Start here: the question on your desk
Every evaluation in this course ends by comparing a result with a limit. Chapter 1 named that limit the laboratory acceptance criterion and kept it separate from ISO requirements, CLSI recommendations and manufacturer claims (Table 1.5). This chapter explains where the limit comes from.
| Who | What they say about the limit |
|---|---|
| ISO 15189:2022 (clauses 7.3.1 b and 7.3.2 b) | The performance specifications of each examination method must relate to its intended use and its effect on patient care; a verification confirms the specifications relevant to that use. It leaves the numbers to the laboratory. |
| The Milan consensus (2014) | The main framework for choosing the numbers: the first Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM), held in Milan in 2014. |
By the end of this chapter you should be able to
- Explain why a specification must be fixed before any data are collected.
- Describe the three Milan models and choose the one that suits a measurand and its use.
- Calculate imprecision, bias and total error specifications from biological variation.
- Judge when state-of-the-art performance, external quality assessment limits or regulatory limits are acceptable sources.
- Turn a specification into acceptance criteria for each study in an evaluation plan.
- Record the choice so that an assessor can trace every number.
Map of this chapter
| # | Study section | Book section(s) | Printed pages |
|---|---|---|---|
| 1 | Start here: the question on your desk | Chapter 4 opening, learning objectives | 73–74 |
| 2 | The number comes first | §4.1 | 74 |
| 3 | Three models for setting specifications | §4.2 (Stockholm and Milan, Table 4.1) | 74–75 |
| 4 | Which model for which measurand, and more than one use | §4.2 (EFLM list, glucose example) | 75–76 |
| 5 | Specifications from biological variation: the formulas | §4.3 (formulas and cautions) | 76 |
| 6 | Three levels, and the quality of the inputs | §4.3 (Table 4.2, note on sources, EFLM database) | 77 |
| 7 | State of the art, EQA limits and regulatory limits | §4.4 (incl. the CLIA example) | 77–78 |
| 8 | From specification to acceptance criteria | §4.5 (Table 4.3, error budgets) | 78–79 |
| 9 | Two details often missed: the floor and the decision level | §4.5 (concentration floor, hCG carryover example) | 79 |
| 10 | What quality control can and cannot do | §4.5 (CLSI C24) | 79–80 |
| 11 | Worked Example 4.1 (part 1): model, inputs and three levels | Worked Example 4.1, Steps 1–4 (Table 4.4) | 80–81 |
| 12 | Worked Example 4.1 (part 2): decide and record | Worked Example 4.1, Step 5 | 81 |
| 13 | Worked Example 4.1 (part 3): the floor rule, budgets and Figure 4.1 | Worked Example 4.1, floor rule (Table 4.5), budgets, record, Figure 4.1 | 81–83 |
| 14 | Common errors | Common errors | 83–84 |
| 15 | Chapter 4 checklist and references | Chapter 4 checklist; Chapter 4 references | 84–85 |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, opening and learning objectives, pp.73–74; How this book is organised, p.viii
4.1 The number comes first
The temptation
To prevent this, the evaluation plan (Chapter 5) states each specification, its source and its formula before the first sample is run, and the report compares the results with that plan.
Claim or specification? Keep the two ideas apart
| Claim | Specification | |
|---|---|---|
| What it says | What a method can do | What patients need |
| Where it comes from | The manufacturer | Clinical evidence |
| Who handles it | It is verified (Chapter 1) | It is chosen by the laboratory |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.1 The number comes first, p.74
4.2 Three models for setting specifications
From five levels to three models
| Conference | What it agreed |
|---|---|
| Stockholm, 1999 | A five-level hierarchy for choosing specifications, from clinical outcome studies at the top to the current state of the art at the bottom |
| Milan, 2014 | Simplified that hierarchy into three models, each with its own principle (Table 4.1), and kept an order of preference: models 1 and 2 are preferred to model 3 |
| Model | Principle | Best suited to | Main limitation |
|---|---|---|---|
| 1. Clinical outcome | How analytical performance affects clinical outcomes, studied directly or indirectly (by simulation or decision analysis) | Measurands central to decisions in a specific disease, with established cut-offs | Outcome data exist for few measurands, and results depend on setting and population |
| 2. Biological variation | Keep analytical "noise" small compared with the biological "signal" | Measurands in a steady state, especially those under tight homeostatic control | Needs reliable, critically appraised biological variation data |
| 3. State of the art | The best performance that current methods achieve | Measurands that cannot be placed in model 1 or 2, including some placed there temporarily while better data are awaited | What is achievable may not be what patients need |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.2 Three models for setting specifications, pp.74–75
4.2 Which model for which measurand, and more than one use
The Milan statement did not allocate individual measurands to models; it recommended that such a list be drawn up. An EFLM working group then proposed criteria and a preliminary list covering some commonly requested measurands.
