1 · Start here — the words of reprocessing
Simple idea: every device that touches a patient can carry germs to the next patient. Reprocessing is everything we do to make a used item safe again: clean it, then disinfect or sterilize it, then keep it safe until it is used. If one step fails, the next patient is exposed.
| Word | Meaning in simple English | What it kills / removes |
|---|---|---|
| Cleaning | Removing visible dirt, blood, tissue and other organic or inorganic material, usually with water, detergent or enzymes and friction | Removes most germs physically (a 4–6 log fall on endoscopes) — kills nothing reliably |
| Decontamination | Making an item safe for staff to handle (cleaning ± a microbicidal step) | Enough to handle without risk |
| Low-level disinfection (LLD) | For items that touch intact skin | Vegetative bacteria, some fungi, enveloped viruses — not mycobacteria or spores |
| Intermediate-level disinfection (ILD) | Product with a tuberculocidal claim | Adds mycobacteria, most viruses and fungi — not spores |
| High-level disinfection (HLD) | For items that touch mucous membranes | All microorganisms except high numbers of bacterial spores |
| Sterilization | For items that enter sterile tissue or blood | All microorganisms, including spores |
| Antiseptic | An antimicrobial applied to skin (for example chlorhexidine) | Reduces skin flora — never used on devices |
Map of this topic
| # | Study section | APIC chapters | CIC domain(s) |
|---|---|---|---|
| 1 | The words of reprocessing | 31 | CSDA |
| 2 | Aseptic technique: surgical vs clean | 30 | CSDA, PCT |
| 3 | Spaulding classification | 31 | CSDA |
| 4 | Germ resistance and disinfection levels | 31 + CDC | CSDA, IDP |
| 5 | Cleaning: point of use, manual and mechanical, verification | 31, 108 | CSDA |
| 6 | High-level disinfectants and liquid chemical sterilants | 31, 108 | CSDA, EOH |
| 7 | Flexible endoscopes and probes | 31, 108 | CSDA, SEI |
| 8 | Sterilization methods | 31, 108 + CDC | CSDA, EOH |
| 9 | The sterile processing department | 108 | CSDA, EC, EOH |
| 10 | Packaging, loading, storage and distribution | 108 + CDC | CSDA |
| 11 | Monitoring sterilization: physical, chemical, biological | 108 + CDC | CSDA |
| 12 | Failures, recalls, IUSS, loaners and records | 31, 108 | CSDA, MC, SEI |
| 13 | Environmental surfaces and no-touch technologies | 31 + CDC | EC, CSDA |
| 14 | Special situations: prions, TASS, biofilm, single-use devices | 31, 32, 108 | CSDA, MC, IDP |
| 15 | Picture atlas | 31, 108 | CSDA |
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — 'Key Concepts', 'Cleaning', Table 31-1; Ch.30 Aseptic Technique — 'Key Concepts'.
2 · Aseptic technique — surgical asepsis vs clean technique
Asepsis means keeping disease-causing germs away from the patient. It protects the patient by keeping contamination as low as possible during care. There are two levels:
| Surgical asepsis (sterile technique) | Medical asepsis (clean technique) | |
|---|---|---|
| Goal | Keep germs at an irreducible minimum — no organisms moved to the site | Reduce the number of germs and stop spread between people or sites |
| Typical tools | Sterile gloves, sterile gown, mask, sterile drapes and a sterile field | Clean exam gloves, clean or disinfected equipment, "no-touch" dressing method |
| Examples | Surgery, central line insertion and central line dressing change, sharp bedside debridement, open tracheal suctioning | Routine wound cleaning, routine dressing changes without debridement, peripheral IV placement, closed-system suctioning; clean intermittent self-catheterization at home |
Surgical wounds
- CDC: a primarily closed incision is protected with a sterile dressing for 24–48 hours after surgery.
- The American College of Surgeons: sterile gloves for dressing changes in the first 24 hours.
- Whether clean or sterile gloves are needed for later routine dressing changes, tracheostomy care and line maintenance is an unresolved issue (the site may already be colonized).
Procedures outside the operating room
When a patient is too unstable to move (for example a trauma patient or a neonate), a procedure may be done at the bedside or in radiology. Reduce germs in the area: close doors, limit traffic and visitors, use screens, stop cleaning activity during the procedure, and use the best ventilation available.
In the operating room
- The sterile field is kept with sterile drapes, sterile gloves and gowns; the OR has positive pressure to corridors and more air changes.
- Scrubs are not PPE. Freshly laundered scrubs, head and facial hair cover and clean shoes are worn in semirestricted and restricted areas; masks in restricted areas when open sterile supplies are present.
- No long, artificial or enhanced nails for staff caring for high-risk patients (ICU, OR, oncology, neonates).
- Skin antisepsis: right agent for the procedure, check allergies and age limits, clean skin first, and allow the full contact and drying time.
Sources: APIC Text, Ch.30 Aseptic Technique — 'Key Concepts', 'Aseptic Technique', 'Clean Technique', 'Surgical Aseptic Technique Outside of the Operating Room', 'Attire', 'Skin Antisepsis', 'Single-use Devices, Equipment, and Supplies', 'Conclusions', Table 30-1.
3 · The Spaulding classification — how clean must it be?
Simple idea: the deeper an item goes into the body, the higher the risk, so the stronger the process must be. Earle Spaulding's system (more than 45 years old) is still the main rule for choosing disinfection or sterilization.
| Category | Touches… | Minimum process | Examples |
|---|---|---|---|
| Critical | Sterile tissue, bone or the vascular system | Sterilization | Surgical instruments, implants, cardiac and urinary catheters, ultrasound probes used in sterile body cavities |
| Semicritical | Mucous membranes or non-intact skin | High-level disinfection (minimum) | GI endoscopes, bronchoscopes, laryngoscopes, respiratory therapy and anesthesia equipment, vaginal and rectal probes, prostate biopsy probes, esophageal and anorectal manometry probes, diaphragm fitting rings |
| Noncritical | Intact skin only | Low-level disinfection (intermediate if visible blood) | Patient equipment: stethoscopes, blood pressure cuffs, bedpans, crutches, monitors, wheelchairs. Environmental surfaces: bed rails, bedside tables, counters, sinks, floors |
- Intact mucous membranes resist common bacterial spores but not bacteria, mycobacteria or viruses — that is why HLD (which may leave a few spores) is enough for semicritical items.
