1 · Ion-Selective Electrodes (ISE) and electrochemistry
📚 Why this section: Calendar (Instrumentation II + QC) items: Ion-Selective Electrode (ISE); Direct ISE; Indirect ISE · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I E 1–2 Electrochemistry, potentiometric methods (p.6) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Electrochemistry (potentiometry, amperometry, coulometry) (p.237)
Potentiometry = measuring a voltage (potential) between two electrodes when no current flows. The voltage tells us the activity of one ion. An Ion-Selective Electrode (ISE) is a potentiometric electrode with a membrane that "prefers" one ion.
The parts of every ISE system
| Part | Job | Examples |
|---|---|---|
| Indicator (measuring) electrode | Voltage changes with the ion | Na⁺ = special glass (aluminum silicate) · pH = glass · K⁺ = valinomycin membrane · Ca²⁺, Li⁺ = organic ion-exchanger (ionophore) · NH₄⁺ = nonactin/monactin |
| Reference electrode | Gives a CONSTANT voltage | Silver/silver chloride (Ag/AgCl) · Calomel (mercury + mercurous chloride + saturated KCl). Ag/AgCl potential depends on the Cl⁻ of its KCl filling solution. |
| Salt bridge / liquid junction | Completes the circuit | Blocked junction → unstable, noisy readings |
| Standard electrode | Zero point for all half-cells | Standard hydrogen electrode = 0.000 V |
An ISE measures activity (the "free, active" ion), and it responds to one ion much more than to other ions — it is not 100% specific. Selectivity depends on the membrane.
💡 Direct vs indirect ISE — the most tested idea
| Direct ISE | Indirect ISE | |
|---|---|---|
| Sample | Undiluted (whole blood, plasma, serum) | Diluted before measurement |
| Where used | Blood gas analyzers, point-of-care | Large automated chemistry analyzers |
| What it measures | Ion in plasma water (mmol/kg water) | Ion in total plasma volume (mmol/L) |
| High lipids / very high protein | NOT affected | Falsely LOW Na⁺ (pseudohyponatremia) — "electrolyte exclusion" |
Blood gas and other electrochemical methods
| Method | Measures | Key point |
|---|---|---|
| pH electrode | H⁺ | Glass electrode, potentiometric. Calibrate with two buffers of known pH at 37 °C. Store in the medium the manufacturer recommends. |
| PCO₂ (Severinghaus) | CO₂ | A pH electrode inside a bicarbonate solution behind a CO₂-permeable membrane. CO₂ → carbonic acid → pH falls. Potentiometric. |
| PO₂ (Clark) | O₂ | Amperometric (CURRENT, not voltage). Platinum cathode reduces O₂; Ag/AgCl anode. Current ∝ PO₂. Based on polarography. |
| Coulometry | Chloride (sweat Cl⁻) | Silver electrode releases Ag⁺ at constant current → AgCl precipitates; endpoint by amperometry. Titration time ∝ Cl⁻. Bromide interferes (falsely high). |
| Amperometric glucose | Glucose | Glucose oxidase → H₂O₂ oxidized at platinum electrode → current |
| Anodic stripping voltammetry | Lead, heavy metals | Metal plated onto electrode, then stripped off → current |
Blood gas analyzers measure pH, PCO₂ and PO₂; they calculate HCO₃⁻ and base excess.
More electrochemistry facts (added in v1.1)
| Principle | What is measured | Examples |
|---|---|---|
| Potentiometry | Voltage at zero current | pH, Na⁺, K⁺, Ca²⁺ ISE, PCO₂ |
| Amperometry / polarography | Current at a fixed applied voltage | PO₂ (Clark), glucose oxidase electrode, HPLC-ECD (glassy carbon) |
| Coulometry | Quantity of charge (coulombs) over time | Chloride titrator (Ag⁺ generated) |
| Anodic stripping voltammetry | Current when plated metal is stripped off | Lead |
- The Ag/AgCl reference potential is set by the chloride (KCl) concentration of its filling solution. Reference electrodes: Ag/AgCl, calomel, hydrogen. Glass = indicator, not reference.
- In the Clark electrode, O₂ is reduced at the platinum cathode; the silver anode supplies the electrons. Current ∝ PO₂.
- Store a pH glass electrode wet, in the manufacturer's recommended solution — never dry, never in tap water.
Sources for v1.1 additions: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q36, Q44, Q54, Q59 (pp.180–184); SUCCESS! in Clinical Laboratory Science, Ch.1, Q27, Q34, Q36 (pp.98–99; rationales pp.150–152); BOC Study Guide, 6th ed., Ch.3, Q323 (p.190); A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1, p.6.
📚 Book check — added from the two review books
- Anode = the electrode that electrons flow FROM. Cathode = the electrode that ACCEPTS the electrons. The two half-cells are joined by a salt bridge. (Hubbard Ch.1, p.6)
- PCO₂ electrode = a gas-sensing electrode: a pH glass electrode covered by a silicone rubber membrane that lets CO₂ pass. CO₂ + H₂O ⇌ H₂CO₃ ⇌ H⁺ + HCO₃⁻ → the H⁺ change is measured. (6.0 p.237)
- Enzyme electrode = an electrode covered by a layer of immobilized enzyme. The enzyme reaction makes a product that the electrode then measures. (6.0 p.237)
📚 How the sections were chosen
Two sources: the fixed calendar list for this day and the chapter headings of A Concise Review of Clinical Laboratory Science (Hubbard). Check book: Clinical Laboratory Science Review, 5th ed. (6.0).
