METHOD EVALUATION COURSE · PART I · CHAPTER 01

Fit for Purpose: Verification, Validation and the Life of a Test

Error and the margin for error · Intended use · Validation vs verification · ISO 15189:2022 clauses 7.3.1–7.3.3 · The life of a test · Verify or validate? · What the evidence must cover · Criteria first · Worked example: the error budget

Course text from Clinical Laboratory Method Evaluation in Practice (1st ed., 2026) · Original practice questions · Works offline · questions · Version 1.0

📘 Study progress

Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 key point · 📌 remember · ⚠️ common error · 💡 understand this · 🧪 at the bench · 🧠 memory aid.

Start here: the question on your desk

🔥 THE QUESTION ON YOUR DESK: Your laboratory has bought a new chemistry analyser. The vendor's engineer has installed it, run the calibrators and handed over a folder of performance claims. The emergency department wants the analyser in service next week. Before the first patient result leaves the laboratory, what must you prove, and how will you know you have proved enough?

Chapter 0 has already made sure that the analyser itself is identified, accepted and in service for evaluation. This chapter answers the question in general terms for the examination methods that run on it. The rest of the course answers it study by study.

Chapter 0: Is the equipment ready? → Chapter 1: Do we verify or validate, and what must we prove? → Chapters 2–20: the studies, one by one
🧪 THE DECISION THIS CHAPTER SUPPORTS: Do we verify or validate, and what must we prove before the first patient result? (Main references named by the book: ISO 15189:2022; ISO 9000:2015; CLSI EP19-Ed3.)

By the end of this chapter you should be able to

  1. State the intended use of a test in enough detail to judge its performance.
  2. Tell verification from validation, and decide which one a situation needs.
  3. Place any evaluation within the life of a test, from development to retirement.
  4. List the performance characteristics that a quantitative or qualitative evaluation must cover.
  5. Explain why acceptance criteria must be written down before any data are collected.

Map of this chapter

#Study sectionBook section(s)Printed pages
1Start here: the question on your deskChapter 1 opening, learning objectives27
2Every result carries error; how much is acceptable?§1.128
3Intended use is the yardstick§1.2 (Table 1.1)28–29
4Two words, two jobs: validation and verification§1.3 (definitions, EP19, ISO 15189 entries)29–30
5What ISO 15189:2022 requires: clauses 7.3.1–7.3.3§1.3 (clauses, Table 1.2, vocabulary)30–31
6The life of a test: two stages, and choosing a candidate§1.4 (first part)31–32
7Each phase ends with a decision (Figure 1.1)§1.4 (Figure 1.1, three features)33–34
8Verify or validate? The decision path (Figure 1.2)§1.5 (Figure 1.2)34
9Common situations and the evaluation each needs§1.5 (Table 1.3)35
10What the evidence must cover§1.6 (Table 1.4)35–38
11Criteria first, data second§1.7 (Table 1.5); Requirements beyond ISO 1518938–39
12Worked Example 1.1 (part 1): screening the claimsWorked Example 1.1, Situation and Step 139–40
13Worked Example 1.1 (part 2): the laboratory's own dataWorked Example 1.1, Step 2 (Figure 1.3), interpretation, decision40–41
14Common errors of interpretationCommon errors of interpretation41–42
15Chapter 1 checklist and referencesChapter 1 checklist; Chapter 1 references42–43
📌 REMEMBER: Two questions run through the whole chapter. Which kind of evaluation does this situation need (verification or validation)? And what must the evidence show, judged against limits written before testing?

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, opening and learning objectives, p.27; How this book is organised, p.viii

1.1 Every result carries error; the evaluation decides how much is acceptable

No measurement is exact. Run the same serum twice and you get two slightly different numbers. Run it on two analysers and the gap may be larger.

🔥 KEY POINT: These differences are not mistakes. They are a property of every measurement procedure, and they never reach zero.

The useful question

Not this questionBut this question
"Is this method free of error?""Is the error small enough that a clinician who acts on the result will make the same decision they would make on the true value?"

A method evaluation exists to answer that second question with data.

The answer depends on how the result will be used

A potassium result far inside the reference interval can carry a fair amount of error without changing anything. The same error in a result close to the level at which a clinician would start treatment can change what happens to the patient.

💡 UNDERSTAND THIS: The method has not changed; the stakes have. That is why every evaluation starts from the intended use (§1.2) and from limits set before testing (§1.7).

