METHOD EVALUATION COURSE · PART I · CHAPTER 02

The Language of Error

Error, mistake and uncertainty · The comparator and traceability · Systematic and random error · Trueness, precision, accuracy · Precision conditions · Sources of bias and imprecision · Total error and measurement uncertainty · The reference change value · Worked example: is the fall real?

Course text from Clinical Laboratory Method Evaluation in Practice (1st ed., 2026) · Original practice questions · Works offline · questions · Version 1.0

📘 Study progress

Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 key point · 📌 remember · ⚠️ common error · 💡 understand this · 🧪 at the bench · 🧠 memory aid.

Start here: the question on your desk

🔥 THE QUESTION ON YOUR DESK: A ward sample is run on both chemistry analysers during a morning check. One reports potassium as 4.4 mmol/L, the other as 4.7 mmol/L. The nurse asks which result is wrong. A colleague says both are "within error". Another says one analyser must be "inaccurate". Before anyone can answer, the team has to agree what those words mean.

Method evaluation runs on a small vocabulary: error, bias, precision, trueness, accuracy, uncertainty. People use these words loosely in conversation, and the looseness causes real mistakes:

  • A precise method is called accurate.
  • A bias is called an error.
  • A difference between two analysers is blamed on the one that happens to be newer.

This chapter fixes the meaning of each term, shows how the terms relate, and connects them to decisions made about patients.

💡 UNDERSTAND THIS: The definitions follow the International Vocabulary of Metrology (VIM, JCGM 200:2012, 3rd edition): measurement accuracy (entry 2.13), measurement trueness (2.14), measurement precision (2.15) and measurement bias (2.18). The CLSI evaluation documents adopt the same terms. The book explains them in plain language rather than quoting them.
🧪 THE DECISION THIS CHAPTER SUPPORTS: Which kind of error are we measuring, and what does each word mean? (Main references named by the book: ISO 15189:2022; CLSI EP21-Ed2, EP29-A.)

By the end of this chapter you should be able to

  1. Tell an error from a mistake, and both from uncertainty.
  2. Name the value a result is compared against, and explain why it is never exactly the true value.
  3. Separate systematic from random error, and say which statistic describes each.
  4. State the conditions behind any precision estimate.
  5. List common causes of bias and of imprecision in a routine laboratory.
  6. Explain how the total error and measurement uncertainty models combine the two kinds of error.
  7. Use analytical and biological variation to judge whether a change in a patient's results is real.

Map of this chapter

#Study sectionBook section(s)Printed pages
1Start here: the question on your deskChapter 2 opening, learning objectives45–46
2Error, mistake and uncertainty are different things§2.146
3The value we compare against, and traceability§2.246–47
4Two kinds of error: systematic and random§2.347–48
5The vocabulary at a glance (Table 2.1, Figure 2.1)§2.3 (Table 2.1, Figure 2.1)49–50
6Precision always depends on the conditions (Figure 2.2)§2.450–51
7Where bias and imprecision come from (Table 2.2)§2.551–53
8Two ways to put the errors together§2.653–54
9Why the vocabulary matters for patients: limits, change and the RCV§2.7 (first part)54–55
10The logarithmic (asymmetric) RCV§2.7 (second part)55
11Worked Example 2.1 (part 1): is the fall in cholesterol real?Worked Example 2.1, Situation, Data, Steps 1–356
12Worked Example 2.1 (part 2): a poorer method, the log check and the decisionWorked Example 2.1, Step 4 (Figure 2.3), Step 5, interpretation, decision56–58
13Common errors of interpretationCommon errors of interpretation58–59
14Chapter 2 checklist and referencesChapter 2 checklist; Chapter 2 references59–60
15Reading the chartsOriginal practice charts built on the chapter's ideas—
📌 REMEMBER: Every number in an evaluation report needs its words to be right: which kind of error it describes, compared with what, and under which conditions.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, opening and learning objectives, pp.45–46; How this book is organised, p.viii

2.1 Error, mistake and uncertainty are different things

Three words are often used as if they meant the same thing. They do not, and an evaluation treats each one differently.

Measurement error

Measurement error is the difference between a measured value and the value it should have been.

error = measured value − reference value
  • Every result has an error, and it has a sign: a result can sit above or below its reference value.
  • For a single patient result the error is never known exactly, because the reference value is not known.
  • What can be studied is the pattern of errors a method produces over many measurements. That is the subject of the book.

Mistake

A mistake is something that should not have happened: the wrong tube, a transposed digit, an expired reagent, a clot that blocks the sample probe.

Mistakes…What it means
are not part of a method's normal errorThey do not follow its statistical pattern; precision or bias figures say nothing about how often they occur.
are prevented by process design…and caught by checks such as specimen-identity procedures, instrument flags and result review.
in an evaluation studyA result spoiled by an identified mistake is excluded, but the exclusion and its reason are written down.
a pattern of mistakes…during an evaluation is itself a finding: the method or its workflow may be prone to them.