The working group's examples, with the clinical use discussed
| Model | Examples (and the use behind them) |
|---|---|
| 1. Clinical outcome | Total, HDL and LDL cholesterol (classifying cardiovascular risk against treatment thresholds) · glucose and HbA1c (diagnosing and monitoring diabetes; for glucose monitoring the group notes that treatment goals are not always established, and other models can then be used) · plasma albumin (nutritional monitoring in dialysis, staging of myeloma, decisions on albumin infusion) · C-reactive protein (separating viral from bacterial infection; severity of acute pancreatitis) · cardiac troponins (diagnostic misclassification at the decision limit) · haemoglobin, platelets and neutrophils (anaemia, transfusion thresholds, infection risk of severe neutropenia) · thyroid-stimulating hormone (diagnosing and monitoring thyroid disease) |
| 2. Biological variation | Plasma electrolytes and minerals · plasma creatinine, urea and cystatin C · urate · total protein · erythrocyte indices · haemoglobin when used for monitoring · prothrombin time and activated partial thromboplastin time when used to monitor anticoagulant treatment |
| 3. State of the art | Measurands that cannot be placed in model 1 or 2, and measurands placed there temporarily while outcome or biological variation data are awaited · mainly urine measurands, such as urinary electrolytes, creatinine and urate |
(HDL = high-density lipoprotein; LDL = low-density lipoprotein.) The paper's summary table and its text differ for a few of these entries, so check both before relying on one.
One measurand, more than one specification
A measurand can have more than one use, so it can have more than one specification. The Milan statement's example is glucose:
| Use | Where the specification might come from |
|---|---|
| Blood glucose in critical care | Simulating the effect of analytical performance on patient outcomes (an indirect outcome study) |
| Self-monitoring of blood glucose in type 1 diabetes | Clinical outcome studies or biological variation |
The statement also allows models to be combined, and asks that every specification state its rationale, its source and the quality of the evidence behind it.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.2 Three models for setting specifications, pp.75–76
4.3 Specifications from biological variation: the formulas
Model 2 can be applied to most measurands for which reliable biological variation data exist, which is why it is the model laboratories meet most often.
The principle
| Specification | Judged against |
|---|---|
| Imprecision | The natural within-subject biological variation (CVI): analytical imprecision should be small compared with it, so that it adds little to the variation the clinician sees |
| Bias | The combined within- and between-subject variation (CVI and CVG) |
The commonly used formulas, at the "desirable" level
| Specification | Formula |
|---|---|
| Imprecision | CVA ≤ 0.5 × CVI |
| Bias | B ≤ 0.25 × √(CVI2 + CVG2) |
| Total allowable error | TEa ≤ 1.65 × 0.5 × CVI + 0.25 × √(CVI2 + CVG2) |
CVA = analytical coefficient of variation (imprecision), %; B = bias, %; TEa = allowable total error, %; CVI = within-subject biological variation, %; CVG = between-subject biological variation, %.
What each formula means, and its limits
| Formula | What the book says about it |
|---|---|
| Imprecision | With imprecision at half the within-subject variation, the total variation a clinician sees rises by only about 12%. |
| Bias | Proposed so that laboratories could share common reference intervals. It was not intended as a universal bias limit, and its use for other purposes should be justified. |
| Total error | Adds the two allowances in a straight line. Critics point out there is no theoretical basis for adding imprecision and bias linearly and that the result overstates the allowable error; it remains widely used because it is simple. |
| Measurement uncertainty | A specification of the same family limits the expanded uncertainty to Z × 0.5 CVI. Chapter 14 returns to uncertainty budgets. |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.3 Specifications from biological variation, p.76
4.3 Three levels, and the quality of the inputs
Three levels are often quoted, scaling the factors up or down. The cited textbook chapter attributes the scheme to Fraser and colleagues.
