1 · Ion-Selective Electrodes (ISE) and electrochemistry
Potentiometry = measuring a voltage (potential) between two electrodes when no current flows. The voltage tells us the activity of one ion. An Ion-Selective Electrode (ISE) is a potentiometric electrode with a membrane that "prefers" one ion.
The parts of every ISE system
| Part | Job | Examples |
|---|---|---|
| Indicator (measuring) electrode | Voltage changes with the ion | Na⁺ = special glass (aluminum silicate) · pH = glass · K⁺ = valinomycin membrane · Ca²⁺, Li⁺ = organic ion-exchanger (ionophore) · NH₄⁺ = nonactin/monactin |
| Reference electrode | Gives a CONSTANT voltage | Silver/silver chloride (Ag/AgCl) · Calomel (mercury + mercurous chloride + saturated KCl). Ag/AgCl potential depends on the Cl⁻ of its KCl filling solution. |
| Salt bridge / liquid junction | Completes the circuit | Blocked junction → unstable, noisy readings |
| Standard electrode | Zero point for all half-cells | Standard hydrogen electrode = 0.000 V |
An ISE measures activity (the "free, active" ion), and it responds to one ion much more than to other ions — it is not 100% specific. Selectivity depends on the membrane.
💡 Direct vs indirect ISE — the most tested idea
| Direct ISE | Indirect ISE | |
|---|---|---|
| Sample | Undiluted (whole blood, plasma, serum) | Diluted before measurement |
| Where used | Blood gas analyzers, point-of-care | Large automated chemistry analyzers |
| What it measures | Ion in plasma water (mmol/kg water) | Ion in total plasma volume (mmol/L) |
| High lipids / very high protein | NOT affected | Falsely LOW Na⁺ (pseudohyponatremia) — "electrolyte exclusion" |
Blood gas and other electrochemical methods
| Method | Measures | Key point |
|---|---|---|
| pH electrode | H⁺ | Glass electrode, potentiometric. Calibrate with two buffers of known pH at 37 °C. Store in the medium the manufacturer recommends. |
| PCO₂ (Severinghaus) | CO₂ | A pH electrode inside a bicarbonate solution behind a CO₂-permeable membrane. CO₂ → carbonic acid → pH falls. Potentiometric. |
| PO₂ (Clark) | O₂ | Amperometric (CURRENT, not voltage). Platinum cathode reduces O₂; Ag/AgCl anode. Current ∝ PO₂. Based on polarography. |
| Coulometry | Chloride (sweat Cl⁻) | Silver electrode releases Ag⁺ at constant current → AgCl precipitates; endpoint by amperometry. Titration time ∝ Cl⁻. Bromide interferes (falsely high). |
| Amperometric glucose | Glucose | Glucose oxidase → H₂O₂ oxidized at platinum electrode → current |
| Anodic stripping voltammetry | Lead, heavy metals | Metal plated onto electrode, then stripped off → current |
Blood gas analyzers measure pH, PCO₂ and PO₂; they calculate HCO₃⁻ and base excess.
More electrochemistry facts (added in v1.1)
| Principle | What is measured | Examples |
|---|---|---|
| Potentiometry | Voltage at zero current | pH, Na⁺, K⁺, Ca²⁺ ISE, PCO₂ |
| Amperometry / polarography | Current at a fixed applied voltage | PO₂ (Clark), glucose oxidase electrode, HPLC-ECD (glassy carbon) |
| Coulometry | Quantity of charge (coulombs) over time | Chloride titrator (Ag⁺ generated) |
| Anodic stripping voltammetry | Current when plated metal is stripped off | Lead |
- The Ag/AgCl reference potential is set by the chloride (KCl) concentration of its filling solution. Reference electrodes: Ag/AgCl, calomel, hydrogen. Glass = indicator, not reference.
- In the Clark electrode, O₂ is reduced at the platinum cathode; the silver anode supplies the electrons. Current ∝ PO₂.