- Intact skin is a good barrier, so sterility is "not critical" for items that touch only skin.
- The Association for the Advancement of Medical Instrumentation (AAMI) adds environmental surfaces as a fourth category.
- Most noncritical items can be cleaned and disinfected where they are used; critical and many semicritical items go to a central processing area.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — 'Key Concepts', 'Background', 'Critical Items', 'Semicritical Items', 'Noncritical Items', 'Human Papilloma Virus (HPV)', Table 31-1.
4 · Germ resistance and the levels of disinfection
Simple idea: germs are not equally hard to kill. A process chosen for an item must kill the hardest germ that matters for that use.
⚑ The full resistance order above is taught from the CDC Guideline for Disinfection and Sterilization (2008); the APIC chapter states the parts used in Table 31-1 (what each level kills).
| Level | Kills | Does NOT reliably kill | Typical products (examples) | Used for |
|---|---|---|---|---|
| Sterilization | Everything incl. spores | Prions need special cycles | Steam, ethylene oxide, hydrogen peroxide gas plasma/vapor, ozone, dry heat, liquid chemical sterilants (long exposure) | Critical items |
| High-level | All microorganisms | High numbers of spores | Glutaraldehyde, ortho-phthalaldehyde (OPA), hydrogen peroxide 7.5%, improved hydrogen peroxide 2%, peracetic acid ± hydrogen peroxide, chlorine 650–675 ppm, pasteurization (65–77 °C, 30 min) | Semicritical items |
| Intermediate | Vegetative bacteria, mycobacteria, most viruses, most fungi | Spores | EPA-registered hospital disinfectant with a tuberculocidal claim (chlorine products, phenolics, improved hydrogen peroxide) | Noncritical items with visible blood |
| Low-level | Vegetative bacteria, some fungi and viruses | Mycobacteria, spores | Quaternary ammonium compounds ("quats"), phenolics, chlorine products, improved hydrogen peroxide, 70–90% alcohol | Noncritical items and surfaces |
Disinfectant resistance like antibiotic resistance is a common worry, but disinfectant concentrations are far above the killing level, and there are no data showing that multidrug-resistant organisms (MDROs) are less susceptible to the usual environmental disinfectants.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — Table 31-1, 'Contact Time for Disinfection of Noncritical Patient Care Equipment and Environmental Surfaces', 'Role of Environmental Surfaces in Disease Transmission', 'Factors that Affect the Manual Disinfection Process' · CDC: Guideline for Disinfection and Sterilization in Healthcare Facilities (2008) — Summary of Recommendations (reviewed Dec 2023); Guideline updates page (Apr 2019 C. difficile / EPA List K update).
5 · Cleaning — the most important step
Cleaning and rinsing is the most important step in reprocessing. It lowers the bioburden and removes the soil (protein, blood, salts) that shields germs from the sterilant. Cleaning alone gives a 4–6 log drop in germs on endoscopes.
1. At the point of use (the user's job)
- Start as soon as possible: wipe off gross soil (a sponge moistened with sterile water) and flush lumens.
- Keep items moist until cleaning — a pretreatment foam/gel, a towel wet with water (not saline), or a moisture-keeping container. Dried soil is hard to remove and makes disinfection or sterilization fail; it also helps biofilm form.
- Open hinged items and take multi-part items apart. Items opened but not used are still treated as contaminated.
- Separate reusable items from disposables, sharps and waste. Transport in a closed, puncture-resistant, leak-proof container labelled as biohazard.
2. Manual cleaning
- Two essential parts: friction (brushing, using the brush type and size in the instructions for use, IFU) and fluidics (fluid under pressure through channels).
- Brush under water to avoid splashing and aerosols. Single-use brushes → discard; reusable brushes → clean and disinfect at least daily.
- Measure the detergent dilution for the sink volume; monitor and record solution temperature and soak time; change the solution after each use.
- Rinse with tap water, then a final rinse of critical (treated) water for critical and semicritical items unless the IFU says otherwise; dry with air or a lint-free cloth.
3. Detergents
| Type | Key points |
|---|---|
| Neutral (pH ≈ 7) ± enzymes | Best material compatibility; best choice for delicate items and flexible endoscopes. Enzymes (proteases, lipases, amylases) break down protein, fat, starch |
| Alkaline | Dissolve protein and fat well but may corrode instruments |
| Enzymatic cleaners | Not disinfectants; must be rinsed off completely (residues can cause reactions such as fever); may cause asthma or allergy in staff |
4. Mechanical cleaning (more consistent than manual)
- Ultrasonic cleaner — sound waves create bubbles that collapse (cavitation) and pull fine soil out of joints and lumens. For fine cleaning after gross soil is removed; not a disinfection step. Used fluid can carry bacteria and endotoxin → change it after each use, empty and clean the tank at least daily, and rinse items well. Test cavitation daily when in use.
- Washer-disinfector / washer-decontaminator — like a dishwasher: wash, rinse, thermal or chemical microbicidal phase, drying. Open hinges, do not stack, heavy items at the bottom. Test the washer each day it is used, after major repairs and when changing detergent.
- Washer-pasteurizer — hot water (≈70 °C) for 30 min, used for respiratory therapy equipment.
5. Inspection and cleaning verification
- Minimum standard: every item is inspected and visibly clean before HLD or sterilization — with lighted magnification; a borescope camera looks inside channels.