| Level | Imprecision (CVA) | Bias |
|---|---|---|
| Optimum | 0.25 × CVI | 0.125 × √(CVI2 + CVG2) |
| Desirable | 0.50 × CVI | 0.250 × √(CVI2 + CVG2) |
| Minimum | 0.75 × CVI | 0.375 × √(CVI2 + CVG2) |
The formulas are only as good as their inputs
- Take current, critically appraised values of CVI and CVG from the EFLM Biological Variation Database.
- Record the version or access date.
- Check that the population studied resembles the patients the laboratory serves.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.3 Specifications from biological variation (Table 4.2), p.77
4.4 State of the art, EQA limits and regulatory limits
Model 3: the state of the art
Model 3 uses the best performance that current methods achieve, read from external quality assessment (EQA) data, manufacturers' information or published evaluations.
| Strength | Weakness |
|---|---|
| Easy to obtain; the only option for many measurands | Describes what is possible, not what patients need |
| A moving target: as methods improve, yesterday's state of the art becomes today's minimum |
EQA and regulatory acceptance limits
EQA schemes and regulators also publish acceptance limits. Many schemes base them on a combination of biological variation and state of the art. Regulatory limits differ between jurisdictions, and a laboratory must meet those that apply to it.
How wide such limits can be: a United States example
Included for international readers. Under the Clinical Laboratory Improvement Amendments (CLIA), the proficiency testing (PT) scoring criterion for creatinine is:
- Each PT sample result is scored as acceptable or not against it.
- Two dates apply: the revised criteria became effective on 11 July 2024, and the Centers for Medicare & Medicaid Services (CMS) set 1 January 2025 as the date from which PT programmes and laboratories implement them.
- The previous criterion was ±15% or ±0.3 mg/dL.
| Creatinine | CLIA criterion allows | Illustrative specification of Worked Example 4.1 allows |
|---|---|---|
| 60 µmol/L | 17.7 µmol/L | 4.4 µmol/L (the CLIA figure is four times larger) |
| 300 µmol/L | 30 µmol/L | 22.2 µmol/L |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.4 State of the art, EQA limits and regulatory limits, pp.77–78
4.5 From specification to acceptance criteria
| Term | What it does |
|---|---|
| Specification | Describes the performance needed |
| Acceptance criterion | Applies it to one study |
Table 4.3 shows how the same specification feeds different studies in a plan.
| Study (chapter) | What is compared | Typical criterion |
|---|---|---|
| Precision (6) | Within-laboratory CV, with its confidence interval | CV no greater than the imprecision specification; claim also checked with a verification limit (Chapter 3) |
| Method comparison (8) | Bias at each decision level, or each difference as a share of the allowable total error | Bias no greater than the bias specification; every difference within the allowable total error |
| Linearity (9) | Deviation from linearity at each level | A stated fraction of the allowable total error or bias, chosen in advance |
| Interference (11), carryover (12) | Effect on a sample near the lowest decision level | An effect smaller than a clinically meaningful change at that level |
| Stability, where verified locally | Change over the storage period | A stated fraction of the allowable total error or bias |
Error "budgets" in evaluation software
| Budget | Meaning |
|---|---|
| Random error budget (for example 25% or 50%) | The allowed SD or CV is that share of the allowable total error |
| Systematic error budget | The same, for bias |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.5 From specification to acceptance criteria (Table 4.3), pp.78–79
4.5 Two details often missed: the floor and the decision level
Detail 1: a percentage needs a concentration floor
A percentage specification becomes impossible near zero, so most laboratories state it as "a concentration or a percentage, whichever is greater". The rule must say:
- which value the percentage is taken of, and
- where the two limits meet;
- and the concentration floor needs its own justification, usually from the method's imprecision near the lowest decision level.
Worked Example 4.1 (study section 13) shows a complete rule.
