1 · Light, wavelength and color — the language of photometry
Photometry = measuring a substance by measuring light. Most chemistry tests make a colored product; the more analyte, the more color, the more light is absorbed.
Wavelength and energy
- Wavelength (λ) = distance between two wave peaks, measured in nanometers (nm).
- Short wavelength = HIGH energy (ultraviolet). Long wavelength = LOW energy (infrared). Energy is inversely proportional to wavelength.
- 📌 Ultraviolet (UV) ≈ below 400 nm · Visible ≈ 400–700 nm (Tietz: about 380–750 nm) · Infrared (IR) ≈ above 700 nm.
| Wavelength (approx.) | Color of this light | A solution that ABSORBS it looks… |
|---|---|---|
| 400–450 nm | Violet | Yellow / yellow-green |
| 450–500 nm | Blue | Yellow / orange (red solutions also absorb here) |
| 500–560 nm | Green | Red / purple |
| 560–600 nm | Yellow–orange | Blue |
| 600–700 nm | Red | Blue-green |
• The color you SEE is the light that is transmitted (not absorbed).
• A solution absorbs the complementary color. We measure at the wavelength the solution absorbs most strongly.
• Example: a red solution transmits 600–650 nm and absorbs strongly at 400–500 nm. A green solution transmits ~525 nm and is measured at a wavelength it absorbs (e.g. ~450 nm or ~620 nm).
Sources: SUCCESS! in Clinical Laboratory Science, Ch.1, §I.A (p.2); Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q4–Q5 rationales (p.173); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 (pp.201–203).
2 · Beer–Lambert law, absorbance (A) and transmittance (%T)
Transmittance (T) = fraction of light that passes through: T = I / I₀ (I₀ = light going in, I = light coming out). %T = T × 100.
Absorbance (A) = how much light the solution "takes": A = −log T = log (1/T) = 2 − log %T.
| Relationship | Type | Graph |
|---|---|---|
| Absorbance vs concentration | Directly proportional (linear) | Straight line on ordinary (linear) graph paper |
| %T vs concentration | Inverse and logarithmic — NON-linear | Curve on linear paper; straight line on semi-log paper |
| Absorbance vs light path | Directly proportional | Double the path → double the A |
📌 Numbers to memorize
| %T | 100% | 50% | 10% | 1% | 0.1% |
|---|---|---|---|---|---|
| A | 0 | 0.301 | 1.0 | 2.0 | 3.0 |
Beer's law works only when…
- The light is monochromatic (one narrow wavelength band).
- The solvent absorbs very little compared with the analyte.
- The solution is dilute enough (within the linear range).
- No optical interferent (turbidity, fluorescence, other colored substances).
- No chemical change of the colored product during reading.
Deviations from Beer's law (curve bends at high concentration)
- Stray light (most common cause of loss of linearity at high absorbance).
- Very high concentration → dilute the sample and multiply by the dilution factor.
- Wide bandpass (polychromatic light), interfering absorbers, chemical changes.
• "Amount of light absorbed is directly proportional to concentration" = Beer's law.
• "%T and concentration" = non-linear (inverse log).
• A = 2 − log %T.
• Absorbance has no unit (it is a log ratio). It is never "0.500 nm".
Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q1–Q3 rationales (p.173); SUCCESS! in Clinical Laboratory Science, Ch.12, Q55 (p.965) rationale; Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 (pp.202–204); A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1, §D.2–3 (p.5).
