Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 key point · 📌 remember · ⚠️ common error · 💡 understand this · 🧪 at the bench · 🧠 memory aid.
Start here: the question on your desk
An evaluation produces numbers, but the laboratory needs decisions. Between the two sits a small set of calculations.
- Most are taught early in a laboratory career, usually as formulas to apply to quality control data.
- What is taught less often is which calculation answers which question, and how much trust each answer deserves. This chapter covers that ground.
- It does not replace a statistics text or evaluation software. It explains what the software is doing, so that you can choose the right calculation, read the output and catch the mistakes that software cannot.
The examples follow the precision, comparison and agreement designs of the CLSI evaluation documents (EP15, EP05, EP09c, EP12). The wider statistical background follows the Tietz Textbook of Laboratory Medicine.
By the end of this chapter you should be able to
- Match each evaluation question to the calculation that answers it.
- Calculate a mean, SD and CV correctly, and say when the median is the better summary.
- Separate repeatability from between-day variation with a one-way analysis of variance.
- Attach a confidence interval to a mean or an SD, and explain what a verification limit does.
- Tell a statistically significant difference from one that matters clinically.
- Explain why a correlation coefficient says nothing about agreement.
- Report agreement between qualitative tests with a confidence interval.
Map of this chapter
| # | Study section | Book section(s) | Printed pages |
|---|---|---|---|
| 1 | Start here: the question on your desk | Chapter 3 opening, learning objectives | 61–62 |
| 2 | Start with the question, not the formula (Table 3.1) | §3.1 | 62 |
| 3 | Describing a set of results: mean, median, SD and CV | §3.2 (first part) | 62–63 |
| 4 | A result far from the others: outliers | §3.2 (second part) | 63 |
| 5 | Variances add; SDs do not | §3.3 (first part) | 63–64 |
| 6 | Separating the components: one-way ANOVA and the standard designs | §3.3 (second part) | 64 |
| 7 | How sure are we? Confidence intervals (Figure 3.1) | §3.4 (first part) | 64–65 |
| 8 | The upper verification limit | §3.4 (second part) | 65–66 |
| 9 | Significant is not the same as important (Table 3.2) | §3.5 | 66 |
| 10 | Two methods: correlation is not agreement | §3.6 | 67 |
| 11 | Counting agreement: qualitative results | §3.7 | 67 |
| 12 | Worked Example 3.1 (part 1): the data and the outlier check (Table 3.3, Figure 3.2) | Worked Example 3.1, Situation, Steps 1–2 | 68–69 |
| 13 | Worked Example 3.1 (part 2): ANOVA, the claims and the confidence intervals (Table 3.4) | Worked Example 3.1, Steps 3–5 | 69–70 |
| 14 | Worked Example 3.1 (part 3): two common slips and the decision | Worked Example 3.1, Step 6, Decision, What this example shows | 70 |
| 15 | Common errors | Common errors | 70 |
| 16 | Chapter 3 checklist and references | Chapter 3 checklist; Chapter 3 references | 71 |
| 17 | Reading the charts | Original practice charts built on the chapter's ideas | — |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, opening and learning objectives, pp.61–62; How this book is organised, p.viii
3.1 Start with the question, not the formula
Every calculation in an evaluation answers one question. Choosing the calculation first and the question afterwards causes many of the errors found in verification reports.