Detail 2: the limit must be able to detect an error that matters at the decision level
Carryover shows why. When a specimen with a very high concentration is followed by one with a low concentration, a small fraction of the first can be carried into the second and raise its result.
| Illustrative figures (human chorionic gonadotropin, hCG) | Value |
|---|---|
| High specimen | 200,000 IU/L |
| Next specimen (non-pregnant patient), true value | 2 IU/L |
| Reported positive at | 5 IU/L or more |
| Carryover of 0.01% adds | 20 IU/L → the low specimen reads 22 IU/L and is reported positive |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.5 From specification to acceptance criteria, p.79
4.5 What quality control can and cannot do
CLSI C24 takes the same position
CLSI C24, the quality control (QC) guideline:
- asks the laboratory director to set the allowable total error from the way results are used in the population the laboratory serves, independently of how the method actually performs, and to agree it with the clinicians;
- lists the three Milan models in the same order of preference;
- notes that no single model suits every measurand, and that when no available method can meet the chosen goal, the laboratory may have to re-examine the goal or use a QC strategy sensitive enough to catch small deteriorations.
Chapter 18 shows how the same specifications drive the choice of control rules.
A more sensitive QC strategy does not make a method better
| A more sensitive strategy (tighter limits, more control results per run, more frequent QC events)… | |
|---|---|
| What it does | Detects smaller shifts or drifts away from the method's stable performance, and detects them sooner |
| What it costs | More false rejections, more repeat testing and slower reporting |
| What it cannot do | Bring any benefit when the problem is the method's own stable imprecision: random error affects control and patient results independently, so a control result inside tighter limits does not show that patient results meet the imprecision goal |
Where the evidence of fitness comes from
| When | Evidence that the method meets the goal for its intended use |
|---|---|
| Before use | The precision, bias and total error estimates from the evaluation, each compared with the specification fixed in advance (Chapters 6, 8 and 14) |
| In use | The long-term QC SD, external quality assessment results and comparability checks that ISO 15189:2022 asks laboratories to monitor (clauses 7.3.7.1 to 7.3.7.4) |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.5 From specification to acceptance criteria, pp.79–80
Worked Example 4.1 (part 1): choosing the model, recording the inputs, calculating the levels
Situation
The laboratory in Worked Example 3.1 must set specifications for its new creatinine method before the evaluation begins. Results are used to:
- estimate glomerular filtration rate,
- monitor patients with kidney disease, and
- adjust drug doses.
Step 1 — Choose the model
Plasma creatinine is held in a steady state by kidney function, and the EFLM working group lists it among the measurands suited to the biological variation model; urine creatinine, by contrast, is in its state-of-the-art group.
Step 2 — Record the inputs
| Input (illustrative) | Value |
|---|---|
| CVI (within-subject) | 4.5% |
| CVG (between-subject) | 14.0% |
| √(CVI2 + CVG2) | 14.7% |
Step 3 — Calculate the three levels
| Level | Imprecision (CVA) | Bias | Total error (1.65 × CVA + bias) |
|---|---|---|---|
| Optimum | 1.13% | 1.84% | 3.7% |
| Desirable | 2.25% | 3.68% | 7.4% |
| Minimum | 3.38% | 5.51% | 11.1% |
The arithmetic, row by row
| Level | Imprecision | Bias | Total error |
|---|---|---|---|
| Optimum | 0.25 × 4.5 = 1.13% | 0.125 × 14.7 = 1.84% | 1.65 × 1.13 + 1.84 = 3.7% |
| Desirable | 0.50 × 4.5 = 2.25% | 0.250 × 14.7 = 3.68% | 1.65 × 2.25 + 3.68 = 7.39 → 7.4% |
| Minimum | 0.75 × 4.5 = 3.38% | 0.375 × 14.7 = 5.51% | 1.65 × 3.38 + 5.51 = 11.1% |
The minimum row uses Fraser's factors of 0.75 for imprecision and 0.375 for bias. Like the other two rows, it is a proposal for comparison, not an acceptance criterion; the limits the laboratory adopts in Step 5 do not change.