Measurement uncertainty

Measurement uncertainty describes how far the reference value could reasonably be from a result. CLSI EP29 defines it as a non-negative parameter that characterises the spread of values that could reasonably be attributed to the measurand.

ErrorUncertainty
A single, unknown, signed number for one resultA range that says how large that error could plausibly be
🧠 MEMORY AID: Error is how far you missed; uncertainty is how far you might have missed; a mistake is aiming at the wrong target.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.1 Error, mistake and uncertainty are different things, p.46

2.2 The value we compare against

Every statement about error compares a result with something. The true value of a measurand in a patient sample is never known exactly.

In practice the comparison is made with a reference quantity value: the best available stand-in for the truth, chosen for a stated purpose. It can come from several places, and each carries a different weight.

ComparatorWhat it isWhat a difference from it shows
Reference measurement procedure or certified reference materialHigh in the hierarchy; comes with a stated uncertaintyThe closest a laboratory gets to a true bias
Conventional quantity valueA value attributed by agreement for a given purpose, accepted as having a suitably small uncertainty (example: international units assigned to a reference preparation)A difference from the agreed value
Consensus value (for example the mean of a peer group in an external quality assessment scheme)Shows how a laboratory compares with others using similar methodsNot the truth: if the whole peer group is biased, the laboratory can agree with its peers and still be biased
Current routine method (the old method as comparator in a method comparison)Often serves as the comparatorHow results will change for clinicians; it does not show which method is closer to the truth
🔥 KEY POINT: The word "bias" always needs a partner: biased compared with what? An evaluation report should name the comparator every time it reports a bias.

Traceability

Routine results link to higher-order references through metrological traceability: an unbroken, documented chain of calibrations from the routine result up to a reference, each link adding its own uncertainty.

Reference (top) → manufacturer builds the upper links (assigns values to calibrators) → laboratory adds the last link (calibrates its analyser) → routine result
⚠️ COMMON ERROR: A claim of traceability tells you the chain exists. It does not tell you that the uncertainty at the end of the chain is small enough for clinical use. Two methods traceable to the same reference can still give different results on patient samples, for example because they react differently to other substances in the sample. Chapters 7 and 8 return to this.
📌 REMEMBER: Comparator weight, from strongest to weakest claim on "truth": reference procedure / certified material → conventional value → consensus value → current method (which shows change, not truth).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.2 The value we compare against (including Traceability), pp.46–47

2.3 Two kinds of error: systematic and random

Repeat a measurement on the same sample many times and look at the results. Two things stand out:

  1. The results cluster around some value, which may or may not be the reference value.
  2. They scatter around that cluster.

These two features are the two components of measurement error.

Systematic errorRandom error
DefinitionThe part of the error that stays the same, or changes in a predictable way, from one replicate to the nextThe part that changes unpredictably from one replicate to the next
Effect on the clusterMoves the whole clusterSpreads the cluster
Its estimateBias: the difference between the average of many results and the reference valueImprecision: the standard deviation (SD) of replicate results, or the coefficient of variation (CV), the SD as a percentage of the mean
Reported asIn the measurement unit (for example +0.20 mmol/L) or as a percentage of the reference valueSD in units, or CV in %
Can it be corrected?Because it is predictable, a known systematic error can in principle be correctedNo; averaging more replicates reduces its effect on a mean
Does averaging help?Averaging does nothing to systematic errorYes, for the mean

Three quality words

QualityWhat it isKind of errorExpressed through
TruenessHow close the average of many results comes to the reference valueSystematic onlyBias
PrecisionHow close replicate results are to one another under stated conditionsRandom onlyImprecision
AccuracyHow close a single result comes to the true valueBoth kinds(depends on both)
🔥 KEY POINT: None of the three is a number in itself. They are qualities, measured through bias and imprecision. The VIM also warns against using "accuracy" to mean trueness, or "precision" to mean accuracy (entry 2.13).
⚠️ COMMON ERROR: A method can be beautifully precise and still give inaccurate results, if every result is shifted by the same amount.

The link to basic statistics

StatisticTracksWhat a change signals
Mean of replicate resultsSystematic errorA shift in the mean of control results signals a change in bias
SDRandom errorA widening of the spread signals a change in imprecision

Chapter 3 covers the calculations.