Table 3.1 Which calculation answers which question
| Question | Calculation | Watch out for | Section |
|---|---|---|---|
| Where is the centre of the results? | Mean; median if the data are skewed or contain an extreme value | One extreme result pulls the mean | 3.2 |
| How scattered are they? | SD (divisor n − 1) and CV | State the conditions; a CV is unstable near zero | 3.2 |
| Where does the scatter come from? | One-way analysis of variance | Variances add; SDs do not | 3.3 |
| How sure are we of an estimate? | Confidence interval (t for a mean, chi-square for an SD) | Small studies give wide intervals | 3.4 |
| Do our results agree with a claim? | Comparison with a verification limit | About half of all studies exceed a true claim by chance | 3.4 |
| Does a difference matter? | The estimate and its interval, compared with an allowable limit | Significant is not the same as important | 3.5 |
| How do two methods relate? | Difference plot; Deming or Passing–Bablok regression | A high r says nothing about bias | 3.6 |
| How well do two qualitative tests agree? | Positive and negative percent agreement with Wilson intervals | Few positives give wide intervals | 3.7 |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.1 Start with the question, not the formula (Table 3.1), p.62
3.2 Describing a set of results: mean, median, SD and CV
The centre: mean and median
| Statistic | What it is | When to use it |
|---|---|---|
| Mean (x̄) | The sum of the results divided by their number. It estimates where the results centre. | The usual summary of the centre. |
| Median | The middle result once they are put in order. It is less affected by a single extreme value. | The better summary when data are skewed or when a set of differences contains an aberrant point. |
The scatter: standard deviation
The standard deviation (SD) describes how widely results scatter around their mean:
s = sample standard deviation · xi = each result · x̄ = the mean of the results · n = the number of results · Σ = sum over all n results.
- The deviations are measured from the sample's own mean, which always sits closer to the results than the true mean does, so their squared sum is slightly too small. Dividing by n − 1 corrects this.
- The correction matters most for small sets: for five results, dividing by n gives an SD about 11% too low.
- Spreadsheets usually offer both versions. The "sample" function (for example STDEV.S) divides by n − 1; the "population" function (STDEV.P) divides by n. Check which one a template uses.
The relative scatter: coefficient of variation
The coefficient of variation (CV) expresses the SD as a percentage of the mean:
- Because the SD of most methods grows with concentration, the CV often stays roughly constant across the measuring interval.
- That makes it convenient for comparing levels and for stating specifications.
- Near zero the CV becomes unstable and misleading, so state the SD at the low end of a method.
The normal (Gaussian) assumption
If results follow a normal (Gaussian) distribution, about 95% lie within 1.96 SD of the mean, the familiar "±2 SD" range. Many evaluation calculations rely on this assumption.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.2 Describing a set of results, pp.62–63
3.2 A result far from the others: outliers
| What EP15 asks | In practice |
|---|---|
| Exclude only results with a known, documented cause | A clot, a sample mix-up, an instrument error |
| Do not reject or repeat a result merely because it looks aberrant | "It looks odd" is not a reason on its own |
| Use a statistical test, such as Grubbs' test, to show whether an extreme result is unusual enough to be treated as an outlier | Worked Example 3.1 applies the test (section 12) |
| At most one such result per sample | …and the analysis is reported both with and without it |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.2 Describing a set of results (outliers), p.63
3.3 Variances add; SDs do not
A within-laboratory SD is built from several sources (Chapter 2, Figure 2.2):
- variation within a run;
- variation between runs and days;
- variation between reagent lots, calibrations and operators.
Independent sources combine by adding their variances, the squares of their SDs:
sWL = within-laboratory SD · sR = repeatability SD · sB = between-run (or between-day) SD.
Two consequences
| Consequence | Example from the book |
|---|---|
| 1. The larger component dominates. | Repeatability SD 1.0 and between-day SD 2.0: the within-laboratory SD is √(1.0² + 2.0²) = √5 = 2.24, not 3.0. |
| 2. Halving the smaller component barely changes the total. | Worth knowing before effort is spent on it. |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.3 Variances add; SDs do not, pp.63–64
3.3 Separating the components: one-way ANOVA and the standard designs
To separate the components, use a design with structure (several runs, each with replicates) and analyse it with a one-way analysis of variance (ANOVA).
ANOVA in plain words: two kinds of scatter
| Scatter | What it estimates |
|---|---|
| Results around their own run mean | Repeatability |
| Run means around the grand mean | The between-run variation plus a share of the repeatability, because each run mean is itself an average of a few results |
Subtracting that share leaves the between-run component:
VR = repeatability variance · VB = between-run variance · MS = the mean squares in the ANOVA table · n0 = the number of replicates per run when every run has the same number.