4 · The spectrophotometer I — light source, slits and monochromator
Light sources
| Lamp | Spectrum | Use |
|---|---|---|
| Tungsten (incandescent) / tungsten–halogen (quartz-halogen) | Continuous: near-UV → visible → near-IR (most energy in IR; weak below ~350 nm) | 🔥 Most common lamp for the VISIBLE range. Halogen vapor (iodine/bromine) gives longer life and a brighter, wider output. |
| Deuterium / hydrogen | Continuous UV (~200–375 nm) | Ultraviolet work |
| Mercury vapor (arc) | Line (discontinuous) spectrum (e.g. 254, 365 nm) | Wavelength calibration, HPLC UV detectors, fluorescence excitation |
| Xenon arc / flash | Continuous UV + visible | Fluorometers, some analyzers |
| Hollow-cathode lamp | Line spectrum of ONE element | Atomic absorption (cathode made of the metal measured) |
| Laser | Coherent, very narrow, intense | Flow cytometry, nephelometry |
| LED | Narrow band, incoherent | Compact photometers, point-of-care |
Monochromator = isolates the wavelength band you want
- Glass (absorption) filters — cheap, wide bandpass (~50 nm). Used in photometers/colorimeters.
- Interference filters — narrow bandpass (~5–15 nm).
- Prism — refraction; non-linear spectrum, very good separation in the UV.
- Diffraction grating — most common in spectrophotometers; gives a uniform, linear spectrum.
- Entrance slit lets light in (narrow beam); exit slit selects the band that reaches the sample. Narrower exit slit → narrower bandpass.
Bandpass (spectral bandwidth)
- = the range of wavelengths passed at half of the peak transmittance ("half-band width").
- Setting 540 nm with a 10 nm bandpass → light 535–545 nm hits the sample.
- Narrower bandpass = better resolution and better agreement with Beer's law (improves linearity).
- For true absorbance, instrument bandpass should be much smaller than the compound's natural absorption band (rule: ≤ about 1/5–1/10 of it).
• Visible lamp = tungsten (halogen). UV lamp = deuterium/hydrogen. AAS lamp = hollow cathode. Flow cytometry = laser.
• Monochromator job = "allow only the desired wavelength to reach the sample".
• After changing a lamp → re-check wavelength calibration (the light hits the monochromator at a new angle).
Sources: SUCCESS! in Clinical Laboratory Science, Ch.1, §I.B (pp.2–3) and Q1–Q2, Q9 rationales (pp.145–146); Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q6–Q8, Q16 rationales (pp.174–176); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 (pp.204–206); Clinical Laboratory Science Review, 5th ed. (LSU), Instrumentation (p.234).
5 · The spectrophotometer II — cuvette, detector, readout and beam design
Cuvette (cuvet, sample cell)
- Glass — visible range only (glass absorbs UV, below ~340 nm).
- Quartz / fused silica — needed for UV work.
- Plastic — disposable, visible (special UV-grade plastics exist).
- Standard light path = 1 cm. Cuvettes must be clean, unscratched, matched, without fingerprints or air bubbles.
Detectors (photodetectors) — convert light → electricity
| Detector | Key facts |
|---|---|
| Barrier-layer cell (photocell, photovoltaic) | Iron plate + selenium layer; no amplification; least sensitive; cheap colorimeters. |
| Phototube | Photo-emissive cathode in a vacuum tube; more sensitive than a photocell. |
| Photomultiplier tube (PMT) | 🔥 Most sensitive to low light; cathode + 10–15 dynodes amplify up to ~10⁶ times; very fast response (can follow a chopped beam); needs a regulated DC power supply; must be shielded from room/stray light. |
| Photodiode | Solid-state, rugged, less sensitive than PMT. |
| Photodiode array | Hundreds of diodes in a line, each reading a narrow band → whole spectrum at once (spectral scans, multi-wavelength analyzers). |
Beam designs
- Single-beam: one light path; blank and sample read one after the other.
- Double-beam (or reference beam): a chopper/mirrors split the light between sample and reference → corrects for lamp fluctuation and detector drift.
- Readout: meter, digital display, computer.
Photometer vs spectrophotometer
Filter photometer (colorimeter) uses filters. Spectrophotometer uses a prism or grating (continuous choice of wavelength). Any colored product of two reacting chemicals is measured with a spectrophotometer/colorimeter.
• Most sensitive detector = PMT. Spectral scan = photodiode array. Iron + selenium = barrier-layer cell.
• UV measurement (e.g. NADH at 340 nm) → quartz cuvette or UV-grade plastic.
