C(ASCPi) CHEMISTRY · LABORATORY OPERATIONS · MODULE 19

Quality Assessment, Troubleshooting & Regulation

The quality system and the three phases of testing · Preanalytical errors (identification, order of draw, tube additives, delay, hemolysis, icterus, lipemia, blood gas specimens) · Quality control, Levey–Jennings charts and Westgard rules · QC troubleshooting, reagent lots, peer comparison and sigma · Method evaluation and reference intervals · Postanalytical phase (delta checks, critical values) · US regulation and accreditation (CLIA, PT, competency, IQCP)

Simple English · Question sources: BOC · Harr · SUCCESS! · CLSR (adapted) + original questions · Answers checked against Tietz Textbook of Laboratory Medicine, 7th ed. (US regulatory items checked against current CLIA rules) · Works offline · questions · Version 1.2

Independent exam-preparation material. Not affiliated with or endorsed by ASCP or the ASCP Board of Certification.

📘 Study progress

Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 very high-yield · 📌 must memorize · ⚠️ exam trap · 🧠 memory trick · 🔬 lab connection · 💡 understand this.

START HERE

1 · The quality system and the three phases of testing

Simple idea: a laboratory result is only useful if the right patient, the right specimen, the right analysis and the right report all come together. Every step can fail, so the laboratory controls every step.

Test ordered → patient prepared → patient identified → specimen collected → transported → processed (PREANALYTICAL — most errors happen here) ↓ Specimen analysed with calibration + QC + maintenance + trained staff (ANALYTICAL — QC, PT, method validation) ↓ Result checked → reported → critical values called → interpreted (POSTANALYTICAL — reporting and interpretation errors)
TermMeaningExample
Quality control (QC)Statistical checks of the analytical process with control materialsTwo controls per run plotted on a Levey–Jennings chart
Quality assurance / assessment (QA)Monitoring of the whole process (pre, analytical, post)Turnaround time, specimen rejection rate, critical-value calls
Quality management system (QMS)Organisation, documents, training, audits and improvement for the whole laboratoryISO 15189, CAP checklists, CLIA rules
External QC = proficiency testing (PT / EQA)Unknown samples from an outside provider, graded against other labsCAP surveys three times a year
🔥 VERY HIGH-YIELD — Most laboratory errors are preanalytical (identification, collection, transport, hemolysis). QC on the analyzer cannot detect them.

Quality improvement tools

  • PDCA — Plan, Do, Check, Act: the basic improvement cycle.
  • Six Sigma (DMAIC) — Define, Measure, Analyze, Improve, Control; reduces variation and defects.
  • Lean — removes waste (waiting, extra movement, rework) → shorter turnaround time.
  • Root cause analysis (RCA) — after a serious error, ask "why?" until the system cause is found.
  • First step in any problem (for example a turnaround-time complaint): collect data (measure) before changing anything.
🧠 MEMORY TRICK: "Lean = less waste; Six Sigma = less variation."

Total quality management (TQM) and ISO standards

  • TQM = a management approach that improves every factor that affects testing, not only the analyzer. Its five parts (the "5 Q" model): quality planning, the quality laboratory process, quality control, quality assessment and quality improvement.
  • ISO 9000 / ISO 9001 = general international quality-management standards for any organisation. ISO 15189 = the version written for medical laboratories (quality and competence). ISO 22870 = point-of-care testing.
  • ISO meaning of quality: the result is fit for its intended use (it meets the stated requirement).
🧠 MEMORY TRICK — TQM wheel: Plan → Process → Control → Assess → Improve, then round again.

Tietz 7th ed., Ch 3 Governance, Risk and Quality Management, p. 60.e1–60.e16; Ch 5, p. 89; SUCCESS! Ch.13 §II; BOC Ch.8 Quality Assessment. · Tietz 7th ed., Ch 3 Governance, Risk and Quality Management, p. 60.e1, 60.e5–60.e6 (gap check v-next)

PREANALYTICAL

2 · The preanalytical phase and specimen errors

Patient identification

  • Use two patient identifiers (for example full name + date of birth or medical record number). A room or bed number is never an identifier.
  • Label tubes at the bedside, after collection, in front of the patient. Never pre-label.