Two standard designs
| Design | Purpose | Layout | Minimum |
|---|---|---|---|
| EP15 | The user's verification design | Each sample in five runs on five days, five replicates per run | At least five runs; total results minus number of runs at least 18 |
| EP05 (single-site) | Meant for establishing precision; says far more about long-term variation | 20 days, two runs a day, two replicates per run | — |
Two warnings
- Ten replicates in a single run measure repeatability only; they cannot verify a within-laboratory claim, however good they look.
- A five-day within-laboratory estimate has usually seen one reagent lot, perhaps one calibration and a handful of operators. It is no substitute for months of quality control data when a long-term CV is needed, for example for a reference change value (Chapter 2) or for measurement uncertainty (Chapter 14).
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.3 Variances add; SDs do not (ANOVA and designs), p.64
3.4 How sure are we? Confidence intervals
Every estimate comes from a sample of results, and a second study would give a slightly different answer. A confidence interval (CI) shows how far the true value might lie from the estimate.
CI for a mean: symmetric
For a mean the interval is symmetric, with s ⁄ √n as the standard error of the mean and t taken from Student's t distribution:
x̄ = mean · s = SD · n = number of results · t0.975, n−1 = the t value for a two-sided 95% interval with n − 1 degrees of freedom.
CI for an SD: not symmetric
For an SD the interval is not symmetric. It is calculated from the chi-square distribution and the degrees of freedom of the estimate, which for a simple set of n results is n − 1.
| Number of results | The true SD could lie between… |
|---|---|
| 10 | 0.69 and 1.83 times the SD you measured |
| 25 | 0.78 and 1.39 times |
| 80 | 0.86 and 1.18 times |
Figure 3.1 The 95% confidence interval for an SD narrows slowly as the number of results grows.
Original chart · calculated from the chi-square distribution with df = n − 1
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.4 How sure are we? Confidence intervals (Figure 3.1), pp.64–65
3.4 The upper verification limit
The uncertainty of an SD explains a rule that surprises many laboratories.
- Judging the study against the claim itself would fail good methods half the time.
- EP15 instead compares the estimate with an upper verification limit: the largest value a study of that size would be expected to produce, 95% of the time, if the claim were true.
- The limit is the claim multiplied by a factor that depends on the degrees of freedom.
χ²0.95 = the 95th percentile of the chi-square distribution for the study's degrees of freedom · df = degrees of freedom of the estimate.
| Degrees of freedom | Factor (single sample) |
|---|---|
| 20 | about 1.25 |
| 8 | 1.39 |
- When several samples are tested the factor grows slightly, to keep the overall chance of a false failure near 5%.
- Evaluation software often prints this limit beside the claim as a "verification value". A report showing a claimed SD of 1.35 and a verification value of 1.69 at 20 degrees of freedom is applying exactly this factor.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.4 How sure are we? Confidence intervals (upper verification limit), pp.65–66
3.5 Significant is not the same as important
| Who asks | The question |
|---|---|
| A statistical test | Is the difference larger than chance alone would explain? |
| The laboratory | Is the difference large enough to matter for patients? |
The two questions have different answers more often than one might expect.
| Study size | What happens |
|---|---|
| Large studies make small differences "significant" | In a comparison of 30 samples spread over a wide measuring interval, the CI of the slope is often so narrow that it excludes 1 for a proportional difference of 1%, far inside any allowable limit. |
| Small studies do the reverse | A bias of 6% can be "not significant" in a study of ten samples because the interval is wide. |
Table 3.2 Reading an estimate and its 95% confidence interval against an allowable limit
| What you see | What it means | Decision |
|---|---|---|
| The whole interval lies inside the allowable limit | Performance is acceptable, whatever a significance test says | Accept |
| The whole interval lies beyond the limit | Performance is unacceptable | Reject, or investigate the cause |
| The estimate is inside the limit but the interval crosses it | The study cannot tell | Accept provisionally and confirm with more data, or extend the study |
| The estimate is beyond the limit but the interval crosses it | The study cannot tell, and the evidence leans against the method | Investigate before use |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.5 Significant is not the same as important (Table 3.2), p.66
3.6 Two methods: correlation is not agreement
The correlation coefficient, r, measures how closely paired results follow a straight line.