• Parts of a colorimetric system: light source, filter/monochromator, cuvette, photodetector, readout — NOT a fuel source (fuel/flame belongs to AAS and flame photometry).
Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q9, Q14–Q15, Q19 rationales (pp.174–176); SUCCESS! in Clinical Laboratory Science, Ch.1, Q3–Q5 rationales (pp.145–146); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 (pp.204–207); A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1, §D.1 (p.4).
6 · Blanks, standards, calibrators and the standard curve
Setting the instrument
Types of blank
| Blank | Contains | Corrects for |
|---|---|---|
| Reagent blank | All reagents, NO sample (water/saline instead of sample) | Color of the reagents only |
| Sample (serum) blank | Sample + diluent/saline instead of the color reagent | Sample's own color/turbidity: hemolysis, icterus, lipemia, drugs |
| Water / solvent blank | Solvent only | Cuvette and solvent absorbance |
| Bichromatic reading | Same cuvette read at a primary (peak) and a secondary wavelength | Interferent that absorbs equally at both → acts as a blank for each sample |
Standards, calibrators and controls
- Primary standard: highly purified substance (≥ 99.95% pure), stable, not hygroscopic, can be dried and weighed directly to make an exact concentration.
- Secondary standard: concentration assigned by comparison with a primary standard or reference method. 🔥 Commercial calibrator sera are secondary standards (serum matrix).
- Calibrator = used to SET the instrument response (build the curve).
- Control = used to CHECK that the run is correct. A control is never used to calibrate.
Standard (calibration) curve
- Plot absorbance (y) vs concentration (x) on linear paper → straight line through the origin if Beer's law holds.
- Read unknowns only within the range of the standards. Above the top standard → dilute and repeat.
- Single-point calibration (one standard) assumes linearity through zero. Two-point/multi-point calibration defines slope and intercept; non-linear methods (many immunoassays) need several calibrators and a curve-fitting model.
- Linearity (reportable range / analytical measurement range) = lowest to highest concentration measured accurately. A linearity check is a non-statistical QC step.
- A wrong calibrator reading shifts ALL results (systematic error). Example: if one high standard reads too low, the slope falls and controls/patients read falsely high.
• Reagent blank ≠ sample blank. Hemolysis/icterus/lipemia need a SAMPLE blank or bichromatic reading.
• Calibrator sera = secondary standards.
• Instrument zero = 0 %T with light blocked; 100 %T with blank.
Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q17–Q18, Q21 rationales (pp.176–177); SUCCESS! in Clinical Laboratory Science, Ch.1, Q7–Q8 rationales (p.146); Ch.12, Q57 (p.966); Ch.13, §I (p.996); Clinical Laboratory Science Review, 5th ed. (LSU), Instrumentation Q1, Q10 (pp.246–249); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 (pp.203–205).
7 · Endpoint vs kinetic (rate) reactions and enzyme kinetics
| Endpoint (equilibrium) | Fixed-time (two-point) | Kinetic (continuous monitoring / rate) | |
|---|---|---|---|
| How read | Reaction goes to completion; read once at the end | Read at two fixed times; use ΔA | Read many times at equal intervals; calculate ΔA/min |
| Typical use | Glucose oxidase–peroxidase, cholesterol, triglyceride, biuret protein, classic Jaffe | Kinetic Jaffe creatinine | Enzyme activity (ALT, AST, CK, LD, ALP, amylase) |
| Blank | Reagent/sample blank needed | Initial reading acts as its own blank | Each sample is its own baseline |
Enzyme reaction curve
- Zero-order kinetics: substrate in great excess (≫ Km, about 100 × Km) → rate depends only on enzyme → used to measure enzyme activity.
- First-order kinetics: rate proportional to substrate concentration → used when the enzyme is a reagent to measure a substrate (e.g. glucose by hexokinase).
- Substrate exhaustion (very high enzyme): rate falls during the reading, analyzer flags "non-linear" → dilute the sample (less enzyme, substrate stays in excess) and repeat.
