Start here: the question on your desk
Chapter 0 has already made sure that the analyser itself is identified, accepted and in service for evaluation. This chapter answers the question in general terms for the examination methods that run on it. The rest of the course answers it study by study.
By the end of this chapter you should be able to
- State the intended use of a test in enough detail to judge its performance.
- Tell verification from validation, and decide which one a situation needs.
- Place any evaluation within the life of a test, from development to retirement.
- List the performance characteristics that a quantitative or qualitative evaluation must cover.
- Explain why acceptance criteria must be written down before any data are collected.
Map of this chapter
| # | Study section | Book section(s) | Printed pages |
|---|---|---|---|
| 1 | Start here: the question on your desk | Chapter 1 opening, learning objectives | 27 |
| 2 | Every result carries error; how much is acceptable? | §1.1 | 28 |
| 3 | Intended use is the yardstick | §1.2 (Table 1.1) | 28–29 |
| 4 | Two words, two jobs: validation and verification | §1.3 (definitions, EP19, ISO 15189 entries) | 29–30 |
| 5 | What ISO 15189:2022 requires: clauses 7.3.1–7.3.3 | §1.3 (clauses, Table 1.2, vocabulary) | 30–31 |
| 6 | The life of a test: two stages, and choosing a candidate | §1.4 (first part) | 31–32 |
| 7 | Each phase ends with a decision (Figure 1.1) | §1.4 (Figure 1.1, three features) | 33–34 |
| 8 | Verify or validate? The decision path (Figure 1.2) | §1.5 (Figure 1.2) | 34 |
| 9 | Common situations and the evaluation each needs | §1.5 (Table 1.3) | 35 |
| 10 | What the evidence must cover | §1.6 (Table 1.4) | 35–38 |
| 11 | Criteria first, data second | §1.7 (Table 1.5); Requirements beyond ISO 15189 | 38–39 |
| 12 | Worked Example 1.1 (part 1): screening the claims | Worked Example 1.1, Situation and Step 1 | 39–40 |
| 13 | Worked Example 1.1 (part 2): the laboratory's own data | Worked Example 1.1, Step 2 (Figure 1.3), interpretation, decision | 40–41 |
| 14 | Common errors of interpretation | Common errors of interpretation | 41–42 |
| 15 | Chapter 1 checklist and references | Chapter 1 checklist; Chapter 1 references | 42–43 |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, opening and learning objectives, p.27; How this book is organised, p.viii
1.1 Every result carries error; the evaluation decides how much is acceptable
No measurement is exact. Run the same serum twice and you get two slightly different numbers. Run it on two analysers and the gap may be larger.
The useful question
| Not this question | But this question |
|---|---|
| "Is this method free of error?" | "Is the error small enough that a clinician who acts on the result will make the same decision they would make on the true value?" |
A method evaluation exists to answer that second question with data.
The answer depends on how the result will be used
A potassium result far inside the reference interval can carry a fair amount of error without changing anything. The same error in a result close to the level at which a clinician would start treatment can change what happens to the patient.
The margin for error
An evaluation also tells you something you will need later. Once you know how much error the method actually produces, you can compare it with how much error the test can tolerate.
| Margin | What it means for routine quality control (QC) |
|---|---|
| Wide | Lets you run a light QC plan |
| Narrow | Needs closer watching |
Worked Example 1.1 (study sections 12–13) shows this margin with numbers, and Chapter 18 turns it into a QC plan.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.1 Every result carries error; the evaluation decides how much is acceptable, p.28
1.2 Intended use is the yardstick
The intended use is a short, specific statement of what the test is for. Every later choice is judged against it:
- which studies to run,
- at which concentrations,
- with which samples, and
- how strict the limits should be.