The margin for error

An evaluation also tells you something you will need later. Once you know how much error the method actually produces, you can compare it with how much error the test can tolerate.

Error the test can tolerate − error the method actually produces = margin for error
MarginWhat it means for routine quality control (QC)
WideLets you run a light QC plan
NarrowNeeds closer watching

Worked Example 1.1 (study sections 12–13) shows this margin with numbers, and Chapter 18 turns it into a QC plan.

📌 REMEMBER: Error is normal. The evaluation asks whether it is small enough for this use, and how much room is left.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.1 Every result carries error; the evaluation decides how much is acceptable, p.28

1.2 Intended use is the yardstick

The intended use is a short, specific statement of what the test is for. Every later choice is judged against it:

  • which studies to run,
  • at which concentrations,
  • with which samples, and
  • how strict the limits should be.
⚠️ COMMON ERROR: A vague intended use ("to measure troponin") gives you nothing to judge against. A specific one tells you where the method must perform best.

Write the intended use so that it answers the questions in Table 1.1.

Table 1.1 Elements of an intended-use statement
ElementQuestion to answerWhy it matters for the evaluation
MeasurandWhat exactly is measured, and in what units?Fixes what "correct" means and which reference materials apply
SpecimenWhich sample types, containers and anticoagulants?Each specimen type may behave differently and may need its own evidence
PopulationAdults, children, neonates, pregnant women, dialysis patients?Interferences, reference intervals and decision points differ by population
Clinical purposeScreening, diagnosis, ruling out, monitoring or prognosis?Sets the decision points and how much error each purpose can tolerate
Decision pointsWhich concentrations trigger a clinical action?The evaluation must show performance at these points, not only on average
Setting and userCentral laboratory, satellite laboratory, ward or clinic; trained or occasional users?Point-of-care use by non-laboratory staff adds operator and environment effects
Reporting needsTurnaround time, measuring interval, units, critical valuesSome needs (for example, a wide measuring interval) are performance requirements in themselves
🧠 MEMORY AID: Measurand · Specimen · Population · Purpose · Decision points · Setting and user · Reporting needs — seven questions, one statement.

Same assay, different evaluation priorities

Two laboratories can install the same troponin assay for different purposes:

Laboratory 1Laboratory 2
UseRule out heart muscle injury in the emergency departmentFollow known cardiac patients over days
Where decisions turnLow concentrations near a cut-offChange between results taken at different times
Evidence it needs mostPrecision and bias at low concentrationsConsistency between results taken at different times

Same assay, same instrument, different evaluation priorities.

A change of intended use is a change of method

🔥 KEY POINT: CLSI EP19 lists a move from adult to paediatric use, or from diagnostic to screening use, among the modifications that send a method back for re-establishment of its performance. Section 1.5 returns to this point.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.2 Intended use is the yardstick (Table 1.1), pp.28–29

1.3 Two words, two jobs: validation and verification

Both words mean confirming something with objective evidence. They differ in what is being confirmed.

WordConfirms that…The question it asksISO 9000:2015 entry
Validationa method meets the requirements of a particular intended useIs this the right method for the job?3.8.13
Verificationspecified requirements have been metDoes this method do what was promised?3.8.12

In a medical laboratory, "specified requirements" usually means the performance claims that the developer of the method has published.

📌 REMEMBER: Verification checks claims. Validation creates or re-creates the evidence on which claims rest.

CLSI EP19 gives the two words a place in time

CLSI EP19 (3rd edition, 2022) places the two words in the life of a method:

Developer establishes the method → ends with validation (extensive testing: analytical and, where needed, clinical performance for the intended use) → End-user laboratory implements it → verification (confirms, under its own conditions, that the method performs as claimed)
🔥 KEY POINT: EP19 is clear that the developer need not be a manufacturer. A laboratory that designs its own test, or changes an approved commercial test, becomes the developer for that test.

ISO 15189:2022 draws the same line

ISO 15189:2022, the accreditation standard for medical laboratories, defines both terms in the language of ISO/IEC 17000:2020 (entries 3.31 and 3.32):

  • Validation confirms that a method is plausible for a specific intended use.
  • Verification confirms that specified requirements have been fulfilled.
💡 UNDERSTAND THIS: A note to the verification entry adds that verification alone may be enough to introduce a new IVD device, provided the examination is performed and used as its instructions direct.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.3 Two words, two jobs: validation and verification, pp.29–30

What ISO 15189:2022 requires: clauses 7.3.1–7.3.3

Clause 7.3.1 — the frame for both

Methods must have been validated for their intended use, and each method's performance specifications must relate to that use and to its effect on patient care.

