Start here: the question on your desk
Method evaluation runs on a small vocabulary: error, bias, precision, trueness, accuracy, uncertainty. People use these words loosely in conversation, and the looseness causes real mistakes:
- A precise method is called accurate.
- A bias is called an error.
- A difference between two analysers is blamed on the one that happens to be newer.
This chapter fixes the meaning of each term, shows how the terms relate, and connects them to decisions made about patients.
By the end of this chapter you should be able to
- Tell an error from a mistake, and both from uncertainty.
- Name the value a result is compared against, and explain why it is never exactly the true value.
- Separate systematic from random error, and say which statistic describes each.
- State the conditions behind any precision estimate.
- List common causes of bias and of imprecision in a routine laboratory.
- Explain how the total error and measurement uncertainty models combine the two kinds of error.
- Use analytical and biological variation to judge whether a change in a patient's results is real.
Map of this chapter
| # | Study section | Book section(s) | Printed pages |
|---|---|---|---|
| 1 | Start here: the question on your desk | Chapter 2 opening, learning objectives | 45–46 |
| 2 | Error, mistake and uncertainty are different things | §2.1 | 46 |
| 3 | The value we compare against, and traceability | §2.2 | 46–47 |
| 4 | Two kinds of error: systematic and random | §2.3 | 47–48 |
| 5 | The vocabulary at a glance (Table 2.1, Figure 2.1) | §2.3 (Table 2.1, Figure 2.1) | 49–50 |
| 6 | Precision always depends on the conditions (Figure 2.2) | §2.4 | 50–51 |
| 7 | Where bias and imprecision come from (Table 2.2) | §2.5 | 51–53 |
| 8 | Two ways to put the errors together | §2.6 | 53–54 |
| 9 | Why the vocabulary matters for patients: limits, change and the RCV | §2.7 (first part) | 54–55 |
| 10 | The logarithmic (asymmetric) RCV | §2.7 (second part) | 55 |
| 11 | Worked Example 2.1 (part 1): is the fall in cholesterol real? | Worked Example 2.1, Situation, Data, Steps 1–3 | 56 |
| 12 | Worked Example 2.1 (part 2): a poorer method, the log check and the decision | Worked Example 2.1, Step 4 (Figure 2.3), Step 5, interpretation, decision | 56–58 |
| 13 | Common errors of interpretation | Common errors of interpretation | 58–59 |
| 14 | Chapter 2 checklist and references | Chapter 2 checklist; Chapter 2 references | 59–60 |
| 15 | Reading the charts | Original practice charts built on the chapter's ideas | — |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, opening and learning objectives, pp.45–46; How this book is organised, p.viii
2.1 Error, mistake and uncertainty are different things
Three words are often used as if they meant the same thing. They do not, and an evaluation treats each one differently.
Measurement error
Measurement error is the difference between a measured value and the value it should have been.
- Every result has an error, and it has a sign: a result can sit above or below its reference value.
- For a single patient result the error is never known exactly, because the reference value is not known.
- What can be studied is the pattern of errors a method produces over many measurements. That is the subject of the book.
Mistake
A mistake is something that should not have happened: the wrong tube, a transposed digit, an expired reagent, a clot that blocks the sample probe.
| Mistakes… | What it means |
|---|---|
| are not part of a method's normal error | They do not follow its statistical pattern; precision or bias figures say nothing about how often they occur. |
| are prevented by process design | …and caught by checks such as specimen-identity procedures, instrument flags and result review. |
| in an evaluation study | A result spoiled by an identified mistake is excluded, but the exclusion and its reason are written down. |
| a pattern of mistakes | …during an evaluation is itself a finding: the method or its workflow may be prone to them. |
Measurement uncertainty
Measurement uncertainty describes how far the reference value could reasonably be from a result. CLSI EP29 defines it as a non-negative parameter that characterises the spread of values that could reasonably be attributed to the measurand.
| Error | Uncertainty |
|---|---|
| A single, unknown, signed number for one result | A range that says how large that error could plausibly be |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.1 Error, mistake and uncertainty are different things, p.46
2.2 The value we compare against
Every statement about error compares a result with something. The true value of a measurand in a patient sample is never known exactly.
