MLS(ASCPi) STUDY TRACK · LABORATORY OPERATIONS · TOPIC 04

Electrochemistry, osmometry, electrophoresis, chromatography*, MS & flow cytometry

ISE & electrochemistry · Osmometry · Electrophoresis · Chromatography & mass spec · Flow cytometry

Simple English · organised by the ASCP BOC MLS content outline (revised 9 June 2026) · BOC6e → Harr → SUCCESS! → Other References → Original · Works offline · questions · Version 1.0

Independent study material — not affiliated with or endorsed by ASCP or the ASCP Board of Certification. Practice questions are not ASCP examination questions.

📘 Study progress

Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 very high-yield · 📌 must memorize · ⚠️ exam trap · 🧠 memory trick · 🔬 lab connection · 💡 understand this.

ASCP EXAM REFERENCE

📏 ASCP exam reference box

The ASCP BOC gives these composite reference ranges in the MLS content guideline (revised 9 June 2026). They are not shown during the exam — learn them. Exam values may be given in conventional or SI units. Where a study card below quotes a different textbook range, use this box for the ASCP exam.

🔬 ASCP questions follow US practice: regulatory and procedural questions use US sources (for example CLIA, CLSI, FDA, AABB). One best answer from four options.

29 more question(s) from the source lesson are not shown yet: they are still being verified or reworded and will be added when they pass.

START HERE

1 · Ion-Selective Electrodes (ISE) and electrochemistry

Potentiometry = measuring a voltage (potential) between two electrodes when no current flows. The voltage tells us the activity of one ion. An Ion-Selective Electrode (ISE) is a potentiometric electrode with a membrane that "prefers" one ion.

Sample touches the ion-selective membrane ↓ The chosen ion (e.g., Na⁺) moves into / binds the membrane ↓ A potential (voltage) develops across the membrane ↓ Compared with a REFERENCE electrode (constant voltage) ↓ Voltage difference ∝ log (ion activity) → Nernst equation

The parts of every ISE system

PartJobExamples
Indicator (measuring) electrodeVoltage changes with the ionNa⁺ = special glass (aluminum silicate) · pH = glass · K⁺ = valinomycin membrane · Ca²⁺, Li⁺ = organic ion-exchanger (ionophore) · NH₄⁺ = nonactin/monactin
Reference electrodeGives a CONSTANT voltageSilver/silver chloride (Ag/AgCl) · Calomel (mercury + mercurous chloride + saturated KCl). Ag/AgCl potential depends on the Cl⁻ of its KCl filling solution.
Salt bridge / liquid junctionCompletes the circuitBlocked junction → unstable, noisy readings
Standard electrodeZero point for all half-cellsStandard hydrogen electrode = 0.000 V
🔥 VERY HIGH-YIELD — Nernst equation: E = E° + (2.303 RT/nF) log(activity). The term RT/nF is the slope. For a univalent ion the slope is about 59–61 mV per 10-fold change (about 60 mV at 37 °C). For a divalent ion (Ca²⁺) it is about 30 mV. A 10-fold rise in Na⁺ → voltage rises about 60 mV.

An ISE measures activity (the "free, active" ion), and it responds to one ion much more than to other ions — it is not 100% specific. Selectivity depends on the membrane.

💡 Direct vs indirect ISE — the most tested idea

Direct ISEIndirect ISE
SampleUndiluted (whole blood, plasma, serum)Diluted before measurement
Where usedBlood gas analyzers, point-of-careLarge automated chemistry analyzers
What it measuresIon in plasma water (mmol/kg water)Ion in total plasma volume (mmol/L)
High lipids / very high proteinNOT affectedFalsely LOW Na⁺ (pseudohyponatremia) — "electrolyte exclusion"
Severe lipemia or paraprotein (e.g., myeloma) ↓ Lipid/protein takes space → LESS water per liter of plasma ↓ Indirect ISE dilutes a fixed VOLUME (includes the "dead" lipid space) ↓ Less sodium in the measured aliquot → Na⁺ falsely LOW Direct ISE reads the water phase only → Na⁺ TRUE
🧠 MEMORY TRICK — "Indirect = Inaccurate in lipemia." Flame photometry behaves like indirect ISE (also falsely low).