Write the intended use so that it answers the questions in Table 1.1.
| Element | Question to answer | Why it matters for the evaluation |
|---|---|---|
| Measurand | What exactly is measured, and in what units? | Fixes what "correct" means and which reference materials apply |
| Specimen | Which sample types, containers and anticoagulants? | Each specimen type may behave differently and may need its own evidence |
| Population | Adults, children, neonates, pregnant women, dialysis patients? | Interferences, reference intervals and decision points differ by population |
| Clinical purpose | Screening, diagnosis, ruling out, monitoring or prognosis? | Sets the decision points and how much error each purpose can tolerate |
| Decision points | Which concentrations trigger a clinical action? | The evaluation must show performance at these points, not only on average |
| Setting and user | Central laboratory, satellite laboratory, ward or clinic; trained or occasional users? | Point-of-care use by non-laboratory staff adds operator and environment effects |
| Reporting needs | Turnaround time, measuring interval, units, critical values | Some needs (for example, a wide measuring interval) are performance requirements in themselves |
Same assay, different evaluation priorities
Two laboratories can install the same troponin assay for different purposes:
| Laboratory 1 | Laboratory 2 | |
|---|---|---|
| Use | Rule out heart muscle injury in the emergency department | Follow known cardiac patients over days |
| Where decisions turn | Low concentrations near a cut-off | Change between results taken at different times |
| Evidence it needs most | Precision and bias at low concentrations | Consistency between results taken at different times |
Same assay, same instrument, different evaluation priorities.
A change of intended use is a change of method
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.2 Intended use is the yardstick (Table 1.1), pp.28–29
1.3 Two words, two jobs: validation and verification
Both words mean confirming something with objective evidence. They differ in what is being confirmed.
| Word | Confirms that… | The question it asks | ISO 9000:2015 entry |
|---|---|---|---|
| Validation | a method meets the requirements of a particular intended use | Is this the right method for the job? | 3.8.13 |
| Verification | specified requirements have been met | Does this method do what was promised? | 3.8.12 |
In a medical laboratory, "specified requirements" usually means the performance claims that the developer of the method has published.
CLSI EP19 gives the two words a place in time
CLSI EP19 (3rd edition, 2022) places the two words in the life of a method:
ISO 15189:2022 draws the same line
ISO 15189:2022, the accreditation standard for medical laboratories, defines both terms in the language of ISO/IEC 17000:2020 (entries 3.31 and 3.32):
- Validation confirms that a method is plausible for a specific intended use.
- Verification confirms that specified requirements have been fulfilled.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.3 Two words, two jobs: validation and verification, pp.29–30
What ISO 15189:2022 requires: clauses 7.3.1–7.3.3
Clause 7.3.1 — the frame for both
Methods must have been validated for their intended use, and each method's performance specifications must relate to that use and to its effect on patient care.
Clause 7.3.2 — methods used as designed (verification)
- Before such a method goes into use, the laboratory follows a procedure showing that it can achieve the performance stated by the manufacturer or by the method.
- The specifications it confirms are the ones that matter for the intended use.
- The work must be extensive enough to support the clinical decisions made from the results.
- Authorised, competent staff review the results and record whether they meet the requirements.
- If the issuing body revises the method, the laboratory repeats verification as far as the revision requires.
- The records show: the specifications to be achieved · the results obtained · whether the specifications were met (and, if not, what was done).
Clause 7.3.3 — three kinds of method (validation)
| # | Kind of method | Examples given |
|---|---|---|
| 1 | A method the laboratory designs or develops | — |
| 2 | A method used outside its originally intended scope | Outside the instructions for use or the validated measuring range; third-party reagents on an instrument they were not intended for, when no validation data exist |
| 3 | A validated method that is later modified | — |
- Validation must go as far as necessary to show, through performance specifications, that the requirements of the intended use are met.
- The clinical impact of any proposed change to a validated method is reviewed before the change is adopted.
- The records include: the procedure used · the requirements for the intended use · the performance specifications determined · the results · a statement of the method's fitness for its intended use.
| Validation | Verification | |
|---|---|---|
| Core question | Is the method fit for this intended use? | Does the method perform as claimed, here? |
| Who normally does it | The developer: a manufacturer, or a laboratory that develops or modifies a method | The end-user laboratory |
| Starting point | Design requirements derived from the intended use | The developer's published performance claims |
| Typical scale | Larger studies, often several reagent lots, instruments and sites | Smaller studies in the laboratory's own conditions |
| Output | Performance claims and the evidence behind them | A documented decision that the claims hold locally |
| ISO 15189:2022 clause | 7.3.3 | 7.3.2 |
One caution on vocabulary
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.3 Two words, two jobs: validation and verification (Table 1.2), pp.30–31
1.4 The life of a test: two stages, and choosing a candidate
Every test method passes through the same broad stages, whoever made it. CLSI EP19 calls this the Test Life Phases Model. Knowing where your work sits in that life tells you what kind of evidence is expected.
| Stage | Belongs to | Phases |
|---|---|---|
| 1. Establishment | The developer | Feasibility and design → development → validation |
| 2. Implementation | The end-user laboratory | Preliminary evaluation → verification → launch → maintenance (a long phase in routine service) → retirement |
Establishment: the design is frozen before validation
At the end of development the design is frozen, so that what is validated is exactly what users will receive. The developer then hands over the method with its instructions for use and its performance claims.