- Rapid tests for residual soil: protein, hemoglobin, adenosine triphosphate (ATP), carbohydrate. AAMI benchmark for residual protein on instruments: < 6.4 µg/cm².
- ATP results are useful for quality monitoring, but each brand uses its own scale and ATP has not been shown to lower infection risk.
- Verify manual cleaning regularly (weekly; daily preferred for flexible endoscopes) using the hardest-to-clean items; report cleaning-monitoring results to the infection prevention committee.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — 'Cleaning'; Ch.108 Sterile Processing — 'Point-of-Use Cleaning', 'Transportation to Sterile Processing', 'Cleaning Used Items', 'Manual Cleaning Methods', 'Mechanical Cleaning Methods', 'Ultrasonic Cleaners', 'Washer-Disinfectors and Washer-Decontaminators', 'Cleaning Verification of Manual and Mechanical Cleaning Processes', 'Inspection of Instruments and Other Items'.
6 · High-level disinfectants and liquid chemical sterilants
Simple idea: the same chemical can be a high-level disinfectant (short exposure) or a sterilant (very long exposure). The time, temperature and concentration on the FDA-cleared label decide the result.
| Agent | HLD exposure (examples) | Good points | Problems / traps |
|---|---|---|---|
| Glutaraldehyde ≥ 2% | Label 20–90 min; professional societies: 20 min at 20 °C after good cleaning | Cheap, many studies, excellent material compatibility | Irritating vapor and odor (respiratory irritation), allergic dermatitis, slow on mycobacteria, coagulates blood and fixes tissue to surfaces |
| Ortho-phthalaldehyde (OPA) 0.55% | 12 min at 20 °C (5 min at 25 °C in an AER) | Fast, no activation, little odor, does not fix tissue | Stains protein grey (skin, mucosa, clothes); eye irritation; slow sporicidal; anaphylaxis after repeated cystoscopy in bladder-cancer patients; cleared as HLD only |
| Hydrogen peroxide 7.5% | 30 min | No odor, no disposal issue, does not fix tissue, kills Cryptosporidium | Can damage brass, zinc, copper, nickel/silver plating; serious eye damage on contact |
| Improved (accelerated) hydrogen peroxide 2% | 8 min at 20 °C | Hydrogen peroxide + surfactants in an acid formula → faster and better cleaning; no odor; non-staining; EPA toxicity category IV (practically nontoxic) | Limited clinical data; low concentrations (0.5–1.4%) are for low-level surface disinfection, 2% for HLD |
| Peracetic acid ± hydrogen peroxide | e.g. 15–25 min (mixtures) | No activation, low odor | Material damage (lead, brass, copper, zinc); eye and skin damage |
| Chlorine | 650–675 ppm, 10 min | Cheap, fast | Corrosive to metals |
- Most HLDs need 8–45 minutes at 20–25 °C. Raising the temperature in an automated endoscope reprocessor (AER) can shorten the time.
- Fill all tubing and channels completely — air pockets block contact.
- Test the minimum effective concentration (MEC) with the correct test strip before each use (manual or automated). If the strip fails, do not use the solution, even within its reuse life. Discard solution with precipitates.
- Monitor manual HLD with a thermometer and timer. Automated equipment without a recording (printout) device should not be used for HLD; review and initial the printout at the end of every cycle.
- Keep records: solution lot, dates opened/activated, reuse life, time–temperature–concentration, operator, scope number and patient.
- Pasteurization (hot water 65–77 °C for at least 30 min) is an acceptable, low-cost HLD for respiratory therapy and anesthesia equipment.
Liquid chemical sterilants (LCS)
- Most need 3–12 hours; the exception is 0.2% peracetic acid at 50–56 °C for about 12 minutes in an automated system (biological indicator: Geobacillus stearothermophilus).
- Limits: items cannot be wrapped, must be rinsed (sterile or filtered water per IFU) and used immediately; sterility is hard to keep in storage.
- So LCS are reserved for heat-sensitive critical items that cannot go through any other sterilization method.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — Table 31-1, Table 31-2, 'Critical Items', 'Semicritical Items', 'Improved Hydrogen Peroxide (HP)', 'Reprocessing of Endoscopes'; Ch.108 Sterile Processing — 'Washer-Pasteurizers', 'Automatic Endoscope Reprocessors', 'High-Level Disinfection' · CDC: Guideline for Disinfection and Sterilization — Summary of Recommendations (glutaraldehyde 20 min at 20 °C).
7 · Flexible endoscopes and probes
Why it matters: endoscopes have been tied to more healthcare outbreaks than any other device, including outbreaks of carbapenem-resistant Enterobacteriaceae (CRE). Scopes are long, narrow, heat-sensitive and full of channels — perfect for retained soil and biofilm. Most failures come from staff not following the reprocessing steps.
| Step | What to do | Why |
|---|---|---|
| 1. Pre-clean at the bedside | Wipe the insertion tube and suction/flush channels right after the procedure | Stops soil drying and biofilm forming |
| 2. Leak test | Before immersion | A leaking scope lets fluid and germs into inner parts |
| 3. Manual cleaning | Brush all channels; flush with enzymatic detergent and water; outside and inside | The most important step — removes the bioburden |
| 4. Visual inspection | Look for soil and damage (magnification, borescope); rapid soil tests if used | A dirty scope cannot be disinfected |
| 5. HLD (or sterilization) | Immerse and perfuse all channels for the labelled time (manual or AER) | Kills remaining germs |
| 6. Rinse | Sterile, filtered or tap water through all channels | Removes chemical residue |
| 7. Dry | Flush channels with 70–90% alcohol, then forced air | Moisture lets bacteria regrow |
| 8. Store | Hang vertically, not coiled, protected from contamination — never in the foam-lined carrying case | Promotes drying; prevents recontamination |
Program requirements (CDC/HICPAC essential elements)
- Initial and annual competency testing for every person who reprocesses scopes.