🧠 MEMORY AID: Bias moves, imprecision spreads. Mean ↔ trueness ↔ bias; SD ↔ precision ↔ imprecision; a single result ↔ accuracy ↔ both.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.3 Two kinds of error: systematic and random, pp.47–48

The vocabulary of error at a glance (Table 2.1, Figure 2.1)

Table 2.1 The vocabulary of error at a glance

TermWhat it describesHow it is expressedKind of error
Measurement errorMeasured value minus reference value, for one resultA signed value in units; unknown for a single patient resultBoth
BiasEstimated systematic errorMean minus reference value, in units or %Systematic
ImprecisionScatter of replicate resultsSD in units, or CV in %Random
TruenessCloseness of the long-run average to the reference valueThrough bias (not a number itself)Systematic
PrecisionCloseness of replicates to one another under stated conditionsThrough imprecision (not a number itself)Random
AccuracyCloseness of a single result to the true valueThrough total error or uncertainty (not a number itself)Both
Measurement uncertaintySpread of values that could reasonably be the true valueStandard uncertainty u, or expanded uncertainty UBoth, once known biases are corrected or included

Figure 2.1 Trueness and precision are separate qualities: four sets of replicate potassium results.

Figure 2.1: four rows of 20 replicate potassium results around a reference value of 4.50 mmol/L. True and precise: bias 0.00, SD 0.03. True but imprecise: bias 0.00, SD 0.12. Precise but biased: bias +0.20, SD 0.03. Biased and imprecise: bias +0.20, SD 0.12.

Synthetic, illustrative data · 20 replicates per set · reference value 4.50 mmol/L

🧪 HOW TO READ THIS FIGURE: Each row shows 20 replicate results on the same sample from one method; each grey dot is one result, spread slightly up and down only so that overlapping dots stay visible. The horizontal axis is the result in mmol/L. The dashed line is the reference value. The blue bar marks the mean of the 20 results, and the text on the left gives the bias (mean minus reference value) and the SD.
RowBiasSDWhat it shows
True and precise0.000.03 mmol/LCentred on the reference value and tight
True but imprecise0.000.12 mmol/LAlso centred (good trueness), but four times as scattered
Precise but biased+0.200.03 mmol/LTight, but sits 0.20 mmol/L too high
Biased and imprecise+0.200.12 mmol/LBoth problems: a single result can be far from the reference value
📌 REMEMBER: Compare the top two rows for precision (same centre, different spread). Compare the first and third rows for trueness (same spread, different centre).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.3 Two kinds of error (Table 2.1, Figure 2.1), pp.49–50

2.4 Precision always depends on the conditions

An SD on its own means little. The same method can give:

How the sample was runSD
One sample run ten times in a row0.03 mmol/L
That sample run once a day for a month0.06 mmol/L

Both figures are correct. They describe precision under different conditions, and every precision estimate must say which conditions it covers. The CLSI documents use three levels, taken from the VIM and ISO statistical standards.

ConditionsWhat stays the same / what changesOther names
RepeatabilitySame procedure, same operator, same measuring system, same operating conditions, same place, over a short time. In practice this is one analytical run.Older documents: within-run precision
Intermediate precisionSame procedure and same laboratory, but some operating conditions change: the day, the calibration, the operator or the equipment.Collected over enough time to include routine changes in reagent and calibrator lots, CLSI EP15 calls it within-laboratory imprecision, formerly "total precision"
ReproducibilityDifferent locations, operators and measuring systems.The scatter seen between laboratories, for example in an external quality assessment scheme

Figure 2.2 The three sets of precision conditions, nested from the narrowest to the widest.

Figure 2.2: three nested boxes. Inner box repeatability conditions (one run; formerly within-run precision); middle box within-laboratory (intermediate precision) conditions where day, operator, calibration and lots change; outer box reproducibility conditions between laboratories. An arrow shows SD and CV usually grow as more conditions change.

Original diagram · condition definitions from CLSI EP05-A3 and EP15-A3 (after JCGM 200:2012)

🧪 HOW TO READ THIS FIGURE: Start in the highlighted inner box: repeatability, where almost nothing is allowed to change. Each box outward keeps everything in the box inside it but allows more factors to change: first the day, operator, calibration and lots within one laboratory, then the laboratory and measuring system themselves. Each extra source of variation adds to the scatter, as the arrow on the right shows, so a repeatability SD is normally the smallest and a reproducibility SD the largest. A precision figure is only meaningful when you know which box it came from.

Three practical rules follow

  1. Compare like with like. A laboratory's within-run SD cannot verify a manufacturer's within-laboratory claim, and vice versa. Check which condition a claim describes before designing the study (Chapter 6).
  2. Choose the condition that matches the use. Patients are usually tested on different days, after recalibrations and lot changes. The within-laboratory SD is therefore the figure that describes the scatter a clinician will actually see between a patient's results.
  3. Expect precision to vary with concentration. The SD often grows as concentration rises, while the CV may be roughly constant over part of the range and rise steeply near the lower limit. Report precision at the concentrations where decisions are made.
🧠 MEMORY AID: Boxes inside boxes: run → laboratory → laboratories. Each step out lets more change, so the SD usually grows. Always ask: which box?

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.4 Precision always depends on the conditions (Figure 2.2), pp.50–51

2.5 Where bias and imprecision come from

Knowing which kind of error a fault produces speeds up troubleshooting and helps an evaluator design studies that will expose it.

