record refrigerator and freezer temperatures daily; document preventive maintenance.
Chemical grades and reagent water
- Analytical reagent (AR) grade = meets American Chemical Society (ACS) specifications; used for most tests. Ultrapure (spectrograde, HPLC grade, nanograde) = extra purification for AAS, chromatography, molecular work. CP, USP and NF grades and "technical/practical" grades are not used for clinical analysis without purification.
| CLRW specification (CLSI) | Limit |
|---|---|
| Resistivity at 25 °C | ≥ 10 MΩ·cm (more ions → lower resistivity) |
| Microbes | < 10 CFU/mL |
| Total organic carbon | < 500 ng/g (ppb) |
| Particles | Passed through a 0.22 µm filter |
- Distillation alone does not meet CLRW (volatile and ionic carry-over).
- Special reagent water = at least CLRW plus extra limits for one use (nuclease-free for DNA/RNA, metal-free for trace metals, sterile for culture).
- Instrument feed water = meets the analyzer maker's specification. Autoclave/wash water (old "type III") = glassware washing; final rinse with the grade the glassware will be used for. Bottled water can grow microbes; manufacturer's diluent water is only for its labelled use.
- Test water for resistivity, microbes, organics and particles at a frequency the lab sets — ideally at the point of use.
Reference materials (standards)
| Type | What it is |
|---|---|
| Primary reference material | Very pure chemical (about 99.98%) that is weighed directly to make a solution of exactly known concentration; comes with a certificate |
| Secondary reference material | Solution whose value is assigned by analysis with a reference method calibrated by a primary material |
| SRM | Standard reference material sold by NIST (USA), with a certificate of its properties |
| CRM | Certified reference material (e.g., from the European IRMM) |
Traceability = a result can be linked back to a higher reference through an unbroken, documented chain of calibrations.
Glassware, plastics and desiccants
| Material | Key property |
|---|---|
| Borosilicate (Pyrex, Kimax) | Most lab glassware; low alkali; resists heat, chemicals and thermal shock |
| Corex (alumina-silicate) | About 6× stronger than borosilicate; centrifuge tubes, thermometers |
| Vycor | Very high temperatures (to about 900 °C) |
| Flint (soda-lime) | Cheap; some disposable glassware |
| Polyolefins (polyethylene, polypropylene) | Chemically inert to most acids, alkalis, salts |
| Polycarbonate | Clear and strong (centrifuge tubes) but less chemical resistance |
| Teflon (fluorocarbon) | Inert, non-wettable, wide temperature range; stir bars, tubing |
Desiccants absorb water to keep chemicals dry in a desiccator; magnesium perchlorate is one of the strongest.
Pipettes, balances and thermometers
| Pipette | Type | Blow out? |
|---|---|---|
| Volumetric (transfer) | TD, one fixed volume of watery liquid; most accurate | No — drains by gravity, touch tip to wall |
| Ostwald–Folin (transfer) | TD, bulb near the tip; for viscous fluids (blood, serum) | Yes (etched ring) |
| Mohr (measuring) | TD, graduated between two marks | No |
| Serologic (measuring) | TD, graduated to the tip | Yes (etched ring or rings) |
| TC micropipettes (e.g., Sahli) | To contain | Rinse out with the diluent |
- Measuring pipettes are for reagents, not for samples or calibrators. Hold pipettes vertical; read the bottom of the meniscus at eye level.
- Semiautomatic pipettes: handheld, disposable tips, user aspirates and dispenses. Automatic pipettes/dilutors: motor-driven syringes.
- Checking pipette volume: gravimetric — weigh the delivered water (all at room temperature; mass ÷ density = volume); or photometric — deliver a dye and read its absorbance.
- Balances: electronic balances use an electromagnetic restoring force proportional to mass; tare the container. Verify with NIST Class S weights (at least monthly).
- Thermometers (liquid-in-glass or thermistor probe) must be checked against a NIST-certified thermometer (e.g., SRM with points at 0, 25, 30 and 37 °C) before use.
Procedure manual, records and CLIA enforcement
- Each written procedure contains: test name and principle, clinical use, patient preparation, specimen requirements and rejection criteria, reagents/calibrators/controls (preparation, storage, labelling), equipment, step-by-step method, calibration, QC and corrective action, calculations, reference intervals, critical values, limitations and interferences, reporting, references.
- Approved, signed and dated by the laboratory director before use; reviewed on a set schedule (the review books say yearly) and re-approved when changed. Keep retired versions.
- Keep QC and other quality records for at least 2 years (CLIA).
- Label every reagent, standard and control (content, concentration, storage, preparation and expiry dates). Do not mix components from different kit lots unless the manufacturer allows it.
- PT grading: a score of at least 80% per analyte is needed for most tests (blood bank ABO/D and compatibility: 100%). PT providers must be approved by CMS. Every unacceptable PT result is investigated (clerical error? method? sample?) and documented.
- PT limits: PT does not check identification, collection or transport, and it fails if PT samples get special treatment.
- CLIA inspections are usually every 2 years and may be unannounced. Sanctions: limits, suspension or revocation of the CLIA certificate, fines.
Point-of-care testing (POCT) compliance
- POCT is still laboratory testing: the same rules apply for QC, PT/EQA, documentation and competency (ISO 22870; CLIA rules follow the test complexity). Most operators are nurses or doctors, not laboratory staff, so the program needs planning and oversight.
- Training covers why the test is done, patient preparation, specimen collection, device preparation, running QC, recording and reporting results, interpretation and safety.
- Competency is proven by questions plus practical tests (QC samples, parallel testing of patient samples against the laboratory) and direct observation, then kept up by regular practice.