- Follow the scope's IFU; policies for loaner scopes; periodic quality audits.
- Record the scope used on each patient (scope ID + patient identifier) so patients can be found after a recall or a reprocessing failure.
- AERs standardize steps, reduce skipped steps and reduce staff exposure; check that the AER reaches every channel (test at least weekly, after repairs or suspected malfunction); test HLD concentration before each use; a failed printout → the scope is reprocessed before use.
- Contamination can persist even with full adherence; rapid indicator tests (blood, protein, carbohydrate) are recommended as part of quality assurance.
Endocavitary probes
Endovaginal and rectal probes are semicritical → HLD between patients even when a cover is used. Probe covers can fail and HPV has been found on covered probes. In one study, aldehyde HLDs showed little activity against HPV types 16 and 18, while a 35% hydrogen peroxide mist system inactivated them; an FDA-cleared automated hydrogen-peroxide mist system is available.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — 'Reprocessing of Endoscopes', 'Human Papilloma Virus (HPV)'; Ch.108 Sterile Processing — 'Manual Cleaning Methods', 'Automatic Endoscope Reprocessors', 'High-Level Disinfection'.
8 · Sterilization methods — which one, and why?
Rule: critical items are bought sterile or steam sterilized. Steam under pressure is the oldest, most effective and preferred method. Heat- or moisture-sensitive items use a low-temperature method.
| Method | Typical cycle | Advantages | Disadvantages / traps | Biological indicator |
|---|---|---|---|---|
| Steam | Gravity 121 °C, 30 min (wrapped instruments); dynamic-air-removal (prevacuum) 132 °C, 4 min — plus drying time | Nontoxic, cheap, fast, penetrates packs and lumens, least affected by soil, easy to monitor | Damages heat-sensitive items; wet packs if not dried; burns | Geobacillus stearothermophilus |
| Ethylene oxide (EO/ETO) | Long (≈15 h incl. aeration) | Penetrates packaging and lumens; compatible with most materials | Toxic, carcinogenic, flammable; needs aeration to remove residue; OSHA-regulated (29 CFR 1910.1047); slow | Bacillus atrophaeus |
| Hydrogen peroxide gas plasma | 28–52 min, < 50 °C | No toxic residue, no aeration, simple to install | No cellulose (paper), linens or liquids; synthetic packaging only (polypropylene wrap, polyolefin pouches); lumen length/diameter limits | G. stearothermophilus |
| Vaporized hydrogen peroxide | ≈55 min | No residue or aeration; heat- and moisture-sensitive items | Lumen limits; no cellulose, linen, liquids, powders | G. stearothermophilus |
| Ozone (± hydrogen peroxide) | ≈4 h | Made from oxygen and water; no aeration | Limited clinical data | G. stearothermophilus |
| Dry heat | 1–6 h (temperature-dependent) | For items steam cannot penetrate or that steam corrodes | Slow; high temperatures | B. atrophaeus |
| 0.2% peracetic acid (liquid, automated) | ≈12 min at 50–56 °C | Fast, low temperature, standardized, harmless by-products | Immersible items only; unwrapped; use immediately (point-of-use) | G. stearothermophilus |
Steam sterilizer types
- Gravity displacement: steam pushes air out through a drain at the bottom — slower; a Bowie-Dick test is NOT used.
- Dynamic air removal (prevacuum or steam-flush pressure-pulse): air is pulled out by vacuum or pulses — faster; needs a daily Bowie-Dick test (Section 11).
- Tabletop sterilizers (≤ 2 cubic feet) are easily overloaded and have their own monitoring rules.
Choosing a method
Consider: is the method validated in the device IFU, turnaround needs, volume, material compatibility, space, safety and cost. If IFUs conflict (instrument vs sterilizer vs packaging), ask the instrument manufacturer — it must validate the method it recommends. Do not change cycles, contents or packaging unless the device maker validated the change.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — Table 31-1, Table 31-3, 'Critical Items'; Ch.108 Sterile Processing — 'Key Terms', 'Background', 'Sterilization', 'Selection of Sterilization Methods', 'Biological Indicators' · CDC: Guideline for Disinfection and Sterilization — Table 7 'Minimum cycle times for steam sterilization cycles' (reviewed Dec 2023).
9 · The sterile processing department (SPD)
Sterile processing (SP) handles, collects, transports, sorts, takes apart, cleans, disinfects, inspects, packages, sterilizes, stores and distributes reusable items. AAMI and the Association of periOperative Registered Nurses (AORN) encourage one centralized department and the same standard everywhere items are reprocessed (OR, clinics, endoscopy, offices).
Design rules
- One-way flow: decontamination (dirty) → preparation/packaging (clean) → sterilization → sterile storage.
- A physical barrier (wall) separates decontamination from clean work. In small settings where separate rooms are not possible, the decontamination sink is at least 4 feet from the clean area, or behind a wall/screen at least 4 feet high above the sink rim (AORN: as wide as the counter).
- Air moves from clean areas toward decontamination (decontamination is negative). Temperature, humidity and air changes follow ANSI/ASHRAE/ASHE Standard 170; variances are reviewed by a team that includes the IP.
- Floors, walls, ceilings and work surfaces: non-shedding, cleanable, non-porous. Decontamination: ideally a three-compartment sink (wash, first rinse, final rinse) and a critical-water source for final rinse.
- Eyewash/shower reachable within 10 seconds wherever corrosive chemicals are used.
- Hand-hygiene sinks and alcohol-based hand rub easily available. A breakout room removes items from outside shipping cartons before they enter clean or sterile areas.
People
- Restricted areas: authorized people only; scrubs laundered by a healthcare-accredited laundry, changed daily or when soiled; head and facial hair covered; dedicated shoes; visitors wear shoe covers.
- No rings, watches or bracelets; nails no longer than 2 mm past the fingertip; no nail polish or artificial nails; no food, drink, lip balm or contact lens handling.