Implementation: from preliminary evaluation to retirement
- Preliminary evaluation — EP19 describes it as the point where the laboratory checks that a method suits its needs and screens for grossly unacceptable performance before the formal verification.
- Verification — the formal confirmation of the claims under the laboratory's own conditions.
- Launch — everything else needed before patient testing: procedures, training and competence, order codes, reference intervals, critical values and information-system set-up.
- Maintenance — the long phase in routine service.
- Retirement — the end of the test's life.
Before any formal study: choose a candidate worth evaluating
EP19 advises selecting against specifications set in advance (such as intended use, performance, cost per test, complexity and throughput) and preferring methods registered as medical devices.
It helps to sort what you know about a candidate into three groups, because each group is judged in a different way:
| Group | The question | Think of | How it is judged |
|---|---|---|---|
| Fit with the laboratory | Can the method run here at all? | Specimen types and volume, turnaround time, workload, staffing and training, space, utilities, safety and cost | From documents and site visits, not from data |
| Soundness of the measurement principle | Is the chemistry or technology likely to give good results? | How specific the reaction is for the measurand; how the method is calibrated and to what reference it is traceable; how robust the procedure is to small changes | From the instructions for use, the literature and expert review |
| Demonstrated performance | How well does the method actually perform? | Precision, trueness, measuring interval, detection capability, interference and the other characteristics in Table 1.4 | Only data can answer these questions — the rest of this course |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.4 The life of a test, pp.31–32
Each phase ends with a decision (Figure 1.1)
Three features of this model that matter in daily practice
- Each phase ends with a decision. The team proceeds, repeats work or stops, by comparing results with criteria set in advance. EP19 asks that the decision, the date and the people who made it are recorded.
- A modification sends the method back. If the laboratory changes a validated method, the method returns to the establishment stage for the changed aspects. The laboratory then acts as developer for those aspects.
- Maintenance is not the end of evaluation. Quality control, external quality assessment (EQA), reagent lot checks and comparisons between analysers keep confirming, in service, what verification first showed.
ISO 15189:2022 clause 7.3.7 — ensuring the validity of examination results
| Clause | What it requires |
|---|---|
| 7.3.7.1 | A planned procedure for monitoring results in a way that shows trends and shifts |
| 7.3.7.2 | Internal quality control |
| 7.3.7.3 | External quality assessment |
| 7.3.7.4 | Where different methods, instruments or sites perform the same examination, a procedure for showing that their results on patient samples are comparable across clinically significant intervals |
Part IV of the course covers that work.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.4 The life of a test (Figure 1.1), pp.33–34
1.5 Verify or validate? A decision path
Most laboratories verify far more often than they validate. The deciding question is simple:
The path as a table
| Question (in order) | Answer that leads to validation | Validation box |
|---|---|---|
| 1. Did the laboratory develop the method? | Yes | Validate in full — the laboratory is the developer |
| 2. Has the laboratory changed the method? | Yes | Validate the change — re-establish the changed aspects |
| 3. Used within its validated scope? | No | Validate for the new scope — specimen, population or purpose |
| All routes → Verify the claims under routine conditions → Record and sign the decision | ||
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.5 Verify or validate? A decision path (Figure 1.2), p.34
Common situations and the evaluation each needs
Table 1.3 applies the path to situations a hospital laboratory meets. The examples are illustrative. Your local regulator or accreditation body may set further rules (§1.7).