- It grows with the range of concentrations studied.
- A single result far from the others can raise it on its own.
- It says nothing about bias: if one method reads exactly 20% higher than the other, r can still be 1.000.
- A high r shows only that the study covered a wide range.
Agreement is judged with two tools
| Tool | What it does |
|---|---|
| Difference plot | Shows each difference, absolute or in percent, against concentration; makes a constant bias, a trend or an aberrant point visible at a glance. |
| Regression | Describes the relationship as a line. |
Choosing the regression
| Regression | Assumption | Notes |
|---|---|---|
| Ordinary least-squares | The comparative method has no error | Rarely true |
| Deming | Allows for error in both methods, either as a constant SD or, for wide ranges, as a constant CV | The constant-CV form is the default that EP09c recommends when scatter grows in proportion to concentration |
| Passing–Bablok | Makes no assumption about how errors are distributed | Resists outliers |
Chapter 8 develops these methods, and its "Reading the report" feature shows how one result can steer a regression line.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.6 Two methods: correlation is not agreement, p.67
3.7 Counting agreement: qualitative results
When a test reports positive or negative, the results of a comparison fit a 2 × 2 table. Three proportions summarise it:
| Proportion | Meaning |
|---|---|
| Overall agreement | The share of all samples on which the two tests agree |
| Positive percent agreement (PPA) | The share of comparator-positive samples that the new test also calls positive |
| Negative percent agreement (NPA) | The same for negatives |
How sure is a proportion?
A proportion from a small number of samples is uncertain, and the uncertainty is easy to underestimate. EP12 recommends the Wilson score method for its confidence interval.
| Evidence | Lower 95% limit |
|---|---|
| 10 out of 10 positives agree | 72% |
| 22 out of 22 overall | 85% |
| To show PPA is at least 90% | About 35 positive samples and no disagreements |
Cohen's kappa
Cohen's kappa, often printed alongside, measures agreement beyond what chance would produce. It depends on how common positives are in the sample set and does not replace PPA and NPA.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.7 Counting agreement: qualitative results, p.67
Worked Example 3.1 (part 1): verifying a precision claim — the data and the outlier check
Situation
| Item | Value |
|---|---|
| Method | A new creatinine method |
| Manufacturer's claims at about 95 µmol/L | Repeatability CV 1.5% · within-laboratory CV 2.5% |
| Laboratory's own specification (Chapter 4) | Within-laboratory CV of no more than 3.0% |
| Material and design | A plasma pool near 95 µmol/L, run in the EP15 design: one run a day for five days, five replicates per run |
Step 1 — Look at the data
Table 3.3 lists the 25 results and Figure 3.2 plots them.
Table 3.3 Creatinine results, µmol/L (synthetic)
| Day | Rep 1 | Rep 2 | Rep 3 | Rep 4 | Rep 5 | Day mean |
|---|---|---|---|---|---|---|
| 1 | 97.1 | 94.7 | 96.5 | 95.7 | 94.4 | 95.7 |
| 2 | 95.9 | 93.8 | 93.8 | 93.1 | 95.1 | 94.3 |
| 3 | 94.3 | 95.4 | 93.5 | 94.7 | 93.1 | 94.2 |
| 4 | 93.3 | 92.2 | 92.5 | 92.6 | 90.3 | 92.2 |
| 5 | 98.5 | 100.5 | 94.6 | 94.5 | 94.8 | 96.6 |
Figure 3.2 Results of the five-day precision study in Worked Example 3.1.