In practice the comparison is made with a reference quantity value: the best available stand-in for the truth, chosen for a stated purpose. It can come from several places, and each carries a different weight.
| Comparator | What it is | What a difference from it shows |
|---|---|---|
| Reference measurement procedure or certified reference material | High in the hierarchy; comes with a stated uncertainty | The closest a laboratory gets to a true bias |
| Conventional quantity value | A value attributed by agreement for a given purpose, accepted as having a suitably small uncertainty (example: international units assigned to a reference preparation) | A difference from the agreed value |
| Consensus value (for example the mean of a peer group in an external quality assessment scheme) | Shows how a laboratory compares with others using similar methods | Not the truth: if the whole peer group is biased, the laboratory can agree with its peers and still be biased |
| Current routine method (the old method as comparator in a method comparison) | Often serves as the comparator | How results will change for clinicians; it does not show which method is closer to the truth |
Traceability
Routine results link to higher-order references through metrological traceability: an unbroken, documented chain of calibrations from the routine result up to a reference, each link adding its own uncertainty.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.2 The value we compare against (including Traceability), pp.46–47
2.3 Two kinds of error: systematic and random
Repeat a measurement on the same sample many times and look at the results. Two things stand out:
- The results cluster around some value, which may or may not be the reference value.
- They scatter around that cluster.
These two features are the two components of measurement error.
| Systematic error | Random error | |
|---|---|---|
| Definition | The part of the error that stays the same, or changes in a predictable way, from one replicate to the next | The part that changes unpredictably from one replicate to the next |
| Effect on the cluster | Moves the whole cluster | Spreads the cluster |
| Its estimate | Bias: the difference between the average of many results and the reference value | Imprecision: the standard deviation (SD) of replicate results, or the coefficient of variation (CV), the SD as a percentage of the mean |
| Reported as | In the measurement unit (for example +0.20 mmol/L) or as a percentage of the reference value | SD in units, or CV in % |
| Can it be corrected? | Because it is predictable, a known systematic error can in principle be corrected | No; averaging more replicates reduces its effect on a mean |
| Does averaging help? | Averaging does nothing to systematic error | Yes, for the mean |
Three quality words
| Quality | What it is | Kind of error | Expressed through |
|---|---|---|---|
| Trueness | How close the average of many results comes to the reference value | Systematic only | Bias |
| Precision | How close replicate results are to one another under stated conditions | Random only | Imprecision |
| Accuracy | How close a single result comes to the true value | Both kinds | (depends on both) |
The link to basic statistics
| Statistic | Tracks | What a change signals |
|---|---|---|
| Mean of replicate results | Systematic error | A shift in the mean of control results signals a change in bias |
| SD | Random error | A widening of the spread signals a change in imprecision |
Chapter 3 covers the calculations.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.3 Two kinds of error: systematic and random, pp.47–48
The vocabulary of error at a glance (Table 2.1, Figure 2.1)
Table 2.1 The vocabulary of error at a glance
| Term | What it describes | How it is expressed | Kind of error |
|---|---|---|---|
| Measurement error | Measured value minus reference value, for one result | A signed value in units; unknown for a single patient result | Both |
| Bias | Estimated systematic error | Mean minus reference value, in units or % | Systematic |
| Imprecision | Scatter of replicate results | SD in units, or CV in % | Random |
| Trueness | Closeness of the long-run average to the reference value | Through bias (not a number itself) | Systematic |
| Precision | Closeness of replicates to one another under stated conditions | Through imprecision (not a number itself) | Random |
| Accuracy | Closeness of a single result to the true value | Through total error or uncertainty (not a number itself) | Both |
| Measurement uncertainty | Spread of values that could reasonably be the true value | Standard uncertainty u, or expanded uncertainty U | Both, once known biases are corrected or included |
Figure 2.1 Trueness and precision are separate qualities: four sets of replicate potassium results.
Synthetic, illustrative data · 20 replicates per set · reference value 4.50 mmol/L
| Row | Bias | SD | What it shows |
|---|---|---|---|
| True and precise | 0.00 | 0.03 mmol/L | Centred on the reference value and tight |
| True but imprecise | 0.00 | 0.12 mmol/L | Also centred (good trueness), but four times as scattered |
| Precise but biased | +0.20 | 0.03 mmol/L | Tight, but sits 0.20 mmol/L too high |
| Biased and imprecise | +0.20 | 0.12 mmol/L | Both problems: a single result can be far from the reference value |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.3 Two kinds of error (Table 2.1, Figure 2.1), pp.49–50
2.4 Precision always depends on the conditions
An SD on its own means little. The same method can give:
| How the sample was run | SD |
|---|---|
| One sample run ten times in a row | 0.03 mmol/L |
| That sample run once a day for a month | 0.06 mmol/L |
Both figures are correct. They describe precision under different conditions, and every precision estimate must say which conditions it covers. The CLSI documents use three levels, taken from the VIM and ISO statistical standards.