Clause 7.3.2 — methods used as designed (verification)

  • Before such a method goes into use, the laboratory follows a procedure showing that it can achieve the performance stated by the manufacturer or by the method.
  • The specifications it confirms are the ones that matter for the intended use.
  • The work must be extensive enough to support the clinical decisions made from the results.
  • Authorised, competent staff review the results and record whether they meet the requirements.
  • If the issuing body revises the method, the laboratory repeats verification as far as the revision requires.
  • The records show: the specifications to be achieved · the results obtained · whether the specifications were met (and, if not, what was done).

Clause 7.3.3 — three kinds of method (validation)

#Kind of methodExamples given
1A method the laboratory designs or develops—
2A method used outside its originally intended scopeOutside the instructions for use or the validated measuring range; third-party reagents on an instrument they were not intended for, when no validation data exist
3A validated method that is later modified—
  • Validation must go as far as necessary to show, through performance specifications, that the requirements of the intended use are met.
  • The clinical impact of any proposed change to a validated method is reviewed before the change is adopted.
  • The records include: the procedure used · the requirements for the intended use · the performance specifications determined · the results · a statement of the method's fitness for its intended use.
Table 1.2 Validation and verification compared
ValidationVerification
Core questionIs the method fit for this intended use?Does the method perform as claimed, here?
Who normally does itThe developer: a manufacturer, or a laboratory that develops or modifies a methodThe end-user laboratory
Starting pointDesign requirements derived from the intended useThe developer's published performance claims
Typical scaleLarger studies, often several reagent lots, instruments and sitesSmaller studies in the laboratory's own conditions
OutputPerformance claims and the evidence behind themA documented decision that the claims hold locally
ISO 15189:2022 clause7.3.37.3.2
🧠 MEMORY AID: Verification = 7.3.2 (the smaller number, the smaller study, checking someone else's claims). Validation = 7.3.3 (creating the claims).

One caution on vocabulary

⚠️ COMMON ERROR: Usage is not uniform. EP19 notes that some laboratories, especially in microbiology, call the routine monitoring of a test in service "validation". Older documents and some training material also assign the words differently. When you read any procedure, check what its author means by each word before you follow it. This course uses the meanings in Table 1.2 throughout.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.3 Two words, two jobs: validation and verification (Table 1.2), pp.30–31

1.4 The life of a test: two stages, and choosing a candidate

Every test method passes through the same broad stages, whoever made it. CLSI EP19 calls this the Test Life Phases Model. Knowing where your work sits in that life tells you what kind of evidence is expected.

StageBelongs toPhases
1. EstablishmentThe developerFeasibility and design → development → validation
2. ImplementationThe end-user laboratoryPreliminary evaluation → verification → launch → maintenance (a long phase in routine service) → retirement

Establishment: the design is frozen before validation

At the end of development the design is frozen, so that what is validated is exactly what users will receive. The developer then hands over the method with its instructions for use and its performance claims.

Implementation: from preliminary evaluation to retirement

  • Preliminary evaluation — EP19 describes it as the point where the laboratory checks that a method suits its needs and screens for grossly unacceptable performance before the formal verification.
  • Verification — the formal confirmation of the claims under the laboratory's own conditions.
  • Launch — everything else needed before patient testing: procedures, training and competence, order codes, reference intervals, critical values and information-system set-up.
  • Maintenance — the long phase in routine service.
  • Retirement — the end of the test's life.

Before any formal study: choose a candidate worth evaluating

EP19 advises selecting against specifications set in advance (such as intended use, performance, cost per test, complexity and throughput) and preferring methods registered as medical devices.

It helps to sort what you know about a candidate into three groups, because each group is judged in a different way:

GroupThe questionThink ofHow it is judged
Fit with the laboratoryCan the method run here at all?Specimen types and volume, turnaround time, workload, staffing and training, space, utilities, safety and costFrom documents and site visits, not from data
Soundness of the measurement principleIs the chemistry or technology likely to give good results?How specific the reaction is for the measurand; how the method is calibrated and to what reference it is traceable; how robust the procedure is to small changesFrom the instructions for use, the literature and expert review
Demonstrated performanceHow well does the method actually perform?Precision, trueness, measuring interval, detection capability, interference and the other characteristics in Table 1.4Only data can answer these questions — the rest of this course
📌 REMEMBER: A method that fails the first group never needs the third. A method that passes the first two still has to earn its place with data.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.4 The life of a test, pp.31–32