Step 4 — Compare with what methods achieve
In this example the best-performing methods in the laboratory's EQA scheme achieve a within-laboratory CV of about 2.5%.
| Level | Imprecision | Compared with the state of the art (about 2.5%) |
|---|---|---|
| Desirable | 2.25% | At the edge of what current methods deliver |
| Minimum | 3.38% | Comfortably achievable |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Worked Example 4.1 Choosing specifications for creatinine, Steps 1–4 (Table 4.4), pp.80–81
Worked Example 4.1 (part 2): Step 5 — decide and record
The laboratory adopts three limits. They are chosen separately. Only the total error limit is calculated from other numbers, and it is calculated from the desirable levels, not from the two adopted limits.
| Limit adopted | Value | How it was chosen |
|---|---|---|
| Imprecision | Within-laboratory CV no greater than 3.0% | A local judgement, not a formula result. The desirable level (2.25%) is below what the best current methods achieve (about 2.5%), so it would fail most methods. The minimum level (3.38%) is easily met but gives away more than necessary. The laboratory chose a round figure between the two and recorded 2.25% as its long-term aim. |
| Bias | No greater than 3.7% at each decision level | The desirable bias level (3.68%) rounded to one decimal place. It applies at the decision levels used for estimating glomerular filtration rate. Because filtration-rate equations assume results traceable to the reference measurement procedure, the laboratory adopts the formula here even though it was designed for shared reference intervals. |
| Allowable total error | 7.4% of the comparative method's result, or 4 µmol/L, whichever is greater | The percentage is the desirable total error, 1.65 × 2.25 + 3.68 = 7.39%, rounded to 7.4%. It is used to judge each patient difference in the comparison study (Chapter 8). |
Limits chosen separately do not automatically agree
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Worked Example 4.1 Choosing specifications for creatinine, Step 5, p.81
Worked Example 4.1 (part 3): the floor rule, budgets and Figure 4.1
The floor rule
For each patient sample, the allowable difference is the larger of 4 µmol/L and 7.4% of the comparative method's result.
| Comparative result | 7.4% of it | Allowable difference | New method result | Difference | Judgement |
|---|---|---|---|---|---|
| 40 µmol/L | 3.0 µmol/L | 4.0 µmol/L (floor) | 43.5 µmol/L | 3.5 µmol/L (8.8%) | Within the limit |
| 120 µmol/L | 8.9 µmol/L | 8.9 µmol/L (percentage) | 130.0 µmol/L | 10.0 µmol/L (8.3%) | Outside the limit |
- The first result would fail a percentage limit alone (8.8% is more than 7.4%), although a difference of 3.5 µmol/L at 40 µmol/L is within what the method's imprecision allows.
- The second shows the percentage limit taking over at higher concentrations.
Justifying the floor
The floor itself is a local choice. Near 45 µmol/L the laboratory expects a within-laboratory SD of about 1.2 µmol/L (illustrative), so:
The clinical users accepted this for the low range, and the plan records their agreement.
Budgets for software
| Budget | Calculation (share of the 7.4% total error) | Result |
|---|---|---|
| Random error budget | 3.0 ⁄ 7.4 | 41% |
| Systematic error budget | 3.7 ⁄ 7.4 | 50% |
Entering these values makes the software reproduce the plan's limits.
Figure 4.1 Candidate specifications for within-laboratory imprecision, and the result of Worked Example 3.1
Illustrative values · Worked Example 4.1 · observed CV from Worked Example 3.1 (synthetic)
What this example shows
- The 3.0% used in Worked Example 3.1 was not chosen after the data were seen. It came from biological variation, was moderated by what current methods achieve, and was written into the plan with its reason.
- Limits chosen separately must be checked together.