Blood gas and other electrochemical methods

MethodMeasuresKey point
pH electrodeH⁺Glass electrode, potentiometric. Calibrate with two buffers of known pH at 37 °C. Store in the medium the manufacturer recommends.
PCO₂ (Severinghaus)CO₂A pH electrode inside a bicarbonate solution behind a CO₂-permeable membrane. CO₂ → carbonic acid → pH falls. Potentiometric.
PO₂ (Clark)O₂Amperometric (CURRENT, not voltage). Platinum cathode reduces O₂; Ag/AgCl anode. Current ∝ PO₂. Based on polarography.
CoulometryChloride (sweat Cl⁻)Silver electrode releases Ag⁺ at constant current → AgCl precipitates; endpoint by amperometry. Titration time ∝ Cl⁻. Bromide interferes (falsely high).
Amperometric glucoseGlucoseGlucose oxidase → H₂O₂ oxidized at platinum electrode → current
Anodic stripping voltammetryLead, heavy metalsMetal plated onto electrode, then stripped off → current

Blood gas analyzers measure pH, PCO₂ and PO₂; they calculate HCO₃⁻ and base excess.

🔬 LAB CONNECTION — ISE troubleshooting: slow response with serum but not with aqueous calibrator = protein coating the membrane. Persistent noise = blocked reference junction (salt crystals). Ionized calcium is measured by Ca²⁺ ISE.
⚠️ EXAM TRAP — Clark PO₂ electrode is amperometric. PCO₂ and pH electrodes are potentiometric. Valinomycin = K⁺ (not Na⁺).

More electrochemistry facts (added in v1.1)

PrincipleWhat is measuredExamples
PotentiometryVoltage at zero currentpH, Na⁺, K⁺, Ca²⁺ ISE, PCO₂
Amperometry / polarographyCurrent at a fixed applied voltagePO₂ (Clark), glucose oxidase electrode, HPLC-ECD (glassy carbon)
CoulometryQuantity of charge (coulombs) over timeChloride titrator (Ag⁺ generated)
Anodic stripping voltammetryCurrent when plated metal is stripped offLead
  • The Ag/AgCl reference potential is set by the chloride (KCl) concentration of its filling solution. Reference electrodes: Ag/AgCl, calomel, hydrogen. Glass = indicator, not reference.
  • In the Clark electrode, O₂ is reduced at the platinum cathode; the silver anode supplies the electrons. Current ∝ PO₂.
  • Store a pH glass electrode wet, in the manufacturer's recommended solution — never dry, never in tap water.
🧠 MEMORY TRICK — Potentio = Potential (volts), Ampero = Amperes (current), Coulo = Coulombs (charge).

Sources for v1.1 additions: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q36, Q44, Q54, Q59 (pp.180–184); SUCCESS! in Clinical Laboratory Science, Ch.1, Q27, Q34, Q36 (pp.98–99; rationales pp.150–152); BOC Study Guide, 6th ed., Ch.3, Q323 (p.190); A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1, p.6.

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.17 Electrochemistry and Chemical Sensors (pp.224.e4–224.e8); Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q34–Q47 (pp.179–182); SUCCESS! in Clinical Laboratory Science, Ch.1, Q24–Q36 rationales (pp.149–152); BOC Study Guide, 6th ed., Ch.3, Q309, Q319–Q328 (pp.188–190).

📌 MUST MEMORIZE

2 · Osmometry — counting dissolved particles

Osmolality = number of dissolved particles per kilogram of water (mOsm/kg). Particle number matters, not size. Lipids and proteins are not dissolved particles, so they add almost nothing.

Colligative properties (change with particle number)

Add 1 osmol of solute to 1 kg of water: • Freezing point ↓ by 1.86 °C ← used by most labs • Vapor pressure ↓ (about 0.3 mm Hg) • Boiling point ↑ by 0.52 °C • Osmotic pressure ↑ (about 22.4 atm)
🧠 MEMORY TRICK — "More particles make water LAZY": it freezes LATER (lower), boils LATER (higher), and evaporates LESS (lower vapor pressure).

Freezing point depression osmometer (cryoscopic)

  • Sample is supercooled, then stirred → ice forms → temperature rises to a plateau = the freezing point.
  • Temperature is sensed by a thermistor (temperature-sensitive resistor) measured with a Wheatstone bridge.
  • Freezing point ↓ is directly proportional to osmolality. Example: −0.558 °C ÷ 1.86 = 0.300 Osm/kg = 300 mOsm/kg.
  • Detects volatile substances such as ethanol and methanol → used for the osmolal gap.

Vapor pressure (dew point) osmometer

  • Uses a thermocouple (Seebeck effect) to find the dew point.
  • ⚠️ Misses volatile solutes (ethanol, methanol) because they evaporate. Do NOT use it to calculate an osmolal gap in poisoning.
🔥 VERY HIGH-YIELD — Calculated osmolality ≈ 2 Na⁺ + glucose/18 + BUN/2.8 (mg/dL). Normal serum 275–295 mOsm/kg. Osmolal gap = measured − calculated; normally <10. A big gap → unmeasured osmoles (ethanol, methanol, ethylene glycol).

Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q48–Q53 (pp.182–183); BOC Study Guide, 6th ed., Ch.3, Q305, Q353–Q354 (pp.188, 195).

🔥 HIGH-YIELD

3 · Electrophoresis — separation by charge (and size)

Electrophoresis = movement of charged molecules in an electric field. Speed depends on net charge, size and shape, the buffer (pH, ionic strength), the support medium, voltage/current and temperature.

Protein in a buffer with pH ABOVE its isoelectric point (pI) ↓ Protein loses H⁺ → net NEGATIVE charge (anion) ↓ Moves toward the ANODE (+) pH BELOW the pI → net POSITIVE → moves to CATHODE (−) pH = pI → net charge ZERO → does NOT move
🔥 VERY HIGH-YIELD — Serum protein electrophoresis (SPE) uses barbital (Veronal) buffer pH 8.6. At pH 8.6 all major serum proteins are negative and move from cathode (−) to anode (+). Sample is applied near the cathode.

Order of serum protein fractions (pH 8.6, from anode)

ANODE (+) Albumin → α1 → α2 → β → γ CATHODE (−) Albumin = fastest (most negative, lowest pI ≈ 4.6) γ-globulin = slowest (highest pI ≈ 7.2)
FractionMain proteins
α1α1-antitrypsin, α1-acid glycoprotein (orosomucoid)
α2α2-macroglobulin, haptoglobin, ceruloplasmin
βTransferrin, β-lipoprotein (LDL), C3, C4 (fibrinogen if plasma is used!)
γImmunoglobulins

Electroendosmosis (electro-osmotic flow)

Agarose and cellulose acetate carry fixed negative charges. Buffer cations (with water) move to the cathode. This flow pushes against protein migration, so γ-globulins can be carried slightly toward the cathode (behind the application point). Capillary electrophoresis uses this flow on purpose (no solid support; high voltage; UV detector).

Support media, stains and problems

  • Media: agarose, cellulose acetate (5–7 fractions); polyacrylamide and starch gel separate by charge AND size (molecular sieving; SDS-PAGE gives molecular weight). Celite is NOT an electrophoresis medium (it is a GC support).
  • Protein stains: Ponceau S, amido black, Coomassie blue; silver stain is most sensitive (CSF). Lipoprotein stains: Oil red O, Fat red 7B, Sudan black. DNA: ethidium bromide.
  • Quantify bands with a densitometer.
  • Raising ionic strength of buffer → slower migration (more heat). Raising voltage → faster.
  • Problems: doughnut-shaped (hollow) band = protein too concentrated → dilute; distorted zones = overloading; no migration = broken contact between buffer and medium; bisalbuminemia/wide albumin = drugs (e.g., penicillin) bound to albumin.
  • Isoelectric focusing: pH gradient made by ampholytes; each protein stops at its pI. Best for proteins with very close pIs (e.g., CSF oligoclonal bands).
  • DNA: agarose gel separates large DNA fragments; polyacrylamide for small ones. Western blot = protein electrophoresis + transfer + antibody detection.
  • Hemoglobin: cellulose acetate pH 8.6 (A fastest; S, D, G together; A2, C, E, O together) → confirm with citrate agar pH 6.0–6.2 (separates S from D, C from A2).
⚠️ EXAM TRAP — "Electrophoresis separates by molecular weight only" is wrong. Basic zone electrophoresis separates by charge (and size/shape); the best answer is "ionic charge and/or molecular weight". A buffer of pH 4.5 (below the pI) sends proteins to the cathode.
  • Silver stain is about 30 times more sensitive than Coomassie blue → CSF and urine proteins can be run without concentrating them.
  • Isoelectric focusing: a protein moves until it reaches the pH equal to its pI; there its net charge is zero and it stops (focuses).

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.18 Electrophoresis (pp.225–227, 235); Medical Laboratory Science Review (Harr), Ch.5, §5.6, Q11–Q20, Q41 (pp.248–249, 255); SUCCESS! in Clinical Laboratory Science, Ch.1, Q37–Q47 (pp.99–100, rationales pp.152–154); BOC Study Guide, 6th ed., Ch.3, Q313–Q318 (p.189).

💡 UNDERSTAND THIS

4 · Chromatography and mass spectrometry

Chromatography separates substances by how they share themselves between a stationary phase and a mobile phase (differential solubility / partitioning). Substances that like the mobile phase move fast; substances that stick to the stationary phase move slowly.