Each phase ends with a decision (Figure 1.1)

Figure 1.1: the life of a test in two bands. Establishment by the developer: feasibility and design, development, validation. Transfer of instructions for use and claims. Implementation by the laboratory: preliminary evaluation, verification, launch, maintenance, retirement, with a decision gate after each phase and a dashed arrow from maintenance back to establishment for a modification.
Figure 1.1 The life of a test: two stages, eight phases, a decision after each. Original diagram · phase names after CLSI EP19-Ed3 (2022).
💡 HOW TO READ THIS FIGURE: Read the upper band from left to right: this is the developer's work, ending in validation and the transfer of claims. The arrow then drops to the lower band, the laboratory's work, which runs from preliminary evaluation to retirement. Each small diamond is a decision point where the team proceeds, repeats the phase or stops. The dashed arrow shows the one way back: a modification made while the method is in service returns it to the establishment stage. The two highlighted boxes (validation and verification) are the phases most of this course is about.

Three features of this model that matter in daily practice

  1. Each phase ends with a decision. The team proceeds, repeats work or stops, by comparing results with criteria set in advance. EP19 asks that the decision, the date and the people who made it are recorded.
  2. A modification sends the method back. If the laboratory changes a validated method, the method returns to the establishment stage for the changed aspects. The laboratory then acts as developer for those aspects.
  3. Maintenance is not the end of evaluation. Quality control, external quality assessment (EQA), reagent lot checks and comparisons between analysers keep confirming, in service, what verification first showed.

ISO 15189:2022 clause 7.3.7 — ensuring the validity of examination results

ClauseWhat it requires
7.3.7.1A planned procedure for monitoring results in a way that shows trends and shifts
7.3.7.2Internal quality control
7.3.7.3External quality assessment
7.3.7.4Where different methods, instruments or sites perform the same examination, a procedure for showing that their results on patient samples are comparable across clinically significant intervals

Part IV of the course covers that work.

🧠 MEMORY AID: Decide · Go back · Keep checking — every phase ends in a decision; a modification goes back to establishment; maintenance keeps confirming.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.4 The life of a test (Figure 1.1), pp.33–34

1.5 Verify or validate? A decision path

Most laboratories verify far more often than they validate. The deciding question is simple:

🔥 KEY POINT: Are you using a method exactly as its developer validated it, for the purpose it was validated for? If yes, you verify. If anything about the method, its specimen, its population or its purpose departs from what was validated, you validate that departure, because no one else has.
Figure 1.2: decision path. Intended use written down; did the laboratory develop the method (yes: validate in full); has the laboratory changed the method (yes: validate the change); used within its validated scope (no: validate for the new scope); all routes lead to verify the claims under routine conditions, then record and sign the decision.
Figure 1.2 Deciding whether a method needs verification or validation. Original diagram · based on CLSI EP19-Ed3 and ISO 15189:2022 clauses 7.3.2–7.3.3.
💡 HOW TO READ THIS FIGURE: Start at the top and answer the three questions in order. A "yes" to either of the first two, or a "no" to the third, leads right to a validation box: the laboratory must create the missing evidence itself. Every route, including the validation routes, ends at verification under routine conditions, because a validated method must still be shown to work in daily practice. Only then is the decision recorded and signed.

The path as a table

Question (in order)Answer that leads to validationValidation box
1. Did the laboratory develop the method?YesValidate in full — the laboratory is the developer
2. Has the laboratory changed the method?YesValidate the change — re-establish the changed aspects
3. Used within its validated scope?NoValidate for the new scope — specimen, population or purpose
All routes → Verify the claims under routine conditions → Record and sign the decision
📌 REMEMBER: Verify what was validated; validate any departure from it — then verify before launch.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.5 Verify or validate? A decision path (Figure 1.2), p.34

Common situations and the evaluation each needs

Table 1.3 applies the path to situations a hospital laboratory meets. The examples are illustrative. Your local regulator or accreditation body may set further rules (§1.7).