| Situation | Evaluation needed | Why |
|---|---|---|
| New analyser running the manufacturer's approved assays as instructed | Verification | The method is used exactly as validated; the laboratory confirms the claims locally |
| A second analyser of the same model added for the same assays | Verification of the new analyser, plus a comparison between analysers | Patients may be tested on either system, so results must agree (Chapter 16) |
| An approved serum assay used on a body fluid not listed in the instructions for use | Validation for the new specimen type | A new specimen matrix is a modification; the developer's evidence does not cover it |
| The laboratory changes the sample volume, dilution or incubation time set by the manufacturer | Validation of the changed method | A modified method returns to the establishment stage |
| An in-house method, such as a liquid chromatography–mass spectrometry drug assay | Full validation, then implementation | The laboratory is the developer; no external claims exist |
| A point-of-care meter used in a patient group the instructions for use do not include | Validation for that group | Use outside the validated scope; interferences and decision points may differ |
| The manufacturer revises the method (for example, a new reagent formulation) | Verification repeated to the extent the change requires | The claims changed; a risk assessment sets how much evidence is needed |
| A new lot of the same reagent | Neither: a reagent lot change check | Routine maintenance work, not a new method (Chapter 16) |
Two points worth stating plainly
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.5 Verify or validate? A decision path (Table 1.3), p.35
1.6 What the evidence must cover
A performance characteristic is one measurable aspect of how a method behaves: its random error, its systematic error, the range over which it reads correctly, and so on.
| What end users commonly verify (EP19) | What a developer's validation adds |
|---|---|
| Precision · trueness or bias · accuracy · analytical sensitivity and specificity · the measuring interval · the reference interval | Characteristics a user rarely has the materials or time to establish, such as detection capability, reagent and sample stability across lots, and, for many tests, clinical performance |
Table 1.4 summarises the characteristics, what each tells you, and the CLSI document that describes how to study it. The last column shows the chapter of this course that covers it.
| Characteristic | What it tells you | In a verification | Main CLSI document | Chapter |
|---|---|---|---|---|
| Precision | How much results scatter on repeated testing of the same sample | Commonly checked | EP05-A3 (establish); EP15-A3 (verify) | 6 |
| Trueness (bias) against reference materials | How far the average result sits from an assigned value | Commonly checked | EP15-A3 | 7 |
| Comparison with patient samples | How patient results differ from the current or a comparative method, across the range and at decision points | Commonly checked | EP09c-Ed3 | 8 |
| Linearity and measuring interval | Over what range results are proportional to concentration and can be reported directly | Commonly checked against the claimed interval | EP06-Ed2; EP34-Ed1 for dilution | 9 |
| Detection capability | The lowest concentrations that can be detected and quantified reliably | Checked when low results drive decisions | EP17-A2 | 10 |
| Analytical specificity (interference) | Which substances or sample conditions change the result, and from what level | Checked when the use or population calls for it; otherwise the claims are reviewed | EP07-Ed3; EP37-Ed1 | 11 |
| Carryover | Whether one sample or reagent affects the next result | Sample carryover often screened early; reagent carryover usually the developer's work | EP10-Ed4 (sample); EP47-Ed1 (reagent) | 12 |
| Reference interval | Whether the interval suits the local population and method | Commonly checked | EP28-A3c | 13 |
| Measurement uncertainty | The range within which the true value is expected to lie | Estimated by the laboratory; reviewed with verification data | EP29-A; ISO 15189:2022 clause 7.3.4 | 14 |
| Qualitative agreement | How often a positive/negative result agrees with the comparator; precision near the cut-off | Commonly checked for binary tests | EP12-Ed3 | 15 |
| Reagent stability | How long reagents, calibrators and controls keep their performance in storage and once opened | Normally established by the developer; a laboratory may confirm an in-use claim | EP25-Ed2 | 5 |
An example: stability
- EP25-Ed2 covers the stability of reagents, calibrators and control materials, not of patient specimens.
- Its intended users are mainly manufacturers and regulators (§1.1 of EP25).
- It notes that laboratories usually check a manufacturer's stability claims only indirectly, through QC trends (§2.2).
- It also describes how a laboratory can confirm a stated in-use claim, for example when a method is first verified or a new lot arrives (§6.1, §6.3).
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.6 What the evidence must cover (Table 1.4), pp.35–38
1.7 Criteria first, data second
The reason is human, not bureaucratic. Once numbers are on the screen, it is very easy to find a reason why a borderline result is acceptable. A limit written in advance cannot be bent to fit the data.