- NADH/NADPH absorb at 340 nm; NAD⁺ does not. NADH formed → A rises; NADH used → A falls.
- Km (Michaelis constant) = substrate concentration giving ½ Vmax. Lineweaver–Burk (1/v vs 1/[S]): y-intercept = 1/Vmax; x-intercept = −1/Km.
- Competitive inhibitor: binds the active site; overcome by more substrate; Km ↑, Vmax same. Non-competitive: binds elsewhere; Vmax ↓, Km same.
- Rate also depends on pH, temperature (each 10 °C ≈ doubles rate until denaturation), cofactors/activators (Mg²⁺ for CK and ALP; Ca²⁺ and Cl⁻ for amylase).
• Kinetic = several evenly spaced readings → ΔA/min. A single read after 5 min is NOT kinetic.
• Enzyme activity is proportional to rate only under zero-order conditions.
• Unit "U" (IU) = µmol/min → it measures enzyme activity.
• LD lactate → pyruvate: NADH produced → A at 340 nm increases. Pyruvate → lactate: NADH used → A decreases.
Sources: BOC Study Guide, 6th ed., Ch.3 Chemistry, Q301, Q342–Q347 (pp.187–193); Medical Laboratory Science Review (Harr), Ch.5, §5.7, Q2, Q5, Q7 rationales (pp.263–264); A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1, enzyme kinetics (pp.30–31).
8 · Spectrophotometer quality checks and photometric troubleshooting
| Check | What it tests | Material |
|---|---|---|
| Wavelength accuracy | Dial wavelength = real wavelength | Holmium oxide glass (sharp peaks, e.g. 361 nm; UV + visible), didymium filter (585 nm; visible, broad-bandpass instruments), deuterium 656 nm line, mercury lines |
| Stray light | Light outside the chosen band reaching the detector | Sharp-cutoff filters or cutoff solutions (e.g. sodium nitrite, nickel sulfate) — reading should be ~0 %T |
| Photometric (absorbance) accuracy and linearity | Correct absorbance readings across the range | Neutral-density filters, potassium dichromate solutions, dilution series of standards |
| Spectral bandwidth | Width at half-peak transmittance | Narrow interference filter or emission line |
Stray light — the classic cause of bad linearity
- Caused by second-order spectra, scratched/dirty optics, a darkened lamp envelope, room-light leaks.
- Effect: falsely LOW absorbance, worst at high concentration and at the ends of the spectrum → standard curve bends toward the x-axis.
Troubleshooting table
| What you see | Most likely cause | Action |
|---|---|---|
| Cannot reach 100 %T at 340 nm (OK in visible) | Aging tungsten lamp (weak in UV) | Replace the lamp, then re-check wavelength |
| High standards read too low; curve flattens | Stray light (or above linear range) | Check with cutoff filter; dilute high samples |
| Peak absorbance at the "wrong" wavelength | Wavelength mis-calibration | Holmium oxide / didymium check |
| Drift during readings | Lamp warm-up, unstable voltage, temperature | Warm up, double-beam/reference, check power |
| Erratic, noisy readings | Air bubbles, particles, fibrin, dirty or scratched cuvette, failing lamp | Clean/replace cuvette, remove bubbles, centrifuge |
| One sample falsely high | Lipemia, hemolysis, icterus (spectral interference) | Sample blank, bichromatic reading, ultracentrifuge/clear lipemia |
| All photometric tests "below detection limit" but electrolytes OK | Fibrin clot in the photometric sample probe | Clean the probe, re-centrifuge, repeat |
| Enzyme "non-linear" flag | Substrate exhaustion | Dilute sample, repeat |
| Controls high, patients high after calibration | Bad calibrator reading (slope too low) | Recalibrate before repeating controls |
• Stray light → sharp-cutoff filter. Wavelength → holmium oxide / didymium. Absorbance accuracy → neutral-density filter / dichromate.
• Stray light = most common cause of loss of linearity at high absorbance.
Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q6, Q10–Q13 rationales (pp.174–175); §5.11, Q40 (p.320); BOC Study Guide, 6th ed., Ch.3 Chemistry, Q307 (p.188); SUCCESS! in Clinical Laboratory Science, Ch.1, Q6 rationale (p.146); Clinical Laboratory Science Review, 5th ed. (LSU), Instrumentation (p.234); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 (pp.208–209).
9 · Other optical methods — fluorescence, luminescence, light scatter, reflectance and atomic absorption
| Method | What is measured | Key exam facts |
|---|---|---|
| Absorption spectrophotometry | Light absorbed by molecules | Beer's law; most chemistry tests |
| Fluorometry | Light emitted after excitation, at a longer wavelength | Detector at 90° to the excitation beam; two monochromators (primary = excitation, secondary = emission); ~1000× more sensitive and more specific than absorbance; problems: quenching, temperature (fluorescence ↓ as temperature ↑), pH, photobleaching, inner-filter effect |
| Fluorescence polarization (FPIA) | Polarized fluorescent light | Homogeneous immunoassay; polarization inversely proportional to drug concentration |
| Chemiluminescence | Light made by a chemical reaction | No light source and no excitation monochromator needed; labels: acridinium, luminol (H₂O₂ / horseradish peroxidase), dioxetane (alkaline phosphatase); extremely sensitive → most common automated immunoassay signal (e.g. high-sensitivity troponin) |
| Bioluminescence | Light from an enzyme reaction | Luciferase–luciferin; more sensitive than direct fluorescence |
| Electrochemiluminescence | Light produced at an electrode | Ruthenium label + tripropylamine |
| Nephelometry | Light scattered by particles, at an angle (often 90° or forward) | Signal directly proportional to particles; more sensitive than turbidimetry; immune complexes (IgG, IgA, IgM, CRP, C3); lasers often used |
| Turbidimetry | Decrease in transmitted light (straight through, 180°) caused by particles | Done on an ordinary spectrophotometer; particle size of standards and samples must match; settling causes error |
| Reflectance photometry | Light reflected from a colored dry pad/slide | Dry slides, urine strips, glucose meters; does not follow Beer's law; white reference = 100% reflectance; algorithm linearizes |
| Atomic absorption (AAS) | Light absorbed by ground-state atoms | Hollow-cathode lamp of the same element; flame breaks chemical bonds → free ground-state atoms and acts as the cuvette; chopper + tuned amplifier ignore flame emission; lanthanum stops phosphate interference in calcium; graphite furnace (flameless) more sensitive → lead; main method for trace and heavy metals (Pb, Cu, Zn, Mg, Ca, Li) |
| Flame emission photometry | Light emitted by atoms excited in a flame | Na (yellow 589 nm), K (violet 766 nm), Li (red 671 nm); lithium/cesium internal standard; largely replaced by ISE |
• Fluorometer principle = measures emitted light after excitation.
• Function of the flame in AAS = break chemical bonds → ground-state atoms.
• Heavy metals = AAS. Most sensitive troponin = chemiluminescence. HPLC detector = UV spectrophotometer. Chemiluminescent label = horseradish peroxidase (with luminol).
• Flow cytometry light source = laser.
Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q20–Q33 rationales (pp.177–179); SUCCESS! in Clinical Laboratory Science, Ch.1, §I.C–F (pp.3–5) and Q12–Q23 rationales (pp.146–149); BOC Study Guide, 6th ed., Ch.3 Chemistry, Q308, Q336, Q352 (pp.188–194); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 (pp.209–221); p.10, Q56–Q58; p.20, Q77.
⚡ Last-Minute Review — Photometry & optical methods
1. Light
- UV < 400 nm · Visible 400–700 nm · IR > 700 nm. Short wavelength = HIGH energy.
- Color seen = light transmitted. Solution absorbs the complementary color. Red solution absorbs 400–500 nm.
- Measure at the wavelength of maximum absorbance.
2. Beer's law
- A = abc (a = absorptivity, b = path cm, c = concentration). A ∝ c (linear).
- A = 2 − log %T. %T vs c = non-linear (inverse log).