Synthetic, illustrative data · Worked Example 3.1 · 25 results
Step 2 — Check for outliers
| Quantity | Value |
|---|---|
| Mean of all 25 results | 94.6 µmol/L |
| SD of all 25 results | 2.09 |
| EP15's Grubbs factor for 25 results | 3.135 |
The highest result, 100.5 µmol/L on day 5, lies inside them. No result is excluded.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, Worked Example 3.1 Verifying a precision claim (Situation, Steps 1–2, Table 3.3, Figure 3.2), pp.68–69
Worked Example 3.1 (part 2): ANOVA, the claims and the confidence intervals
Step 3 — Separate the components
A one-way ANOVA with day as the factor gives:
Table 3.4 Analysis of variance for Worked Example 3.1
| Source | Sum of squares | Degrees of freedom | Mean square |
|---|---|---|---|
| Between days | 55.85 | 4 | 13.96 |
| Within days | 49.36 | 20 | 2.468 |
| Total | 105.21 | 24 |
| Component | Calculation | SD (µmol/L) | CV |
|---|---|---|---|
| Repeatability | Variance = within-day mean square = 2.468 | 1.57 | 1.66% |
| Between-day | Variance = (13.96 − 2.468) ⁄ 5 = 2.30 | 1.52 | — |
| Within-laboratory | √(2.468 + 2.30) | 2.18 | 2.31% |
Step 4 — Compare with the claims
| Claim | Estimate | How it is judged | Result |
|---|---|---|---|
| Within-laboratory CV 2.5% | 2.31% | Below the claim, so consistent with it without further calculation | Consistent |
| Repeatability CV 1.5% | 1.66% | Above the claim, which is not yet a failure. 25 − 5 = 20 degrees of freedom; this pool is analysed on its own, so the one-sample factor 1.25 applies: upper verification limit = 1.5% × 1.25 = 1.88% | 1.66% < 1.88% → consistent |
Step 5 — Ask how sure the answer is
| Estimate | Degrees of freedom | 95% CI (SD, µmol/L) | 95% CI (CV) |
|---|---|---|---|
| Repeatability SD 1.57 | 20 | 1.20 to 2.27 | 1.27% to 2.40% |
| Within-laboratory SD 2.18 | about 10.6 effective (Satterthwaite approximation) | 1.54 to 3.75 | 1.63% to 3.97% |
For its comparison with the claim, EP15 takes the degrees of freedom from the ratio of the two claims; for a ratio of 1.67 and five runs that gives about 8.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, Worked Example 3.1 Verifying a precision claim (Steps 3–5, Table 3.4), pp.69–70
Worked Example 3.1 (part 3): two common slips and the decision
Step 6 — Two common slips
| Slip | What it gives | Why it is wrong |
|---|---|---|
| Dividing by n instead of n − 1 | An SD of 1.03 rather than 1.15 for day 1 alone | 11% too low |
| One SD from all 25 results | 2.09 µmol/L | Neither the repeatability SD (1.57) nor the within-laboratory SD (2.18). It treats structured data as a single source of variation, and here it understates within-laboratory imprecision by about 4%. |
Decision
Record both estimates with their confidence intervals, the ANOVA and the outlier check.
Because the within-laboratory interval extends beyond the laboratory's own 3.0% specification, review the cumulative quality control CV after three months (Chapter 18).