| # | Study section | Calendar items it covers | Book heading (Hubbard, page) |
|---|---|---|---|
| 1 | ISE & electrochemistry | ISE; Direct ISE; Indirect ISE | Ch.1 §I E 1–2 Electrochemistry (p.6) |
| 2 | Osmometry | Osmometry; Freezing point depression | Ch.1 §I E 3–4 Osmometry (pp.6–7) |
| 3 | Electrophoresis | Electrophoresis | Ch.1 §II A Electrophoresis (pp.7–8) |
| 4 | Chromatography & mass spec | Chromatography | Ch.1 §II C Chromatography (pp.8–9) |
| 5 | Immunoassay principles | Immunoassay; Competitive; Sandwich; Chemiluminescence | Ch.1 §II B Immunoassay, Table 1–2 (pp.8–9) |
| 6 | Immunoassay errors | Immunoassay interferences; QC troubleshooting | Ch.1 §II B 1 Immunoassay components (p.8) |
| 7 | Flow cytometry | Flow cytometry concepts | Ch.3 §II D Laser scatter counting and flow cytometry (p.91) |
| 8 | QC statistics | Mean; SD; CV; Precision; Accuracy; Bias; Random error; Systematic error | Ch.1 §I B Statistical concepts (pp.2–3) |
| 9 | Levey-Jennings & Westgard | Levey-Jennings charts; Westgard rules | Ch.11 §III B 6 QC program — shifts and trends (p.372) |
| 10 | QC action, QA & method evaluation | QC troubleshooting; Bias | Ch.11 §III Quality assurance (pp.371–372); §IV A 4–5 QC and PT (p.375) |
| 11 | Top exam traps | — (review) | Ch.1 §II B Immunoassay (p.8) |
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.17 Electrochemistry and Chemical Sensors (pp.224.e4–224.e8); Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q34–Q47 (pp.179–182); SUCCESS! in Clinical Laboratory Science, Ch.1, Q24–Q36 rationales (pp.149–152); BOC Study Guide, 6th ed., Ch.3, Q309, Q319–Q328 (pp.188–190).
2 · Osmometry — counting dissolved particles
📚 Why this section: Calendar (Instrumentation II + QC) items: Osmometry; Freezing point depression · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I E 3–4 Osmometry and osmolality (pp.6–7) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Osmometry (pp.235–236)
Osmolality = number of dissolved particles per kilogram of water (mOsm/kg). Particle number matters, not size. Lipids and proteins are not dissolved particles, so they add almost nothing.
Colligative properties (change with particle number)
Freezing point depression osmometer (cryoscopic)
- Sample is supercooled, then stirred → ice forms → temperature rises to a plateau = the freezing point.
- Temperature is sensed by a thermistor (temperature-sensitive resistor) measured with a Wheatstone bridge.
- Freezing point ↓ is directly proportional to osmolality. Example: −0.558 °C ÷ 1.86 = 0.300 Osm/kg = 300 mOsm/kg.
- Detects volatile substances such as ethanol and methanol → used for the osmolal gap.
Vapor pressure (dew point) osmometer
- Uses a thermocouple (Seebeck effect) to find the dew point.
- ⚠️ Misses volatile solutes (ethanol, methanol) because they evaporate. Do NOT use it to calculate an osmolal gap in poisoning.
📚 Book check — added from the two review books
- Freezing point osmometer errors: cooling bath too cold or too warm; particles in the sample stop proper supercooling (the sample freezes too early). (6.0 pp.235–236)
- Same sample, freezing point osmolality much HIGHER than vapor pressure osmolality → a volatile solute (e.g., ethanol, methanol) is present. The vapor pressure method misses it, so it reads lower. (6.0 pp.235–236)
Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q48–Q53 (pp.182–183); BOC Study Guide, 6th ed., Ch.3, Q305, Q353–Q354 (pp.188, 195); DoH Past Paper Exam — Chemistry, p.8, Q29.
3 · Electrophoresis — separation by charge (and size)
📚 Why this section: Calendar (Instrumentation II + QC) items: Electrophoresis · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §II A Electrophoresis (pp.7–8) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Electrophoresis and Sources of Error (pp.236–237)
Electrophoresis = movement of charged molecules in an electric field. Speed depends on net charge, size and shape, the buffer (pH, ionic strength), the support medium, voltage/current and temperature.
Order of serum protein fractions (pH 8.6, from anode)
| Fraction | Main proteins |
|---|---|
| α1 | α1-antitrypsin, α1-acid glycoprotein (orosomucoid) |
| α2 | α2-macroglobulin, haptoglobin, ceruloplasmin |
| β | Transferrin, β-lipoprotein (LDL), C3, C4 (fibrinogen if plasma is used!) |
| γ | Immunoglobulins |
Electroendosmosis (electro-osmotic flow)
Agarose and cellulose acetate carry fixed negative charges. Buffer cations (with water) move to the cathode. This flow pushes against protein migration, so γ-globulins can be carried slightly toward the cathode (behind the application point). Capillary electrophoresis uses this flow on purpose (no solid support; high voltage; UV detector).
Support media, stains and problems
- Media: agarose, cellulose acetate (5–7 fractions); polyacrylamide and starch gel separate by charge AND size (molecular sieving; SDS-PAGE gives molecular weight). Celite is NOT an electrophoresis medium (it is a GC support).
- Protein stains: Ponceau S, amido black, Coomassie blue; silver stain is most sensitive (CSF). Lipoprotein stains: Oil red O, Fat red 7B, Sudan black. DNA: ethidium bromide.
- Quantify bands with a densitometer.
- Raising ionic strength of buffer → slower migration (more heat). Raising voltage → faster.
- Problems: doughnut-shaped (hollow) band = protein too concentrated → dilute; distorted zones = overloading; no migration = broken contact between buffer and medium; bisalbuminemia/wide albumin = drugs (e.g., penicillin) bound to albumin.