- Decontamination PPE: fluid-resistant gown with sleeves (or jumpsuit), liquid-resistant shoe covers or boots, fluid-resistant mask, eye protection (goggles or face shield), and heavy-duty, waterproof utility gloves with cuffs that extend over the gown cuffs.
- Education, competency testing, and certification of SP staff (mandated in some states, e.g. New Jersey first in 2004).
Cleaning the department
SP should be as clean as the OR: work surfaces and floors cleaned and disinfected at least daily; case carts and transport carts after each use; walls, vents and shelves on a schedule.
Sources: APIC Text, Ch.108 Sterile Processing — 'Key Concepts', 'Basic Principles', 'The Physical Environment for Sterile Processing', 'Safety Requirements for Sterile Processing Staff', 'Disinfecting Sterile Processing Work Areas'.
10 · Packaging, loading, storage and distribution
Packaging
- Packaging must let air out and sterilant in, then block germs until use. Options: woven or nonwoven wraps, peel pouches (small, light items), rigid sterilization containers. Use packaging cleared for that sterilization method and cycle.
- Instruments open, unlocked, disassembled and dry. Basins all facing the same way with non-linting absorbent material between nested basins.
- Instrument sets (including the container or wrap) should weigh no more than 11 kg (25 lb).
- Close wraps with sterilization indicator tape — never pins, clips or staples. Write only on the tape or label, never on the wrap.
- Peel pouches: seal properly, expel air; write on the plastic side (not the paper); no pouches inside trays or containers and no double-pouching unless the manufacturer validated it (if allowed: paper to paper, plastic to plastic, inner pouch not folded).
- Each package has a lot-control label (sterilizer number, cycle/load number, date) and an event-related expiry statement ("sterile unless the package is opened or damaged").
Loading a steam sterilizer
- One load = items with the same cycle parameters.
- Textile packs on the upper shelf; metal trays and rigid containers below, so condensate does not drip onto absorbent packs.
- Trays and containers flat (horizontal); basins tilted on edge, all one way; containers stacked only if the same maker allows it.
- Peel pouches on edge in a rack, paper side facing plastic side — never laid flat.
- EO: full loads of items with the same aeration time, loose in baskets, not touching the walls.
Unloading
- Verify the cycle printout. Let the load cool on the cart in a low-traffic area away from air-conditioning vents before touching (warm packs + cool air = condensation).
- Wet, torn, stained, dropped or unsealed packs and packs whose external indicator did not change are not used — they go back to decontamination for full reprocessing.
Storage and distribution
- Shelf life is event-related: an item stays sterile until something happens to the package (tear, moisture, soil, tampering) — not because a date passes. CDC: packaged sterile items can be used indefinitely unless the packaging is compromised.
- Store in a limited-access, enclosed area next to sterilization, with controlled humidity; on shelves with a solid bottom shelf; closed or covered shelving preferred; away from floor, ceiling, outside walls, sinks and exposed pipes; heavy trays on middle shelves; first in, first out.
- Transport in covered or enclosed carts with a solid bottom shelf; never drag or crush packs; clean carts after each use; keep sterile items separate from contaminated items; never leave loaded vehicles unattended.
Sources: APIC Text, Ch.108 Sterile Processing — 'Preparation and Packaging', 'Packaging of Healthcare Facility–Prepared Sets or Trays', 'Product Identification and Traceability', 'Loading the Sterilizer', 'Unloading the Sterilizer and Sterile Storage', 'Storage and Shelf Life', 'Event-Related Sterility', 'Distribution' · CDC: Guideline for Disinfection and Sterilization — Summary of Recommendations (storage of sterile items).
11 · Monitoring sterilization — physical, chemical, biological
Simple idea: we cannot see sterility, so we check it three ways. Each monitor answers a different question.
| Monitor | Question it answers | Key rules |
|---|---|---|
| Physical (time, temperature, pressure printouts, gauges) | Did the machine reach the cycle settings? | Read and initial every load; the first tool to detect a failure; a sterilizer with no recording device must not be used; wrong readings → do not release the load, tell the supervisor |
| Chemical indicator (CI) | Was this pack exposed to the right conditions? | Detects packaging, loading or machine problems. A pass does not prove the item is sterile |
| Biological indicator (BI) — spores | Did the process actually kill highly resistant spores? | The only monitor that directly measures killing (lethality); cannot be replaced by any CI |
Chemical indicator types (AAMI)
| Type | Name | Use |
|---|---|---|
| 1 | Process indicator | External (e.g. indicator tape) on every pack unless the internal CI is visible — shows the pack was processed |
| 2 | Bowie-Dick | Air-removal test for dynamic-air-removal steam sterilizers |
| 3 | Single-variable | Responds to one critical variable |
| 4 | Multivariable | Two or more critical variables |
| 5 | Integrating | All critical variables; correlates with BI performance — used inside PCDs and for implant loads |
| 6 | Emulating | All critical variables of a specific cycle |
- Internal CI in every facility-prepared pack, placed where the sterilant reaches last (for EO not necessarily the centre). AORN prefers Type 5 or 6 inside every package. Read it before the contents go on the sterile field; a failed internal CI → return the unused pack (with CI and label) to SP.
- External CIs are read when unloading, when issuing, and before opening.
Bowie-Dick test (Type 2)
- Checks air removal in dynamic-air-removal (prevacuum) steam sterilizers, including immediate-use cycles — not gravity sterilizers.
- Every day the sterilizer is used, before the first load, in an empty, pre-heated chamber, near the drain; 3.5–4 min at 134 °C.
- Fail → take out of service; repeat once (air in the line can cause a false fail). Pass on retest → back in service. Fails twice → shut down, service, qualification testing.
Biological indicators
- Geobacillus stearothermophilus: steam, hydrogen peroxide vapor/plasma, hydrogen peroxide + ozone, liquid peracetic acid. Bacillus atrophaeus: EO and dry heat.