Patient preparation and collection

FactorEffect
Recent mealTriglycerides ↑↑, glucose ↑, phosphate ↑ (ALP may rise after fat). Cholesterol hardly changes.
Prolonged tourniquet (> 1 min)Hemoconcentration: proteins, protein-bound substances (calcium, lipids, enzymes) ↑; K ↑
Fist pumpingPotassium ↑ (can be > 1 mmol/L), phosphate and lactate ↑
Drawing above an IV lineDilution: analytes in the fluid ↑ (glucose, Na, Cl), others ↓ (K, CO2, Ca, proteins)
Wrong order of draw (EDTA carryover)K ↑, Ca ↓, Mg ↓, ALP ↓
ORDER OF DRAW (CLSI) blood culture → light blue (citrate) → red / gold (serum, gel) → green (heparin) → lavender (EDTA) → gray (fluoride/oxalate)
🧠 MEMORY TRICK: "Can Come Clean, Heparin Eats Fruit" — Culture, Citrate, Clot, Heparin, EDTA, Fluoride.

Tube additives

TopAdditiveChemistry use / warning
Red / gold (SST)None or clot activator ± gelMost chemistry; gel may absorb some drugs
Green / light greenLithium or sodium heparin (± gel)Plasma chemistry, blood gases; not for lithium if lithium heparin
LavenderK2/K3 EDTAHbA1c, CBC; not for K, Ca, Mg, ALP
GraySodium fluoride + potassium oxalateGlucose, lactate, alcohol; fluoride inhibits glycolysis (enolase) and some enzymes
Light blueSodium citrateCoagulation

Handling and delay

  • Unseparated blood keeps metabolising: glucose falls about 10 mg/dL per hour (about 5–7%) at room temperature; potassium, phosphate, LD and AST leak out of cells; CO2 is lost. Separate serum within about 2 hours.
  • Refrigerating unseparated blood stops the Na/K pump → potassium rises even without visible hemolysis.
  • Bilirubin is destroyed by light — protect from light.
  • Stable after delay: urea, creatinine, uric acid.

Hemolysis, icterus and lipemia (HIL)

ProblemMain effectsWhat to do
Hemolysis (commonest preanalytical error)K, LD, AST, Mg, phosphate, iron ↑; spectral interferenceIn vitro: recollect (or report unaffected tests with comment). In vivo: report with comment.
IcterusBilirubin absorbs light and reacts in peroxidase methods (creatinine, cholesterol, uric acid)Check method limits; dilution or other method
LipemiaTurbidity (light scatter); volume displacement (pseudohyponatremia by indirect ISE); non-homogeneous sampleUltracentrifugation (CLSI recommended); never for lipid tests
⚠️ EXAM TRAP: In in vivo hemolysis haptoglobin is low and potassium is usually normal. In in vitro hemolysis haptoglobin is normal and potassium is high.
🔥 Very high K + very low Ca = EDTA contamination. Very high glucose + high Na/Cl + low K/CO2 = IV-fluid contamination. Lithium 14 mmol/L = lithium-heparin tube.

Blood gas specimens

  • Heparinized syringe, no clots, no air bubbles, mixed gently.
  • Air bubble: pO2 ↑, pCO2 ↓, pH ↑. Delay (cell metabolism): pO2 ↓, pCO2 ↑, pH ↓.
  • CLSI: plastic syringe at room temperature, analyse within 30 min; ice slurry only for glass syringes when delay is expected. Excess liquid heparin dilutes the sample.
🔬 LAB CONNECTION: whole-blood glucose is about 11–15% lower than plasma glucose because red cells hold less water; the difference depends on the hematocrit.

