METHOD EVALUATION COURSE · PART I · CHAPTER 04

Setting Performance Specifications Before Testing

The number comes first · Claim versus specification · The three Milan models · Which model for which measurand · Biological variation formulas and three levels · State of the art, EQA and regulatory limits · From specification to acceptance criteria · Concentration floors · What QC can and cannot do · Worked example: choosing specifications for creatinine

Course text from Clinical Laboratory Method Evaluation in Practice (1st ed., 2026) · Original practice questions · Works offline · questions · Version 1.0

📘 Study progress

Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 key point · 📌 remember · ⚠️ common error · 💡 understand this · 🧪 at the bench · 🧠 memory aid.

Start here: the question on your desk

🔥 THE QUESTION ON YOUR DESK: Two laboratories in the same network verify the same glucose analyser. One accepts a bias of up to 6%, the other up to 10%. Both say they follow international guidance. The analyser passes in one laboratory and fails in the other. Where do allowable-error figures come from, and who decides which one applies?

Every evaluation in this course ends by comparing a result with a limit. Chapter 1 named that limit the laboratory acceptance criterion and kept it separate from ISO requirements, CLSI recommendations and manufacturer claims (Table 1.5). This chapter explains where the limit comes from.

WhoWhat they say about the limit
ISO 15189:2022 (clauses 7.3.1 b and 7.3.2 b)The performance specifications of each examination method must relate to its intended use and its effect on patient care; a verification confirms the specifications relevant to that use. It leaves the numbers to the laboratory.
The Milan consensus (2014)The main framework for choosing the numbers: the first Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM), held in Milan in 2014.
Chapter 3: which calculation, how sure? → Chapter 4: how much error can the test carry? → Chapter 5: the written plan
🧪 THE DECISION THIS CHAPTER SUPPORTS: How much error can this test carry and still serve its intended use? (Main references named by the book: EFLM Milan consensus (2015); EFLM Biological Variation Database; CLSI EP21-Ed2.)

By the end of this chapter you should be able to

  1. Explain why a specification must be fixed before any data are collected.
  2. Describe the three Milan models and choose the one that suits a measurand and its use.
  3. Calculate imprecision, bias and total error specifications from biological variation.
  4. Judge when state-of-the-art performance, external quality assessment limits or regulatory limits are acceptable sources.
  5. Turn a specification into acceptance criteria for each study in an evaluation plan.
  6. Record the choice so that an assessor can trace every number.

Map of this chapter

#Study sectionBook section(s)Printed pages
1Start here: the question on your deskChapter 4 opening, learning objectives73–74
2The number comes first§4.174
3Three models for setting specifications§4.2 (Stockholm and Milan, Table 4.1)74–75
4Which model for which measurand, and more than one use§4.2 (EFLM list, glucose example)75–76
5Specifications from biological variation: the formulas§4.3 (formulas and cautions)76
6Three levels, and the quality of the inputs§4.3 (Table 4.2, note on sources, EFLM database)77
7State of the art, EQA limits and regulatory limits§4.4 (incl. the CLIA example)77–78
8From specification to acceptance criteria§4.5 (Table 4.3, error budgets)78–79
9Two details often missed: the floor and the decision level§4.5 (concentration floor, hCG carryover example)79
10What quality control can and cannot do§4.5 (CLSI C24)79–80
11Worked Example 4.1 (part 1): model, inputs and three levelsWorked Example 4.1, Steps 1–4 (Table 4.4)80–81
12Worked Example 4.1 (part 2): decide and recordWorked Example 4.1, Step 581
13Worked Example 4.1 (part 3): the floor rule, budgets and Figure 4.1Worked Example 4.1, floor rule (Table 4.5), budgets, record, Figure 4.181–83
14Common errorsCommon errors83–84
15Chapter 4 checklist and referencesChapter 4 checklist; Chapter 4 references84–85
📌 REMEMBER: The number comes first. A specification has a model, a source, a formula and a reason, and it is written into the plan before the first sample is run.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, opening and learning objectives, pp.73–74; How this book is organised, p.viii

4.1 The number comes first

🔥 KEY POINT: A limit chosen after the results are known is not a criterion; it is a description of the results.