- Each day a test BI is incubated, incubate a positive control from the same lot (not sterilized) — it must grow. If the control does not grow, the test result is not valid.
- BIs are used inside a process challenge device (PCD) (formerly "test pack"), placed in the hardest spot — usually the front, bottom of the chamber near the drain.
| Situation | BI frequency |
|---|---|
| Steam — routine sterilizer efficacy | At least weekly (CDC and AAMI); daily preferred (AAMI/AORN). Test the shortest cycle of each cycle type |
| Any load containing an implant | Every load: BI PCD with a Type 5 CI; quarantine the implant until the BI is negative — early release only in a documented emergency, traceable to the patient |
| Ethylene oxide | Every load |
| Hydrogen peroxide vapor / + ozone | At least daily for each cycle type; every load preferred |
| Qualification testing (after installation, relocation, major repair, malfunction, process failure, utility change) | Steam dynamic-air-removal: BI PCD in an empty chamber for 3 consecutive cycles, then 3 consecutive Bowie-Dick tests. Tabletop: 3 cycles in a full load. Return to service only when all BIs negative, controls positive, CIs and physical monitors correct |
Sources: APIC Text, Ch.108 Sterile Processing — 'Sterilization Process Monitoring', 'Physical Monitors', 'Chemical Indicators', 'Bowie-Dick Testing for Dynamic-Air-Removal Steam Sterilizers', 'External Chemical Indicators', 'Internal Chemical Indicators', 'Biological Indicators', 'Routine Load Release of Implant Loads', 'Routine Sterilizer Efficacy Monitoring', 'Qualification Testing', 'EO Sterilizers', 'Low-Temperature Gaseous Chemical Sterilizers', 'Reporting Results of Sterilization Process Monitoring' · CDC: Guideline for Disinfection and Sterilization — Summary of Recommendations (biological indicators at least weekly; every load with implants).
12 · Failures, recalls, immediate-use sterilization, loaners and records
When a monitor shows a failure
| Situation | Action |
|---|---|
| Cause found at once and limited to one load or one item (wrong cycle chosen, one non-responsive internal CI) | Correct the error; inspect, re-clean if needed, rewrap with new indicators, reprocess that load/item |
| Cause not immediately known (most failures) — positive BI, failed CI or physical monitor | Take the sterilizer out of service; recall and reprocess everything processed since the last negative BI; find the root cause; qualification testing before reuse; written recall report; surveillance of exposed patients |
Note: the CDC guideline adds that for steam, a single positive BI with no sterilizer or user problem found may not require recall of non-implant items; repeat the test — if it is positive again, recall. Implants are always quarantined until the BI is negative.
The recall team usually includes the IP, risk manager, surgical services manager and the surgeons/physicians of the involved patients.
Exposure investigation after a reprocessing failure (APIC Table 31-5, simplified)
Immediate-use steam sterilization (IUSS — formerly "flash")
- Riskier than SP: under time pressure, cleaning (the most critical and most often shortened step) and drying get skipped.
- Only for urgent clinical situations (e.g. a dropped instrument needed now) — not for convenience and not to replace buying enough instruments.
- Only if the instrument IFU includes an IUSS cycle; clean exactly as in SP; use a rigid container validated for IUSS; transfer aseptically and use immediately — never store or hold for another case.
- Not for implants except a documented emergency with no other option; record in the patient's chart. Document the reason for every IUSS and use the log to reduce IUSS.
Loaner instruments
- Written loaner program with the vendor. Existing loaner trays arrive at least 2 working days (48 h) before the case; a set borrowed for the first time 3 working days before.
- Late arrival is not a reason for IUSS. Do not add instruments to loaner trays.
Records and traceability
- For each cycle: lot number, date, time, contents, exposure time and temperature, operator, BI and Bowie-Dick results, CI in the PCD, aeration (EO). Keep maintenance and repair logs.
- Every implant must be traceable to the patient; scope HLD records must be traceable to the patient.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — Table 31-5 'Protocol for Exposure Investigation after the Failure to Follow Disinfection and Sterilization Principles'; Ch.108 Sterile Processing — 'Recalls', 'Immediate-Use Steam Sterilization', 'Loaner Instruments', 'Documentation/Record Keeping' · CDC: Guideline for Disinfection and Sterilization — Summary of Recommendations (positive spore tests).
13 · Environmental surfaces and no-touch technologies
Why it matters: MRSA, VRE, Acinetobacter, norovirus, C. difficile and now Candida auris survive on surfaces for days to months, contaminate hands and cause outbreaks. Touching the environment contaminates hands about as often as touching the patient. Being admitted to a room whose previous patient carried one of these germs raises the new patient's risk.
- Patient rooms are cleaned and disinfected on a regular schedule, when visibly soiled, and at discharge (terminal cleaning), usually with an EPA-registered hospital disinfectant such as a quat.
- Thoroughness matters most: studies using fluorescent marking found only about half of room surfaces were cleaned at terminal cleaning (49%; ICU ≈ 57%). Wipe all hand-contact surfaces, not only a short "high-touch" list.
- Assign responsibility (e.g. nursing cleans patient equipment such as monitors; environmental services cleans surfaces) so nothing is missed.
- Improvement programs: education + monitoring (fluorescent marker or ATP) + feedback + checklists. A room that only looks clean may still carry germs.
- Problems that reduce disinfection: wrong dilution, organic soil, poor distribution (rags vs microfiber), dirty buckets, solution reused in too many rooms, too little time.
- Improved hydrogen peroxide surface products: fast (EPA contact times of 30 s–1 min for bacteria), low toxicity; used for surfaces and privacy curtains.