Biological variation and patient preparation

FactorWhat happensExam example
Circadian / diurnal (24-hour rhythm, linked to sleep–wake)Some analytes peak at a fixed time of dayCortisol highest in the early morning, lowest late evening; potassium lower in the afternoon. Draw at the time used for the reference interval (usually 7–9 am).
Seasonal (circannual) and monthly cyclesRepeating changes over the year or the menstrual cycle25-OH vitamin D higher in summer; total T3 about 20% lower in summer; reproductive hormones follow the menstrual cycle
PostureStanding up moves plasma water into tissues → large molecules and cells concentrateSupine → upright: proteins, cells, protein-bound substances ↑; total calcium ↑ 5–10%, free calcium unchanged. Rest at least 15 min before drawing.
ExerciseMuscle releases contents; mostly transient (back to baseline in about 3 h)K ↑ (even fist clenching), CK, LD, ALT, troponin ↑, WBC ↑; regular heavy training keeps CK/LD high for longer
Stress (physical or mental)Stress hormones are releasedCortisol ↑ (and total cholesterol ↑); other pituitary hormones and aldosterone may change
Standard preparation for routine blood tests (EFLM) Draw in the morning (7–9 am) → fast 12 h (water allowed) → no alcohol for 24 h → no smoking or caffeine that morning
⚠️ EXAM TRAP — "Circannual" means a yearly (seasonal) cycle, not "twice a year". Posture and exercise are biological (real in-vivo) changes, not analytical errors — QC will be normal.

Venipuncture technique — key facts

  • Identification: inpatient → compare the ID band with the requisition (or scan it). No band → do not draw until the patient's nurse or physician identifies the patient; record that person's name. Outpatient → patient states two identifiers (e.g., name + date of birth).
  • Site: median cubital vein is preferred (large, near the surface); cephalic next; basilic last (close to nerve and brachial artery). Avoid an arm with an IV line, hematoma, fistula or on the side of a mastectomy.
  • Needle gauge: the bigger the number, the smaller the bore. Adult usual = 20–21 G; small or fragile veins / children = 23 G, often with a butterfly set.
  • Needle enters bevel up at a low angle (about 15–30°). Release the tourniquet as soon as blood flows (within 1 min).
  • Draw the smallest volume needed — repeated draws cause iatrogenic anemia (about 7 g/L hemoglobin lost per 100 mL drawn).
  • Invert additive tubes gently (CLSI: citrate 3–4 times, serum tubes 5, heparin/EDTA/fluoride 8). Never shake.
  • Activate the needle safety device and discard into a sharps container; never recap. Label at the bedside: patient name, ID number, date and time, collector's initials.

How the anticoagulants work

AdditiveMechanism and amountExam point
EDTA (lavender)Chelates (binds) Ca²⁺ and Mg²⁺; about 1.5 mg per mL blood (1–2 g/L)Also inhibits ALP and CK (cofactor removal); too much EDTA (short draw) shrinks RBCs
Heparin (green)Speeds up antithrombin → thrombin is neutralised; about 0.2 mg (≈18 units) per mL bloodAmmonium heparin → false high ammonia; lithium heparin → false lithium
Sodium citrate 3.2% (light blue)Binds calcium; 9 parts blood : 1 part citrateFill to the mark (±10%). Underfilled tube → too much citrate → falsely long PT/aPTT
Gel separator (SST/PST)Thixotropic gel moves between cells and serum/plasma during centrifugationNot for blood bank; can absorb some drugs
⚠️ EXAM TRAP — With a butterfly set, if the citrate tube is the first tube, draw a discard tube first. The tubing air would otherwise underfill the citrate tube.

Transport, storage and chain of custody

  • Ice and speed: ammonia (analyse within about 30 min) and lactate, pyruvate, ACTH, renin, gastrin → send on ice. Cells keep making ammonia and lactate at room temperature.
  • Keep the cap on during centrifugation: prevents loss of CO₂ (pH ↑, ionized Ca changes) and of volatiles (ethanol ↓), reduces evaporation and aerosols.
  • If a whole-blood sample cannot be spun within 2 h, keep it at room temperature, not in the fridge.
  • Separated serum/plasma: refrigerate (2–8 °C) for short storage (about 2–3 days for many analytes); freeze at −20 °C or colder for long storage. Avoid frost-free freezers (temperature swings).
  • Chain of custody (forensic drug or alcohol testing): sealed container, signatures and a document that follows the specimen through every hand-over, so it can be defended in court.