Four kinds of numbers — keep them apart
A plan usually contains numbers from four different sources. Keep them apart, because they carry different weight.
| Kind | Example | What it is | How to use it |
|---|---|---|---|
| ISO 15189 requirement | Verify an unmodified method before routine use (clause 7.3.2) | A "shall" statement that an accredited laboratory must meet | Meet it, and cite the clause in the plan |
| Manufacturer's performance claim | "Within-laboratory CV 1.2% at 5.0 mmol/L" in the instructions for use | What the developer showed under its own conditions | The thing a verification tests; not a statement of what the patient needs |
| CLSI recommendation | A study design or statistical approach in an EP document | Consensus good practice, written as "should" | Follow it, or record why you departed from it |
| Laboratory acceptance criterion | "Bias at 7.0 mmol/L no more than 0.2 mmol/L" | The laboratory's own limit, derived from the intended use | Set it before testing; record where it came from (Chapter 4) |
Decide in advance what happens if a criterion is missed
The plan should also say what happens if a criterion is not met. Options include:
Deciding this in advance avoids pressure to launch a method that has not earned it.
Record the decision
EP19 lists what the record should hold: the study results, the acceptance criteria, the decision, the date and the people who made it, kept as a permanent record. Chapter 20 and Templates A–C give a structure.
Requirements beyond ISO 15189
ISO 15189 is the international baseline, and it is the requirement document used throughout this course. It is rarely the only one. Depending on where a laboratory works, health regulators, licensing authorities and accreditation bodies may add rules of their own:
- which standard they accredit against,
- minimum study designs,
- extra evidence for particular tests, or
- how long records must be kept.
These rules differ between jurisdictions and change over time, so no book can list them reliably.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, §1.7 Criteria first, data second (Table 1.5); Requirements beyond ISO 15189, pp.38–39
Worked Example 1.1 (part 1): How much room does the method leave?
All data and limits in this example are illustrative and synthetic.
Situation
The laboratory is choosing a sodium method for the new analyser. For this intended use it has set, in advance, an allowable total error of 4.0 mmol/L at 140 mmol/L (2.9%). Chapter 4 explains how to set such a limit; here it is simply given.
Step 1 — Screen the claims (preliminary evaluation)
The instructions for use claim a within-laboratory CV of 0.9% and a bias of +0.5% at 140 mmol/L. Convert both to mmol/L:
| Quantity | Calculation | Result |
|---|---|---|
| SD | 0.9% × 140 | 1.26 mmol/L |
| Bias | 0.5% × 140 | 0.70 mmol/L |
A simple estimate of the largest error expected in 95% of single results combines the two. CLSI EP21 describes this model, in which the absolute bias is added to 1.96 times the within-laboratory SD, and notes its limits.
TAE = 0.70 + 1.96 × 1.26 = 0.70 + 2.47 = 3.17 mmol/L
TAE = estimated total analytical error (mmol/L) · |bias| = absolute bias (mmol/L) · SD = within-laboratory standard deviation (mmol/L) · 1.96 = the multiplier for 95% of single results in this model.
The margin for error is the allowable limit minus the estimate:
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, Worked Example 1.1 How much room does the method leave? (Situation, Step 1), pp.39–40
Worked Example 1.1 (part 2): the laboratory's own data
Step 2 — Compare with the laboratory's own data (verification)
Two outcomes are possible. Suppose the precision and bias studies (Chapters 6–8) give the results below.
| SD (mmol/L) | Bias (mmol/L) | 1.96 × SD (mmol/L) | Estimated TAE (mmol/L) | Margin (mmol/L) | |
|---|---|---|---|---|---|
| Claims | 1.26 | +0.70 | 2.47 | 3.17 | 0.83 |
| Scenario A | 1.20 | +0.90 | 2.35 | 3.25 | 0.75 |
| Scenario B | 1.75 | +0.90 | 3.43 | 4.33 | −0.33 |
The arithmetic: Scenario A — 1.96 × 1.20 = 2.35; 0.90 + 2.35 = 3.25; 4.0 − 3.25 = 0.75. Scenario B — 1.96 × 1.75 = 3.43; 0.90 + 3.43 = 4.33; 4.0 − 4.33 = −0.33.
Interpretation
- Scenario A: the local bias is a little larger than claimed, but the total still fits inside 4.0 mmol/L. The margin of 0.75 mmol/L is narrow, which tells the laboratory to plan its quality control with care (Chapter 18).
- Scenario B: the bias is the same, but the scatter is larger. The estimate of 4.33 mmol/L exceeds the limit, so a meaningful share of single results could be wrong by more than the intended use allows.