- Isoelectric focusing: pH gradient made by ampholytes; each protein stops at its pI. Best for proteins with very close pIs (e.g., CSF oligoclonal bands).
- DNA: agarose gel separates large DNA fragments; polyacrylamide for small ones. Western blot = protein electrophoresis + transfer + antibody detection.
- Hemoglobin: cellulose acetate pH 8.6 (A fastest; S, D, G together; A2, C, E, O together) → confirm with citrate agar pH 6.0–6.2 (separates S from D, C from A2).
- Silver stain is about 30 times more sensitive than Coomassie blue → CSF and urine proteins can be run without concentrating them.
- Isoelectric focusing: a protein moves until it reaches the pH equal to its pI; there its net charge is zero and it stops (focuses).
📚 Book check — added from the two review books
- Migration speed: more charge → faster · bigger molecule → slower · more viscous medium → slower · higher buffer ionic strength → slower · lower buffer pH → slower · too much heat → proteins denature and migrate slower. (Hubbard Ch.1, p.7; 6.0 p.236)
- Prealbumin moves even faster than albumin. Six bands from the anode: prealbumin, albumin, α1, α2, β, γ. (6.0 p.248, Q4 rationale)
- Crescent-shaped bands → bent applicator or sample overload. Split/broken bands → too much pressure on the wire applicator. (6.0 p.237; p.248, Q5 rationale)
- No bands at all after staining, on patients AND controls → voltage too high (proteins denatured). No migration → instrument not connected or wrong buffer pH. Sample precipitates in the support → pH too high/low or voltage too high. (6.0 p.237; p.246, Q8; p.248)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.18 Electrophoresis (pp.225–227, 235); Medical Laboratory Science Review (Harr), Ch.5, §5.6, Q11–Q20, Q41 (pp.248–249, 255); SUCCESS! in Clinical Laboratory Science, Ch.1, Q37–Q47 (pp.99–100, rationales pp.152–154); BOC Study Guide, 6th ed., Ch.3, Q313–Q318 (p.189).
4 · Chromatography and mass spectrometry
📚 Why this section: Calendar (Instrumentation II + QC) items: Chromatography · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §II C Chromatography (pp.8–9) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Chromatography (p.236)
Chromatography separates substances by how they share themselves between a stationary phase and a mobile phase (differential solubility / partitioning). Substances that like the mobile phase move fast; substances that stick to the stationary phase move slowly.
| Type | Mobile phase | Stationary phase | Key exam facts |
|---|---|---|---|
| Thin-layer (TLC) | Liquid (solvent) | Silica gel on glass/plastic | Qualitative drug screen; non-volatile mixtures. Rf = distance of spot ÷ distance of solvent front (from origin). All spots at the solvent front → wrong solvent mixture. |
| Gas–liquid (GLC/GC) | Gas (carrier, e.g., nitrogen, helium) | Liquid coated on column packing | Sample must be volatile or made volatile. Elution order ≈ boiling point. Detector: flame ionization. Alcohols, drugs. |
| High-performance liquid (HPLC) | Liquid pumped under pressure | Packed column (e.g., C18) | Parts: injector, pump(solvent delivery), column, detector (UV, fluorescence, electrochemical). Reverse phase = stationary phase nonpolar (C18), mobile phase more polar. Isocratic = constant solvent; gradient = changing solvent. High pressure + bad peaks → obstructed line. |
| Ion-exchange | Liquid | Charged resin | Separates by sign and size of charge (e.g., HbA1c, amino acids, CK isoenzymes historically) |
| Gel filtration (size exclusion) | Liquid | Porous gel | Separates by molecular size/weight |
Internal standard (added to every sample and standard) corrects for variation in extraction and injection.
Mass spectrometry (MS)
- Molecules are ionized (usually to cations) and fragments are separated by their mass-to-charge ratio (m/z). The fragment pattern is a "fingerprint".
- Needs a vacuum in the mass filter (stops ions colliding).
- GC-MS = confirmation test for drugs of abuse (gold standard). GC-MS uses electron (impact) ionization. LC-MS uses electrospray ionization.
- Tandem MS (MS/MS) = newborn screening for amino acid, organic acid and fatty acid disorders; the first mass filter acts like a chromatography column.
- ICP-MS = trace elements and metals.
More chromatography and MS facts (added in v1.1)
- HPLC fractions are usually detected by a UV spectrophotometer. Other HPLC detectors: fluorescence, electrochemical (glassy carbon electrode — catecholamines), mass spectrometer. GC detectors: flame ionization, thermal conductivity, MS.
- MS measures cations by m/z. Base peak = the most abundant ion (set at 100%).
- GC-MS ionizes with an electron source (electron impact, under vacuum). LC-MS uses electrospray. ICP-MS (argon plasma torch) = trace metals. MALDI = proteins/bacteria.
- Vacuum in the mass filter prevents ions colliding with gas molecules.
- Tandem MS: the first mass filter acts like the chromatography column (selects parent ions); collision cell makes daughter ions.
- Troubleshooting: peaks at m/z 16, 17, 18 = oxygen (and its isotopes) → leak / contaminated carrier gas.
- Screen positive by immunoassay but GC-MS negative → the screen was a false positive (cross-reactivity); report negative.