Tietz 7th ed., Ch 4 Specimen Collection and Processing, p. 64–70; Ch 5 Preanalytical Variation, p. 81–94 and p. 116–119; Harr §5.11; BOC Ch.3 Quality Assessment. · Tietz 7th ed., Ch 5 Preanalytical Variation, p. 81–82, 87–88 (gap check v-next) · Tietz 7th ed., Ch 4 Specimen Collection and Processing, p. 61, 63–66 (Table 4.2) (gap check v-next) · Tietz 7th ed., Ch 4, p. 65 (Table 4.2), 69–70; Ch 5, p. 120 (gap check v-next) · Tietz 7th ed., Ch 4, p. 72, 76–77 (gap check v-next)

🔥 QUALITY CONTROL

3 · Quality control: materials, Levey–Jennings charts and Westgard rules

Simple idea: a control is a sample with a known expected value. If the control result is where it should be, the analyzer is probably working and patient results can be reported.

Control materials

  • At least two levels (normal and abnormal), near medical decision points; matrix like patient samples; stable (one lot for about a year); go through the same pretreatment as samples.
  • Calibrator = sets the signal–concentration relationship. Control = checks it. Never calibrate with a control.
  • New control lot: target from the lab's own results over at least about 10 days; keep the established long-term SD. Package-insert and peer-group SDs are too wide.
  • CLIA minimum for quantitative tests: two levels every day of testing (or an approved IQCP). Many labs run controls every shift.

Levey–Jennings chart

+3 SD ─────────── reject (13s) +2 SD ─────────── warning (12s) +1 SD ··········· Mean ─────────── results scatter on both sides −1 SD ··········· −2 SD ─────────── −3 SD ─────────── Gaussian: ±1 SD = 68% · ±2 SD = 95% (1 in 20 outside; 1 in 40 above) · ±3 SD = 99.7%

Westgard multirules

RuleMeaningError type
12sOne result beyond ±2 SDWARNING only (5% false alarms)
13sOne result beyond ±3 SDRandom (or large systematic)
22sTwo consecutive results (or both levels in one run) beyond 2 SD on the same sideSystematic
R4sRange between two results in a run (or two consecutive) > 4 SDRandom
41sFour consecutive results beyond 1 SD on the same side (across runs and levels)Systematic
10xTen consecutive results on the same side of the meanSystematic (high false-alert rate)
🧠 MEMORY TRICK: "R for Random" — 13s and R4s catch random error. The "stacked" rules 22s, 41s, 10x catch systematic error.

Shift, trend, random error

PatternLooks likeTypical causes
ShiftSudden jump; then 6+ results on one side of the meanNew reagent or calibrator lot, wrongly made calibrator, recalibration, new lamp, temperature or wavelength change
Trend6+ results moving steadily in one directionReagent or calibrator deterioration, evaporation, weakening lamp, dirty optics, slow temperature drift
Random errorScatter in both directions, 13s or R4sBubbles, clots, pipetting error, unstable power, poor mixing, operator technique
⚠️ EXAM TRAP: a calibrator made too weak makes patient and control results read too high (upward shift) — the analyzer treats a low signal as the full labelled value.

Other QC charts

  • CUSUM — sums the deviations; earliest warning of small shifts and trends.
  • Youden (Tonks–Youden) plot — level 1 against level 2 for many labs; points far along the diagonal = systematic error.
  • Moving average of patient results — patient-based QC between control runs.

More about control materials

  • Assayed control = the manufacturer lists expected values for each method. Unassayed control = no values given; the lab sets its own mean and SD (cheaper). Even with assayed controls, the lab should establish its own target.
  • Lyophilized (freeze-dried) controls: reconstitute exactly (same volumetric pipette or device, ideally the same person) to avoid vial-to-vial variation. Respect stability after reconstitution or opening.
  • Qualitative tests: run a negative and a positive control; graded tests (e.g., urine dipstick) also need a graded (weakly positive) level.
  • On a Levey–Jennings chart, loss of precision looks like wider scatter of points around a normal mean.

Tietz 7th ed., Ch 6 Quality Control of the Analytical Examination Process, p. 131–142 (Table 6.3); Harr §5.4; SUCCESS! Ch.13 §II. · Tietz 7th ed., Ch 6 Quality Control of the Analytical Examination Process, p. 135–136 (gap check v-next)