The temptation

A study finishes → the coefficient of variation (CV) is 4.1% → someone finds a published figure of 5% → the method "passes"

To prevent this, the evaluation plan (Chapter 5) states each specification, its source and its formula before the first sample is run, and the report compares the results with that plan.

Claim or specification? Keep the two ideas apart

ClaimSpecification
What it saysWhat a method can doWhat patients need
Where it comes fromThe manufacturerClinical evidence
Who handles itIt is verified (Chapter 1)It is chosen by the laboratory
💡 UNDERSTAND THIS: A method can meet its claims and still fail a specification, or the reverse. A verification usually checks both.
⚠️ COMMON ERROR: Searching the literature for a figure that makes a finished study pass. The source and the formula must already be in the plan.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.1 The number comes first, p.74

4.2 Three models for setting specifications

From five levels to three models

ConferenceWhat it agreed
Stockholm, 1999A five-level hierarchy for choosing specifications, from clinical outcome studies at the top to the current state of the art at the bottom
Milan, 2014Simplified that hierarchy into three models, each with its own principle (Table 4.1), and kept an order of preference: models 1 and 2 are preferred to model 3
Table 4.1 The three Milan models for analytical performance specifications
ModelPrincipleBest suited toMain limitation
1. Clinical outcomeHow analytical performance affects clinical outcomes, studied directly or indirectly (by simulation or decision analysis)Measurands central to decisions in a specific disease, with established cut-offsOutcome data exist for few measurands, and results depend on setting and population
2. Biological variationKeep analytical "noise" small compared with the biological "signal"Measurands in a steady state, especially those under tight homeostatic controlNeeds reliable, critically appraised biological variation data
3. State of the artThe best performance that current methods achieveMeasurands that cannot be placed in model 1 or 2, including some placed there temporarily while better data are awaitedWhat is achievable may not be what patients need
🧠 MEMORY AID: Outcome · Biology · Best available. The first two describe what patients need; the third describes what methods can do, which is why it comes last.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.2 Three models for setting specifications, pp.74–75

4.2 Which model for which measurand, and more than one use

The Milan statement did not allocate individual measurands to models; it recommended that such a list be drawn up. An EFLM working group then proposed criteria and a preliminary list covering some commonly requested measurands.

⚠️ READ WITH CARE: The authors present the list as a preliminary proposal and state that they did not themselves review the outcome or biological variation evidence for each measurand.

The working group's examples, with the clinical use discussed

ModelExamples (and the use behind them)
1. Clinical outcomeTotal, HDL and LDL cholesterol (classifying cardiovascular risk against treatment thresholds) · glucose and HbA1c (diagnosing and monitoring diabetes; for glucose monitoring the group notes that treatment goals are not always established, and other models can then be used) · plasma albumin (nutritional monitoring in dialysis, staging of myeloma, decisions on albumin infusion) · C-reactive protein (separating viral from bacterial infection; severity of acute pancreatitis) · cardiac troponins (diagnostic misclassification at the decision limit) · haemoglobin, platelets and neutrophils (anaemia, transfusion thresholds, infection risk of severe neutropenia) · thyroid-stimulating hormone (diagnosing and monitoring thyroid disease)
2. Biological variationPlasma electrolytes and minerals · plasma creatinine, urea and cystatin C · urate · total protein · erythrocyte indices · haemoglobin when used for monitoring · prothrombin time and activated partial thromboplastin time when used to monitor anticoagulant treatment
3. State of the artMeasurands that cannot be placed in model 1 or 2, and measurands placed there temporarily while outcome or biological variation data are awaited · mainly urine measurands, such as urinary electrolytes, creatinine and urate

(HDL = high-density lipoprotein; LDL = low-density lipoprotein.) The paper's summary table and its text differ for a few of these entries, so check both before relying on one.

🔥 KEY POINT: The same substance can fall under different models depending on the sample and the use: plasma creatinine → model 2, urine creatinine → model 3.