No-touch room decontamination (after cleaning, at terminal clean only)
| Ultraviolet-C (UV-C, ≈254 nm) | Hydrogen peroxide vapor / mist | Copper surfaces | |
|---|---|---|---|
| Speed | Faster (minutes; longer for spores) | Slower (> 4 times longer than routine cleaning) | Continuous, but kills over hours |
| Strengths | No residue; no need to seal room or shut HVAC | Most sporicidal; uniform distribution; shown to lower C. difficile and MDRO acquisition | No equipment to move; works all the time on the covered surface |
| Limits | Line of sight — shadowed surfaces get less; organic matter lowers effect; no published proof of fewer HAIs | HVAC off, doors sealed; sensitive to concentration | Only the copper surface; lower log kill |
| Common to UV and HP | Room must be empty of patients and staff (so terminal only); cleaning must come first (they do not remove dust or soil); high capital cost — they supplement, never replace manual cleaning | ||
COVID-19 and surfaces
Coronaviruses are enveloped and easy to kill. Use a product on EPA List N at its labelled contact time; clean and disinfect high-touch surfaces daily and portable equipment per its IFU.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — 'Role of Environmental Surfaces in Disease Transmission', 'Factors that Affect the Manual Disinfection Process', 'Contact Time for Disinfection of Noncritical Patient Care Equipment and Environmental Surfaces', 'Improving Room Cleaning and Disinfection', 'New Technologies for Environmental Surface Disinfection', Table 31-4, 'Coronavirus Disease 2019 (COVID-19)' · CDC: Guideline for Disinfection and Sterilization — Updates (Dec 2009 fogging clarification; page reviewed Jun 2024).
14 · Special situations — prions, TASS, biofilm and single-use devices
Prions (Creutzfeldt-Jakob disease, CJD)
- Prions resist normal disinfection and sterilization; CJD is always fatal (about 1–1.5 cases per million people per year in the US). FDA does not allow prion claims on labels.
- Special processing only when BOTH are present: a high-risk patient (known or suspected CJD/prion disease) AND a critical or semicritical device contaminated with high-risk tissue (brain, spinal cord, eye). Ch.108 adds: treat instruments contaminated with cerebrospinal fluid as highly infectious.
- Keep instruments moist after use; clean thoroughly; then, for example, steam at 134 °C for 18 min (prevacuum) or 132 °C for 1 h (gravity); sodium hydroxide (1N NaOH) combinations follow strict protocols — never put instruments soaking in NaOH straight into a steam sterilizer (staff hazard, sterilizer damage).
- Devices that are impossible or hard to clean → discard. Cover surfaces with plastic-backed paper; spot-clean contaminated surfaces with 1:10 bleach.
Toxic anterior segment syndrome (TASS)
- An acute, noninfectious inflammation of the front of the eye, usually 12–48 hours after uneventful cataract surgery; looks like endophthalmitis but improves with steroids.
- Causes: detergent residues, endotoxin, impurities in steam, antiseptics, talc, ultrasonic bath contaminants, irritants on intraocular instruments.
- Prevention: follow IFUs exactly; keep instruments moist; clean them separately from other instruments; final rinse with sterile distilled/deionized water; discard cleaning tools and solution after each use; surveillance for TASS cases.
Biofilm
- Masses of bacteria in a slime matrix stuck to wet surfaces (dental waterlines, dialysis systems, catheters, endoscope channels). Bacteria in biofilm can be up to 1,000 times more resistant than free bacteria.
- Best prevention: prompt cleaning and reprocessing so biofilm never has time to form. No products are FDA-cleared or EPA-registered specifically to remove biofilm.
Reprocessing single-use devices (SUDs)
- Under the FDA's 2000 guidance, anyone who reprocesses an SUD (hospital or third party) must meet every regulatory requirement that applies to the original device maker (registration, reporting, quality system, labelling, premarket review).
- Because this is costly, most hospitals use FDA-regulated third-party reprocessors; the Centers for Medicare & Medicaid Services (CMS) encourages third-party reprocessing over in-house programs.
- Choose a reprocessor after reviewing its FDA clearance and visiting its site. The FDA reviews validation data (cleaning, sterilization, function) and decides whether the reprocessed device is "substantially equivalent".
- Device classes: Class I lowest risk/least regulated → Class III highest risk/most regulated. "Opened-but-unused" devices are a gray area left to the hospital.
- A 2008 Government Accountability Office report found reprocessed SUDs posed no greater risk than new devices. IPs should be part of any reprocessing decision.
Independent study material written from the APIC Text and public CDC guidance. Not affiliated with or endorsed by CBIC or APIC. Questions are original practice items, not examination questions.
Sources: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — 'Inactivation of Creutzfeldt-Jakob Disease Agent (CJD)', 'Toxic Anterior Segment Syndrome (TASS)', 'Biofilms'; Ch.108 Sterile Processing — 'Managing Items Contaminated with Prions', 'Special Precautions to Prevent Toxic Anterior Segment Syndrome', 'Third-Party Reprocessing'; Ch.32 Reprocessing Single-Use Devices — 'Key Concepts', 'Background', 'Conclusions'.
15 · Picture atlas — reading reprocessing at a glance
Much of reprocessing is checked by looking: the category of an item, the Bowie-Dick sheet, the integrator strip, the spore vials, the way a sterilizer cart is loaded, the air flow in the department and the state of a pack. Learn the key feature of each picture; the picture questions in the quiz use the same drawings.
Classify and choose
The department and the load
Monitors
Packs
Drawings: original teaching diagrams made for this page (no copyrighted images). Facts verified with: APIC Text, Ch.31 Cleaning, Disinfection, and Sterilization — 'Critical Items', 'Semicritical Items', 'Noncritical Items'; Ch.108 Sterile Processing — 'The Physical Environment for Sterile Processing', 'Loading the Sterilizer', 'Unloading the Sterilizer and Sterile Storage', 'Bowie-Dick Testing for Dynamic-Air-Removal Steam Sterilizers', 'Internal Chemical Indicators', 'Biological Indicators', 'Recalls' · CDC: Guideline for Disinfection and Sterilization (2008).