TROUBLESHOOTING

4 · QC failure, troubleshooting, reagent lots, peer comparison and sigma

QC rule broken (13s, 22s, R4s, 41s, 10x) ↓ 1. STOP reporting patient results ↓ 2. Test a NEW vial of control (rules out deteriorated control material) ↓ 3. Look for the cause: reagents, calibration, maintenance, temperature, probe, lamp ↓ 4. Correct it (often recalibrate) → rerun BOTH controls ↓ 5. Repeat patient samples back to the last acceptable QC; issue corrected reports ↓ 6. Document everything
⚠️ EXAM TRAP: Do not simply repeat the same control until it "passes". With a real error the repeat may fall inside the limits by chance while the error continues.
  • All parameters fail together (e.g., blood gas analyzer) → common cause → recalibrate first.
  • Reagent carryover: errors follow a particular test (e.g., triglyceride after lipase). Sample carryover: a low sample reads high after a very high sample.
  • Probe obstruction: all tests from one probe read very low or undetectable.

New reagent lots

QC materials may react differently with a new reagent lot (noncommutability). So run patient samples on the old and new lots; accept the lot if patient results agree, and then update the QC target if needed (keep the SD).

Peer comparison and PT

SDI (standard deviation index) = (lab result or mean − peer mean) ÷ peer SD |SDI| ≤ 1 good · 1–2 acceptable · > 2 investigate (bias)
  • Same difference at both control levels = constant bias (often calibrator value).
  • Difference growing with concentration = proportional bias.

Sigma metric

Sigma = (TEa − |bias|) ÷ CV (all in %) Example: TEa 10%, bias 2%, CV 2% → (10 − 2) ÷ 2 = 4 sigma 6 sigma ≈ 3.4 defects per million (world class, simple QC) · 3 sigma = marginal (strict QC, still risky)
🔥 VERY HIGH-YIELD — Required calculations: mean, SD, CV; Westgard limits (mean ± 1, 2, 3 SD); bias % = (mean − target) ÷ target × 100.

Point-of-care QC

Electronic QC checks only the reader (electronics, optics). It does not check the reagents in cartridges or strips — liquid QC is still needed (new lots, shipments, set intervals or an IQCP).

Tietz 7th ed., Ch 6, p. 133–158 (Fig. 6.10, Fig. 6.13); Ch 30 Point-of-Care Testing, p. 340–341; Harr §5.4 and §5.11.

METHOD EVALUATION

5 · Method evaluation and reference intervals

PropertyMeaningHow it is checked
Precision (random error)Closeness of replicate results to each otherReplicates within run and between days → SD, CV
Trueness / bias (systematic error)Closeness of the mean to the true valueReference material, comparison of methods, PT
AccuracyCloseness of one result to the true value (bias + imprecision)—
Recovery% of added analyte that is measuredDetects proportional error
InterferenceEffect of another substanceDetects constant error
Linearity / AMRRange measured directly without dilutionLinearity material, dilutions
Analytical sensitivitySmallest amount detectedLimit of detection
Analytical specificityMeasures only the intended analyteInterference studies
ReliabilityKeeps accuracy and precision over timeQC history
Recovery % = (spiked result − baseline result) ÷ amount added (after dilution) × 100 Example: 0.9 mL serum + 0.1 mL of 800 mEq/L standard: added = 80 mEq/L; found 161 − 89 = 72 → 90%

Comparison of methods (regression y = a + bx)

  • Intercept a ≠ 0 → constant systematic error.
  • Slope b ≠ 1 → proportional systematic error (slope 0.90 = −10%).
  • Sy/x (standard error of the estimate) → random error around the line.
  • r (correlation coefficient) → association only; depends on the range; not agreement.
  • Paired t-test → difference of means (bias); F-test → difference in precision (variances).
  • Bias (difference) plot: y − x against x; points below zero = negative bias; widening = proportional.
⚠️ EXAM TRAP: "r = 0.99 so the methods agree" is WRONG. A method can correlate perfectly and still read 2 units too high.

Target diagram

Tight cluster away from the centre = precise, not accurate. Scatter around the centre = accurate on average, imprecise. Tight cluster in the centre = precise and accurate.

Reference intervals

  • Establish: at least 120 healthy reference subjects per partition; central 95% (2.5th–97.5th percentile, nonparametric).
  • Verify (transfer) a manufacturer's interval: 20 subjects; accept if ≤ 2 are outside. If more fail, test 20 more; then consider a full study.
  • Reference interval ≠ decision limit (e.g., HbA1c ≥ 6.5% for diabetes).