📚 Book check — added from the two review books
- Adsorption (liquid–solid) chromatography: sample and mobile phase compete for binding sites on a solid. Partition (liquid–liquid): separation by solubility in a polar (water) vs a nonpolar (organic) liquid. (Hubbard Ch.1, p.9; 6.0 p.236)
- Ion-exchange: ions with the greatest charge density are held most strongly. Anion + cation exchange resins mixed together are used to deionize water. (Hubbard Ch.1, p.9; 6.0 p.236)
- TLC: the solvent moves up the plate by capillary action; TLC is a semi-quantitative screening test (drugs, lipids, carbohydrates, amino acids). (Hubbard Ch.1, p.9; 6.0 p.236)
- GC: the retention time identifies each compound (methanol, isopropanol, drugs, organic acids). GC is the best method for blood alcohol. (6.0 p.236)
- HPLC troubleshooting: noisy baseline → bubbles or particles in the flow cell, or a leaking filter · drifting baseline → overloaded column or contaminated solvent · loss of resolution → column overload or degraded column packing. (6.0 p.236)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.19 Chromatography (p.268) and Ch.20 Mass Spectrometry; Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q55–Q73 (pp.183–187); SUCCESS! in Clinical Laboratory Science, Ch.1, Q48–Q58 (pp.100–101, rationales pp.154–155); BOC Study Guide, 6th ed., Ch.3, Q310–Q312 (pp.188–189) and Ch.8, Q213–Q220 (pp.493–494); DoH Past Paper Exam — Chemistry, p.8, Q28.
5 · Immunoassays — competitive vs sandwich
📚 Why this section: Calendar (Instrumentation II + QC) items: Immunoassay; Competitive immunoassay; Sandwich immunoassay; Chemiluminescence concepts · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §II B Immunoassay and Table 1–2 (pp.8–9) · Check book: Clinical Laboratory Science Review, 5th ed., Immunology and Serology — RIA and EIA/ELISA (p.40)
An immunoassay uses an antibody to bind the analyte, and a label to make the binding measurable. Used for hormones, drugs, tumor markers, vitamins B12/folate, troponin, infectious serology. Very sensitive (pg–ng amounts). Electrolytes are NOT measured by immunoassay (they use ISE).
Labels
| Label | Name | Notes |
|---|---|---|
| Radioisotope (¹²⁵I) | Radioimmunoassay (RIA) | Historical; gamma counter; radiation hazard, short shelf life |
| Enzyme (HRP, ALP, G6PD, β-galactosidase) | EIA / ELISA / EMIT / CEDIA | Product read by spectrophotometer; cheaper and safer than RIA |
| Fluorophore | FIA, FPIA | Fluorescence polarization (FPIA) for drugs |
| Chemiluminescent (acridinium ester, luminol + HRP, dioxetane + ALP) | CLIA | 🔥 Most common automated immunoassay; most sensitive (e.g., high-sensitivity troponin); no lamp needed |
| Ruthenium complex | Electrochemiluminescence (ECL) | Light produced at an electrode; very low detection limit |
💡 Competitive vs noncompetitive (sandwich)
| Competitive (limited reagent) | Sandwich (two-site, excess reagent, "immunometric") | |
|---|---|---|
| How | Labeled antigen and patient antigen compete for a LIMITED number of antibody sites | Capture antibody on the solid phase binds antigen; a second labeled antibody binds another epitope |
| Signal vs concentration | 🔥 Bound signal is INVERSELY proportional to analyte | Signal is DIRECTLY proportional to analyte |
| Best for | Small molecules (drugs, T4, steroid hormones) | Large molecules with ≥2 epitopes (TSH, hCG, troponin, PSA) |
| Special error | — | Hook (prozone) effect at extremely high antigen → falsely low |
Homogeneous vs heterogeneous
- Heterogeneous = must separate bound from free label (washing): RIA, ELISA, IRMA, most CLIA.
- Homogeneous = no separation needed: EMIT, FPIA, CEDIA, LOCI, particle-enhanced turbidimetric inhibition.
- EMIT: enzyme (G6PD) is coupled to the drug (antigen). Antibody binding blocks the enzyme. Free enzyme-drug is active → NAD⁺ → NADH read at 340 nm. Enzyme activity is directly proportional to drug concentration.
ELISA formats
- Indirect ELISA detects ANTIBODY (e.g., to a virus or parasite): antigen on the well → patient antibody → enzyme-labeled anti-human globulin → substrate → color.
- Sandwich ELISA detects ANTIGEN: the specific antibody is attached to the solid phase.
- Immunoprecipitation, agglutination and immunodiffusion are unlabeled techniques (not labeled immunoassays).
Labels and light (added in v1.1)
| Signal | Label / substrate | Read with |
|---|---|---|
| Chemiluminescence (direct label) | Acridinium ester, isoluminol | Luminometer (PMT) |
| Enzyme-enhanced chemiluminescence | Horseradish peroxidase + luminol + H₂O₂; alkaline phosphatase + dioxetane phosphate | Luminometer |
| Electrochemiluminescence | Ruthenium label + tripropylamine at an electrode | PMT (620 nm) |
| LOCI (homogeneous) | Light → sensitizer bead → singlet oxygen → chemiluminescer/fluorophore bead | Light emission |
| Radioactive | ¹²⁵I — RIA labels antigen, IRMA labels antibody | Gamma counter |
Sources for v1.1 additions: Medical Laboratory Science Review (Harr), Ch.5, §5.1, Q24–Q25 rationales (pp.177–178); SUCCESS! in Clinical Laboratory Science, Ch.1, Q75–Q76 (p.103; rationale p.159); A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1, pp.8–9 (Table 1–2); DoH Past Paper Exam — Chemistry, p.9, Q50.