A fault that acts the same way on every measurement → bias · A fault that acts differently each time → imprecision · Many faults can do either, depending on how they behave

Table 2.2 Common sources of systematic and random error

SourceTypically shows as systematic error (bias) when…Typically shows as random error (imprecision) when…
CalibrationA calibrator value is wrong, a calibrator has deteriorated, or a new calibrator lot is assigned differentlyCalibration is repeated often and each calibration lands slightly differently
ReagentsA new reagent lot behaves differently, or a reagent slowly deteriorates on board (a drift)Reagent is poorly mixed, or its performance varies across the time it is open
Control materialIt is reconstituted wrongly, a new lot has different values, or it agesReconstitution or thawing varies from vial to vial
InstrumentWavelength, temperature or a pipetting volume is set or calibrated wronglyPipetting, bubbles, a worn part or an intermittent fault varies from test to test
SampleA substance in the sample adds to or suppresses the signal in every sample that contains itCarryover from the previous sample, or clots and bubbles, affect some tests and not others
EnvironmentRoom temperature or water quality shifts and stays shiftedTemperature or power fluctuates through the day
OperatorA step in the procedure is done consistently wrongManual steps, such as timing or pipetting, are done slightly differently each time

Three further points matter for evaluation

1. Bias need not be a single number.

Constant biasProportional bias
Shifts every result by the same amount, for example 0.1 mmol/LShifts results by the same percentage, so the shift grows with concentration

Many methods show some of both. That is why bias must be estimated at each decision point, not only on average. Chapter 8 shows how a method comparison separates the two.

2. The line between the two kinds of error depends on the time scale. CLSI EP29 notes that a bias which changes with each reagent or calibrator lot behaves like random variation when viewed over a long period.

One-week study → sees a fixed lot bias · Six-month record of quality control → sees it as part of the scatter

State the period an estimate covers, and interpret it in that light.

3. Part of a method's bias can belong to the individual sample rather than to the calibration. The Tietz Textbook of Laboratory Medicine splits the error of a single patient result into three parts:

PartWhat it isSeen by control and reference materials?
Calibration biasShared by all samplesYes
Sample-related random biasDiffers from one patient sample to the next, because of interfering substances or matrix effectsNo
Random measurement errorWhat replicates showYes
⚠️ COMMON ERROR: Control materials and reference materials reveal only the first and the last part, so a bias estimated from them alone can understate how wrong a patient result may be. A method comparison on patient samples exposes the middle term as extra scatter, and occasionally as an outlier.
🔥 KEY POINT: An outlier is a finding to investigate, not a point to delete: a heterophilic antibody, for example, may mark a real limitation of the method (Chapters 8 and 11).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.5 Where bias and imprecision come from (Table 2.2), pp.51–53

2.6 Two ways to put the errors together

A clinician receives one result, not a bias and an SD. At some point the two kinds of error have to be combined into a single statement about how far a result might be wrong. Laboratory medicine uses two models for this. CLSI EP29 describes both.

The error model (total analytical error)

  • Treats the true value as a single unknown number and keeps the two errors separate until the end.
  • It then adds them: the bias plus a multiple of the SD gives an upper limit on the error of a single result, assuming random errors follow a normal distribution.
  • Sources differ on the multiplier: a traditional form adds 2 SD, a one-sided view uses 1.65 SD, and a smaller chance of exceeding the limit needs a larger one.
  • This is the total analytical error used in Worked Example 1.1, and CLSI EP21 develops it further, including an approach that needs no distributional assumption.
StrengthWeakness (EP29)
Simplicity: one figure that can be set against an allowable limitA total error cannot be carried into a calculation that uses the result, such as a ratio or a clearance. For that, it must first be split back into its parts.

The uncertainty model

  • Describes an interval within which the true value is expected to lie, with a stated level of confidence.
  • Assumes that known biases are corrected, or at least identified, and treats the uncertainty of each correction as one more component to combine with the random components.
  • The combined standard uncertainty is multiplied by a coverage factor to give the expanded uncertainty.
  • This is the approach of the international guide to the expression of uncertainty in measurement (GUM).

The two models answer slightly different questions

QuestionUsually answered with
Is this method good enough for its intended use?Evaluations: the error model, estimated total error against an allowable total error set in advance
How sure can we be about this result?Interpretation and reporting: the uncertainty model

ISO 15189:2022 clause 7.3.4

ISO 15189:2022 treats the two models as complementary. Under clause 7.3.4 the laboratory must:

  • evaluate the measurement uncertainty of its measured values where relevant and maintain it for the intended use;
  • compare that uncertainty with performance specifications;
  • document it, review it regularly and record why any examination is left out.
💡 UNDERSTAND THIS: The same clause recommends, rather than requires, that uncertainty be considered during verification and validation when relevant. When users ask about uncertainty, the laboratory's answer must also allow for other sources, such as biological variation.
🔥 KEY POINT: A modern evaluation therefore produces the data for both models: bias and imprecision estimates feed the total error decision now, and the uncertainty estimate later. Chapter 14 shows how to do both with the same data set.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.6 Two ways to put the errors together, pp.53–54

2.7 Why the vocabulary matters for patients: limits, change and the RCV

A clinician uses a numerical result in one of two ways. The kind of error that matters most differs between them.