Choosing a new instrument

1. NEEDS: test menu (now and future) · workload and throughput (STAT turnaround) · batch or random access · LIS connectivity · FDA-cleared tests 2. PRACTICAL: space, power, water and drain needs · heat and waste · downtime and service support · ease of troubleshooting · training · safety features 3. COST: instrument + reagents + service contract + interface ↓ 4. EVALUATE in your lab: precision · accuracy (method comparison) · linearity/AMR · sensitivity and specificity · carryover · calibration stability
🔥 VERY HIGH-YIELD — Selection first asks "what do we need?"; evaluation then proves the analyzer meets precision, accuracy and linearity before patient use.

Tietz 7th ed., Ch 2 Statistical Methodologies, p. 17, 30–35; Ch 9 Reference Intervals, p. 178, 186; Harr §5.4; BOC Ch.8. · Book only: SUCCESS! 4th ed., Ch 13 §II.D, p. 997; Concise Review 2nd ed., Ch 11 §VII, p. 382–383 (gap check v-next)

POSTANALYTICAL

6 · The postanalytical phase

  • Delta check: compares a result with the patient's previous result (usually within 24–72 h). A failure means: wrong patient / mislabel, IV contamination, interference, analytical error — or a real change. Look at the pattern before deciding.
  • Critical (panic) values: life-threatening results → phone promptly to the responsible caregiver, read-back, document who, date and time (TJC National Patient Safety Goal).
  • Autoverification: software releases results that pass validated rules; holds results with flags (QC, HIL, delta, critical).
  • Corrected reports: if a wrong result was released, correct it, notify the clinician, keep the original traceable.
  • Turnaround time: from order (or collection) to report — a key quality indicator.
  • Report contents (CLIA): patient ID, testing laboratory name and address, test, result and units, reference interval, specimen condition comments.
  • The anion gap is a handy internal check: a gap near zero or negative suggests an electrolyte error (often sodium).
🧠 MEMORY TRICK: "Delta = Did the patient change, or did the specimen?"

Tietz 7th ed., Ch 6, p. 130 and 149–150 (Table 6.7); Ch 3, p. 60.e4; Harr §5.4 and §5.11; SUCCESS! Ch.13.

US REGULATION

7 · Regulation, accreditation and laboratory materials (US-based items)

Body / lawRole
CLIA '88US law covering every lab that tests human specimens for patient care
CMSAdministers CLIA: certificates, inspections, PT approval, sanctions
FDAClears tests and categorizes complexity (waived, moderate, high)
CDCScientific support for CLIA
CAP, TJC, COLAAccrediting organizations with deemed status (meet or exceed CLIA)
CLSIWrites consensus standards (e.g., order of draw, reference intervals)
HIPAAPrivacy of patient health information
ASCP BOCCertifies individuals (certification ≠ accreditation ≠ licensure)

Key CLIA requirements (nonwaived testing)

  • QC: at least two control levels per day of testing for quantitative tests, or an approved IQCP (risk assessment → QC plan → quality assessment).
  • Proficiency testing: three events per year, usually five challenges each; test like patients; no referral and no inter-lab communication before submission; review ungraded results too. No PT available → alternative assessment twice a year.
  • Competency: six elements (direct observation, monitoring reporting, record review, maintenance observation, blind or PT samples, problem solving); at 6 and 12 months in the first year, then annually.
  • Calibration verification / AMR: AMR = range measured directly without dilution.
  • Procedure manual: approved written procedures; package inserts may be referenced but are not procedures.
  • Waived tests: Certificate of Waiver, follow manufacturer's instructions. A modified test becomes high complexity.

Laboratory materials and equipment

  • Class A volumetric glassware for standards; check reusable glassware for detergent residue.
  • Class S weights check analytical balances; Class M weights calibrate weights.
  • TD blow-out pipettes have two etched rings at the top; TC pipettes are rinsed out.
  • CLRW (clinical laboratory reagent water) for quantitative tests, standards and buffers.
  • Record refrigerator and freezer temperatures daily; document preventive maintenance.
⚠️ EXAM TRAP: FDA categorizes complexity; CMS runs CLIA. Accreditation = an institution; certification = a person; licensure = state permission.