Decision
| Scenario | Decision | Next action |
|---|---|---|
| A | Accept | Record the narrow margin as a risk to monitor |
| B | Do not accept yet | Investigate why the precision is worse than claimed (Chapter 17). Common causes: a fault on the analyser, a problem with the control material, or a study design that captured more sources of variation than the manufacturer's did |
What this example does not prove
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, Worked Example 1.1 How much room does the method leave? (Step 2, Figure 1.3, interpretation, decision), pp.40–41
Common errors of interpretation
| The error | Why it is wrong |
|---|---|
| ⚠️ "The manufacturer validated it, so we have nothing to do." | The developer's validation shows what the method can do under the developer's conditions. Only your verification shows what it does with your staff, samples and environment. |
| ⚠️ "It met the claims, so it is fit for purpose." | Claims describe the method, not the patient's need. Judge the results against criteria derived from the intended use as well (Table 1.5). |
| ⚠️ Treating a modification as a verification. | A new specimen type, a changed dilution or use in a new patient group is outside the validated scope. A verification-sized study cannot cover it. |
| ⚠️ Setting the limits after seeing the data. | A limit chosen once the numbers are known is not an acceptance criterion. It is a description of the result. |
| ⚠️ Judging performance only on average. | A method can be acceptable overall and unacceptable at the one concentration that drives a clinical decision. Check performance at each decision point. |
| ⚠️ Reading another document's vocabulary as your own. | "Validation" and "verification" are used in different senses in different documents and departments. Confirm the meaning before following a procedure. |
| ⚠️ Assuming one evaluation lasts forever. | Revised methods, relocated analysers and changes in use all call for a fresh look, scaled by a risk assessment. |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, Common errors of interpretation, pp.41–42
Chapter 1 checklist and references
Before planning any study, confirm that you can tick each item.
| ☐ | Checklist item | See |
|---|---|---|
| ☐ | The intended use is written down: measurand, specimen, population, clinical purpose, decision points, setting and reporting needs | Table 1.1 |
| ☐ | You have decided whether the method is used exactly as validated (verify) or departs from its validated scope (validate), and recorded why | Figure 1.2, Table 1.3 |
| ☐ | Any modification has been identified, and the laboratory accepts that it is the developer for the changed aspects | §1.4, §1.5 |
| ☐ | The performance characteristics to study are chosen from the intended use and a risk assessment | Table 1.4 |
| ☐ | Each number in the plan is labelled by its source: requirement, manufacturer's claim, CLSI recommendation or laboratory criterion | Table 1.5 |
| ☐ | Acceptance criteria and the action to take if one is missed are written, signed and dated before data collection | §1.7 |
| ☐ | Regulatory and accreditation requirements that apply to the laboratory have been checked and cited by title and version | §1.7 |
| ☐ | The person who will make and sign the final decision is named | §1.7 |
Chapter 1 references (as cited by the book)
- Clinical and Laboratory Standards Institute. A Framework for Using CLSI Documents to Evaluate Medical Laboratory Test Methods. 3rd ed. CLSI report EP19-Ed3. CLSI; 2022. Sections checked: Introduction to the Test Life Phases Model; Modifications to Established Test Methods; Acceptance Criteria; Documentation of Decisions and Actions; Validation Phase; Implementation Stage (Preliminary Evaluation, Test Method Selection, Verification, Launch, Maintenance); Definitions.
- International Organization for Standardization. ISO 9000:2015 Quality management systems — Fundamentals and vocabulary. ISO; 2015. Entries 3.8.12 (verification) and 3.8.13 (validation). Entry numbers confirmed from the IEC Electropedia record for validation (IEV 192-01-18).
- International Organization for Standardization. ISO 15189:2022 Medical laboratories — Requirements for quality and competence. 4th ed. ISO; 2022 (December). Entries 3.31 and 3.32; clauses 7.3.1–7.3.4 and 7.3.7.
- Clinical and Laboratory Standards Institute. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures. 2nd ed. CLSI guideline EP21-Ed2. CLSI; 2016. Chapter 1, background to the total analytical error model and its limitations.
CLSI documents named in Table 1.4 (all published by CLSI)
- EP05-A3. Evaluation of Precision of Quantitative Measurement Procedures. 3rd ed. 2014.
- EP06-Ed2. Evaluation of Linearity of Quantitative Measurement Procedures. 2nd ed. 2020.