One measurand, more than one specification

A measurand can have more than one use, so it can have more than one specification. The Milan statement's example is glucose:

UseWhere the specification might come from
Blood glucose in critical careSimulating the effect of analytical performance on patient outcomes (an indirect outcome study)
Self-monitoring of blood glucose in type 1 diabetesClinical outcome studies or biological variation

The statement also allows models to be combined, and asks that every specification state its rationale, its source and the quality of the evidence behind it.

📌 REMEMBER: The intended use, written down as in Table 1.1, decides which model and which specification applies.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.2 Three models for setting specifications, pp.75–76

4.3 Specifications from biological variation: the formulas

Model 2 can be applied to most measurands for which reliable biological variation data exist, which is why it is the model laboratories meet most often.

The principle

SpecificationJudged against
ImprecisionThe natural within-subject biological variation (CVI): analytical imprecision should be small compared with it, so that it adds little to the variation the clinician sees
BiasThe combined within- and between-subject variation (CVI and CVG)

The commonly used formulas, at the "desirable" level

SpecificationFormula
ImprecisionCVA ≤ 0.5 × CVI
BiasB ≤ 0.25 × √(CVI2 + CVG2)
Total allowable errorTEa ≤ 1.65 × 0.5 × CVI + 0.25 × √(CVI2 + CVG2)

CVA = analytical coefficient of variation (imprecision), %; B = bias, %; TEa = allowable total error, %; CVI = within-subject biological variation, %; CVG = between-subject biological variation, %.

What each formula means, and its limits

FormulaWhat the book says about it
ImprecisionWith imprecision at half the within-subject variation, the total variation a clinician sees rises by only about 12%.
BiasProposed so that laboratories could share common reference intervals. It was not intended as a universal bias limit, and its use for other purposes should be justified.
Total errorAdds the two allowances in a straight line. Critics point out there is no theoretical basis for adding imprecision and bias linearly and that the result overstates the allowable error; it remains widely used because it is simple.
Measurement uncertaintyA specification of the same family limits the expanded uncertainty to Z × 0.5 CVI. Chapter 14 returns to uncertainty budgets.
⚠️ TWO CAUTIONS FROM THE EFLM TASK GROUP ON BIOLOGICAL VARIATION: (1) The bias formula assumes almost no imprecision, and when the imprecision and bias specifications are used side by side, neither controls the total error. (2) The total error formula was designed for single results in external quality assessment.
🧠 MEMORY AID: Imprecision → half of CVI · bias → a quarter of the combined variation · total error → 1.65 × imprecision + bias.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.3 Specifications from biological variation, p.76

4.3 Three levels, and the quality of the inputs

Three levels are often quoted, scaling the factors up or down. The cited textbook chapter attributes the scheme to Fraser and colleagues.

Table 4.2 Three levels of specification derived from biological variation
LevelImprecision (CVA)Bias
Optimum0.25 × CVI0.125 × √(CVI2 + CVG2)
Desirable0.50 × CVI0.250 × √(CVI2 + CVG2)
Minimum0.75 × CVI0.375 × √(CVI2 + CVG2)
💡 NOTE ON SOURCES: The minimum bias factor of 0.375 is taken from Fraser and colleagues' original paper (page 10, under "Allowable bias"). The textbook chapter cited above prints 0.325 for this factor, which appears to be a misprint. The imprecision factors and the optimum and desirable bias factors are as given in that chapter, which attributes the three-level scheme to the same paper and notes that it has no firm theoretical basis.
🔥 KEY POINT: These three levels are Fraser's proposals. They become acceptance criteria only when a laboratory adopts one of them, with its reasons, as Worked Example 4.1 shows.