TypeMobile phaseStationary phaseKey exam facts
Thin-layer (TLC)Liquid (solvent)Silica gel on glass/plasticQualitative drug screen; non-volatile mixtures. Rf = distance of spot ÷ distance of solvent front (from origin). All spots at the solvent front → wrong solvent mixture.
Gas–liquid (GLC/GC)Gas (carrier, e.g., nitrogen, helium)Liquid coated on column packingSample must be volatile or made volatile. Elution order ≈ boiling point. Detector: flame ionization. Alcohols, drugs.
High-performance liquid (HPLC)Liquid pumped under pressurePacked column (e.g., C18)Parts: injector, pump(solvent delivery), column, detector (UV, fluorescence, electrochemical). Reverse phase = stationary phase nonpolar (C18), mobile phase more polar. Isocratic = constant solvent; gradient = changing solvent. High pressure + bad peaks → obstructed line.
Ion-exchangeLiquidCharged resinSeparates by sign and size of charge (e.g., HbA1c, amino acids, CK isoenzymes historically)
Gel filtration (size exclusion)LiquidPorous gelSeparates by molecular size/weight

Internal standard (added to every sample and standard) corrects for variation in extraction and injection.

Mass spectrometry (MS)

  • Molecules are ionized (usually to cations) and fragments are separated by their mass-to-charge ratio (m/z). The fragment pattern is a "fingerprint".
  • Needs a vacuum in the mass filter (stops ions colliding).
  • GC-MS = confirmation test for drugs of abuse (gold standard). GC-MS uses electron (impact) ionization. LC-MS uses electrospray ionization.
  • Tandem MS (MS/MS) = newborn screening for amino acid, organic acid and fatty acid disorders; the first mass filter acts like a chromatography column.
  • ICP-MS = trace elements and metals.
🧠 MEMORY TRICK — "Gas needs Gone-to-vapor (volatile). TLC is for things that will Take a seat (non-volatile)." Screen by immunoassay → confirm by GC-MS.

More chromatography and MS facts (added in v1.1)

  • HPLC fractions are usually detected by a UV spectrophotometer. Other HPLC detectors: fluorescence, electrochemical (glassy carbon electrode — catecholamines), mass spectrometer. GC detectors: flame ionization, thermal conductivity, MS.
  • MS measures cations by m/z. Base peak = the most abundant ion (set at 100%).
  • GC-MS ionizes with an electron source (electron impact, under vacuum). LC-MS uses electrospray. ICP-MS (argon plasma torch) = trace metals. MALDI = proteins/bacteria.
  • Vacuum in the mass filter prevents ions colliding with gas molecules.
  • Tandem MS: the first mass filter acts like the chromatography column (selects parent ions); collision cell makes daughter ions.
  • Troubleshooting: peaks at m/z 16, 17, 18 = oxygen (and its isotopes) → leak / contaminated carrier gas.
  • Screen positive by immunoassay but GC-MS negative → the screen was a false positive (cross-reactivity); report negative.

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.19 Chromatography (p.268) and Ch.20 Mass Spectrometry; Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q55–Q73 (pp.183–187); SUCCESS! in Clinical Laboratory Science, Ch.1, Q48–Q58 (pp.100–101, rationales pp.154–155); BOC Study Guide, 6th ed., Ch.3, Q310–Q312 (pp.188–189) and Ch.8, Q213–Q220 (pp.493–494).

🔬 LAB CONNECTION

7 · Flow cytometry

Cells flow one by one in a stream (hydrodynamic focusing) through a laser beam. Each cell scatters light and, if stained with fluorescent antibodies, emits fluorescence.

SignalTells you
Forward scatter (FSC)Cell size
Side scatter (SSC, 90°)Internal complexity / granularity
Fluorescence (fluorochrome-labeled monoclonal antibodies, CD markers)Which antigens the cell carries (immunophenotype)
  • Light source: laser (e.g., 488-nm argon/solid-state). Detectors: photodiode (FSC) and photomultiplier tubes (SSC, fluorescence).
  • Gating = selecting the cell population to analyze. Wrong gate → the desired cells are not counted.
  • Uses: CD4 T-cell counts in HIV (best instrument), leukemia/lymphoma immunophenotyping, PNH (CD55/CD59), stem cell (CD34) counts, DNA ploidy, reticulocyte counts.
  • Basic workflow: sample preparation → (Fc) blocking → antibody incubation → data acquisition → analysis/gating.
  • Scatter plot: lymphocytes = low FSC/low SSC; granulocytes = high SSC. A dense population between lymphoid and myeloid regions → abnormal cells (often blasts).
⚠️ EXAM TRAP — Flow cytometry light source is a LASER, not a mercury arc lamp. Flow cytometry both scatters light AND detects emitted fluorescence.

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 Optical Techniques (pp.214, 217–218); Medical Laboratory Science Review (Harr), Ch.3, §3.2, Q21–Q23 (p.87); p.4, Q58; p.5, Q81.