Table 1.3 Common situations and the evaluation each needs
SituationEvaluation neededWhy
New analyser running the manufacturer's approved assays as instructedVerificationThe method is used exactly as validated; the laboratory confirms the claims locally
A second analyser of the same model added for the same assaysVerification of the new analyser, plus a comparison between analysersPatients may be tested on either system, so results must agree (Chapter 16)
An approved serum assay used on a body fluid not listed in the instructions for useValidation for the new specimen typeA new specimen matrix is a modification; the developer's evidence does not cover it
The laboratory changes the sample volume, dilution or incubation time set by the manufacturerValidation of the changed methodA modified method returns to the establishment stage
An in-house method, such as a liquid chromatography–mass spectrometry drug assayFull validation, then implementationThe laboratory is the developer; no external claims exist
A point-of-care meter used in a patient group the instructions for use do not includeValidation for that groupUse outside the validated scope; interferences and decision points may differ
The manufacturer revises the method (for example, a new reagent formulation)Verification repeated to the extent the change requiresThe claims changed; a risk assessment sets how much evidence is needed
A new lot of the same reagentNeither: a reagent lot change checkRoutine maintenance work, not a new method (Chapter 16)

Two points worth stating plainly

🔥 KEY POINT 1: Validation does not replace verification. A laboratory that validates its own method still has to show the method works in routine conditions, with routine staff, before launch.
🔥 KEY POINT 2: The decision is per method and per change. One analyser can host verified assays and a validated in-house assay side by side.
⚠️ COMMON ERROR: Notice who made the change. When the laboratory changes a method (row 4), it validates the change. When the manufacturer revises it (row 7), the laboratory repeats verification to the extent the change requires. A new lot of the same reagent (row 8) is neither.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.5 Verify or validate? A decision path (Table 1.3), p.35

1.6 What the evidence must cover

A performance characteristic is one measurable aspect of how a method behaves: its random error, its systematic error, the range over which it reads correctly, and so on.

🔥 KEY POINT: An evaluation does not have to measure every characteristic. It has to measure the ones the intended use depends on, to the depth the situation needs.
What end users commonly verify (EP19)What a developer's validation adds
Precision · trueness or bias · accuracy · analytical sensitivity and specificity · the measuring interval · the reference intervalCharacteristics a user rarely has the materials or time to establish, such as detection capability, reagent and sample stability across lots, and, for many tests, clinical performance

Table 1.4 summarises the characteristics, what each tells you, and the CLSI document that describes how to study it. The last column shows the chapter of this course that covers it.

Table 1.4 Performance characteristics and where they are studied
CharacteristicWhat it tells youIn a verificationMain CLSI documentChapter
PrecisionHow much results scatter on repeated testing of the same sampleCommonly checkedEP05-A3 (establish); EP15-A3 (verify)6
Trueness (bias) against reference materialsHow far the average result sits from an assigned valueCommonly checkedEP15-A37
Comparison with patient samplesHow patient results differ from the current or a comparative method, across the range and at decision pointsCommonly checkedEP09c-Ed38
Linearity and measuring intervalOver what range results are proportional to concentration and can be reported directlyCommonly checked against the claimed intervalEP06-Ed2; EP34-Ed1 for dilution9
Detection capabilityThe lowest concentrations that can be detected and quantified reliablyChecked when low results drive decisionsEP17-A210
Analytical specificity (interference)Which substances or sample conditions change the result, and from what levelChecked when the use or population calls for it; otherwise the claims are reviewedEP07-Ed3; EP37-Ed111
CarryoverWhether one sample or reagent affects the next resultSample carryover often screened early; reagent carryover usually the developer's workEP10-Ed4 (sample); EP47-Ed1 (reagent)12
Reference intervalWhether the interval suits the local population and methodCommonly checkedEP28-A3c13
Measurement uncertaintyThe range within which the true value is expected to lieEstimated by the laboratory; reviewed with verification dataEP29-A; ISO 15189:2022 clause 7.3.414
Qualitative agreementHow often a positive/negative result agrees with the comparator; precision near the cut-offCommonly checked for binary testsEP12-Ed315
Reagent stabilityHow long reagents, calibrators and controls keep their performance in storage and once openedNormally established by the developer; a laboratory may confirm an in-use claimEP25-Ed25
⚠️ COMMON ERROR: The column headed "In a verification" describes common practice, not a requirement. The laboratory decides the scope of each verification from the intended use and from a risk assessment, then records its reasons.

An example: stability

  • EP25-Ed2 covers the stability of reagents, calibrators and control materials, not of patient specimens.
  • Its intended users are mainly manufacturers and regulators (§1.1 of EP25).
  • It notes that laboratories usually check a manufacturer's stability claims only indirectly, through QC trends (§2.2).
  • It also describes how a laboratory can confirm a stated in-use claim, for example when a method is first verified or a new lot arrives (§6.1, §6.3).
📌 REMEMBER: Choose the characteristics from the intended use and a risk assessment, and write down why.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.6 What the evidence must cover (Table 1.4), pp.35–38

1.7 Criteria first, data second

🔥 KEY POINT: Write the acceptance criteria before you collect a single result. EP19 asks that every evaluation start from a written plan that states how the study will be done and the criteria for judging it.