What this example shows
| Colleague in the opening question | Verdict |
|---|---|
| Wanted to fail the study | Had forgotten that about half of all studies exceed a true claim by chance. |
| Called it "just statistics" | Was right about the outcome, but the reason is the verification limit, which must be calculated and recorded. |
| Got a different SD on a spreadsheet | The spreadsheet probably divided by n, or calculated one SD from all 25 results. |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, Worked Example 3.1 Verifying a precision claim (Step 6, Decision, What this example shows), p.70
Common errors
| ⚠️ Common error | Why it is wrong / what to do |
|---|---|
| Dividing by n. | For small sets this understates the SD; use n − 1. |
| One SD from structured data. | Results from several runs must be analysed by ANOVA, with repeatability and within-laboratory SDs reported separately. |
| Ten replicates as a within-laboratory study. | A single run measures repeatability only. |
| Judging against the claim instead of the verification limit, or passing a study without noticing how wide its confidence interval is. | About half of studies exceed a true claim by chance; and a pass does not prove the claim. |
| Treating statistical significance as clinical failure. | A slope interval that excludes 1 is not, by itself, a reason to reject a method. |
| Using r as evidence of agreement. | It reflects the range studied, not the bias. |
| Reporting 100% agreement without its interval. | Ten out of ten is consistent with a true agreement of 72%. |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, Common errors, p.70
Chapter 3 checklist and references
Chapter 3 checklist
Before approving the statistics in an evaluation plan or report, confirm that:
- ☐ The question each calculation answers is written down.
- ☐ SDs use n − 1, and the conditions behind each SD are stated.
- ☐ Structured precision data are analysed by ANOVA, with repeatability and within-laboratory SDs reported separately.
- ☐ Key estimates carry confidence intervals.
- ☐ Claims are judged with verification limits; acceptance criteria are judged with the estimate and its interval.
- ☐ Suspected outliers are tested, documented and analysed with and without exclusion.
- ☐ Agreement between methods is judged from difference plots and bias at decision levels, not from r.
- ☐ Qualitative agreement is reported as PPA and NPA with Wilson intervals.
Chapter 3 references (as listed in the book)
- Clinical and Laboratory Standards Institute. User Verification of Precision and Estimation of Bias. 3rd ed. CLSI guideline EP15-A3. CLSI; 2014. Sections 2.3 (5 × 5 design), 2.3.3 (data integrity; minimum runs and results; no rejection merely because a result looks aberrant), 2.3.4 (outliers; Grubbs' test), 2.3.5 (variance components by one-way ANOVA) and 2.3.6 (comparison with claims; upper verification limits).
- Clinical and Laboratory Standards Institute. Evaluation of Precision of Quantitative Measurement Procedures. 3rd ed. CLSI guideline EP05-A3. CLSI; 2014. Sections 1.5.4.2 (confidence intervals and degrees of freedom) and 2.3.1 (20 × 2 × 2 design).
- Clinical and Laboratory Standards Institute. Measurement Procedure Comparison and Bias Estimation Using Patient Samples. 3rd ed. CLSI guideline EP09c-Ed3. CLSI; 2018. Section 6.2 (regression analysis), including 6.2.2 (constant-CV Deming regression).
- Clinical and Laboratory Standards Institute. Evaluation of Qualitative, Binary Output Examination Performance. 3rd ed. CLSI guideline EP12-Ed3. CLSI; 2023. Sections 4.2.4.1 (Wilson score intervals) and 4.2.5 (positive and negative percent agreement).
- Linnet K, Moons KGM, Boyd JC. Statistical methodologies in laboratory medicine: analytical and clinical evaluation of laboratory tests. In: Rifai N, Chiu RWK, Young I, Burnham CD, Wittwer CT, editors. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023. Chapter 2, pp. 15 (Student t distribution, confidence intervals, median), 34–35 (the correlation coefficient) and 36 (nonparametric regression).
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, Chapter 3 checklist and Chapter 3 references, p.71
Training course, not a certification. Based on the author's textbook; all examples use synthetic data. It does not replace the CLSI and ISO documents it cites, and it is not affiliated with or endorsed by CLSI or ISO.
© 2026 Hassan Shawki. Course text adapted from the author's textbook Clinical Laboratory Method Evaluation in Practice.
17 · Reading the charts — precision designs, intervals, limits and agreement
The calculations in this chapter usually reach the evaluator as charts: replicates by day, variance components, a claim beside its verification limit, confidence intervals against an allowable limit, a comparison scatter and a 2 × 2 table. Each card below shows one pattern with new synthetic data (not the book's figures). The quiz then asks you to read similar charts and decide.