| Conditions | What stays the same / what changes | Other names |
|---|---|---|
| Repeatability | Same procedure, same operator, same measuring system, same operating conditions, same place, over a short time. In practice this is one analytical run. | Older documents: within-run precision |
| Intermediate precision | Same procedure and same laboratory, but some operating conditions change: the day, the calibration, the operator or the equipment. | Collected over enough time to include routine changes in reagent and calibrator lots, CLSI EP15 calls it within-laboratory imprecision, formerly "total precision" |
| Reproducibility | Different locations, operators and measuring systems. | The scatter seen between laboratories, for example in an external quality assessment scheme |
Figure 2.2 The three sets of precision conditions, nested from the narrowest to the widest.
Original diagram · condition definitions from CLSI EP05-A3 and EP15-A3 (after JCGM 200:2012)
Three practical rules follow
- Compare like with like. A laboratory's within-run SD cannot verify a manufacturer's within-laboratory claim, and vice versa. Check which condition a claim describes before designing the study (Chapter 6).
- Choose the condition that matches the use. Patients are usually tested on different days, after recalibrations and lot changes. The within-laboratory SD is therefore the figure that describes the scatter a clinician will actually see between a patient's results.
- Expect precision to vary with concentration. The SD often grows as concentration rises, while the CV may be roughly constant over part of the range and rise steeply near the lower limit. Report precision at the concentrations where decisions are made.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.4 Precision always depends on the conditions (Figure 2.2), pp.50–51
2.5 Where bias and imprecision come from
Knowing which kind of error a fault produces speeds up troubleshooting and helps an evaluator design studies that will expose it.
Table 2.2 Common sources of systematic and random error
| Source | Typically shows as systematic error (bias) when… | Typically shows as random error (imprecision) when… |
|---|---|---|
| Calibration | A calibrator value is wrong, a calibrator has deteriorated, or a new calibrator lot is assigned differently | Calibration is repeated often and each calibration lands slightly differently |
| Reagents | A new reagent lot behaves differently, or a reagent slowly deteriorates on board (a drift) | Reagent is poorly mixed, or its performance varies across the time it is open |
| Control material | It is reconstituted wrongly, a new lot has different values, or it ages | Reconstitution or thawing varies from vial to vial |
| Instrument | Wavelength, temperature or a pipetting volume is set or calibrated wrongly | Pipetting, bubbles, a worn part or an intermittent fault varies from test to test |
| Sample | A substance in the sample adds to or suppresses the signal in every sample that contains it | Carryover from the previous sample, or clots and bubbles, affect some tests and not others |
| Environment | Room temperature or water quality shifts and stays shifted | Temperature or power fluctuates through the day |
| Operator | A step in the procedure is done consistently wrong | Manual steps, such as timing or pipetting, are done slightly differently each time |
Three further points matter for evaluation
1. Bias need not be a single number.
| Constant bias | Proportional bias |
|---|---|
| Shifts every result by the same amount, for example 0.1 mmol/L | Shifts results by the same percentage, so the shift grows with concentration |
Many methods show some of both. That is why bias must be estimated at each decision point, not only on average. Chapter 8 shows how a method comparison separates the two.
2. The line between the two kinds of error depends on the time scale. CLSI EP29 notes that a bias which changes with each reagent or calibrator lot behaves like random variation when viewed over a long period.
State the period an estimate covers, and interpret it in that light.
3. Part of a method's bias can belong to the individual sample rather than to the calibration. The Tietz Textbook of Laboratory Medicine splits the error of a single patient result into three parts:
| Part | What it is | Seen by control and reference materials? |
|---|---|---|
| Calibration bias | Shared by all samples | Yes |
| Sample-related random bias | Differs from one patient sample to the next, because of interfering substances or matrix effects | No |
| Random measurement error | What replicates show | Yes |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.5 Where bias and imprecision come from (Table 2.2), pp.51–53
2.6 Two ways to put the errors together
A clinician receives one result, not a bias and an SD. At some point the two kinds of error have to be combined into a single statement about how far a result might be wrong. Laboratory medicine uses two models for this. CLSI EP29 describes both.
The error model (total analytical error)
- Treats the true value as a single unknown number and keeps the two errors separate until the end.
- It then adds them: the bias plus a multiple of the SD gives an upper limit on the error of a single result, assuming random errors follow a normal distribution.
- Sources differ on the multiplier: a traditional form adds 2 SD, a one-sided view uses 1.65 SD, and a smaller chance of exceeding the limit needs a larger one.