Chemical grades and reagent water

  • Analytical reagent (AR) grade = meets American Chemical Society (ACS) specifications; used for most tests. Ultrapure (spectrograde, HPLC grade, nanograde) = extra purification for AAS, chromatography, molecular work. CP, USP and NF grades and "technical/practical" grades are not used for clinical analysis without purification.
CLRW specification (CLSI)Limit
Resistivity at 25 °C≥ 10 MΩ·cm (more ions → lower resistivity)
Microbes< 10 CFU/mL
Total organic carbon< 500 ng/g (ppb)
ParticlesPassed through a 0.22 µm filter
Typical purification train Prefilter (removes most particles) → activated carbon (organics, chlorine) → reverse osmosis (95–99% of organics and bacteria) → deionization by ion-exchange resins (mixed beds in series → > 10 MΩ·cm) → UV light 254 nm (kills bacteria, breaks organics) → 0.22 µm submicron filter
  • Distillation alone does not meet CLRW (volatile and ionic carry-over).
  • Special reagent water = at least CLRW plus extra limits for one use (nuclease-free for DNA/RNA, metal-free for trace metals, sterile for culture).
  • Instrument feed water = meets the analyzer maker's specification. Autoclave/wash water (old "type III") = glassware washing; final rinse with the grade the glassware will be used for. Bottled water can grow microbes; manufacturer's diluent water is only for its labelled use.
  • Test water for resistivity, microbes, organics and particles at a frequency the lab sets — ideally at the point of use.
⚠️ EXAM TRAP — "Deionized" and "distilled" describe the method; CLRW describes the quality. A falling resistivity reading means the deionizer resin is exhausted.

Reference materials (standards)

TypeWhat it is
Primary reference materialVery pure chemical (about 99.98%) that is weighed directly to make a solution of exactly known concentration; comes with a certificate
Secondary reference materialSolution whose value is assigned by analysis with a reference method calibrated by a primary material
SRMStandard reference material sold by NIST (USA), with a certificate of its properties
CRMCertified reference material (e.g., from the European IRMM)

Traceability = a result can be linked back to a higher reference through an unbroken, documented chain of calibrations.

Glassware, plastics and desiccants

MaterialKey property
Borosilicate (Pyrex, Kimax)Most lab glassware; low alkali; resists heat, chemicals and thermal shock
Corex (alumina-silicate)About 6× stronger than borosilicate; centrifuge tubes, thermometers
VycorVery high temperatures (to about 900 °C)
Flint (soda-lime)Cheap; some disposable glassware
Polyolefins (polyethylene, polypropylene)Chemically inert to most acids, alkalis, salts
PolycarbonateClear and strong (centrifuge tubes) but less chemical resistance
Teflon (fluorocarbon)Inert, non-wettable, wide temperature range; stir bars, tubing

Desiccants absorb water to keep chemicals dry in a desiccator; magnesium perchlorate is one of the strongest.

Pipettes, balances and thermometers

PipetteTypeBlow out?
Volumetric (transfer)TD, one fixed volume of watery liquid; most accurateNo — drains by gravity, touch tip to wall
Ostwald–Folin (transfer)TD, bulb near the tip; for viscous fluids (blood, serum)Yes (etched ring)
Mohr (measuring)TD, graduated between two marksNo
Serologic (measuring)TD, graduated to the tipYes (etched ring or rings)
TC micropipettes (e.g., Sahli)To containRinse out with the diluent
  • Measuring pipettes are for reagents, not for samples or calibrators. Hold pipettes vertical; read the bottom of the meniscus at eye level.
  • Semiautomatic pipettes: handheld, disposable tips, user aspirates and dispenses. Automatic pipettes/dilutors: motor-driven syringes.
  • Checking pipette volume: gravimetric — weigh the delivered water (all at room temperature; mass ÷ density = volume); or photometric — deliver a dye and read its absorbance.
  • Balances: electronic balances use an electromagnetic restoring force proportional to mass; tare the container. Verify with NIST Class S weights (at least monthly).
  • Thermometers (liquid-in-glass or thermistor probe) must be checked against a NIST-certified thermometer (e.g., SRM with points at 0, 25, 30 and 37 °C) before use.