Precision designs and their components
Claims, limits and how sure we are
Two methods and two qualitative tests
Drawings: original teaching diagrams made for this page (no copyrighted images). Facts verified with: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, §3.2–§3.7 and Worked Example 3.1, pp.62–70.
⚡ Last-Minute Review — The Statistics an Evaluator Uses
Read these lists before you approve the statistics in an evaluation plan or report. Then open Progress / Results to see your weak areas.
The decision
- Which calculation answers our question, and how sure are we of the answer?
- Numbers become decisions only through the right calculation and an honest statement of its uncertainty. Software calculates; the evaluator chooses, reads and catches its mistakes.
Question → calculation (§3.1, Table 3.1)
- Centre → mean (median if skewed or one extreme value) · Scatter → SD (n − 1) and CV · Source of scatter → one-way ANOVA · How sure → CI (t for a mean, chi-square for an SD) · Claim → verification limit · Does it matter → estimate + CI vs allowable limit · Two methods → difference plot, Deming or Passing–Bablok · Qualitative → PPA/NPA with Wilson intervals.
Describing results (§3.2)
- s = √[Σ(xi − x̄)² ⁄ (n − 1)]. Divide by n and five results give an SD about 11% too low. STDEV.S = n − 1; STDEV.P = n.
- CV = 100 × s ⁄ x̄; roughly constant across the range for most methods; unstable near zero → state the SD at the low end.
- Normal data: about 95% within 1.96 SD ("±2 SD"). Check with a plot; median and percentiles when skewed.
- Outlier = a question, not a rejection. Exclude only with a known, documented cause (clot, mix-up, instrument error). Grubbs' test; at most one per sample; analyse with and without it.
Variances add (§3.3)
- sWL = √(sR² + sB²). 1.0 and 2.0 → 2.24, not 3.0. The larger component dominates; halving the smaller barely helps.
- ANOVA: VR = MSwithin; VB = (MSbetween − MSwithin) ⁄ n0; negative → set to zero.
- EP15 (verify): 5 runs on 5 days × 5 replicates; ≥ 5 runs; results − runs ≥ 18. EP05 (establish): 20 days × 2 runs × 2 replicates.
- Ten replicates in one run = repeatability only. Five days ≠ months of QC for a long-term CV (RCV, measurement uncertainty).
Confidence intervals and verification limits (§3.4)
- Mean: x̄ ± t0.975, n−1 × s ⁄ √n (symmetric). SD: chi-square, df = n − 1 (not symmetric).
- True SD vs measured SD: n = 10 → 0.69–1.83× · n = 25 → 0.78–1.39× · n = 80 → 0.86–1.18×. Band narrows fast to about 30 results, slowly after.
- A true claim is exceeded by about half of all studies. Judge against the upper verification limit = claim × √(χ²0.95 ⁄ df): 1.25 at 20 df, 1.39 at 8 df; slightly larger with several samples. Claimed SD 1.35 → verification value 1.69 (20 df).
- Passing the limit protects against false failure; it does not prove the claim.
Significant ≠ important (§3.5, Table 3.2)
- Large study: a 1% slope difference can be "significant". Small study: a 6% bias can be "not significant". The reference point is the clinical requirement, not zero.
- Whole CI inside limit → accept · whole CI beyond → reject or investigate · estimate inside, CI crosses → accept provisionally, confirm or extend · estimate beyond, CI crosses → investigate before use.
Two methods and qualitative agreement (§3.6–§3.7)
- r grows with range; one far result can raise it; 20% higher throughout can still give r = 1.000. High r = wide range, not agreement.
- Difference plot (constant bias, trend, aberrant point) + regression: OLS assumes no error in the comparator; Deming allows error in both (constant SD, or constant CV — EP09c default when scatter grows with concentration); Passing–Bablok no distribution assumption, resists outliers. Report predicted bias at each decision level with its CI.