The reason is human, not bureaucratic. Once numbers are on the screen, it is very easy to find a reason why a borderline result is acceptable. A limit written in advance cannot be bent to fit the data.

Four kinds of numbers — keep them apart

A plan usually contains numbers from four different sources. Keep them apart, because they carry different weight.

Table 1.5 Four kinds of numbers in an evaluation plan
KindExampleWhat it isHow to use it
ISO 15189 requirementVerify an unmodified method before routine use (clause 7.3.2)A "shall" statement that an accredited laboratory must meetMeet it, and cite the clause in the plan
Manufacturer's performance claim"Within-laboratory CV 1.2% at 5.0 mmol/L" in the instructions for useWhat the developer showed under its own conditionsThe thing a verification tests; not a statement of what the patient needs
CLSI recommendationA study design or statistical approach in an EP documentConsensus good practice, written as "should"Follow it, or record why you departed from it
Laboratory acceptance criterion"Bias at 7.0 mmol/L no more than 0.2 mmol/L"The laboratory's own limit, derived from the intended useSet it before testing; record where it came from (Chapter 4)
💡 UNDERSTAND THIS: Note the difference between the second and fourth rows. A method can meet every manufacturer's claim and still fail the laboratory's criterion, because the claim describes the method while the criterion describes the patient's need. A sound plan tests both.
🧠 MEMORY AID: Shall · Claim · Should · Our limit — requirement, manufacturer, CLSI, laboratory.

Decide in advance what happens if a criterion is missed

The plan should also say what happens if a criterion is not met. Options include:

Repeat a study after finding a fault · contact the manufacturer · restrict the method's use · reject it

Deciding this in advance avoids pressure to launch a method that has not earned it.

Record the decision

EP19 lists what the record should hold: the study results, the acceptance criteria, the decision, the date and the people who made it, kept as a permanent record. Chapter 20 and Templates A–C give a structure.

Requirements beyond ISO 15189

ISO 15189 is the international baseline, and it is the requirement document used throughout this course. It is rarely the only one. Depending on where a laboratory works, health regulators, licensing authorities and accreditation bodies may add rules of their own:

  • which standard they accredit against,
  • minimum study designs,
  • extra evidence for particular tests, or
  • how long records must be kept.

These rules differ between jurisdictions and change over time, so no book can list them reliably.

🧪 AT THE BENCH: Before planning an evaluation, find out which rules apply to your laboratory. Cite each by title and version in the plan, and treat it as a requirement in the sense of Table 1.5. Where a local rule is stricter than anything in this course, the local rule applies.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.7 Criteria first, data second (Table 1.5); Requirements beyond ISO 15189, pp.38–39

Worked Example 1.1 (part 1): How much room does the method leave?

All data and limits in this example are illustrative and synthetic.

Situation

The laboratory is choosing a sodium method for the new analyser. For this intended use it has set, in advance, an allowable total error of 4.0 mmol/L at 140 mmol/L (2.9%). Chapter 4 explains how to set such a limit; here it is simply given.

Step 1 — Screen the claims (preliminary evaluation)

The instructions for use claim a within-laboratory CV of 0.9% and a bias of +0.5% at 140 mmol/L. Convert both to mmol/L:

QuantityCalculationResult
SD0.9% × 1401.26 mmol/L
Bias0.5% × 1400.70 mmol/L

A simple estimate of the largest error expected in 95% of single results combines the two. CLSI EP21 describes this model, in which the absolute bias is added to 1.96 times the within-laboratory SD, and notes its limits.

TAE = |bias| + 1.96 × SD
TAE = 0.70 + 1.96 × 1.26 = 0.70 + 2.47 = 3.17 mmol/L

TAE = estimated total analytical error (mmol/L) · |bias| = absolute bias (mmol/L) · SD = within-laboratory standard deviation (mmol/L) · 1.96 = the multiplier for 95% of single results in this model.