- This is the total analytical error used in Worked Example 1.1, and CLSI EP21 develops it further, including an approach that needs no distributional assumption.
| Strength | Weakness (EP29) |
|---|---|
| Simplicity: one figure that can be set against an allowable limit | A total error cannot be carried into a calculation that uses the result, such as a ratio or a clearance. For that, it must first be split back into its parts. |
The uncertainty model
- Describes an interval within which the true value is expected to lie, with a stated level of confidence.
- Assumes that known biases are corrected, or at least identified, and treats the uncertainty of each correction as one more component to combine with the random components.
- The combined standard uncertainty is multiplied by a coverage factor to give the expanded uncertainty.
- This is the approach of the international guide to the expression of uncertainty in measurement (GUM).
The two models answer slightly different questions
| Question | Usually answered with |
|---|---|
| Is this method good enough for its intended use? | Evaluations: the error model, estimated total error against an allowable total error set in advance |
| How sure can we be about this result? | Interpretation and reporting: the uncertainty model |
ISO 15189:2022 clause 7.3.4
ISO 15189:2022 treats the two models as complementary. Under clause 7.3.4 the laboratory must:
- evaluate the measurement uncertainty of its measured values where relevant and maintain it for the intended use;
- compare that uncertainty with performance specifications;
- document it, review it regularly and record why any examination is left out.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.6 Two ways to put the errors together, pp.53–54
2.7 Why the vocabulary matters for patients: limits, change and the RCV
A clinician uses a numerical result in one of two ways. The kind of error that matters most differs between them.
1. Comparing a result with a limit
Is this troponin above the cut-off? Is this glucose in the diabetic range? Here both bias and imprecision matter.
| Error | Effect near a limit |
|---|---|
| Bias | Moves every result towards or away from the limit, so a whole population of patients is classified differently |
| Imprecision | Scatters individual results across the limit: a patient whose true value sits just below it may be reported just above it on one occasion and below it on the next |
The closer the true values cluster around a cut-off, the more patients are affected. This is why the imprecision of a test used at a low diagnostic cut-off, such as cardiac troponin, has to be judged at that cut-off rather than in the middle of the range. Chapters 10 and 14 return to this case.
2. Comparing a result with the patient's previous result
Has the cholesterol fallen since treatment started? Is the creatinine rising? If both results come from the same method, a constant bias largely cancels out. What remains is:
- imprecision, plus
- something the laboratory cannot control: the natural day-to-day variation within the patient, called within-subject biological variation (CVI).
A change is meaningful only if it is larger than these two sources of variation could produce together.
The reference change value (RCV)
The reference change value (RCV) puts this into a formula. It gives the percentage change between two results that is unlikely to arise from analytical and within-subject variation alone:
| Symbol | Meaning |
|---|---|
| CVA | The laboratory's analytical CV (its within-laboratory imprecision) |
| CVI | The within-subject biological CV |
| Z | Sets the level of confidence: 1.96 for "has the result changed in either direction?" at 95% confidence; 1.65 for "has it changed in the expected direction?" at the same confidence |
| √2 | Appears because two results, each carrying its own variation, are being compared |
Published, quality-appraised estimates of CVI are collected in the EFLM Biological Variation Database.
Three conditions make the RCV trustworthy
- The CVA should come from the laboratory's own long-term internal quality control, at a concentration near the decision point, over a period comparable to the interval between the two samples. A long-term CVA already contains the small shifts caused by recalibration and lot changes.
- Preanalytical conditions (patient preparation, collection, transport and handling) are assumed to be controlled and constant.
- The CVI must come from a population similar to the patient's.
Notice what the formula teaches
| When… | Then… |
|---|---|
| CVI is much larger than CVA | Improving the method changes the RCV only a little |
| The two are similar | Imprecision matters a great deal |
That is one reason analytical performance specifications are often set as a fraction of biological variation (Chapter 4).
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.7 Why the vocabulary matters for patients, pp.54–55
2.7 The logarithmic (asymmetric) RCV
The formula in the previous section treats a rise and a fall alike. Strictly, a percentage change is a ratio of two results, and ratios are not normally distributed.