- 2 × 2 table → overall agreement, PPA, NPA (not sensitivity/specificity when the comparator is another test). Wilson interval; perfect agreement lower limit = n ⁄ (n + 3.84): 10/10 → 72%, 22/22 → 85%; PPA ≥ 90% needs about 35 positives, no disagreements. Kappa depends on prevalence; it does not replace PPA/NPA.
Worked Example 3.1 in numbers
- Creatinine pool ≈ 95 µmol/L, EP15 5 × 5. Claims: repeatability CV 1.5%, within-laboratory CV 2.5%; own specification ≤ 3.0%.
- All 25: mean 94.6, SD 2.09; Grubbs 3.135 → limits 88.0–101.2; 100.5 kept.
- ANOVA: MS between 13.96 (4 df), MS within 2.468 (20 df). sR 1.57 (CV 1.66%) · sB 1.52 (VB 2.30) · sWL 2.18 (CV 2.31%).
- Repeatability 1.66% < verification limit 1.5 × 1.25 = 1.88% · within-laboratory 2.31% < claim 2.5% → both verified.
- CIs: sR 1.20–2.27 (CV 1.27–2.40%); sWL (≈ 10.6 df, Satterthwaite) 1.54–3.75 (CV 1.63–3.97%) → interval passes 3.0% → review the cumulative QC CV after three months.
- Slips: n instead of n − 1 (day 1: 1.03 vs 1.15); one SD from all 25 (2.09: neither 1.57 nor 2.18).
Common errors
- Dividing by n · one SD from structured data · ten replicates as a within-laboratory study · judging against the claim, or ignoring a wide CI · significance treated as clinical failure · r as agreement · 100% agreement without its interval.
Checklist
- Question written for each calculation · n − 1 and conditions stated · ANOVA for structured data, both SDs reported · CIs on key estimates · claims vs verification limits, criteria vs estimate + CI · outliers tested, documented, analysed with and without · agreement from difference plots and bias at decision levels · PPA/NPA with Wilson intervals.
Key terms (book glossary, Chapter 3 and terms used in it)
- Coefficient of variation (CV) — The standard deviation expressed as a percentage of the mean.
- Confidence interval — A range, calculated from the data, that shows how precisely a study has estimated a quantity such as a bias or an SD.
- Deming regression — A regression that allows for measurement error in both methods. The constant-CV form weights results when the CV, rather than the SD, is constant.
- Positive and negative percent agreement — The proportions of positive and negative comparator results that a new qualitative method matches, used when the comparator is not a reference standard.
- Repeatability — Precision under the same conditions over a short time: same operator, analyser, run and reagents.
- Standard deviation (SD) — A measure of the scatter of results about their mean.
- Upper verification limit — The largest SD that a study can produce while still being consistent with the manufacturer's claim, allowing for the study's own uncertainty.
- Within-laboratory precision — Precision over days, runs, calibrations and operators in one laboratory; also called intermediate precision.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.3, pp.61–71; Glossary, pp.411–419
🖼️ Picture clues (Study Section 17)
- Replicates by day: scatter around each day's bar = repeatability; spread of the bars = between-day variation.
- Variance components: the largest share dominates the within-laboratory SD; improve it first.
- Claim and verification limit: an estimate between the two lines is still consistent with the claim.
- SD interval band: with few results the true SD can be far above the estimate.
- Estimate + CI vs allowable limit: inside → accept · crosses with estimate inside → provisional · crosses with estimate outside → investigate · beyond → reject.
- Comparison scatter: points far from the identity line show bias, however high r is.
- 2 × 2 table: PPA = both positive ÷ comparator positive; perfect agreement still has a lower Wilson limit n ⁄ (n + 3.84).
🃏 Flashcards — quick recall
Tap the card (or press Space) to see the answer · 1 = Review again · 2 = Known · ← → move. Flashcards never change your quiz score.