- Store a pH glass electrode wet, in the manufacturer's recommended solution — never dry, never in tap water.
Sources for v1.1 additions: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q36, Q44, Q54, Q59 (pp.180–184); SUCCESS! in Clinical Laboratory Science, Ch.1, Q27, Q34, Q36 (pp.98–99; rationales pp.150–152); BOC Study Guide, 6th ed., Ch.3, Q323 (p.190); A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1, p.6.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.17 Electrochemistry and Chemical Sensors (pp.224.e4–224.e8); Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q34–Q47 (pp.179–182); SUCCESS! in Clinical Laboratory Science, Ch.1, Q24–Q36 rationales (pp.149–152); BOC Study Guide, 6th ed., Ch.3, Q309, Q319–Q328 (pp.188–190).
2 · Osmometry — counting dissolved particles
Osmolality = number of dissolved particles per kilogram of water (mOsm/kg). Particle number matters, not size. Lipids and proteins are not dissolved particles, so they add almost nothing.
Colligative properties (change with particle number)
Freezing point depression osmometer (cryoscopic)
- Sample is supercooled, then stirred → ice forms → temperature rises to a plateau = the freezing point.
- Temperature is sensed by a thermistor (temperature-sensitive resistor) measured with a Wheatstone bridge.
- Freezing point ↓ is directly proportional to osmolality. Example: −0.558 °C ÷ 1.86 = 0.300 Osm/kg = 300 mOsm/kg.
- Detects volatile substances such as ethanol and methanol → used for the osmolal gap.
Vapor pressure (dew point) osmometer
- Uses a thermocouple (Seebeck effect) to find the dew point.
- ⚠️ Misses volatile solutes (ethanol, methanol) because they evaporate. Do NOT use it to calculate an osmolal gap in poisoning.
Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q48–Q53 (pp.182–183); BOC Study Guide, 6th ed., Ch.3, Q305, Q353–Q354 (pp.188, 195).
3 · Electrophoresis — separation by charge (and size)
Electrophoresis = movement of charged molecules in an electric field. Speed depends on net charge, size and shape, the buffer (pH, ionic strength), the support medium, voltage/current and temperature.
Order of serum protein fractions (pH 8.6, from anode)
| Fraction | Main proteins |
|---|---|
| α1 | α1-antitrypsin, α1-acid glycoprotein (orosomucoid) |
| α2 | α2-macroglobulin, haptoglobin, ceruloplasmin |
| β | Transferrin, β-lipoprotein (LDL), C3, C4 (fibrinogen if plasma is used!) |
| γ | Immunoglobulins |
Electroendosmosis (electro-osmotic flow)
Agarose and cellulose acetate carry fixed negative charges. Buffer cations (with water) move to the cathode. This flow pushes against protein migration, so γ-globulins can be carried slightly toward the cathode (behind the application point). Capillary electrophoresis uses this flow on purpose (no solid support; high voltage; UV detector).
Support media, stains and problems
- Media: agarose, cellulose acetate (5–7 fractions); polyacrylamide and starch gel separate by charge AND size (molecular sieving; SDS-PAGE gives molecular weight). Celite is NOT an electrophoresis medium (it is a GC support).
- Protein stains: Ponceau S, amido black, Coomassie blue; silver stain is most sensitive (CSF). Lipoprotein stains: Oil red O, Fat red 7B, Sudan black. DNA: ethidium bromide.
- Quantify bands with a densitometer.
- Raising ionic strength of buffer → slower migration (more heat). Raising voltage → faster.
- Problems: doughnut-shaped (hollow) band = protein too concentrated → dilute; distorted zones = overloading; no migration = broken contact between buffer and medium; bisalbuminemia/wide albumin = drugs (e.g., penicillin) bound to albumin.
- Isoelectric focusing: pH gradient made by ampholytes; each protein stops at its pI. Best for proteins with very close pIs (e.g., CSF oligoclonal bands).