The two steps
Step 1. Convert each CV to a log-scale variance:
Step 2. Calculate the two limits:
| Symbol | Meaning |
|---|---|
| σA², σI² | Log-scale variances of the analytical and within-subject CVs (each CV used as a fraction in ln(CV² + 1)) |
| ∓ | The minus sign gives the limit for a fall; the plus sign gives the limit for a rise |
| Z | As before: 1.65 one-sided, 1.96 two-sided, at 95% confidence |
When does it matter?
| CVs | Difference from the simple formula |
|---|---|
| A few percent | Small |
| Tens of percent (as for some hormones and urine measurands) | Large |
Turning the equation round
The same equation can also be turned round: solving it for Z with the observed change gives the probability that analytical and biological variation alone would produce a change that large. Some laboratories report that probability instead of a yes-or-no flag.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.7 Why the vocabulary matters for patients (logarithmic form), p.55
Worked Example 2.1 (part 1): is the fall in cholesterol real?
Situation
A patient's total cholesterol is 6.5 mmol/L before treatment. Three months after starting a lipid-lowering drug it is 5.7 mmol/L. Both samples were analysed by the same laboratory with the same method. The clinician asks whether the fall is real.
Data
| Input | Value | Where it comes from |
|---|---|---|
| Analytical CV (CVA) | 3.0% | Within-laboratory, from internal quality control near this level |
| Within-subject biological CV (CVI) | 6.0% | Illustrative |
Step 1 — Express the change as a percentage of the first result
Step 2 — The tempting shortcut: consider analytical variation only
The fall of 12.3% is larger than 7.0%. On this basis the change looks real.
Step 3 — Include biological variation: the reference change value
The fall of 12.3% is smaller than 15.7%.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, Worked Example 2.1 Is the fall in cholesterol real? (Situation, Data, Steps 1–3), p.56
Worked Example 2.1 (part 2): a poorer method, the log check and the decision
Step 4 — See how method imprecision moves the threshold
Suppose the laboratory's CVA were 5.0% instead of 3.0%.
| Threshold (Z = 1.65) | CVA (%) | CVI (%) | Threshold (%) | Observed fall exceeds it? |
|---|---|---|---|---|
| Analytical variation only | 3.0 | — | 7.0 | Yes |
| Reference change value | 3.0 | 6.0 | 15.7 | No |
| Reference change value, poorer method | 5.0 | 6.0 | 18.2 | No |
Figure 2.3 The observed fall in cholesterol compared with three thresholds for a meaningful change.
Synthetic, illustrative data · Worked Example 2.1 · formula after Fraser (2011)
Step 5 — Check with the logarithmic form
| CVA = 3.0% | CVA = 5.0% | |
|---|---|---|
| σA² | ln(1.0009) = 0.00090 | ln(1.0025) = 0.00250 |
| σI² (CVI = 6.0%) | ln(1.0036) = 0.00359 | ln(1.0036) = 0.00359 |
| Limit for a fall (Z = 1.65) | 14.5% | 16.6% |
| Limit for a rise (Z = 1.65) | 16.9% | 20.0% |
The observed fall of 12.3% is inside the limit either way, so the conclusion does not change.
Solving for Z instead gives Z = 1.39 with CVA = 3.0%: analytical and within-subject variation alone would produce a fall of at least 12.3% about 8 times in 100.
Interpretation
Taking both the method's scatter and the patient's natural variation into account, a fall of 12.3% could happen by chance more often than 1 time in 20 (about 8 times in 100, Step 5). The data do not show, at 95% confidence, that cholesterol has fallen.
Two lessons follow
- Ignoring biological variation makes changes look significant when they are not.
- Imprecision still matters: the poorer method raises the threshold from 15.7% to 18.2%, so a real change of, say, 17% would be detectable with one method and not the other.
Decision
Had the question been non-directional (has it changed either way?), Z = 1.96 would give an RCV of 18.6% with CVA = 3.0% (√2 × 1.96 = 2.772; 2.772 × 6.708 = 18.6%).
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, Worked Example 2.1 Is the fall in cholesterol real? (Step 4, Figure 2.3, Step 5, Interpretation, Decision), pp.56–58
Common errors of interpretation
| ⚠️ Common error | Why it is wrong / what to do |
|---|---|
| "Precise, therefore accurate." | Tight replicates show low imprecision only. A precise method can be consistently wrong. |
| Reporting a bias without its comparator. | A difference from a peer-group mean, from the old method and from a reference material are three different findings. Name the comparator. |
| Quoting an SD without its conditions. | A within-run SD and a within-laboratory SD are not interchangeable, and neither can verify a claim made under the other. |
| Treating a mistake as method error. | A spoiled sample says nothing about precision. Exclude it, record why, and look for the cause. |
| Judging bias only on average. | A bias that is small at one concentration can be large at another if part of it is proportional. |
| Deciding whether two results differ by checking if their intervals overlap. | Overlap of two uncertainty intervals is not the right test for a change. Use a critical difference or, for patient monitoring, the reference change value. |
| Ignoring biological variation when interpreting change. | Analytical variation alone makes many changes look significant that are within a patient's normal fluctuation. |
| Using one RCV for every question. | A factor of 2.77 assumes a two-sided question. Monitoring usually asks about a rise or a fall, which calls for Z = 1.65, and the CVA should match the interval between the samples. |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, Common errors of interpretation, pp.58–59
Chapter 2 checklist and references
Chapter 2 checklist
Before writing an evaluation plan or report, confirm that:
- ☐ Every bias figure names its comparator: reference material, reference procedure, consensus value or current method.