⚡ Last-Minute Review — Asepsis, Cleaning, Disinfection, Sterilization & Sterile Processing
1. Key concepts
- Cleaning always comes first — soil blocks disinfectants and sterilants; cleaning is the most important step.
- Spaulding: critical (sterile tissue/blood) → sterilize · semicritical (mucous membranes, non-intact skin) → HLD minimum · noncritical (intact skin) → LLD. AAMI adds environmental surfaces.
- Failure to disinfect or sterilize properly = transmission by devices; environmental cleaning lowers bioburden.
2. Levels and resistance
- Prions → spores → mycobacteria → small non-enveloped viruses → fungi → vegetative bacteria → enveloped viruses.
- LLD: no TB, no spores · ILD: tuberculocidal claim · HLD: all but high numbers of spores · Sterilization: all.
- Surfaces: ≥ 1 min contact (studies), follow EPA label; C. diff → sporicide (List K); COVID → List N.
3. Aseptic technique
- Surgical asepsis = no organisms moved; clean technique = fewer. Invasive device placement → aseptic technique.
- Sterile dressing 24–48 h on closed incisions; scrubs are not PPE; check packs for moisture, tears, discoloration, expiry.
4. Cleaning
- Point of use: wipe, flush, keep moist with water (not saline); closed biohazard container.
- Manual = friction + fluidics; neutral enzymatic detergent for scopes; enzymes are not disinfectants; rinse well.
- Ultrasonic = cavitation, fine cleaning only; change fluid each use (endotoxin). Washers tested every day of use.
- Visibly clean minimum; protein < 6.4 µg/cm²; ATP scales vary. Never clean instruments in a hand sink.
5. HLD and endoscopes
- Glut 2%: 20 min at 20 °C (fixes tissue, irritant) · OPA 12 min (grey stain) · H2O2 7.5% 30 min · improved H2O2 2% 8 min · pasteurization 65–77 °C ≥ 30 min.
- Test MEC before each use; failed strip → discard. AER printout failure → reprocess.
- Scopes = most outbreaks. Pre-clean → leak test → manual clean → inspect → HLD → rinse → alcohol + air dry → hang vertically. Competency initial + annual; scope ↔ patient log.
- Endovaginal probe with cover = still semicritical → HLD.
6. Sterilization methods
- Steam preferred: gravity 121 °C 30 min; prevacuum 132 °C 4 min (+ drying).
- EO: toxic, carcinogen, flammable, aeration, OSHA. H2O2 plasma: no cellulose/linen/liquids, synthetic wrap, lumen limits. 0.2% peracetic acid 12 min at 50–56 °C (unwrapped, use now).
- BI: G. stearothermophilus (steam, peroxide, ozone, peracetic) · B. atrophaeus (EO, dry heat).
7. SPD, packaging, storage
- One-way flow; wall between decon and clean; air toward decon; 4 ft / 4-ft screen in small settings; eyewash ≤ 10 s; nails ≤ 2 mm, no jewelry.
- Sets ≤ 25 lb; tape not pins; write on plastic side of pouches; no pouch-in-tray unless validated.
- Cart: linen top, metal bottom, pouches on edge; cool before handling; wet/torn/dropped → reprocess.
- Event-related shelf life; solid bottom shelf; away from sinks; FIFO; covered carts.
8. Monitoring
- Physical = machine (first alert) · CI = pack (Type 1 external, Type 2 Bowie-Dick, Type 5 integrator) · BI = kill (only lethality monitor).
- Bowie-Dick: prevacuum only, daily, empty chamber; fail twice → service + qualification.
- Steam BI ≥ weekly (daily preferred); every implant load (BI PCD + Type 5, quarantine); every EO load; positive control same lot must grow.
- Qualification: 3 BI PCD cycles empty + 3 Bowie-Dick.
9. Failures, IUSS, loaners
- Unknown cause → out of service, recall back to last negative BI. Known single-load error → fix and reprocess.
- Failure protocol: confirm → embargo items → stop the unit → stakeholders → cause → line list → risk → … → notify? → after-action.
- IUSS: urgent only, cleaning is the weak step, rigid container, use at once, document reason; no implants except emergencies.
- Loaners 2 working days early (3 if new); late loaner ≠ reason for IUSS.
10. Special situations
- Prions: high-risk patient + high-risk tissue → 134 °C 18 min prevac or 132 °C 1 h gravity; hard to clean → discard; surfaces 1:10 bleach.
- TASS: sterile eye inflammation 12–48 h after cataract surgery → review cleaning/rinsing/steam.
- No-touch UV/HP: empty room, after cleaning, supplement only. Terminal cleaning thoroughness ≈ 50%.
- SUD reprocessing: reprocessor = manufacturer (FDA); most use third-party reprocessors.
Sources: APIC Text, Ch.30, 31, 32 and 108 (Key Concepts and sections cited in Study Mode) · CDC Guideline for Disinfection and Sterilization (2008): Summary of Recommendations (reviewed Dec 2023), Table 7 steam cycle times (reviewed Dec 2023), Updates page (Jun 2024).
Independent study material written from the APIC Text and public CDC guidance. Not affiliated with or endorsed by CBIC or APIC. Questions are original practice items, not examination questions.
🖼️ Picture clues (Study Section 15)
- Spaulding ladder: sterile tissue → sterilization; mucous membrane → HLD; intact skin → LLD.
- Bowie-Dick: even colour = air removed; pale centre = air pocket → out of service, retest.
- Integrator: bar in ACCEPT → pack conditions met; bar in REJECT → do not use, return to SP.
- Spore vials: test no growth + control growth = pass; test growth = failure; control no growth = invalid test.
- Cart: linen and on-edge pouches on top, metal trays flat below.
- SPD: one-way flow dirty → clean → sterile; air flows toward decontamination.
- Packs: wet, torn or unchanged indicator tape → do not use; reprocess.