📚 Book check — added from the two review books
- The antibody is the most important part of an immunoassay: it sets the sensitivity (detects small amounts) and the specificity (unique binding). Immunoassay is chosen when the analyte concentration is very low. (Hubbard Ch.1, p.8)
- Fluorescent labels: the main drawback is the natural autofluorescence of serum (high background). (Hubbard Ch.1, p.8)
- FPIA labels the antigen; besides drugs and hormones it was used for the surfactant/albumin ratio (fetal lung maturity). (Hubbard Ch.1, p.9, Table 1–2)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.26 Immunochemical Techniques (pp.309–310); Medical Laboratory Science Review (Harr), Ch.3, §3.2, Q6–Q10 (p.84) and Ch.5, §5.9, Q25–Q28 (pp.299–300); SUCCESS! in Clinical Laboratory Science, Ch.1, Q69–Q75 (pp.102–103, rationales pp.157–159); BOC Study Guide, 6th ed., Ch.3, Q352 (p.194) and Ch.8, Q230–Q232 (p.495); DoH Past Paper Exam — Chemistry, p.8, Q23, Q26; p.9, Q38; p.11, Q79.
6 · Immunoassay interferences and technical errors
📚 Why this section: Calendar (Instrumentation II + QC) items: Immunoassay (interferences); Quality Control troubleshooting · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §II B 1 Immunoassay components — antibody sets sensitivity and specificity (p.8) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Q10 EIA calibration and rationale (pp.247, 249)
| Problem | What happens | Fix / clue |
|---|---|---|
| Hook effect (sandwich assay, extremely high antigen, e.g., hCG in molar pregnancy, huge prolactin, tumor markers) | Antigen saturates BOTH antibodies → no sandwich → falsely LOW | Dilute the sample and repeat — the diluted result is HIGHER |
| Heterophile antibodies / human anti-mouse antibodies (HAMA) | Bridge capture and label antibodies without analyte → usually falsely HIGH | Blocking reagent, different platform, serial dilution not linear |
| Poor washing after enzyme conjugate (ELISA) | Unbound enzyme stays → falsely HIGH | Repeat with correct washing |
| Color reaction allowed too long | Falsely HIGH | Stop on time |
| Absorbance above highest standard | Result outside the calibrated range | Dilute and re-assay (never extrapolate) |
| Biotin (high-dose supplements) in streptavidin–biotin assays | Falsely low (sandwich) or falsely high (competitive) | Ask about supplements; stop biotin; re-test |
| Prozone in precipitation/agglutination | Antibody excess → false negative | Dilute the serum |
📚 Book check — added from the two review books
- Calibration error: if one standard reads too LOW in absorbance, the slope falls → controls and patients read falsely HIGH. A standard reading too HIGH → results falsely LOW. Both controls off in the same direction → check the standards. (6.0 p.247, Q10; p.249)
Sources: Medical Laboratory Science Review (Harr), Ch.3, §3.2, Q3, Q8–Q10, Q13 (pp.83–85) and Ch.5, §5.11, Q47 (p.323); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.26 (pp.309–310); DoH Past Paper Exam — Chemistry, p.12 (second set), Q10.
7 · Flow cytometry
📚 Why this section: Calendar (Instrumentation II + QC) items: Flow cytometry concepts · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.3 §II D Laser scatter counting and flow cytometry (p.91) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Flow Cytometry (pp.242–245)
Cells flow one by one in a stream (hydrodynamic focusing) through a laser beam. Each cell scatters light and, if stained with fluorescent antibodies, emits fluorescence.
| Signal | Tells you |
|---|---|
| Forward scatter (FSC) | Cell size |
| Side scatter (SSC, 90°) | Internal complexity / granularity |
| Fluorescence (fluorochrome-labeled monoclonal antibodies, CD markers) | Which antigens the cell carries (immunophenotype) |
- Light source: laser (e.g., 488-nm argon/solid-state). Detectors: photodiode (FSC) and photomultiplier tubes (SSC, fluorescence).
- Gating = selecting the cell population to analyze. Wrong gate → the desired cells are not counted.
- Uses: CD4 T-cell counts in HIV (best instrument), leukemia/lymphoma immunophenotyping, PNH (CD55/CD59), stem cell (CD34) counts, DNA ploidy, reticulocyte counts.
- Basic workflow: sample preparation → (Fc) blocking → antibody incubation → data acquisition → analysis/gating.
- Scatter plot: lymphocytes = low FSC/low SSC; granulocytes = high SSC. A dense population between lymphoid and myeloid regions → abnormal cells (often blasts).
📚 Book check — added from the two review books
- Fluorochromes on monoclonal antibodies: FITC = green (500–530 nm); PE (phycoerythrin) = yellow (550–590 nm); tandem dyes (PE-Texas red, PE-Cy5, PE-Cy7) give orange to infrared colors. Fluorescence: absorb a short wavelength, emit a longer one. (6.0 p.243)
- Isotypic control = antibody of the same isotype but no specific target → measures nonspecific (background, Fc-receptor) staining; used to set the positive cursor. (6.0 p.243)
- Lymphocyte markers: CD45 = all leukocytes · CD3 = all T cells · CD4 = helper T · CD8 = suppressor/cytotoxic T · CD19 = B cells · CD16 and/or CD56 = NK cells. Multicolor panels gate on CD45 vs side scatter. (6.0 pp.243–245)
- T-gating: CD3 is paired with CD4 because monocytes also carry CD4 (but not CD3). Only CD3⁺CD4⁺ cells are counted as helper T cells. (6.0 p.247, Q14; p.249)
- 🔥 Absolute count = % positive cells × % lymphocytes × WBC. Example: WBC 5,500/µL, lymphocytes 20%, CD4 25% → 5,500 × 0.20 × 0.25 = 275/µL. Dual platform = flow % + hematology analyzer WBC/differential; single platform = counting beads. (6.0 pp.244–245; p.247, Q15; p.249)
- Flow QC: lymphosum %T + %B + %NK ≈ 100% (±5%); %CD3 ≈ %CD4 + %CD8. CD4 normal ≈ 450–1500/µL; AIDS = CD4 < 200/µL. (6.0 pp.244–245)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16 Optical Techniques (pp.214, 217–218); Medical Laboratory Science Review (Harr), Ch.3, §3.2, Q21–Q23 (p.87); DoH Past Paper Exam — Chemistry, p.1, Q9; p.4, Q58; p.5, Q81.