- The same process explains the opening question: both glucose limits may be defensible, but only if each laboratory can show where its number came from and why it suits its use.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Worked Example 4.1 Choosing specifications for creatinine, floor rule (Table 4.5), budgets, record and Figure 4.1, pp.81–83
Common errors
| ⚠️ Common error | What to do instead |
|---|---|
| Choosing the limit after seeing the results | A specification must be in the plan before the first sample is run. |
| Confusing a claim with a specification | Meeting the manufacturer's claim does not show that the method meets patients' needs. |
| Using outdated biological variation tables | Take appraised values from the EFLM database and record their date. |
| Applying the bias formula everywhere without thought | It was designed for shared reference intervals. |
| Borrowing EQA or regulatory limits without saying so | They judge single results and are often wide. |
| Stating only a percentage | Near zero a percentage becomes impossible; give a justified concentration floor. |
| Budget fractions that contradict the specifications | Software budgets must reproduce the limits in the plan. |
| Treating separately chosen limits as a guarantee of total error | Imprecision and bias limits set one at a time do not control total error; check the combination as well. |
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Common errors, pp.83–84
Chapter 4 checklist and references
Before approving an evaluation plan, confirm that for each measurand:
| Checklist item | Where taught | |
|---|---|---|
| ☐ | The intended use is written down and the specification model is chosen to suit it. | §4.2 |
| ☐ | The source of every input is recorded, with version or access date. | §4.3 |
| ☐ | Imprecision, bias and total error specifications are stated, with the level chosen for each and the reason. | §4.3, Worked Example 4.1 |
| ☐ | A concentration floor is stated where a percentage alone would fail at low concentrations, with its justification. | §4.5 |
| ☐ | The floor rule states which result the percentage is taken of and the concentration at which the absolute and percentage limits meet. | §4.5, Worked Example 4.1 |
| ☐ | Any EQA, regulatory or state-of-the-art limit used is identified as such. | §4.4 |
| ☐ | Each study in the plan has an acceptance criterion derived from these specifications. | §4.5 |
| ☐ | The specifications were approved before data collection began. | §4.1 |
Chapter 4 references (as cited by the book)
- Sandberg S, Fraser CG, Horvath AR, et al. Defining analytical performance specifications: consensus statement from the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine. Clin Chem Lab Med. 2015;53(6):833–5. doi:10.1515/cclm-2015-0067.
- Ceriotti F, Fernandez-Calle P, Klee GG, et al. Criteria for assigning laboratory measurands to models for analytical performance specifications defined in the 1st EFLM Strategic Conference. Clin Chem Lab Med. 2017;55(2):189–94. doi:10.1515/cclm-2016-0091.
- Sandberg S, Røraas T, Aarsand AK. Biological variation and analytical performance specifications. In: Rifai N, Chiu RWK, Young I, Burnham CD, Wittwer CT, editors. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023. Chapter 8 (online), pp. 165.e14–165.e17.
- Fraser CG, Hyltoft Petersen P, Libeer JC, Ricos C. Proposals for setting generally applicable quality goals solely based on biology. Ann Clin Biochem. 1997;34(Pt 1):8–12. doi:10.1177/000456329703400103.
- European Federation of Clinical Chemistry and Laboratory Medicine. EFLM Biological Variation Database. biologicalvariation.eu (accessed November 2026).
- International Organization for Standardization. ISO 15189:2022 Medical laboratories — Requirements for quality and competence. 4th ed. ISO; 2022 (December). Clauses 7.3.1 b), 7.3.2 b) and 7.3.7.1–7.3.7.4.
- Sandberg S, Coskun A, Carobene A, et al. Analytical performance specifications based on biological variation data – considerations, strengths and limitations. Clin Chem Lab Med. 2024;62(8):1483–9. doi:10.1515/cclm-2024-0108.
- Clinical and Laboratory Standards Institute. Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions. 4th ed. CLSI guideline C24-Ed4. CLSI; 2016. Sections 3.2, 5.1 and 5.4.2.
- Centers for Medicare & Medicaid Services, HHS. Clinical Laboratory Improvement Amendments of 1988 (CLIA) proficiency testing regulations related to analytes and acceptable performance. Final rule (CMS-3355-F). Federal Register 11 July 2022;87(131):41194; effective 11 July 2024. Codified at 42 CFR §493.931 (routine chemistry: creatinine).
- Centers for Medicare & Medicaid Services, Quality, Safety & Oversight Group. Implementation Notification for Final Rule CMS-3355-F, Clinical Laboratory Improvement Amendments of 1988 (CLIA) Proficiency Testing — Analytes and Acceptable Performance Final Rule. Ref. Admin Info-23-07-CLIA. 3 May 2023.
CLSI and ISO documents are cited for attribution only. Readers should obtain the current edition for the full text. Citation does not imply endorsement by CLSI, ISO, EFLM, any accreditation body or any manufacturer.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Chapter 4 checklist; Chapter 4 references, pp.84–85
Training course, not a certification. Based on the author's textbook; all examples use synthetic data. It does not replace the CLSI and ISO documents it cites, and it is not affiliated with or endorsed by CLSI or ISO.