The margin for error is the allowable limit minus the estimate:

Margin = 4.0 − 3.17 = 0.83 mmol/L, about 21% of the limit
🧪 DECISION (Step 1): The claims fit inside the limit with a modest margin. Proceed to verification.
📌 REMEMBER: Convert percentages to the result's unit at the concentration of interest before you combine them: SD = CV × concentration; bias (mmol/L) = bias (%) × concentration.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, Worked Example 1.1 How much room does the method leave? (Situation, Step 1), pp.39–40

Worked Example 1.1 (part 2): the laboratory's own data

Step 2 — Compare with the laboratory's own data (verification)

Two outcomes are possible. Suppose the precision and bias studies (Chapters 6–8) give the results below.

Worked Example 1.1 — error budget for sodium at 140 mmol/L (allowable total error 4.0 mmol/L)
SD (mmol/L)Bias (mmol/L)1.96 × SD (mmol/L)Estimated TAE (mmol/L)Margin (mmol/L)
Claims1.26+0.702.473.170.83
Scenario A1.20+0.902.353.250.75
Scenario B1.75+0.903.434.33−0.33

The arithmetic: Scenario A — 1.96 × 1.20 = 2.35; 0.90 + 2.35 = 3.25; 4.0 − 3.25 = 0.75. Scenario B — 1.96 × 1.75 = 3.43; 0.90 + 3.43 = 4.33; 4.0 − 4.33 = −0.33.

Figure 1.3: horizontal bar chart of the error budget for sodium at 140 mmol/L. Manufacturer claims: bias 0.70 plus 2.47, total 3.17, margin 0.83. Scenario A: 0.90 plus 2.35, total 3.25, margin 0.75. Scenario B: 0.90 plus 3.43, total 4.33, exceeding the 4.0 mmol/L limit by 0.33.
Figure 1.3 The error budget for sodium at 140 mmol/L: bias plus random error against the allowable limit. Synthetic, illustrative data · Worked Example 1.1 · sodium at 140 mmol/L.
💡 HOW TO READ THIS FIGURE: Each bar is the estimated largest error in 95% of single results, built from two parts. The dark segment is the absolute bias; the light segment is 1.96 times the within-laboratory SD. The horizontal axis is in mmol/L, the same unit as the result. The dashed vertical line is the limit set before testing, 4.0 mmol/L. The distance from the end of a bar to that line is the margin for error. Scenario B's bar crosses the line, and the red tip shows by how much. Notice that bias is the same in A and B: the failure comes entirely from the wider scatter.

Interpretation

  • Scenario A: the local bias is a little larger than claimed, but the total still fits inside 4.0 mmol/L. The margin of 0.75 mmol/L is narrow, which tells the laboratory to plan its quality control with care (Chapter 18).
  • Scenario B: the bias is the same, but the scatter is larger. The estimate of 4.33 mmol/L exceeds the limit, so a meaningful share of single results could be wrong by more than the intended use allows.

Decision

ScenarioDecisionNext action
AAcceptRecord the narrow margin as a risk to monitor
BDo not accept yetInvestigate why the precision is worse than claimed (Chapter 17). Common causes: a fault on the analyser, a problem with the control material, or a study design that captured more sources of variation than the manufacturer's did

What this example does not prove

⚠️ LIMITS OF THE MODEL: The simple model assumes that bias is the same at every concentration and that errors follow a normal distribution. It also ignores interference and non-linearity. Whether an observed SD of 1.75 mmol/L really differs from a claim of 1.26 mmol/L is a statistical question that Chapter 6 answers. Different sources also use different multipliers for the SD; whichever you choose, state it in the plan (Chapter 14).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, Worked Example 1.1 How much room does the method leave? (Step 2, Figure 1.3, interpretation, decision), pp.40–41

Common errors of interpretation

The errorWhy it is wrong
⚠️ "The manufacturer validated it, so we have nothing to do."The developer's validation shows what the method can do under the developer's conditions. Only your verification shows what it does with your staff, samples and environment.
⚠️ "It met the claims, so it is fit for purpose."Claims describe the method, not the patient's need. Judge the results against criteria derived from the intended use as well (Table 1.5).
⚠️ Treating a modification as a verification.A new specimen type, a changed dilution or use in a new patient group is outside the validated scope. A verification-sized study cannot cover it.
⚠️ Setting the limits after seeing the data.A limit chosen once the numbers are known is not an acceptance criterion. It is a description of the result.
⚠️ Judging performance only on average.A method can be acceptable overall and unacceptable at the one concentration that drives a clinical decision. Check performance at each decision point.
⚠️ Reading another document's vocabulary as your own."Validation" and "verification" are used in different senses in different documents and departments. Confirm the meaning before following a procedure.
⚠️ Assuming one evaluation lasts forever.Revised methods, relocated analysers and changes in use all call for a fresh look, scaled by a risk assessment.
📌 REMEMBER: Most of these errors come from mixing up what the method can do (claims, validation) with what the patient needs (intended use, criteria set in advance).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, Common errors of interpretation, pp.41–42

Chapter 1 checklist and references

Before planning any study, confirm that you can tick each item.