8 · Quality control statistics — mean, SD, CV, accuracy, precision
📚 Why this section: Calendar (Instrumentation II + QC) items: Mean; Standard Deviation (SD); Coefficient of Variation (CV); Precision; Accuracy; Bias; Random error; Systematic error · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I B Statistical concepts (pp.2–3) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Quality Control definitions; Sources of variation or error (pp.227–228)
| Term | Meaning | Formula / note |
|---|---|---|
| Mean (x̄) | Average | Σx ÷ n |
| Median | Middle value when ranked | Not affected by outliers |
| Mode | Most frequent value | — |
| Standard deviation (SD, s) | Spread (scatter) of values around the mean → precision | SD = √[Σ(x − x̄)² ÷ (n − 1)] |
| Variance | Spread | SD² = Σ(x − x̄)² ÷ (n − 1) |
| Coefficient of variation (CV) | Relative precision (%) — compares precision of tests with different means | CV% = SD ÷ mean × 100. Lower CV = better precision |
| SD index (SDI) | How far your lab mean is from the peer mean | (lab mean − peer mean) ÷ peer SD; >2 is significant |
The Gaussian (normal) curve
Accuracy vs precision
| Term | Simple meaning | Affected by |
|---|---|---|
| Accuracy (trueness) | Close to the true value | Systematic error (bias) |
| Precision (reproducibility, repeatability) | Repeated results close to each other | Random error. Measured by SD/CV. Validate by running the same sample many times. |
| Reliability | Keeps accuracy AND precision over a long time | — |
| Analytical specificity | Measures ONLY the analyte it claims to measure | Interferences |
Types of error
| Random error (RE) | Systematic error (SE) |
|---|---|
| Unpredictable, either direction; affects precision | Predictable, one direction; affects accuracy |
| Causes: bubbles, pipetting variation, unstable voltage/temperature, clots, operator technique | Causes: wrong calibrator, reagent deterioration, wrong wavelength/temperature, new reagent lot, lamp aging |
| QC rules: 1₃s, R₄s | QC rules: 2₂s, 4₁s, 10x̄ (shifts and trends) |
Constant systematic error = same amount at all levels (regression y-intercept). Proportional error = grows with concentration (regression slope ≠ 1). Correlation coefficient (r) shows association only; r = 0 → no correlation.
📚 Book check — added from the two review books
- Range = highest value − lowest value: the simplest measure of spread. (Hubbard Ch.1, p.2)
- CV is also called the relative standard deviation. At least 20 results are needed to calculate a reliable SD. (6.0 p.227)
- Primary standard = chemically pure, weighed or measured directly. Secondary standard = value assigned by a reference method or by comparison with a known solution (cheaper). (6.0 p.228)
- Assayed control = target values given by the manufacturer. Unassayed control = the lab sets the values itself (run ≥20 times → mean ± 2 s). (6.0 p.228)
- More error examples — random: voltage fluctuation, wrong sample volume, lipemia, hemolysis, drug interference · systematic: dirty photometer, faulty ISE, deteriorated reagent. (6.0 p.228)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.2 (pp.14, 17); Medical Laboratory Science Review (Harr), Ch.5, §5.4, Q17–Q27, Q41 (pp.226–228, 232); SUCCESS! in Clinical Laboratory Science, Ch.13, Q16–Q35 (pp.1010–1013, rationales pp.1022–1025); BOC Study Guide, 6th ed., Ch.8, Q7–Q17 (pp.466–468).
9 · Levey-Jennings charts and Westgard multirules
📚 Why this section: Calendar (Instrumentation II + QC) items: Levey-Jennings charts; Westgard rules · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.11 §III B 6 Quality control program — shifts and trends (p.372) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Interpretation of QC; Westgard multi-rules (p.229)
Building a Levey-Jennings chart
- Run the new control lot for about 20 days (all shifts, all staff) → calculate mean and SD (exclude results beyond ±3 SD) → draw lines at mean, ±1, ±2, ±3 SD.
- Plot each control value (y-axis) against run/day (x-axis). Values should scatter randomly around the mean.
- Do NOT throw away values between 2 and 3 SD when you set new limits (limits would shrink).
| Pattern | Looks like | Usual cause |
|---|---|---|
| Shift | Sudden jump; ≥6 consecutive points on one side of the mean | New reagent or calibrator lot, mis-made calibrator, new lamp, temperature/wavelength change, recalibration |
| Trend | Gradual, steady rise or fall over ≥6 runs | Deterioration: reagent, calibrator, lamp, evaporation, bacterial contamination |
| Random scatter beyond limits | Points far out, both directions | Random error: bubbles, pipetting, clot, poor technique |
Westgard rules (written "NL")
| Rule | Meaning | Detects | Action |
|---|---|---|---|
| 1₂s | 1 control beyond ±2 SD | Warning only | Check other rules (not alone a rejection in multirule) |
| 1₃s | 1 control beyond ±3 SD | Random (or large systematic) | Reject |
| 2₂s | 2 consecutive controls beyond 2 SD on the same side (same level across runs OR two levels in one run) | Systematic | Reject |
| R₄s | Range between two controls in the same run > 4 SD (e.g., one +2.5 SD and one −2 SD) | Random | Reject |
| 4₁s | 4 consecutive controls beyond 1 SD on the same side | Systematic | Reject |
| 10x̄ | 10 consecutive controls on the same side of the mean | Systematic (shift) | Reject (high false-alert rate) |
Other QC plots
- Cusum (cumulative sum) — earliest detection of a shift or trend.