- EP07-Ed3. Interference Testing in Clinical Chemistry. 3rd ed. 2018.
- EP09c-Ed3. Measurement Procedure Comparison and Bias Estimation Using Patient Samples. 3rd ed. 2018.
- EP10-Ed4. Preliminary Evaluation of Quantitative Medical Laboratory Measurement Procedures. 4th ed. 2024.
- EP12-Ed3. Evaluation of Qualitative, Binary Output Examination Performance. 3rd ed. 2023.
- EP15-A3. User Verification of Precision and Estimation of Bias. 3rd ed. 2014.
- EP17-A2. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. 2nd ed. 2012.
- EP28-A3c. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory. 3rd ed. 2010.
- EP29-A. Expression of Measurement Uncertainty in Laboratory Medicine. 1st ed. 2012.
- EP34-Ed1. Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking. 1st ed. 2018.
- EP47-Ed1. Evaluation of Reagent Carryover Effects on Test Results. 1st ed. 2024.
- EP25-Ed2. Evaluation of Stability of In Vitro Medical Laboratory Test Reagents. 2nd ed. 2023. Sections 1.1, 2.2, 6.1 and 6.3 read.
- EP37-Ed1. Supplemental Tables for Interference Testing in Clinical Chemistry. 1st ed. 2018.
CLSI documents are cited for attribution only. Readers should obtain the current edition from CLSI for the full procedures. Citation does not imply endorsement by CLSI, ISO, any accreditation body or any manufacturer.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, Chapter 1 checklist; Chapter 1 references, pp.42–43
Training course, not a certification. Based on the author's textbook; all examples use synthetic data. It does not replace the CLSI and ISO documents it cites, and it is not affiliated with or endorsed by CLSI or ISO.
© 2026 Hassan Shawki. Course text adapted from the author's textbook Clinical Laboratory Method Evaluation in Practice.
⚡ Last-Minute Review — Fit for Purpose: Verification, Validation and the Life of a Test
Read these lists before you plan any method evaluation. Then open Progress / Results to see your weak areas.
The decision
- Do we verify or validate, and what must we prove before the first patient result?
- Every result carries error; it never reaches zero. The question is whether the error is small enough that the clinician makes the same decision as on the true value.
- Same error, different stakes: near a decision point it can change patient care. The method has not changed; the stakes have.
- Margin for error = error the test can tolerate − error the method produces. Wide → light QC plan · narrow → closer watching (Chapter 18).
Intended use (Table 1.1)
- Seven elements: measurand · specimen · population · clinical purpose · decision points · setting and user · reporting needs
- "To measure troponin" is too vague; a specific statement shows where the method must perform best
- Same troponin assay: ED rule-out → precision and bias at low concentrations · monitoring → consistency over time
- A change of intended use is a change of method (EP19: adult → paediatric, diagnostic → screening)
Validation vs verification (Table 1.2)
- Validation: meets the requirements of a particular intended use — "right method for the job?" (ISO 9000:2015, 3.8.13) · developer · design requirements → claims · ISO 15189:2022 7.3.3
- Verification: specified requirements met — "does it do what was promised, here?" (3.8.12) · end-user laboratory · published claims → local decision · 7.3.2
- Verification checks claims; validation creates or re-creates the evidence on which claims rest
- The developer need not be a manufacturer: a laboratory that designs a test or changes an approved one becomes the developer
- ISO 15189:2022 entries 3.31/3.32; note: verification alone may be enough for a new IVD device used as its instructions direct
- 7.3.1: methods validated for intended use; specifications relate to the use and to patient care
- 7.3.2: procedure · specifications that matter for the use · extensive enough for the clinical decisions · authorised competent staff review · repeat as far as a revision requires · records (specifications, results, met or not, action)
- 7.3.3: laboratory-developed · used outside intended scope (outside IFU or validated range; third-party reagents with no validation data) · validated method later modified · clinical impact reviewed before a change · statement of fitness
- Vocabulary is not uniform (some microbiology laboratories call routine monitoring "validation"): check what the author means
The life of a test (Figure 1.1)
- Establishment (developer): feasibility and design → development (design frozen) → validation → transfer of instructions for use and claims
- Implementation (laboratory): preliminary evaluation (suits needs; screens for grossly unacceptable performance) → verification → launch (procedures, training, order codes, reference intervals, critical values, LIS) → maintenance → retirement