1. Comparing a result with a limit

Is this troponin above the cut-off? Is this glucose in the diabetic range? Here both bias and imprecision matter.

ErrorEffect near a limit
BiasMoves every result towards or away from the limit, so a whole population of patients is classified differently
ImprecisionScatters individual results across the limit: a patient whose true value sits just below it may be reported just above it on one occasion and below it on the next

The closer the true values cluster around a cut-off, the more patients are affected. This is why the imprecision of a test used at a low diagnostic cut-off, such as cardiac troponin, has to be judged at that cut-off rather than in the middle of the range. Chapters 10 and 14 return to this case.

2. Comparing a result with the patient's previous result

Has the cholesterol fallen since treatment started? Is the creatinine rising? If both results come from the same method, a constant bias largely cancels out. What remains is:

  • imprecision, plus
  • something the laboratory cannot control: the natural day-to-day variation within the patient, called within-subject biological variation (CVI).

A change is meaningful only if it is larger than these two sources of variation could produce together.

The reference change value (RCV)

The reference change value (RCV) puts this into a formula. It gives the percentage change between two results that is unlikely to arise from analytical and within-subject variation alone:

RCV (%) = √2 × Z × √(CVA² + CVI²)
SymbolMeaning
CVAThe laboratory's analytical CV (its within-laboratory imprecision)
CVIThe within-subject biological CV
ZSets the level of confidence: 1.96 for "has the result changed in either direction?" at 95% confidence; 1.65 for "has it changed in the expected direction?" at the same confidence
√2Appears because two results, each carrying its own variation, are being compared

Published, quality-appraised estimates of CVI are collected in the EFLM Biological Variation Database.

Three conditions make the RCV trustworthy

  1. The CVA should come from the laboratory's own long-term internal quality control, at a concentration near the decision point, over a period comparable to the interval between the two samples. A long-term CVA already contains the small shifts caused by recalibration and lot changes.
  2. Preanalytical conditions (patient preparation, collection, transport and handling) are assumed to be controlled and constant.
  3. The CVI must come from a population similar to the patient's.
⚠️ COMMON ERROR: A common habit, multiplying √(CVA² + CVI²) by 2.77 for every comparison, builds in the two-sided Z of 1.96. Most monitoring questions ask about a rise or a fall, not about a change in either direction, so the one-sided Z of 1.65 is usually the right choice.

Notice what the formula teaches

When…Then…
CVI is much larger than CVAImproving the method changes the RCV only a little
The two are similarImprecision matters a great deal

That is one reason analytical performance specifications are often set as a fraction of biological variation (Chapter 4).

📌 REMEMBER: Against a limit, bias and imprecision both matter. Against the previous result on the same method, constant bias cancels; imprecision and CVI remain, and the RCV combines them.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.7 Why the vocabulary matters for patients, pp.54–55

2.7 The logarithmic (asymmetric) RCV

The formula in the previous section treats a rise and a fall alike. Strictly, a percentage change is a ratio of two results, and ratios are not normally distributed.

💡 UNDERSTAND THIS: Working on a logarithmic scale corrects this and gives asymmetric limits: a fall needs a slightly smaller percentage change than a rise to reach the same confidence.

The two steps

Step 1. Convert each CV to a log-scale variance:

σ² = ln(CV² + 1)

Step 2. Calculate the two limits:

RCVfall, rise = 100% × [ exp( ∓ Z × √2 × √(σA² + σI²) ) − 1 ]
SymbolMeaning
σA², σI²Log-scale variances of the analytical and within-subject CVs (each CV used as a fraction in ln(CV² + 1))
∓The minus sign gives the limit for a fall; the plus sign gives the limit for a rise
ZAs before: 1.65 one-sided, 1.96 two-sided, at 95% confidence

When does it matter?

CVsDifference from the simple formula
A few percentSmall
Tens of percent (as for some hormones and urine measurands)Large

Turning the equation round

The same equation can also be turned round: solving it for Z with the observed change gives the probability that analytical and biological variation alone would produce a change that large. Some laboratories report that probability instead of a yes-or-no flag.

🧠 MEMORY AID: Log scale = lopsided limits. A fall reaches significance a little sooner than a rise of the same size.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.7 Why the vocabulary matters for patients (logarithmic form), p.55

Worked Example 2.1 (part 1): is the fall in cholesterol real?

⚠️ All patient data, imprecision figures and the biological variation value below are illustrative and synthetic. For real use, take the within-subject CV from the current EFLM Biological Variation Database and the analytical CV from your own quality control data.