The formulas are only as good as their inputs

Older tables and circulating spreadsheets ✗ (collected before modern quality criteria) → EFLM Biological Variation Database ✓ (current, critically appraised values)
  • Take current, critically appraised values of CVI and CVG from the EFLM Biological Variation Database.
  • Record the version or access date.
  • Check that the population studied resembles the patients the laboratory serves.
⚠️ COMMON ERROR: Using a biological variation spreadsheet "everyone has". Values in older tables were collected before modern quality criteria for biological variation studies existed.
🧠 MEMORY AID: Imprecision factors ¼ · ½ · ¾ of CVI; bias factors are half of those: 0.125 · 0.250 · 0.375.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.3 Specifications from biological variation (Table 4.2), p.77

4.4 State of the art, EQA limits and regulatory limits

Model 3: the state of the art

Model 3 uses the best performance that current methods achieve, read from external quality assessment (EQA) data, manufacturers' information or published evaluations.

StrengthWeakness
Easy to obtain; the only option for many measurandsDescribes what is possible, not what patients need
A moving target: as methods improve, yesterday's state of the art becomes today's minimum

EQA and regulatory acceptance limits

EQA schemes and regulators also publish acceptance limits. Many schemes base them on a combination of biological variation and state of the art. Regulatory limits differ between jurisdictions, and a laboratory must meet those that apply to it.

⚠️ TWO CAUTIONS WHEN THESE LIMITS ARE BORROWED FOR A VERIFICATION: (1) They were designed to judge a single EQA result, so they are often wide. (2) They are fixed for a whole scheme or country, not for the laboratory's intended use.
📌 REMEMBER: A laboratory may use them as an outer boundary, but it should say so in the plan and explain why a tighter specification was not chosen.

How wide such limits can be: a United States example

Included for international readers. Under the Clinical Laboratory Improvement Amendments (CLIA), the proficiency testing (PT) scoring criterion for creatinine is:

Target value ±10% or ±0.2 mg/dL (about 17.7 µmol/L), whichever is greater (42 CFR §493.931)
  • Each PT sample result is scored as acceptable or not against it.
  • Two dates apply: the revised criteria became effective on 11 July 2024, and the Centers for Medicare & Medicaid Services (CMS) set 1 January 2025 as the date from which PT programmes and laboratories implement them.
  • The previous criterion was ±15% or ±0.3 mg/dL.
CreatinineCLIA criterion allowsIllustrative specification of Worked Example 4.1 allows
60 µmol/L17.7 µmol/L4.4 µmol/L (the CLIA figure is four times larger)
300 µmol/L30 µmol/L22.2 µmol/L
🔥 KEY POINT: A method could pass every PT event and still fail that specification. The CLIA figure scores single PT results in one country; it is not a performance specification for a verification, and laboratories elsewhere look to the schemes and regulations that apply to them.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.4 State of the art, EQA limits and regulatory limits, pp.77–78

4.5 From specification to acceptance criteria

TermWhat it does
SpecificationDescribes the performance needed
Acceptance criterionApplies it to one study

Table 4.3 shows how the same specification feeds different studies in a plan.

Table 4.3 Turning a specification into acceptance criteria
Study (chapter)What is comparedTypical criterion
Precision (6)Within-laboratory CV, with its confidence intervalCV no greater than the imprecision specification; claim also checked with a verification limit (Chapter 3)
Method comparison (8)Bias at each decision level, or each difference as a share of the allowable total errorBias no greater than the bias specification; every difference within the allowable total error
Linearity (9)Deviation from linearity at each levelA stated fraction of the allowable total error or bias, chosen in advance
Interference (11), carryover (12)Effect on a sample near the lowest decision levelAn effect smaller than a clinically meaningful change at that level
Stability, where verified locallyChange over the storage periodA stated fraction of the allowable total error or bias

Error "budgets" in evaluation software

BudgetMeaning
Random error budget (for example 25% or 50%)The allowed SD or CV is that share of the allowable total error
Systematic error budgetThe same, for bias
⚠️ COMMON ERROR: Treating budget shares as fixed rules. They are conventions, not standards. Choose them on purpose, write them in the plan, and make sure the resulting limits agree with the specifications in Table 4.2 rather than contradicting them.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.5 From specification to acceptance criteria (Table 4.3), pp.78–79

4.5 Two details often missed: the floor and the decision level

Detail 1: a percentage needs a concentration floor

A percentage specification becomes impossible near zero, so most laboratories state it as "a concentration or a percentage, whichever is greater". The rule must say:

  • which value the percentage is taken of, and
  • where the two limits meet;
  • and the concentration floor needs its own justification, usually from the method's imprecision near the lowest decision level.