- Tonks–Youden plot — interlaboratory comparison of two control levels (both in one quadrant → systematic error).
- Patient-based QC: moving averages; delta check (compares a patient's current result with the previous one → detects specimen mix-up).
Which QC plot for which job? (added in v1.1)
| Plot | Best for |
|---|---|
| Levey-Jennings | Daily QC — detects all error types (multirules, shifts, trends) |
| Cusum | Earliest sign of a trend (small systematic change) |
| Tonks-Youden | Comparing labs (monthly means of 2 levels) — systematic error |
| Linear regression | Method comparison (new vs old) |
- About 1 point in 20 outside ±2 s is expected by chance (warning). Any point outside ±3 s → reject and correct.
📚 Book check — added from the two review books
- Trend (6.0 definition): one control rising or falling for 6 consecutive days — OR two control levels rising/falling together for 3 consecutive days. (6.0 p.229; p.248, Q7 rationale)
- ⚠️ Books differ on the number for a shift: Hubbard = 4 or more consecutive values; 6.0 = 6 or more. Decide by the PATTERN: sudden jump to a new level = shift; slow drift = trend. (Hubbard Ch.11, p.372; 6.0 p.229)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.6 Quality Control of the Analytical Examination Process (p.132, Levey-Jennings; p.140, Table 6.3 QC rules); Medical Laboratory Science Review (Harr), Ch.5, §5.4, Q18–Q36, Q52 (pp.226–231, 236); SUCCESS! in Clinical Laboratory Science, Ch.13, Q18–Q27, Q41–Q44 (pp.1011–1014); BOC Study Guide, 6th ed., Ch.8, Q6 (p.466).
10 · What to do when QC fails, QA and method evaluation
📚 Why this section: Calendar (Instrumentation II + QC) items: Quality Control troubleshooting; Bias · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.11 §III Quality assurance (pp.371–372) and §IV A 4–5 QC and proficiency testing under CLIA 88 (p.375) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Quality Assurance; Comparison to external QC; t-test and F-test; Reference ranges (pp.227–230)
QC material vs calibrator
- Calibrator (secondary standard) = used to SET the instrument (assign the standard curve). Control = used to CHECK the run. They are not interchangeable.
- Good control material: stable for a long time, same matrix as patient samples, concentrations at medical decision levels (e.g., gentamicin controls at ~1.5 (trough) and ~6 µg/mL (peak)), target value set by YOUR method.
- Minimum for most chemistry tests (CLIA): 2 levels every 24 hours; blood gases: more often.
- QC monitors the analytical phase. Preanalytical = specimen collection, identification, acceptability. Postanalytical = reporting, turnaround time, critical-value calls.
Quality assurance tools
- Proficiency testing (PT) / external quality assessment (EQA), e.g., CAP surveys: unknown samples from outside → checks accuracy and compares laboratories. Never share PT results with another lab before submitting.
- Preventive maintenance and function verification (temperature, calibration, electronics) on all instruments; record them.
- Six Sigma: sigma = (allowable total error − bias) ÷ CV; 6 sigma ≈ 3.4 errors per million.
- New reagent lot → run controls AND patient samples (validation) before use.
Method comparison and reference intervals
- Compare a new method with the old one on ≥40 patient samples → linear regression: y-intercept = constant error, slope = proportional error, standard error of estimate = random error. A high r alone does not prove agreement.
- Bias = mean of new method − mean of reference method. t-test compares means (bias); F-test compares variances (precision).
- Total error ≈ systematic error + 1.96 × SD.
- Reference interval: 120 healthy subjects, central 95%. To verify a manufacturer's interval: 20 subjects; if ≤2 fall outside, accept.
- Diagnostic sensitivity = TP ÷ (TP + FN). Specificity = TN ÷ (TN + FP). They are inversely related when you move the cutoff. Predictive values depend on prevalence. A ROC curve picks the best cutoff (upper-left point).
Equipment, ranges and quality indicators (added in v1.1)
- Centrifuge: speed checked with a tachometer (or strobe) at least every 3 months; record RCF. RCF = 1.118 × 10⁻⁵ × radius (cm) × rpm². Disinfect weekly.
- Preventive maintenance is performed and documented for all equipment (balances, centrifuges, refrigerators, water baths, pipettes, analyzers).
- AMR (analytical measurement range) = values measured directly, without dilution. Clinically reportable range = AMR extended by validated dilution.
- Controls are never used as calibrators (calibrate with standards, check with controls).
- Turnaround time = key quality indicator for late STAT results. Staff scheduling is management, not QA.
- Proficiency testing: samples are unknowns, tested like patient samples (CAP, AAB). Ranges come only in the graded report.
- Discrete analyzer = each sample in its own cuvette (random access). Continuous flow = samples in sequence through tubing.
- F-test: F = larger variance ÷ smaller variance; below the critical value → no significant difference in precision.
- Worked example — PV+: sensitivity 80%, specificity 84%, prevalence 4% → TP = 0.032, FP = 0.154 → PV+ = 17.2%. Rare disease → low PV+.
- Worked example — total error: ŷ = −0.3 + 0.90x at 50 → SE = −5.3; RE = 1.96 × 1.12 = 2.2 → TE = −7.5 mg/dL.
Sources for v1.1 additions: SUCCESS! in Clinical Laboratory Science, Ch.1, Q61 (p.102) and Ch.13, Q12, Q25–Q26, Q37–Q39, Q45, Q50, Q52 (pp.1010–1015; rationales pp.1022–1028); Urinalysis and Body Fluids, 5th ed. (Strasinger & Di Lorenzo), Ch.7, p.131; Medical Laboratory Science Review (Harr), Ch.5, §5.4, Q28, Q39, Q45, Q48–Q49, Q53 (pp.228–236); BOC Study Guide, 6th ed., Ch.8, Q12, Q16, Q29 (pp.467–469); A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1, p.3.