Situation

A patient's total cholesterol is 6.5 mmol/L before treatment. Three months after starting a lipid-lowering drug it is 5.7 mmol/L. Both samples were analysed by the same laboratory with the same method. The clinician asks whether the fall is real.

💡 UNDERSTAND THIS: The question is directional (has it fallen?), so Z = 1.65 for 95% confidence.

Data

InputValueWhere it comes from
Analytical CV (CVA)3.0%Within-laboratory, from internal quality control near this level
Within-subject biological CV (CVI)6.0%Illustrative

Step 1 — Express the change as a percentage of the first result

(5.7 − 6.5) ⁄ 6.5 × 100 = −12.3%, a fall of 12.3%

Step 2 — The tempting shortcut: consider analytical variation only

Critical difference = √2 × 1.65 × 3.0 = 2.333 × 3.0 = 7.0%

The fall of 12.3% is larger than 7.0%. On this basis the change looks real.

⚠️ COMMON ERROR: This shortcut ignores the fact that the patient's own cholesterol varies from week to week even with no treatment at all.

Step 3 — Include biological variation: the reference change value

RCV = √2 × 1.65 × √(3.0² + 6.0²) = 2.333 × √45 = 2.333 × 6.708 = 15.7%

The fall of 12.3% is smaller than 15.7%.

📌 REMEMBER: Analytical variation alone: the fall looks real (12.3% > 7.0%). Add the patient's own variation: it does not (12.3% < 15.7%).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, Worked Example 2.1 Is the fall in cholesterol real? (Situation, Data, Steps 1–3), p.56

Worked Example 2.1 (part 2): a poorer method, the log check and the decision

Step 4 — See how method imprecision moves the threshold

Suppose the laboratory's CVA were 5.0% instead of 3.0%.

RCV = 2.333 × √(5.0² + 6.0²) = 2.333 × √61 = 2.333 × 7.810 = 18.2%
Threshold (Z = 1.65)CVA (%)CVI (%)Threshold (%)Observed fall exceeds it?
Analytical variation only3.0—7.0Yes
Reference change value3.06.015.7No
Reference change value, poorer method5.06.018.2No

Figure 2.3 The observed fall in cholesterol compared with three thresholds for a meaningful change.

Figure 2.3: horizontal bars for three thresholds against an observed fall of 12.3%. Analytical variation only 7.0% (exceeded); reference change value 15.7% with CVA 3.0% and CVI 6.0% (not exceeded); poorer method 18.2% with CVA 5.0% (not exceeded).

Synthetic, illustrative data · Worked Example 2.1 · formula after Fraser (2011)

🧪 HOW TO READ THIS FIGURE: Each grey bar is a threshold: the smallest percentage change that the stated sources of variation are unlikely to produce by chance. The horizontal axis is the change between two results as a percentage of the first. The blue vertical line is the fall actually observed, 12.3%. Where the line passes beyond the end of a bar, the change exceeds that threshold; where the bar extends past the line, the change is within normal variation. Only the top bar, which ignores the patient's own biological variation, is exceeded. Comparing the lower two bars shows how a less precise method raises the threshold.

Step 5 — Check with the logarithmic form

CVA = 3.0%CVA = 5.0%
σA²ln(1.0009) = 0.00090ln(1.0025) = 0.00250
σI² (CVI = 6.0%)ln(1.0036) = 0.00359ln(1.0036) = 0.00359
Limit for a fall (Z = 1.65)14.5%16.6%
Limit for a rise (Z = 1.65)16.9%20.0%

The observed fall of 12.3% is inside the limit either way, so the conclusion does not change.

Solving for Z instead gives Z = 1.39 with CVA = 3.0%: analytical and within-subject variation alone would produce a fall of at least 12.3% about 8 times in 100.

Interpretation

Taking both the method's scatter and the patient's natural variation into account, a fall of 12.3% could happen by chance more often than 1 time in 20 (about 8 times in 100, Step 5). The data do not show, at 95% confidence, that cholesterol has fallen.

💡 UNDERSTAND THIS: This does not mean the drug has failed. It means two results are not enough to tell. A further result, or a trend across several results, would settle the question.

Two lessons follow

  1. Ignoring biological variation makes changes look significant when they are not.
  2. Imprecision still matters: the poorer method raises the threshold from 15.7% to 18.2%, so a real change of, say, 17% would be detectable with one method and not the other.

Decision

🧪 Report the result without a "significant change" comment. If the laboratory adds interpretive comments, a comment such as "the change from the previous result is within the expected variation" is supportable. The clinical decision remains the clinician's.