- Choosing a candidate: fit with the laboratory (documents, site visits) · soundness of the principle (IFU, literature, expert review) · demonstrated performance (only data)
- Three features: each phase ends with a recorded decision (decision, date, people) · a modification sends the method back (laboratory acts as developer) · maintenance keeps confirming (ISO 15189 7.3.7.1 trends and shifts, 7.3.7.2 IQC, 7.3.7.3 EQA, 7.3.7.4 comparability)
Verify or validate? (Figure 1.2, Table 1.3)
- Used exactly as validated, for the validated purpose → verify · any departure → validate that departure
- Three questions in order: developed by the lab? (yes → validate in full) · changed by the lab? (yes → validate the change) · within validated scope? (no → validate for the new scope) · all routes → verify under routine conditions → record and sign
- New analyser, approved assays as instructed → verification · second identical analyser → verification + comparison between analysers
- Body fluid not in IFU → validation for the specimen · lab changes volume/dilution/incubation → validation · in-house LC–MS → full validation · POC meter in a group not in IFU → validation for that group
- Manufacturer revises the method → verification repeated to the extent required (risk assessment) · new reagent lot → neither: lot change check
- Validation does not replace verification · the decision is per method and per change
What the evidence must cover (Table 1.4)
- Measure the characteristics the intended use depends on, to the depth the situation needs; scope from intended use + risk assessment, reasons recorded
- Precision EP05-A3/EP15-A3 (Ch 6) · trueness EP15-A3 (Ch 7) · comparison EP09c-Ed3 (Ch 8) · linearity EP06-Ed2, dilution EP34-Ed1 (Ch 9) · detection EP17-A2 (Ch 10) · interference EP07-Ed3/EP37-Ed1 (Ch 11) · carryover EP10-Ed4 sample, EP47-Ed1 reagent (Ch 12) · reference interval EP28-A3c (Ch 13) · uncertainty EP29-A, ISO 15189 7.3.4 (Ch 14) · qualitative EP12-Ed3 (Ch 15) · reagent stability EP25-Ed2 (Ch 5)
- Developer adds: detection capability, stability across lots, clinical performance
- EP25-Ed2 = reagents, calibrators, controls — not patient specimens; labs check indirectly through QC trends, or confirm an in-use claim
Criteria first, data second (Table 1.5)
- Criteria written before a single result; a limit set after seeing the data is a description of the result
- Four kinds of number: ISO requirement ("shall" — meet and cite) · manufacturer's claim (what verification tests; not the patient's need) · CLSI recommendation ("should" — follow or record why not) · laboratory criterion (own limit from intended use; set before testing; record its source)
- A method can meet every claim and still fail the laboratory's criterion
- Plan states the action if a criterion is missed: repeat after a fault · contact manufacturer · restrict · reject
- Record: results · criteria · decision · date · people · permanent
- Local regulators may add rules: cite by title and version; the stricter local rule applies
Formula (Worked Example 1.1)
- SD = CV × concentration · bias (unit) = bias % × concentration
- TAE = |bias| + 1.96 × SD (EP21 model: largest error in 95% of single results)
- Margin = allowable total error − TAE
- Sodium at 140 mmol/L, ATE 4.0 mmol/L: claims TAE 3.17, margin 0.83 (≈21%) → proceed · A: TAE 3.25, margin 0.75 → accept, monitor the narrow margin · B: TAE 4.33, margin −0.33 → do not accept yet, investigate (same bias; wider scatter)
- Limits: constant bias, normal errors, ignores interference and non-linearity; state the SD multiplier in the plan
Common errors
- "Manufacturer validated it, nothing to do" · "met the claims, so fit for purpose" · modification treated as verification · limits set after the data · judged only on average · another document's vocabulary · one evaluation forever
Key terms (book glossary)
- Acceptance criterion — A rule, written before the study, that turns a result into a decision: meets, does not meet, or needs more evidence.
- Intended use — What a result will be used for, in which patients and with which specimen. It is the yardstick for every criterion.
- Validation — Evidence that a method meets the requirements of its intended use, needed when the laboratory develops or modifies a method or uses it outside its validated scope.
- Verification — Evidence that a validated method performs as claimed in the laboratory's own hands, for its intended use.
- Total analytical error — The combined effect of bias and imprecision on a single result.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.1, pp.27–43; Glossary, pp.411–419