Worked Example 4.1 (study section 13) shows a complete rule.

Detail 2: the limit must be able to detect an error that matters at the decision level

Carryover shows why. When a specimen with a very high concentration is followed by one with a low concentration, a small fraction of the first can be carried into the second and raise its result.

Illustrative figures (human chorionic gonadotropin, hCG)Value
High specimen200,000 IU/L
Next specimen (non-pregnant patient), true value2 IU/L
Reported positive at5 IU/L or more
Carryover of 0.01% adds20 IU/L → the low specimen reads 22 IU/L and is reported positive
An acceptance limit of 25 IU/L would accept that effect → it cannot protect the decision it is meant to protect
🔥 KEY POINT: The allowable effect must be smaller than the gap between the low specimen and the threshold (3 IU/L here), with room left for the method's imprecision at that level. A limit of 1 IU/L, chosen locally, does that.
🧪 AT THE BENCH: The high specimen should be as high as patient specimens realistically get, because the effect grows with it. Chapter 12 covers carryover studies.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.5 From specification to acceptance criteria, p.79

4.5 What quality control can and cannot do

CLSI C24 takes the same position

CLSI C24, the quality control (QC) guideline:

  • asks the laboratory director to set the allowable total error from the way results are used in the population the laboratory serves, independently of how the method actually performs, and to agree it with the clinicians;
  • lists the three Milan models in the same order of preference;
  • notes that no single model suits every measurand, and that when no available method can meet the chosen goal, the laboratory may have to re-examine the goal or use a QC strategy sensitive enough to catch small deteriorations.

Chapter 18 shows how the same specifications drive the choice of control rules.

A more sensitive QC strategy does not make a method better

A more sensitive strategy (tighter limits, more control results per run, more frequent QC events)…
What it doesDetects smaller shifts or drifts away from the method's stable performance, and detects them sooner
What it costsMore false rejections, more repeat testing and slower reporting
What it cannot doBring any benefit when the problem is the method's own stable imprecision: random error affects control and patient results independently, so a control result inside tighter limits does not show that patient results meet the imprecision goal

Where the evidence of fitness comes from

WhenEvidence that the method meets the goal for its intended use
Before useThe precision, bias and total error estimates from the evaluation, each compared with the specification fixed in advance (Chapters 6, 8 and 14)
In useThe long-term QC SD, external quality assessment results and comparability checks that ISO 15189:2022 asks laboratories to monitor (clauses 7.3.7.1 to 7.3.7.4)
📌 REMEMBER: Quality control watches a method that has been shown to be fit for purpose; it cannot replace that demonstration.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, §4.5 From specification to acceptance criteria, pp.79–80

Worked Example 4.1 (part 1): choosing the model, recording the inputs, calculating the levels

⚠️ ILLUSTRATIVE VALUES: The biological variation values in this example are illustrative, chosen to show the calculation. They are not quoted from the EFLM database; take current values from the database when setting real specifications. The state-of-the-art figure is also illustrative.

Situation

The laboratory in Worked Example 3.1 must set specifications for its new creatinine method before the evaluation begins. Results are used to:

  • estimate glomerular filtration rate,
  • monitor patients with kidney disease, and
  • adjust drug doses.

Step 1 — Choose the model

Plasma creatinine is held in a steady state by kidney function, and the EFLM working group lists it among the measurands suited to the biological variation model; urine creatinine, by contrast, is in its state-of-the-art group.