Had the question been non-directional (has it changed either way?), Z = 1.96 would give an RCV of 18.6% with CVA = 3.0% (√2 × 1.96 = 2.772; 2.772 × 6.708 = 18.6%).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, Worked Example 2.1 Is the fall in cholesterol real? (Step 4, Figure 2.3, Step 5, Interpretation, Decision), pp.56–58

Common errors of interpretation

⚠️ Common errorWhy it is wrong / what to do
"Precise, therefore accurate."Tight replicates show low imprecision only. A precise method can be consistently wrong.
Reporting a bias without its comparator.A difference from a peer-group mean, from the old method and from a reference material are three different findings. Name the comparator.
Quoting an SD without its conditions.A within-run SD and a within-laboratory SD are not interchangeable, and neither can verify a claim made under the other.
Treating a mistake as method error.A spoiled sample says nothing about precision. Exclude it, record why, and look for the cause.
Judging bias only on average.A bias that is small at one concentration can be large at another if part of it is proportional.
Deciding whether two results differ by checking if their intervals overlap.Overlap of two uncertainty intervals is not the right test for a change. Use a critical difference or, for patient monitoring, the reference change value.
Ignoring biological variation when interpreting change.Analytical variation alone makes many changes look significant that are within a patient's normal fluctuation.
Using one RCV for every question.A factor of 2.77 assumes a two-sided question. Monitoring usually asks about a rise or a fall, which calls for Z = 1.65, and the CVA should match the interval between the samples.
📌 REMEMBER: Every common error here is a vocabulary error: a word or a figure used without its partner (comparator, conditions, concentration, direction, biological variation).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, Common errors of interpretation, pp.58–59

Chapter 2 checklist and references

Chapter 2 checklist

Before writing an evaluation plan or report, confirm that:

  • ☐ Every bias figure names its comparator: reference material, reference procedure, consensus value or current method.
  • ☐ Every precision figure states its conditions (repeatability, within-laboratory or reproducibility), the period covered and the concentration.
  • ☐ Bias and imprecision are reported at each clinical decision point, not only as overall averages.
  • ☐ Results spoiled by identified mistakes are excluded, listed and explained.
  • ☐ "Accuracy", "trueness" and "precision" are used in their defined senses (Table 2.1).
  • ☐ The plan states whether performance will be judged by total error, by measurement uncertainty, or by both.
  • ☐ Where results will be used for monitoring, the analytical CV and a source for the within-subject CV are on record.

Chapter 2 references (as listed in the book)

  1. Clinical and Laboratory Standards Institute. Expression of Measurement Uncertainty in Laboratory Medicine. 1st ed. CLSI guideline EP29-A. CLSI; 2012. Sections checked: 3.3 (definitions of measurement error, systematic and random error, bias, trueness, accuracy, uncertainty, metrological traceability), 4.3 (error and uncertainty models), 9.3 (clinical use).
  2. Clinical and Laboratory Standards Institute. User Verification of Precision and Estimation of Bias. 3rd ed. CLSI guideline EP15-A3. CLSI; 2014. Section 1.4 (terminology: trueness, bias, conventional quantity value, repeatability and intermediate precision conditions, within-laboratory imprecision).
  3. Clinical and Laboratory Standards Institute. Evaluation of Precision of Quantitative Measurement Procedures. 3rd ed. CLSI guideline EP05-A3. CLSI; 2014. Terminology section (precision, imprecision, repeatability, reproducibility).
  4. Clinical and Laboratory Standards Institute. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures. 2nd ed. CLSI guideline EP21-Ed2. CLSI; 2016. Terminology (accuracy, total analytical error) and Chapter 1.
  5. International Organization for Standardization. ISO 15189:2022 Medical laboratories — Requirements for quality and competence. 4th ed. ISO; 2022 (December). Entries 3.1, 3.18, 3.19 and 3.29; clause 7.3.4.
  6. Fraser CG. Reference change values. Clin Chem Lab Med. 2011;50(5):807–12. doi:10.1515/CCLM.2011.733 (record retrieved from PubMed).
  7. European Federation of Clinical Chemistry and Laboratory Medicine. EFLM Biological Variation Database. biologicalvariation.eu (accessed November 2026).
  8. Linnet K, Moons KGM, Boyd JC. Statistical methodologies in laboratory medicine: analytical and clinical evaluation of laboratory tests. In: Rifai N, Chiu RWK, Young I, Burnham CD, Wittwer CT, editors. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023. Chapter 2, pp. 12 (Box 2.1), 17 (trueness, accuracy, precision; Table 2.2) and 22–23 (basic error model; calibration and random bias; total error).
  9. Sandberg S, Røraas T, Aarsand AK. Biological variation and analytical performance specifications. In: Rifai N, et al., editors. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023. Chapter 8 (online), pp. 165.e11–165.e12 (reference change values).
  10. Joint Committee for Guides in Metrology. JCGM 200:2012 International vocabulary of metrology — Basic and general concepts and associated terms (VIM). 3rd ed. BIPM; 2012. doi:10.59161/JCGM200-2012. Entries 2.13 (measurement accuracy), 2.14 (measurement trueness), 2.15 (measurement precision) and 2.18 (measurement bias).