© 2026 Hassan Shawki. Course text adapted from the author's textbook Clinical Laboratory Method Evaluation in Practice.
⚡ Last-Minute Review — Setting Performance Specifications Before Testing
Read these lists before you write the limits into an evaluation plan. Then open Progress / Results to see your weak areas.
The decision
- How much error can this test carry and still serve its intended use?
- ISO 15189:2022 (7.3.1 b, 7.3.2 b): specifications relate to the intended use and patient care; the numbers are left to the laboratory
- Main framework: the Milan consensus (first EFLM Strategic Conference, 2014)
§4.1 The number comes first
- A limit chosen after the results is a description of the results, not a criterion
- The plan states each specification, its source and its formula before the first sample; the report compares results with that plan
- Claim = what a method can do (manufacturer; verified) · specification = what patients need (clinical evidence; chosen by the laboratory) · a method can meet one and fail the other; a verification usually checks both
§4.2 The three Milan models (Table 4.1)
- Stockholm 1999: five-level hierarchy → Milan 2014: three models; models 1 and 2 preferred to model 3
- 1 Clinical outcome: effect on outcomes, direct or indirect (simulation, decision analysis) · for measurands central to decisions with established cut-offs · few outcome data, setting- and population-dependent
- 2 Biological variation: keep analytical noise small against the biological signal · steady-state, tightly controlled measurands · needs reliable, appraised data
- 3 State of the art: best current performance · measurands not placeable in 1 or 2 (some temporarily) · achievable ≠ what patients need
- EFLM working-group list is preliminary; summary table and text differ for a few entries — check both
- Model 1: cholesterol (total, HDL, LDL), glucose, HbA1c, plasma albumin, C-reactive protein, troponins, haemoglobin/platelets/neutrophils, thyroid-stimulating hormone · Model 2: electrolytes and minerals, plasma creatinine, urea, cystatin C, urate, total protein, erythrocyte indices, haemoglobin for monitoring, prothrombin time and activated partial thromboplastin time for anticoagulant monitoring · Model 3: mainly urine measurands
- Plasma creatinine → model 2; urine creatinine → model 3 · one measurand can have more than one specification (glucose: critical care vs self-monitoring) · models may be combined; state rationale, source, quality of evidence
§4.3 Biological variation formulas (desirable level)
- CVA ≤ 0.5 × CVI → total variation seen rises by only about 12%
- B ≤ 0.25 × √(CVI2 + CVG2) → designed for shared reference intervals; not a universal bias limit
- TEa ≤ 1.65 × 0.5 CVI + 0.25 √(CVI2 + CVG2) → linear addition, no theoretical basis, overstates; used because simple
- Expanded uncertainty ≤ Z × 0.5 CVI (Chapter 14)
- EFLM task group: bias formula assumes almost no imprecision; imprecision and bias specifications side by side do not control total error; TE formula designed for single EQA results
- Three levels (Table 4.2): imprecision 0.25 · 0.50 · 0.75 × CVI; bias 0.125 · 0.250 · 0.375 × √(CVI2 + CVG2) (optimum · desirable · minimum) · Fraser's proposals; criteria only when adopted with reasons
- Inputs from the EFLM Biological Variation Database: current, appraised, version/access date recorded, population like your patients
§4.4 State of the art, EQA and regulatory limits
- State of the art: easy, often the only option, describes what is possible, a moving target
- EQA/regulatory limits judge single results (often wide) and are fixed for a whole scheme or country → use only as an outer boundary, say so and explain why no tighter limit
- CLIA PT creatinine: ±10% or ±0.2 mg/dL (about 17.7 µmol/L), whichever is greater (effective 11 July 2024; implemented from 1 January 2025; previously ±15% or ±0.3 mg/dL) · at 60 µmol/L allows 17.7 vs 4.4 µmol/L (four times) · at 300 µmol/L 30 vs 22.2 · not a verification specification
§4.5 From specification to acceptance criteria (Table 4.3)
- Specification = performance needed · acceptance criterion = applied to one study
- Precision: CV with CI ≤ imprecision spec; claim with verification limit · comparison: bias at each decision level ≤ bias spec; every difference within ATE · linearity and stability: a stated fraction of ATE or bias · interference and carryover: effect near the lowest decision level smaller than a clinically meaningful change