- DNA: agarose gel separates large DNA fragments; polyacrylamide for small ones. Western blot = protein electrophoresis + transfer + antibody detection.
- Hemoglobin: cellulose acetate pH 8.6 (A fastest; S, D, G together; A2, C, E, O together) → confirm with citrate agar pH 6.0–6.2 (separates S from D, C from A2).
- Silver stain is about 30 times more sensitive than Coomassie blue → CSF and urine proteins can be run without concentrating them.
- Isoelectric focusing: a protein moves until it reaches the pH equal to its pI; there its net charge is zero and it stops (focuses).
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.18 Electrophoresis (pp.225–227, 235); Medical Laboratory Science Review (Harr), Ch.5, §5.6, Q11–Q20, Q41 (pp.248–249, 255); SUCCESS! in Clinical Laboratory Science, Ch.1, Q37–Q47 (pp.99–100, rationales pp.152–154); BOC Study Guide, 6th ed., Ch.3, Q313–Q318 (p.189).
4 · Chromatography and mass spectrometry
Chromatography separates substances by how they share themselves between a stationary phase and a mobile phase (differential solubility / partitioning). Substances that like the mobile phase move fast; substances that stick to the stationary phase move slowly.
| Type | Mobile phase | Stationary phase | Key exam facts |
|---|---|---|---|
| Thin-layer (TLC) | Liquid (solvent) | Silica gel on glass/plastic | Qualitative drug screen; non-volatile mixtures. Rf = distance of spot ÷ distance of solvent front (from origin). All spots at the solvent front → wrong solvent mixture. |
| Gas–liquid (GLC/GC) | Gas (carrier, e.g., nitrogen, helium) | Liquid coated on column packing | Sample must be volatile or made volatile. Elution order ≈ boiling point. Detector: flame ionization. Alcohols, drugs. |
| High-performance liquid (HPLC) | Liquid pumped under pressure | Packed column (e.g., C18) | Parts: injector, pump(solvent delivery), column, detector (UV, fluorescence, electrochemical). Reverse phase = stationary phase nonpolar (C18), mobile phase more polar. Isocratic = constant solvent; gradient = changing solvent. High pressure + bad peaks → obstructed line. |
| Ion-exchange | Liquid | Charged resin | Separates by sign and size of charge (e.g., HbA1c, amino acids, CK isoenzymes historically) |
| Gel filtration (size exclusion) | Liquid | Porous gel | Separates by molecular size/weight |
Internal standard (added to every sample and standard) corrects for variation in extraction and injection.
Mass spectrometry (MS)
- Molecules are ionized (usually to cations) and fragments are separated by their mass-to-charge ratio (m/z). The fragment pattern is a "fingerprint".
- Needs a vacuum in the mass filter (stops ions colliding).
- GC-MS = confirmation test for drugs of abuse (gold standard). GC-MS uses electron (impact) ionization. LC-MS uses electrospray ionization.
- Tandem MS (MS/MS) = newborn screening for amino acid, organic acid and fatty acid disorders; the first mass filter acts like a chromatography column.
- ICP-MS = trace elements and metals.
More chromatography and MS facts (added in v1.1)
- HPLC fractions are usually detected by a UV spectrophotometer. Other HPLC detectors: fluorescence, electrochemical (glassy carbon electrode — catecholamines), mass spectrometer. GC detectors: flame ionization, thermal conductivity, MS.
- MS measures cations by m/z. Base peak = the most abundant ion (set at 100%).
- GC-MS ionizes with an electron source (electron impact, under vacuum). LC-MS uses electrospray. ICP-MS (argon plasma torch) = trace metals. MALDI = proteins/bacteria.
- Vacuum in the mass filter prevents ions colliding with gas molecules.
- Tandem MS: the first mass filter acts like the chromatography column (selects parent ions); collision cell makes daughter ions.
- Troubleshooting: peaks at m/z 16, 17, 18 = oxygen (and its isotopes) → leak / contaminated carrier gas.