Procedure manual, records and CLIA enforcement

  • Each written procedure contains: test name and principle, clinical use, patient preparation, specimen requirements and rejection criteria, reagents/calibrators/controls (preparation, storage, labelling), equipment, step-by-step method, calibration, QC and corrective action, calculations, reference intervals, critical values, limitations and interferences, reporting, references.
  • Approved, signed and dated by the laboratory director before use; reviewed on a set schedule (the review books say yearly) and re-approved when changed. Keep retired versions.
  • Keep QC and other quality records for at least 2 years (CLIA).
  • Label every reagent, standard and control (content, concentration, storage, preparation and expiry dates). Do not mix components from different kit lots unless the manufacturer allows it.
  • PT grading: a score of at least 80% per analyte is needed for most tests (blood bank ABO/D and compatibility: 100%). PT providers must be approved by CMS. Every unacceptable PT result is investigated (clerical error? method? sample?) and documented.
  • PT limits: PT does not check identification, collection or transport, and it fails if PT samples get special treatment.
  • CLIA inspections are usually every 2 years and may be unannounced. Sanctions: limits, suspension or revocation of the CLIA certificate, fines.

Point-of-care testing (POCT) compliance

  • POCT is still laboratory testing: the same rules apply for QC, PT/EQA, documentation and competency (ISO 22870; CLIA rules follow the test complexity). Most operators are nurses or doctors, not laboratory staff, so the program needs planning and oversight.
  • Training covers why the test is done, patient preparation, specimen collection, device preparation, running QC, recording and reporting results, interpretation and safety.
  • Competency is proven by questions plus practical tests (QC samples, parallel testing of patient samples against the laboratory) and direct observation, then kept up by regular practice.
  • QC frequency is never less than the manufacturer requires; extra QC with new lots, shipments or new operators. Blood gas analyzers can lock out users when QC fails.
🔬 LAB CONNECTION — Connectivity (device → LIS) records the operator ID, QC status and result, and can block untrained operators.

Tietz 7th ed., Ch 3, p. 60.e11; Ch 6, p. 156–158; Ch 30, p. 340–341; CLIA 42 CFR 493 (current US rules); BOC Ch.8; SUCCESS! Ch.13. · Tietz 7th ed., Ch 15 Principles of Basic Techniques and Laboratory Safety, p. 199.e3–199.e4 (Table 15.4) (gap check v-next) · Tietz 7th ed., Ch 15, p. 199.e4–199.e5 (gap check v-next) · Book only: SUCCESS! 4th ed., Ch 13 §I.A–B, p. 989 (gap check v-next) · Tietz 7th ed., Ch 15, p. 199.e6–199.e9, 199.e13–199.e14 (gap check v-next) · Tietz 7th ed., Ch 6, p. 153, 156, 158–160 (Box 6.1); book only: Concise Review 2nd ed., Ch 11, p. 375–377; SUCCESS! 4th ed., Ch 13 §II.C, §II.H (gap check v-next) · Tietz 7th ed., Ch 30 Point-of-Care Testing, p. 338, 340–341 (Box 30.6) (gap check v-next)

⚠️ PUT IT TOGETHER

8 · Exam traps — choose the BEST answer

TrapCorrect thinking
12s broken → reject12s is a warning; check the other rules.
22s → repeat the controlFind and fix the cause, verify, rerun patients.
Both controls high = randomSame side = systematic; opposite sides = random (R4s).
Gradual change called a shiftGradual = trend; sudden = shift.
Weak calibrator → low resultsWeak calibrator → HIGH results.
High r = agreementr shows association; look at slope and intercept.
Recovery detects constant errorRecovery → proportional; interference → constant.
Reference interval from 40 people, mean ± 2 SD≥ 120 people, central 95%; verification uses 20 (≤ 2 outside).
Delta failure = always a mislabelCould be real change, IV fluid, diet, interference.
Hemolyzed sample → always rejectIn vivo hemolysis is real — report with a comment.
Blood gas always on icePlastic syringe: room temperature, within 30 min.
FDA runs CLIACMS runs CLIA; FDA categorizes tests.
Electronic QC replaces liquid QCIt checks the reader only.
Run PT in triplicate / ask another labPT is handled exactly like a patient sample.
🧠 MEMORY TRICK: In every QC question ask two things — Random or systematic? and Report, or stop and fix?

Summary of Study Sections 1–7.