- ☐ Every precision figure states its conditions (repeatability, within-laboratory or reproducibility), the period covered and the concentration.
- ☐ Bias and imprecision are reported at each clinical decision point, not only as overall averages.
- ☐ Results spoiled by identified mistakes are excluded, listed and explained.
- ☐ "Accuracy", "trueness" and "precision" are used in their defined senses (Table 2.1).
- ☐ The plan states whether performance will be judged by total error, by measurement uncertainty, or by both.
- ☐ Where results will be used for monitoring, the analytical CV and a source for the within-subject CV are on record.
Chapter 2 references (as listed in the book)
- Clinical and Laboratory Standards Institute. Expression of Measurement Uncertainty in Laboratory Medicine. 1st ed. CLSI guideline EP29-A. CLSI; 2012. Sections checked: 3.3 (definitions of measurement error, systematic and random error, bias, trueness, accuracy, uncertainty, metrological traceability), 4.3 (error and uncertainty models), 9.3 (clinical use).
- Clinical and Laboratory Standards Institute. User Verification of Precision and Estimation of Bias. 3rd ed. CLSI guideline EP15-A3. CLSI; 2014. Section 1.4 (terminology: trueness, bias, conventional quantity value, repeatability and intermediate precision conditions, within-laboratory imprecision).
- Clinical and Laboratory Standards Institute. Evaluation of Precision of Quantitative Measurement Procedures. 3rd ed. CLSI guideline EP05-A3. CLSI; 2014. Terminology section (precision, imprecision, repeatability, reproducibility).
- Clinical and Laboratory Standards Institute. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures. 2nd ed. CLSI guideline EP21-Ed2. CLSI; 2016. Terminology (accuracy, total analytical error) and Chapter 1.
- International Organization for Standardization. ISO 15189:2022 Medical laboratories — Requirements for quality and competence. 4th ed. ISO; 2022 (December). Entries 3.1, 3.18, 3.19 and 3.29; clause 7.3.4.
- Fraser CG. Reference change values. Clin Chem Lab Med. 2011;50(5):807–12. doi:10.1515/CCLM.2011.733 (record retrieved from PubMed).
- European Federation of Clinical Chemistry and Laboratory Medicine. EFLM Biological Variation Database. biologicalvariation.eu (accessed November 2026).
- Linnet K, Moons KGM, Boyd JC. Statistical methodologies in laboratory medicine: analytical and clinical evaluation of laboratory tests. In: Rifai N, Chiu RWK, Young I, Burnham CD, Wittwer CT, editors. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023. Chapter 2, pp. 12 (Box 2.1), 17 (trueness, accuracy, precision; Table 2.2) and 22–23 (basic error model; calibration and random bias; total error).
- Sandberg S, Røraas T, Aarsand AK. Biological variation and analytical performance specifications. In: Rifai N, et al., editors. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023. Chapter 8 (online), pp. 165.e11–165.e12 (reference change values).
- Joint Committee for Guides in Metrology. JCGM 200:2012 International vocabulary of metrology — Basic and general concepts and associated terms (VIM). 3rd ed. BIPM; 2012. doi:10.59161/JCGM200-2012. Entries 2.13 (measurement accuracy), 2.14 (measurement trueness), 2.15 (measurement precision) and 2.18 (measurement bias).
The vocabulary definitions in CLSI references 1–3 are taken from the International Vocabulary of Metrology (JCGM 200:2012). The VIM is now cited directly for the four core terms (reference 10). ISO 15189:2022 adopts the VIM entries for bias, measurement accuracy, measurement uncertainty and trueness, with notes added for medical laboratories (entries 3.1, 3.18, 3.19 and 3.29; reference 5).
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, Chapter 2 checklist and Chapter 2 references, pp.59–60
Training course, not a certification. Based on the author's textbook; all examples use synthetic data. It does not replace the CLSI and ISO documents it cites, and it is not affiliated with or endorsed by CLSI or ISO.