The vocabulary definitions in CLSI references 1–3 are taken from the International Vocabulary of Metrology (JCGM 200:2012). The VIM is now cited directly for the four core terms (reference 10). ISO 15189:2022 adopts the VIM entries for bias, measurement accuracy, measurement uncertainty and trueness, with notes added for medical laboratories (entries 3.1, 3.18, 3.19 and 3.29; reference 5).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, Chapter 2 checklist and Chapter 2 references, pp.59–60

Training course, not a certification. Based on the author's textbook; all examples use synthetic data. It does not replace the CLSI and ISO documents it cites, and it is not affiliated with or endorsed by CLSI or ISO.

© 2026 Hassan Shawki. Course text adapted from the author's textbook Clinical Laboratory Method Evaluation in Practice.

🖼️ PICTURE ATLAS

15 · Reading the charts — seeing bias, imprecision and meaningful change

The ideas of this chapter show up on the charts an evaluator reads every day: replicate strips, precision summaries, difference plots, control charts and threshold bars. Each card below shows one pattern with new synthetic data (not the book's figures). The quiz then asks you to read similar charts.

🧪 HOW TO READ A CHART: (1) Read the title and the axes: what is measured, in which unit? (2) Find the reference: the reference value, the zero-difference line, the target mean or the threshold. (3) Ask the two questions of the chapter: has the centre moved (bias, systematic error) or has the spread changed (imprecision, random error)? (4) Check where it happens: at which concentration or decision level, and under which conditions.

Bias and imprecision in replicate results

Twenty replicates per method5.65.866.2result (mmol/L)reference value 5.8True and precisebias +0.00 · SD 0.04True, imprecisebias +0.00 · SD 0.16Precise, biasedbias +0.25 · SD 0.04Biased, imprecisebias +0.25 · SD 0.16mean of the rowone result
Replicate strip: trueness vs precision🔎 Look for: where each row's mean (blue tick) sits against the dashed reference line, and how wide each row spreads💡 Think of: centre moved = bias (systematic error); wide row = imprecision (random error); both = single results far from the reference

Precision and its conditions

Three SDs for one control, three sets of conditions00.050.10.15SD (mmol/L)0.05repeatability0.08within-laboratory0.14reproducibility
SDs under three sets of conditions🔎 Look for: the three bars rise from repeatability to within-laboratory to reproducibility💡 Think of: each wider set of conditions lets more factors change, so the SD usually grows; compare like with like
Within-laboratory CV across the range0501001502000510concentration (units)CV (%)decision level 50
Precision profile across the range🔎 Look for: the CV climbing steeply at the low end and flattening higher up💡 Think of: report precision at the decision levels; a mid-range CV can hide poor precision at a low cut-off

Constant and proportional bias

Difference plot: new method minus comparator100200300400-505mean of the two methods (µmol/L)difference (µmol/L)mean 3.653
Difference plot: constant bias🔎 Look for: differences scattered around one level above zero, the same at low and high concentrations💡 Think of: every result shifted by about the same amount
Difference plot: new method minus comparator100200300400-20-1001020mean of the two methods (µmol/L)difference (µmol/L)mean 8.613
Difference plot: proportional bias🔎 Look for: differences that grow as concentration rises (a fan or slope)💡 Think of: shift by the same percentage; bias must be estimated at each decision level, not only on average

Control charts over time

Control results, 20 runs246810121416182098100102runresult (µmol/L)+3 SD+2 SD+1 SDmean-1 SD-2 SD-3 SD
Control chart: the mean shifts🔎 Look for: the points move to a new level while their scatter stays similar💡 Think of: a change in bias (systematic error), e.g. a calibrator or reagent lot behaving differently
Control results, 20 runs246810121416182097.5100102.5105107.5runresult (µmol/L)+3 SD+2 SD+1 SDmean-1 SD-2 SD-3 SD
Control chart: the spread widens🔎 Look for: the points stay centred but scatter more widely💡 Think of: a change in imprecision (random error), e.g. pipetting, bubbles or a worn part

Is a patient's change real?

Is the change larger than the thresholds?0510152025change between two results (% of the first result)Analytical onlyCVA 2.0%4.7%Reference change valueCVA 2.0% · CVI 8.0%19.2%observed change 14.3%
Threshold bars: analytical-only vs RCV🔎 Look for: whether the blue observed-change line passes beyond the end of each grey bar💡 Think of: the analytical-only bar is too short for monitoring; the reference change value adds the patient's own variation
⚠️ These charts are original teaching drawings made with synthetic data. Real reports look different from one system to another. Learn the PATTERN of each chart, because that is what stays the same.
🧠 ONE-LINE ATLAS: centre moved = bias · spread wider = imprecision · differences fan out with concentration = proportional bias · bigger box of conditions = bigger SD · line past the bar = change larger than the threshold.

Drawings: original teaching diagrams made for this page (no copyrighted images). Facts verified with: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.3–§2.7 and Worked Example 2.1, pp.47–58.