Plasma creatinine → steady state → Model 2 applies

Step 2 — Record the inputs

Input (illustrative)Value
CVI (within-subject)4.5%
CVG (between-subject)14.0%
√(CVI2 + CVG2)14.7%

Step 3 — Calculate the three levels

Table 4.4 Specifications for creatinine from the illustrative values
LevelImprecision (CVA)BiasTotal error (1.65 × CVA + bias)
Optimum1.13%1.84%3.7%
Desirable2.25%3.68%7.4%
Minimum3.38%5.51%11.1%

The arithmetic, row by row

LevelImprecisionBiasTotal error
Optimum0.25 × 4.5 = 1.13%0.125 × 14.7 = 1.84%1.65 × 1.13 + 1.84 = 3.7%
Desirable0.50 × 4.5 = 2.25%0.250 × 14.7 = 3.68%1.65 × 2.25 + 3.68 = 7.39 → 7.4%
Minimum0.75 × 4.5 = 3.38%0.375 × 14.7 = 5.51%1.65 × 3.38 + 5.51 = 11.1%

The minimum row uses Fraser's factors of 0.75 for imprecision and 0.375 for bias. Like the other two rows, it is a proposal for comparison, not an acceptance criterion; the limits the laboratory adopts in Step 5 do not change.

Step 4 — Compare with what methods achieve

In this example the best-performing methods in the laboratory's EQA scheme achieve a within-laboratory CV of about 2.5%.

LevelImprecisionCompared with the state of the art (about 2.5%)
Desirable2.25%At the edge of what current methods deliver
Minimum3.38%Comfortably achievable
📌 REMEMBER: The three levels are calculated first, then moderated by what current methods achieve. The decision comes in Step 5.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Worked Example 4.1 Choosing specifications for creatinine, Steps 1–4 (Table 4.4), pp.80–81

Worked Example 4.1 (part 2): Step 5 — decide and record

The laboratory adopts three limits. They are chosen separately. Only the total error limit is calculated from other numbers, and it is calculated from the desirable levels, not from the two adopted limits.

Limit adoptedValueHow it was chosen
ImprecisionWithin-laboratory CV no greater than 3.0%A local judgement, not a formula result. The desirable level (2.25%) is below what the best current methods achieve (about 2.5%), so it would fail most methods. The minimum level (3.38%) is easily met but gives away more than necessary. The laboratory chose a round figure between the two and recorded 2.25% as its long-term aim.
BiasNo greater than 3.7% at each decision levelThe desirable bias level (3.68%) rounded to one decimal place. It applies at the decision levels used for estimating glomerular filtration rate. Because filtration-rate equations assume results traceable to the reference measurement procedure, the laboratory adopts the formula here even though it was designed for shared reference intervals.
Allowable total error7.4% of the comparative method's result, or 4 µmol/L, whichever is greaterThe percentage is the desirable total error, 1.65 × 2.25 + 3.68 = 7.39%, rounded to 7.4%. It is used to judge each patient difference in the comparison study (Chapter 8).

Limits chosen separately do not automatically agree

A method sitting exactly at the adopted limits: 1.65 × 3.0 + 3.7 = 8.65% → more than 7.4%
🔥 KEY POINT: Passing the precision and bias studies separately does not prove that total error is acceptable, a point the EFLM task group on biological variation also makes. The plan therefore also compares the total error estimated from the precision and bias results with 7.4% (Chapter 14).
⚠️ COMMON ERROR: Calculating the allowable total error from the two adopted limits (3.0% and 3.7%). In the example it comes from the desirable levels (2.25% and 3.68%).

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Worked Example 4.1 Choosing specifications for creatinine, Step 5, p.81

Worked Example 4.1 (part 3): the floor rule, budgets and Figure 4.1

The floor rule

For each patient sample, the allowable difference is the larger of 4 µmol/L and 7.4% of the comparative method's result.

The two are equal at 4 ⁄ 0.074 = 54 µmol/L → below 54 µmol/L the absolute limit of 4 µmol/L applies; at 54 µmol/L and above, the percentage limit applies
Table 4.5 Applying the floor rule to two results (illustrative)
Comparative result7.4% of itAllowable differenceNew method resultDifferenceJudgement
40 µmol/L3.0 µmol/L4.0 µmol/L (floor)43.5 µmol/L3.5 µmol/L (8.8%)Within the limit
120 µmol/L8.9 µmol/L8.9 µmol/L (percentage)130.0 µmol/L10.0 µmol/L (8.3%)Outside the limit
  • The first result would fail a percentage limit alone (8.8% is more than 7.4%), although a difference of 3.5 µmol/L at 40 µmol/L is within what the method's imprecision allows.
  • The second shows the percentage limit taking over at higher concentrations.