- Screen positive by immunoassay but GC-MS negative → the screen was a false positive (cross-reactivity); report negative.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.19 Chromatography (p.268) and Ch.20 Mass Spectrometry; Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q55–Q73 (pp.183–187); SUCCESS! in Clinical Laboratory Science, Ch.1, Q48–Q58 (pp.100–101, rationales pp.154–155); BOC Study Guide, 6th ed., Ch.3, Q310–Q312 (pp.188–189) and Ch.8, Q213–Q220 (pp.493–494).
7 · Flow cytometry
Cells flow one by one in a stream (hydrodynamic focusing) through a laser beam. Each cell scatters light and, if stained with fluorescent antibodies, emits fluorescence.
| Signal | Tells you |
|---|---|
| Forward scatter (FSC) | Cell size |
| Side scatter (SSC, 90°) | Internal complexity / granularity |
| Fluorescence (fluorochrome-labeled monoclonal antibodies, CD markers) | Which antigens the cell carries (immunophenotype) |
- Light source: laser (e.g., 488-nm argon/solid-state). Detectors: photodiode (FSC) and photomultiplier tubes (SSC, fluorescence).
- Gating = selecting the cell population to analyze. Wrong gate → the desired cells are not counted.
- Uses: CD4 T-cell counts in HIV (best instrument), leukemia/lymphoma immunophenotyping, PNH (CD55/CD59), stem cell (CD34) counts, DNA ploidy, reticulocyte counts.
- Basic workflow: sample preparation → (Fc) blocking → antibody incubation → data acquisition → analysis/gating.
- Scatter plot: lymphocytes = low FSC/low SSC; granulocytes = high SSC. A dense population between lymphoid and myeloid regions → abnormal cells (often blasts).
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 Optical Techniques (pp.214, 217–218); Medical Laboratory Science Review (Harr), Ch.3, §3.2, Q21–Q23 (p.87); p.4, Q58; p.5, Q81.
⚡ Last-Minute Review — Electrochemistry, osmometry, electrophoresis, chromatography*, MS & flow cytometry
1. ISE & electrochemistry
- Potentiometry = voltage, no current. ISE measures ion activity; Nernst slope ≈ 60 mV per 10-fold change (univalent), ≈30 mV (divalent).
- Na⁺ = glass · pH = glass · K⁺ = valinomycin · NH₄⁺ = nonactin · Ca²⁺ = ionophore. Reference = Ag/AgCl or calomel. Standard = hydrogen electrode (0 V).
- Indirect ISE (diluted) + lipemia/high protein → falsely low Na⁺ (pseudohyponatremia). Direct ISE (undiluted) = true.
- PCO₂ = pH electrode in bicarbonate (potentiometric). PO₂ = Clark, amperometric. Analyzer measures pH, PCO₂, PO₂; calculates HCO₃⁻.
- Coulometry = chloride (sweat); Ag⁺ generated; time ∝ Cl⁻. Protein on membrane = slow response; blocked junction = noise.
2. Osmometry
- 1 Osm/kg: freezing point ↓1.86 °C, boiling ↑0.52 °C, vapor pressure ↓.
- Freezing point osmometer: thermistor + Wheatstone bridge; detects alcohols. Vapor pressure: thermocouple; misses ethanol/methanol.
- Calc osm = 2Na + Glu/18 + BUN/2.8. Gap = measured − calc (<10).
3. Electrophoresis
- pH 8.6 barbital: proteins negative → anode. pH below pI → positive → cathode. At pI → no movement.
- Anode → albumin, α1, α2, β, γ → cathode. Transferrin & C3 = β. Haptoglobin = α2.
- Electroendosmosis pushes γ toward cathode. Capillary electrophoresis: no support, EOF, high voltage.
- Stains: Ponceau S, amido black, Coomassie (protein); Oil red O (lipoprotein); ethidium bromide (DNA). Agarose = large DNA.