- QC frequency is never less than the manufacturer requires; extra QC with new lots, shipments or new operators. Blood gas analyzers can lock out users when QC fails.
Tietz 7th ed., Ch 3, p. 60.e11; Ch 6, p. 156–158; Ch 30, p. 340–341; CLIA 42 CFR 493 (current US rules); BOC Ch.8; SUCCESS! Ch.13. · Tietz 7th ed., Ch 15 Principles of Basic Techniques and Laboratory Safety, p. 199.e3–199.e4 (Table 15.4) (gap check v-next) · Tietz 7th ed., Ch 15, p. 199.e4–199.e5 (gap check v-next) · Book only: SUCCESS! 4th ed., Ch 13 §I.A–B, p. 989 (gap check v-next) · Tietz 7th ed., Ch 15, p. 199.e6–199.e9, 199.e13–199.e14 (gap check v-next) · Tietz 7th ed., Ch 6, p. 153, 156, 158–160 (Box 6.1); book only: Concise Review 2nd ed., Ch 11, p. 375–377; SUCCESS! 4th ed., Ch 13 §II.C, §II.H (gap check v-next) · Tietz 7th ed., Ch 30 Point-of-Care Testing, p. 338, 340–341 (Box 30.6) (gap check v-next)
8 · Exam traps — choose the BEST answer
| Trap | Correct thinking |
|---|---|
| 12s broken → reject | 12s is a warning; check the other rules. |
| 22s → repeat the control | Find and fix the cause, verify, rerun patients. |
| Both controls high = random | Same side = systematic; opposite sides = random (R4s). |
| Gradual change called a shift | Gradual = trend; sudden = shift. |
| Weak calibrator → low results | Weak calibrator → HIGH results. |
| High r = agreement | r shows association; look at slope and intercept. |
| Recovery detects constant error | Recovery → proportional; interference → constant. |
| Reference interval from 40 people, mean ± 2 SD | ≥ 120 people, central 95%; verification uses 20 (≤ 2 outside). |
| Delta failure = always a mislabel | Could be real change, IV fluid, diet, interference. |
| Hemolyzed sample → always reject | In vivo hemolysis is real — report with a comment. |
| Blood gas always on ice | Plastic syringe: room temperature, within 30 min. |
| FDA runs CLIA | CMS runs CLIA; FDA categorizes tests. |
| Electronic QC replaces liquid QC | It checks the reader only. |
| Run PT in triplicate / ask another lab | PT is handled exactly like a patient sample. |
Summary of Study Sections 1–7.
⚡ Last-Minute Review — Quality Assessment, Troubleshooting & Regulation
1. Phases
- Pre (most errors: ID, collection, transport, hemolysis) → analytical (QC, PT) → post (reporting, critical values, delta).
- QC = analytical; QA = whole process; PT = external, blind, like a patient.
2. Specimens
- Two identifiers (never room number); label at bedside after collection.
- Order: culture → citrate → serum → heparin → EDTA → fluoride. EDTA carryover: K ↑ Ca ↓.
- Fist pumping / long tourniquet: K ↑. Meal: TG ↑, glucose ↑, phosphate ↑. IV line: glucose/Na/Cl ↑, K/CO2 ↓.
- Delay: glucose ↓ ~10 mg/dL/h; K, phosphate, LD, AST ↑; CO2 ↓. Stable: urea, creatinine, uric acid.
- In vivo hemolysis: haptoglobin ↓, K normal → report with comment. Lipemia → ultracentrifuge.
- Blood gas: no clot, no air; air bubble → pO2 ↑ pCO2 ↓ pH ↑; delay → pO2 ↓ pCO2 ↑ pH ↓; plastic syringe RT ≤ 30 min.
3. QC
- ±1/2/3 SD = 68/95/99.7%. Outside 2 SD = 1 in 20; above +2 SD = 1 in 40.
- 12s warning · 13s, R4s = random · 22s, 41s, 10x = systematic.
- Shift = sudden (new lot, calibrator, lamp) · trend = gradual (deterioration). Weak calibrator → results HIGH.
- Fail → stop, new vial of control, find and fix (recalibrate), rerun controls, rerun patients, document. Never "repeat until it passes".
- New control lot: own mean (≥ 10 days), keep the old SD. New reagent lot: compare PATIENT samples.
- CUSUM = earliest small shift/trend · Youden = labs compared · SDI = (lab − peer mean)/peer SD.
- Sigma = (TEa − bias)/CV. Bias % = (mean − target)/target × 100.
4. Method evaluation
- Precision = replicates (SD, CV) · accuracy = closeness to truth · recovery → proportional · interference → constant.
- Regression: intercept = constant error; slope = proportional; Sy/x = random; r = association only.
- t-test = means (bias) · F-test = variances (precision).
- Reference interval: ≥ 120 subjects, central 95%; verify with 20, ≤ 2 outside.
5. Post-analytical
- Delta check → mislabel, IV, interference, error or real change. Critical value → call, read-back, document.
- Corrected report for released errors. TAT = key indicator. Report must name the testing lab.
6. Regulation (US)
- CLIA: all patient-testing labs · CMS runs it · FDA categorizes complexity · CAP/TJC/COLA deemed accreditors.
- QC minimum 2 levels/day or IQCP (risk assessment → QC plan → QA).
- PT: 3 events × 5 challenges; no referral, no sharing. No PT → alternative assessment twice a year.
- Competency: 6 elements; at 6 and 12 months in year 1, then yearly. HIPAA = privacy. ICD = diagnosis; CPT = procedure.
- Class A glassware · Class S weights · CLRW water · fridge temperatures daily · detergent check on glassware.