© 2026 Hassan Shawki. Course text adapted from the author's textbook Clinical Laboratory Method Evaluation in Practice.
15 · Reading the charts — seeing bias, imprecision and meaningful change
The ideas of this chapter show up on the charts an evaluator reads every day: replicate strips, precision summaries, difference plots, control charts and threshold bars. Each card below shows one pattern with new synthetic data (not the book's figures). The quiz then asks you to read similar charts.
Bias and imprecision in replicate results
Precision and its conditions
Constant and proportional bias
Control charts over time
Is a patient's change real?
Drawings: original teaching diagrams made for this page (no copyrighted images). Facts verified with: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, §2.3–§2.7 and Worked Example 2.1, pp.47–58.
⚡ Last-Minute Review — The Language of Error
Read these lists before you write an evaluation plan or report. Then open Progress / Results to see your weak areas.
The decision
- Which kind of error are we measuring, and what does each word mean?
- The vocabulary: error, bias, precision, trueness, accuracy, uncertainty. Loose use causes real mistakes: precise called accurate; a bias called an error; the newer analyser blamed.
- Definitions follow the VIM (JCGM 200:2012): accuracy 2.13 · trueness 2.14 · precision 2.15 · bias 2.18.
Error, mistake, uncertainty (§2.1)
- Error = measured value − reference value; every result has one, with a sign; never known exactly for one patient result. Studies look at the pattern of errors.
- Mistake = should not have happened (wrong tube, transposed digit, expired reagent, clot). Not part of method error; prevented by process design, caught by checks. Exclude a spoiled result and record why; a pattern of mistakes is a finding.
- Uncertainty (EP29) = non-negative parameter for the spread of values reasonably attributed to the measurand: a range, not one signed number.
- Error = how far you missed · uncertainty = how far you might have missed · mistake = aiming at the wrong target.
The comparator (§2.2)
- True value never known → reference quantity value, chosen for a stated purpose.
- Reference procedure / certified reference material (closest to true bias) · conventional value (by agreement, e.g. international units) · consensus value (peer group: not the truth) · current method (shows change for clinicians, not truth).
- Biased compared with what? Name the comparator every time.
- Traceability = unbroken documented chain of calibrations, each link adding uncertainty. It shows the chain exists, not that the end uncertainty is small enough; two traceable methods can still disagree on patient samples.
Systematic vs random (§2.3, Table 2.1)
- Systematic error: same or predictable → moves the cluster → bias (mean − reference value, units or %); a known one can be corrected; averaging does nothing.
- Random error: unpredictable → spreads the cluster → imprecision (SD, or CV = SD as % of mean); averaging reduces its effect on a mean.
- Trueness (average vs reference; systematic; via bias) · Precision (replicates vs each other, stated conditions; random; via imprecision) · Accuracy (a single result vs true value; both). None is a number in itself.
- Uncertainty expressed as u (standard) or U (expanded); covers both kinds once known biases are corrected or included.
- Shift in control mean → change in bias · widening spread → change in imprecision.
Precision conditions (§2.4, Figure 2.2)
- Repeatability (one run; formerly within-run) ⊂ intermediate / within-laboratory (day, calibration, operator, equipment, lots change; EP15 within-laboratory imprecision, formerly "total precision") ⊂ reproducibility (between laboratories, e.g. EQA).
- Repeatability SD normally smallest, reproducibility SD largest.
- Rules: compare like with like · choose the condition that matches the use (within-laboratory SD = what the clinician sees) · expect precision to vary with concentration (SD grows; CV roughly constant, rising steeply near the lower limit) — report at decision concentrations.
Sources of error (§2.5, Table 2.2)
- Acts the same way every time → bias · acts differently each time → imprecision · many sources can do either.
- Calibration · reagents · control material · instrument · sample · environment · operator.
- Constant bias (same amount) vs proportional bias (same %; grows with concentration) → estimate bias at each decision point.
- Time scale: a lot-to-lot bias looks fixed in one week, like scatter over six months (EP29). State the period.
- Tietz: calibration bias (all samples) + sample-related random bias (interferents, matrix) + random measurement error. Controls and reference materials miss the middle term; patient comparisons expose it. An outlier is a finding to investigate, not a point to delete.
Combining the errors (§2.6)
- Error model: bias + multiple of SD (2 SD traditional, 1.65 SD one-sided; smaller chance → larger multiplier) = total analytical error vs allowable limit; EP21 (incl. a distribution-free approach). Cannot be carried into a ratio or clearance without splitting into parts.