- Hollow band = too concentrated → dilute. Distorted = overload. No migration = lost buffer contact. Isoelectric focusing = ampholytes.
4. Chromatography & MS
- Principle: differential solubility/partition between stationary and mobile phases.
- TLC: non-volatile; Rf = spot ÷ solvent front. GC: volatile, elution by boiling point, FID detector. HPLC: pumped liquid, UV detector; reverse phase = polar mobile, nonpolar C18; isocratic vs gradient.
- Ion-exchange = charge. Gel filtration = size. Internal standard corrects extraction/injection.
- MS = mass-to-charge (m/z); vacuum. GC-MS = confirmation of drugs. LC-MS = electrospray. Tandem MS = newborn screening.
5. Immunoassays
- Chemiluminescence = most common automated & most sensitive (troponin). ELISA beats RIA: cost, no radiation.
- Competitive: signal ∝ 1/[analyte] (small molecules). Sandwich: signal ∝ [analyte]; specific Ab on solid phase; hook effect → dilute.
- Homogeneous (no separation): EMIT (enzyme on drug, G6PD, 340 nm), FPIA, CEDIA. Heterogeneous: ELISA, RIA.
- Indirect ELISA = antibody detection. Immunoprecipitation is NOT a labeled immunoassay.
- Poor washing/long color time → falsely high. Above top standard → dilute. Heterophile/HAMA → usually falsely high.
6. Flow cytometry
- Laser. FSC = size. SSC = granularity. Fluorescent CD antibodies = phenotype. Best for CD4 counts. Gating selects cells.
7. QC statistics
- CV% = SD/mean × 100 (compare precision at different means). SD = √[Σ(x−x̄)²/(n−1)]; variance = SD².
- ±1 SD 68% · ±2 SD 95% · ±3 SD 99.7%. Outside ±2 SD by chance 1/20; above +2 SD 1/40; two in a row 1/1600.
- Precision = reproducibility (random error). Accuracy = trueness (systematic error, bias). Reliability = accuracy + precision over time.
- Regression: intercept = constant error; slope = proportional error. r = association only.
8. Levey-Jennings & Westgard
- Set limits from ~20 days of data. Shift = sudden, ≥6 on one side (new lot, bad calibrator). Trend = gradual (deterioration).
- 1₂s warning · 1₃s random · 2₂s systematic · R₄s random (within run) · 4₁s systematic · 10x̄ systematic.
- Cusum = earliest trend detection. Youden = interlab. Delta check = specimen mix-up.
9. QC action & QA
- Out of control → don't report → follow procedure → find/fix cause → rerun controls → rerun patients → document.
- Calibrator ≠ control. 2 levels / 24 h minimum. QC = analytical phase. PT/EQA = accuracy vs other labs.
- Reference interval: 120 subjects (central 95%); verify with 20 (≤2 outside). Sensitivity TP/(TP+FN); specificity TN/(TN+FP); inversely related.
v1.1 · Extra high-yield points
- Potentiometry = volts · amperometry = current (PO₂, glucose, HPLC-ECD glassy carbon) · coulometry = charge (Cl⁻) · ASV = current (lead).
- Ag/AgCl reference potential set by KCl (Cl⁻). Clark: O₂ reduced at Pt cathode.
- Chemiluminescence: HRP + luminol; ALP + dioxetane; acridinium direct; ruthenium = ECL; singlet oxygen = LOCI. RIA labels antigen, IRMA labels antibody.
- HPLC detector = UV. Base peak = most abundant. ICP-MS = metals. m/z 16-17-18 = oxygen leak.
- Index of precision = CV. Different means → compare CV.
- L-J = all errors; Cusum = earliest trend; Youden = between labs; regression = method comparison.
- PV− = TN/(TN+FN). PV+ falls when prevalence is low. Six Sigma = error rate (3.4 per million).
- Centrifuge: tachometer, every 3 months. AMR = no dilution. Controls ≠ calibrators. PT samples = unknowns.