- 100 %T = 0 A · 50 %T = 0.301 · 10 %T = 1.0 · 1 %T = 2.0.
- Needs monochromatic light, dilute solution, no interferents, solvent not absorbing.
- Same solution, more volume → same A. Absorbance has no unit.
3. Calculations
- Cu = Au/As × Cs. With serum blank: (Au − Asb)/As × Cs.
- Double concentration → double A. Above top standard → dilute, re-assay, × dilution factor.
- NADH ε = 6.22 × 10³ at 340 nm. U/L = ΔA/min × TV × 10⁶ / (6220 × b × SV).
- 1 IU = 1 µmol/min. Katal = mol/s. 1 U = 16.67 nkat.
4. Components
- Source → entrance slit → monochromator → exit slit → cuvette → detector → readout.
- Visible lamp: tungsten / tungsten-halogen. UV: deuterium/hydrogen. Line source: mercury. AAS: hollow-cathode. Flow cytometry: laser.
- Monochromator: filters (photometer), prism (best UV), diffraction grating (linear, most common). Job: only desired wavelength reaches sample.
- Bandpass = width at ½ peak; 540 nm ± 5 → 535–545. Narrower = better resolution/linearity.
- Glass cuvette = visible; quartz = UV. 1 cm path.
- Detectors: barrier-layer (Fe + Se, no amplification) < phototube < PMT (dynodes, most sensitive, DC power). Diode array = whole spectrum.
- Double beam / reference beam = corrects lamp fluctuation.
5. Blanks & calibration
- 0 %T with light blocked; 100 %T with blank.
- Reagent blank = reagent color only. Sample blank / bichromatic = hemolysis, icterus, lipemia.
- Primary standard ≥ 99.95% pure, not hygroscopic. Calibrator sera = secondary standards. Control ≠ calibrator.
- Standard curve: A (y) vs concentration (x). Low calibrator slope → controls/patients read high.
6. Endpoint vs kinetic
- Endpoint: one reading after completion (glucose oxidase, cholesterol, biuret). Fixed-time: two readings (kinetic Jaffe).
- Kinetic: many readings → ΔA/min; enzymes.
- Zero-order (substrate excess, ≫ Km) → rate ∝ enzyme → measure enzyme activity. First-order → rate ∝ substrate.
- Curve: lag → linear → substrate exhaustion → dilute sample.
- Km = [S] at ½ Vmax. Competitive inhibitor: Km ↑, Vmax same. Lineweaver–Burk x-intercept = −1/Km.
7. Checks & troubleshooting
- Wavelength: holmium oxide (UV/vis), didymium (585 nm). Stray light: sharp-cutoff filter. Absorbance accuracy: neutral-density filter, potassium dichromate.
- Stray light → falsely low A at high concentration → loss of linearity.
- Can't set 100 %T at 340 nm → old tungsten lamp → replace.
- Photometric tests "below detection" but electrolytes OK → fibrin in photometric probe.
8. Other optical methods
- Fluorometry: emitted longer-wavelength light, detector 90°, two monochromators, ~1000× more sensitive; quenching; ↑ temperature ↓ fluorescence.
- Chemiluminescence: light from chemical reaction, no lamp; acridinium, luminol + HRP, dioxetane + ALP; most sensitive troponin. ECL = ruthenium.
- Nephelometry = scattered light (angle), ∝ particles. Turbidimetry = loss of transmitted light.
- Reflectance (dry slides, strips): not Beer's law; white reference.
- AAS: ground-state atoms absorb hollow-cathode light; flame breaks bonds, acts as cuvette; chopper removes flame emission; lanthanum for Ca (phosphate); graphite furnace for lead; heavy metals.
- Flame emission: Na 589 yellow, K 766 violet, Li 671 red.
9. Performance terms
- Analytical sensitivity = detect small differences (slope + precision). LoD = lowest detectable. Analytical specificity = no interference.
- Diagnostic sensitivity = TP/(TP+FN); specificity = TN/(TN+FP).