- Software budgets (e.g. 25% or 50%) = share of ATE allowed to SD/CV (random) or bias (systematic) · conventions, not standards · must agree with the specifications
- Floor: "concentration or percentage, whichever is greater" · say which value the percentage is of and where the limits meet · justify the floor from imprecision near the lowest decision level
- The limit must detect an error that matters at the decision level: hCG 200,000 IU/L, 0.01% carryover adds 20 IU/L → 2 reads 22 IU/L (positive at ≥5) · a 25 IU/L limit accepts it · allowable effect < gap (3 IU/L) with room for imprecision → 1 IU/L · use a realistically high specimen
- CLSI C24: director sets ATE from clinical use, independently of method performance, agreed with clinicians; same Milan order · if no method meets the goal: re-examine the goal or use a more sensitive QC strategy
- More sensitive QC finds smaller shifts sooner (more false rejections, repeats, slower reporting) but cannot fix stable imprecision · fitness shown before use by the evaluation (Ch 6, 8, 14) and in use by long-term QC SD, EQA, comparability (ISO 7.3.7.1–7.3.7.4) · QC watches; it cannot replace the demonstration
Worked Example 4.1 — creatinine (illustrative)
- Model 2 · CVI 4.5%, CVG 14.0%, √ = 14.7%
- Table 4.4: optimum 1.13 / 1.84 / 3.7% · desirable 2.25 / 3.68 / 7.4% · minimum 3.38 / 5.51 / 11.1% (imprecision / bias / TE = 1.65 × CVA + bias)
- State of the art about 2.5%: desirable at the edge, minimum comfortably achievable
- Adopted: CV ≤ 3.0% (local judgement; 2.25% long-term aim) · bias ≤ 3.7% at each decision level (desirable, rounded) · ATE 7.4% or 4 µmol/L, whichever greater (from desirable levels: 1.65 × 2.25 + 3.68 = 7.39)
- Check together: 1.65 × 3.0 + 3.7 = 8.65% > 7.4% → also compare estimated total error with 7.4% (Chapter 14)
- Floor meets percentage at 4 ⁄ 0.074 = 54 µmol/L · 40 → 43.5: 3.5 µmol/L within (floor 4.0) · 120 → 130.0: 10.0 µmol/L outside (8.9) · floor justified: SD 1.2 near 45 → 2.0 used by random error, 2 left for bias; clinical users agreed
- Budgets: random 3.0 ⁄ 7.4 = 41% · systematic 3.7 ⁄ 7.4 = 50%
- Record: model, database version and date, formulas, level and reason, floor rule, approver · Figure 4.1: observed CV 2.31% below 3.0% → accepted; CI reaches past 3.0% and the minimum → QC review after three months
Common errors
- Limit after results · claim mistaken for specification · outdated biological variation tables · bias formula everywhere · borrowed EQA/regulatory limits not declared · percentage only, no floor · budgets that contradict the specifications · separate limits treated as a total error guarantee
Key terms (book glossary, Chapter 4 and terms used in it)
- Acceptance criterion — A rule, written before the study, that turns a result into a decision: meets, does not meet, or needs more evidence.
- Allowable total error (ATE) — The largest difference from the true value that a single result may have and still serve its clinical purpose. In the creatinine example, 7.4% (or 4 µmol/L below 54 µmol/L).
- Analytical performance specification — The quality a method needs, stated as a number (for example, a maximum bias or CV) and chosen from clinical outcome, biological variation or the state of the art.
- Biological variation — The natural variation of a measurand within one person over time (within-subject) and between people (between-subject). It is one basis for performance specifications and for judging change in a patient.
- Decision level — A concentration at which clinical decisions are made, and at which bias and total error are judged. In the creatinine example, 60, 110 and 300 µmol/L.
- State of the art — Performance currently achievable by most laboratories or methods, used as a basis for specifications when clinical or biological models cannot be applied.
- Total analytical error — The combined effect of bias and imprecision on a single result.
- External quality assessment (EQA) — Comparison of a laboratory's results on samples from an external provider with a target value and with other laboratories. Proficiency testing is one form.
Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, pp.73–85; Glossary, pp.411–419