- Uncertainty model (GUM): known biases corrected or identified; combined standard uncertainty × coverage factor = expanded uncertainty.
- Good enough for intended use? → error model · how sure about this result? → uncertainty model.
- ISO 15189:2022 clause 7.3.4: evaluate and maintain MU where relevant, compare with specifications, document, review, record exclusions; considering MU during verification/validation is recommended, not required; answers to users allow for biological variation.
Patients: limits and change (§2.7, Worked Example 2.1)
- Against a limit: bias reclassifies a population; imprecision scatters individuals across the cut-off → judge imprecision at the cut-off (e.g. troponin).
- Against the previous result (same method): constant bias largely cancels; imprecision + CVI remain.
- RCV (%) = √2 × Z × √(CVA² + CVI²) · Z = 1.96 either direction · Z = 1.65 expected direction (95%).
- Trustworthy RCV: CVA from long-term IQC near the decision point over a comparable interval · controlled, constant preanalytical conditions · CVI from a similar population (EFLM database).
- The habit "× 2.77" = two-sided Z 1.96; monitoring usually needs one-sided Z 1.65.
- CVI ≫ CVA → improving the method changes RCV little · similar → imprecision matters a lot.
- Log form: σ² = ln(CV² + 1); RCVfall, rise = 100% × [exp(∓Z√2√(σA² + σI²)) − 1]; asymmetric (a fall needs a slightly smaller %); matters when CVs reach tens of percent; solve for Z → probability.
- Worked Example 2.1: 6.5 → 5.7 mmol/L = −12.3% · analytical only 7.0% (exceeded) · RCV 15.7% (not exceeded) · CVA 5.0% → 18.2% · log limits fall 14.5% / rise 16.9% (16.6% / 20.0%) · Z = 1.39, about 8 in 100 · two-sided RCV 18.6%.
- Decision: no "significant change" comment; "within the expected variation" is supportable; two results are not enough to tell; the clinical decision is the clinician's.
Common errors
- "Precise, therefore accurate" · bias without comparator · SD without conditions · mistake treated as method error · bias judged only on average · overlapping intervals used to judge change · biological variation ignored · one RCV (2.77) for every question.
Checklist
- Comparator named for every bias · conditions, period and concentration for every SD · bias and imprecision at each decision point · mistakes excluded, listed, explained · words used in defined senses · total error, MU or both stated in the plan · CVA and a CVI source on record for monitoring.
Key terms (book glossary, Chapter 2)
- Accuracy — How close a single result is to the true value. It combines trueness and precision, so it is a quality, not a number to report.
- Bias — The average difference between the results of a method and a reference value. It is an estimate, and has its own confidence interval.
- Biological variation — The natural variation of a measurand within one person over time (within-subject) and between people (between-subject). It is one basis for performance specifications and for judging change in a patient.
- Imprecision — The scatter of repeated results, usually expressed as an SD or CV. Precision is the closeness of repeated results; imprecision is the number that measures its lack.
- Measurand — The quantity intended to be measured, such as the substance concentration of creatinine in plasma.
- Measurement uncertainty — A parameter, such as a standard deviation, that describes the spread of values that could reasonably be attributed to the measurand.
- Metrological traceability — The link, through an unbroken chain of calibrations with stated uncertainties, between a result and a reference.
- Reference change value — The smallest difference between two results from one patient that is unlikely to be explained by analytical and within-subject variation alone.
- Repeatability — Precision under the same conditions over a short time: same operator, analyser, run and reagents.
- Reproducibility — Precision under conditions that include different laboratories, operators and measuring systems.
- Systematic and random error — Error that is constant or proportional across results (systematic), and error that varies unpredictably from result to result (random).
- Total analytical error — The combined effect of bias and imprecision on a single result.
- Trueness — How close the average of many results is to the true value. Bias is the usual measure of its lack.
- Within-laboratory precision — Precision over days, runs, calibrations and operators in one laboratory; also called intermediate precision.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.2, pp.45–60; Glossary, pp.411–419
🖼️ Picture clues (Study Section 15)
- Replicate strip: centre away from the reference line = bias; wide strip = imprecision; both = poor single-result accuracy.
- SDs from different conditions: repeatability smallest, within-laboratory larger, reproducibility largest.
- Precision profile: CV often climbs at the low end; judge it at the decision level.
- Difference plot: flat offset = constant bias; differences growing with concentration = proportional bias.
- Control chart: mean shifts = change in bias; spread widens = change in imprecision.
- Threshold bars: the observed change counts only if the line passes beyond the end of the RCV bar.