Justifying the floor

The floor itself is a local choice. Near 45 µmol/L the laboratory expects a within-laboratory SD of about 1.2 µmol/L (illustrative), so:

Random error alone uses about 1.65 × 1.2 = 2.0 µmol/L → a floor of 4 µmol/L leaves about 2 µmol/L for bias

The clinical users accepted this for the low range, and the plan records their agreement.

Budgets for software

BudgetCalculation (share of the 7.4% total error)Result
Random error budget3.0 ⁄ 7.441%
Systematic error budget3.7 ⁄ 7.450%

Entering these values makes the software reproduce the plan's limits.

🧪 THE RECORD names: the model, the EFLM database version and access date, the formulas, the level and reason for each choice, the floor rule and who approved them.

Figure 4.1 Candidate specifications for within-laboratory imprecision, and the result of Worked Example 3.1

Figure 4.1: candidate specifications for the within-laboratory CV on an axis from 0% to 4.5%. Dashed lines mark optimum 1.13%, desirable 2.25%, illustrative state of the art 2.50% and minimum 3.38%; a solid black line marks the laboratory specification of 3.00%. A blue dot at 2.31% with a bar from 1.63% to 3.97% shows the CV observed in Worked Example 3.1 and its 95% confidence interval. Title: the method meets the chosen 3.0% specification, but its interval reaches past the minimum.

Illustrative values · Worked Example 4.1 · observed CV from Worked Example 3.1 (synthetic)

💡 HOW TO READ THIS FIGURE: Each dashed line is a candidate specification for the within-laboratory CV: the three biological variation levels and the illustrative state of the art. The solid black line is the specification the laboratory adopted, 3.0%. The blue dot is the CV observed in Worked Example 3.1 and the blue bar its 95% confidence interval. The estimate lies below the adopted specification, which is why the method was accepted. The interval reaches past both the specification and the minimum level, which is why the laboratory will review its quality control data after three months.

What this example shows

  • The 3.0% used in Worked Example 3.1 was not chosen after the data were seen. It came from biological variation, was moderated by what current methods achieve, and was written into the plan with its reason.
  • Limits chosen separately must be checked together.
  • The same process explains the opening question: both glucose limits may be defensible, but only if each laboratory can show where its number came from and why it suits its use.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Worked Example 4.1 Choosing specifications for creatinine, floor rule (Table 4.5), budgets, record and Figure 4.1, pp.81–83

Common errors

⚠️ Common errorWhat to do instead
Choosing the limit after seeing the resultsA specification must be in the plan before the first sample is run.
Confusing a claim with a specificationMeeting the manufacturer's claim does not show that the method meets patients' needs.
Using outdated biological variation tablesTake appraised values from the EFLM database and record their date.
Applying the bias formula everywhere without thoughtIt was designed for shared reference intervals.
Borrowing EQA or regulatory limits without saying soThey judge single results and are often wide.
Stating only a percentageNear zero a percentage becomes impossible; give a justified concentration floor.
Budget fractions that contradict the specificationsSoftware budgets must reproduce the limits in the plan.
Treating separately chosen limits as a guarantee of total errorImprecision and bias limits set one at a time do not control total error; check the combination as well.
🧠 MEMORY AID: Before · Patients · Current · Purpose · Declared · Floor · Consistent · Combined — one word for each error, in the book's order.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Common errors, pp.83–84

Chapter 4 checklist and references

Before approving an evaluation plan, confirm that for each measurand:

Checklist itemWhere taught
☐The intended use is written down and the specification model is chosen to suit it.§4.2
☐The source of every input is recorded, with version or access date.§4.3
☐Imprecision, bias and total error specifications are stated, with the level chosen for each and the reason.§4.3, Worked Example 4.1
☐A concentration floor is stated where a percentage alone would fail at low concentrations, with its justification.§4.5
☐The floor rule states which result the percentage is taken of and the concentration at which the absolute and percentage limits meet.§4.5, Worked Example 4.1
☐Any EQA, regulatory or state-of-the-art limit used is identified as such.§4.4
☐Each study in the plan has an acceptance criterion derived from these specifications.§4.5
☐The specifications were approved before data collection began.§4.1
Chapter 4: how much error can the test carry? → Chapter 5: which studies, materials, sample sizes and criteria go into the written plan

Chapter 4 references (as cited by the book)

  1. Sandberg S, Fraser CG, Horvath AR, et al. Defining analytical performance specifications: consensus statement from the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine. Clin Chem Lab Med. 2015;53(6):833–5. doi:10.1515/cclm-2015-0067.
  2. Ceriotti F, Fernandez-Calle P, Klee GG, et al. Criteria for assigning laboratory measurands to models for analytical performance specifications defined in the 1st EFLM Strategic Conference. Clin Chem Lab Med. 2017;55(2):189–94. doi:10.1515/cclm-2016-0091.
  3. Sandberg S, Røraas T, Aarsand AK. Biological variation and analytical performance specifications. In: Rifai N, Chiu RWK, Young I, Burnham CD, Wittwer CT, editors. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023. Chapter 8 (online), pp. 165.e14–165.e17.
  4. Fraser CG, Hyltoft Petersen P, Libeer JC, Ricos C. Proposals for setting generally applicable quality goals solely based on biology. Ann Clin Biochem. 1997;34(Pt 1):8–12. doi:10.1177/000456329703400103.
  5. European Federation of Clinical Chemistry and Laboratory Medicine. EFLM Biological Variation Database. biologicalvariation.eu (accessed November 2026).
  6. International Organization for Standardization. ISO 15189:2022 Medical laboratories — Requirements for quality and competence. 4th ed. ISO; 2022 (December). Clauses 7.3.1 b), 7.3.2 b) and 7.3.7.1–7.3.7.4.
  7. Sandberg S, Coskun A, Carobene A, et al. Analytical performance specifications based on biological variation data – considerations, strengths and limitations. Clin Chem Lab Med. 2024;62(8):1483–9. doi:10.1515/cclm-2024-0108.
  8. Clinical and Laboratory Standards Institute. Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions. 4th ed. CLSI guideline C24-Ed4. CLSI; 2016. Sections 3.2, 5.1 and 5.4.2.
  9. Centers for Medicare & Medicaid Services, HHS. Clinical Laboratory Improvement Amendments of 1988 (CLIA) proficiency testing regulations related to analytes and acceptable performance. Final rule (CMS-3355-F). Federal Register 11 July 2022;87(131):41194; effective 11 July 2024. Codified at 42 CFR §493.931 (routine chemistry: creatinine).
  10. Centers for Medicare & Medicaid Services, Quality, Safety & Oversight Group. Implementation Notification for Final Rule CMS-3355-F, Clinical Laboratory Improvement Amendments of 1988 (CLIA) Proficiency Testing — Analytes and Acceptable Performance Final Rule. Ref. Admin Info-23-07-CLIA. 3 May 2023.

CLSI and ISO documents are cited for attribution only. Readers should obtain the current edition for the full text. Citation does not imply endorsement by CLSI, ISO, EFLM, any accreditation body or any manufacturer.

Source: Clinical Laboratory Method Evaluation in Practice (Shawki, 1st ed. 2026), Ch.4, Chapter 4 checklist; Chapter 4 references, pp.84–85

Training course, not a certification. Based on the author's textbook; all examples use synthetic data. It does not replace the CLSI and ISO documents it cites, and it is not affiliated with or endorsed by CLSI or ISO.

© 2026 Hassan Shawki. Course text adapted from the author's textbook Clinical Laboratory Method Evaluation in Practice.