☐Checklist itemSee
☐The intended use is written down: measurand, specimen, population, clinical purpose, decision points, setting and reporting needsTable 1.1
☐You have decided whether the method is used exactly as validated (verify) or departs from its validated scope (validate), and recorded whyFigure 1.2, Table 1.3
☐Any modification has been identified, and the laboratory accepts that it is the developer for the changed aspects§1.4, §1.5
☐The performance characteristics to study are chosen from the intended use and a risk assessmentTable 1.4
☐Each number in the plan is labelled by its source: requirement, manufacturer's claim, CLSI recommendation or laboratory criterionTable 1.5
☐Acceptance criteria and the action to take if one is missed are written, signed and dated before data collection§1.7
☐Regulatory and accreditation requirements that apply to the laboratory have been checked and cited by title and version§1.7
☐The person who will make and sign the final decision is named§1.7
Chapter 1: Verify or validate, and what must we prove? → Chapter 2: the language of error the studies use

Chapter 1 references (as cited by the book)

  1. Clinical and Laboratory Standards Institute. A Framework for Using CLSI Documents to Evaluate Medical Laboratory Test Methods. 3rd ed. CLSI report EP19-Ed3. CLSI; 2022. Sections checked: Introduction to the Test Life Phases Model; Modifications to Established Test Methods; Acceptance Criteria; Documentation of Decisions and Actions; Validation Phase; Implementation Stage (Preliminary Evaluation, Test Method Selection, Verification, Launch, Maintenance); Definitions.
  2. International Organization for Standardization. ISO 9000:2015 Quality management systems — Fundamentals and vocabulary. ISO; 2015. Entries 3.8.12 (verification) and 3.8.13 (validation). Entry numbers confirmed from the IEC Electropedia record for validation (IEV 192-01-18).
  3. International Organization for Standardization. ISO 15189:2022 Medical laboratories — Requirements for quality and competence. 4th ed. ISO; 2022 (December). Entries 3.31 and 3.32; clauses 7.3.1–7.3.4 and 7.3.7.
  4. Clinical and Laboratory Standards Institute. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures. 2nd ed. CLSI guideline EP21-Ed2. CLSI; 2016. Chapter 1, background to the total analytical error model and its limitations.

CLSI documents named in Table 1.4 (all published by CLSI)

  • EP05-A3. Evaluation of Precision of Quantitative Measurement Procedures. 3rd ed. 2014.
  • EP06-Ed2. Evaluation of Linearity of Quantitative Measurement Procedures. 2nd ed. 2020.
  • EP07-Ed3. Interference Testing in Clinical Chemistry. 3rd ed. 2018.
  • EP09c-Ed3. Measurement Procedure Comparison and Bias Estimation Using Patient Samples. 3rd ed. 2018.
  • EP10-Ed4. Preliminary Evaluation of Quantitative Medical Laboratory Measurement Procedures. 4th ed. 2024.
  • EP12-Ed3. Evaluation of Qualitative, Binary Output Examination Performance. 3rd ed. 2023.
  • EP15-A3. User Verification of Precision and Estimation of Bias. 3rd ed. 2014.
  • EP17-A2. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. 2nd ed. 2012.
  • EP28-A3c. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory. 3rd ed. 2010.
  • EP29-A. Expression of Measurement Uncertainty in Laboratory Medicine. 1st ed. 2012.
  • EP34-Ed1. Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking. 1st ed. 2018.
  • EP47-Ed1. Evaluation of Reagent Carryover Effects on Test Results. 1st ed. 2024.
  • EP25-Ed2. Evaluation of Stability of In Vitro Medical Laboratory Test Reagents. 2nd ed. 2023. Sections 1.1, 2.2, 6.1 and 6.3 read.
  • EP37-Ed1. Supplemental Tables for Interference Testing in Clinical Chemistry. 1st ed. 2018.

CLSI documents are cited for attribution only. Readers should obtain the current edition from CLSI for the full procedures. Citation does not imply endorsement by CLSI, ISO, any accreditation body or any manufacturer.