📚 Book check — added from the two review books
- QA = all policies and practices that make results reliable (specimen collection, records, maintenance, proficiency testing, QC, staff training, reporting). QC is only one part of QA — it checks that one test method is working. (Hubbard Ch.11, pp.371–372)
- A written QC program must state: how often controls are run, how many, what type, the acceptable limits, and the corrective action when controls fail. (Hubbard Ch.11, p.372)
- CLIA 88 QC rules: ≥2 control levels on each day of testing · document all calibrations · keep QC records 2 years · do not interchange reagent lot numbers unless the manufacturer allows · label all reagents, controls and standards. (Hubbard Ch.11, p.375)
- Proficiency testing (CLIA): 3 events per year, 3–5 challenges per analyte; passing score 80% (blood bank 100%); done by the staff who routinely run the test. (Hubbard Ch.11, p.375)
Sources: Medical Laboratory Science Review (Harr), Ch.5, §5.4, Q24–Q25, Q31, Q37–Q54 (pp.227–237); SUCCESS! in Clinical Laboratory Science, Ch.1, Q59–Q62 (pp.101–102) and Ch.13, Q19, Q24–Q25, Q37–Q47 (pp.1011–1014, rationales pp.1023–1027); BOC Study Guide, 6th ed., Ch.8, Q3–Q5, Q10 (pp.465–467); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.6 (pp.132, 140); DoH Past Paper Exam — Chemistry, p.8, Q20, Q36.
11 · Top exam traps
📚 Why this section: Calendar (Instrumentation II + QC) items: — (book heading: review of traps) · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §II B Immunoassay — used when analyte concentration is very low (p.8) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — QC definitions: accuracy, precision (p.227)
• Reliability = the ability of a method to keep its accuracy and precision over a long period (consistency). Sensitivity is a different idea (detecting disease, or detecting small amounts of analyte).
• Hormones are measured by immunoassay (today mainly chemiluminescent/enzyme immunoassay; historically radioimmunoassay). They are present at pg–ng amounts; radial immunodiffusion measures only mg/dL proteins and is far too insensitive.
• Centrifuge speed is checked with a tachometer or strobe light at least every 3 months, and the RCF for each setting is recorded. Centrifuges are disinfected weekly.
Top traps for this topic
- Indirect ISE + lipemia/paraprotein → falsely LOW Na⁺. Direct ISE is NOT affected.
- Valinomycin = K⁺. Glass = Na⁺ and pH. Clark PO₂ = amperometric. Nernst slope ≈ 60 mV per decade (univalent).
- Freezing point drops 1.86 °C per Osm/kg. Vapor pressure osmometer misses alcohols.
- pH 8.6: proteins negative → anode. Albumin fastest, γ slowest; γ may move toward the cathode (electroendosmosis).
- TLC Rf = spot distance ÷ solvent-front distance. GC = volatile. Reverse-phase HPLC: mobile phase more polar. GC-MS = confirmation.
- Competitive immunoassay: signal ∝ 1/concentration. Sandwich: signal ∝ concentration; hook effect → dilute.
- Chemiluminescence = most common automated immunoassay. EMIT = homogeneous; enzyme attached to drug.
- Flow cytometry: laser; FSC = size; SSC = granularity; CD4 counts.
- CV = SD/mean × 100 — use it to compare precision at different means.
- 1₃s and R₄s = random error. 2₂s, 4₁s, 10x̄ = systematic error. Shift = sudden; trend = gradual deterioration.
- Precision = reproducibility (random error); accuracy = trueness (systematic error).
- QC monitors the analytical phase. PT/EQA checks accuracy between laboratories.
- Centrifuge speed: tachometer, at least every 3 months. Controls ≠ calibrators. AMR = no dilution.
- Base peak = most abundant ion; ICP-MS = metals; GC-MS electron source, LC-MS electrospray.
Sources: Urinalysis and Body Fluids, 5th ed. (Strasinger & Di Lorenzo), Ch.7, p.131 (reliability; centrifuge calibration every 3 months, weekly disinfection); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.26, pp.308–309 (radial immunodiffusion mostly replaced; sensitive labeled immunoassays). Supporting: SUCCESS! in Clinical Laboratory Science, Ch.13 (pp.1010, 1013); Medical Laboratory Science Review (Harr), Ch.3, §3.2 (p.84). Other traps: see Sections 1–10.
⚡ Last-Minute Review — Instrumentation & Analytical Principles II + Quality Control
1. ISE & electrochemistry
- Potentiometry = voltage, no current. ISE measures ion activity; Nernst slope ≈ 60 mV per 10-fold change (univalent), ≈30 mV (divalent).
- Na⁺ = glass · pH = glass · K⁺ = valinomycin · NH₄⁺ = nonactin · Ca²⁺ = ionophore. Reference = Ag/AgCl or calomel. Standard = hydrogen electrode (0 V).
- Indirect ISE (diluted) + lipemia/high protein → falsely low Na⁺ (pseudohyponatremia). Direct ISE (undiluted) = true.
- PCO₂ = pH electrode in bicarbonate (potentiometric). PO₂ = Clark, amperometric. Analyzer measures pH, PCO₂, PO₂; calculates HCO₃⁻.
- Coulometry = chloride (sweat); Ag⁺ generated; time ∝ Cl⁻. Protein on membrane = slow response; blocked junction = noise.
2. Osmometry
- 1 Osm/kg: freezing point ↓1.86 °C, boiling ↑0.52 °C, vapor pressure ↓.
- Freezing point osmometer: thermistor + Wheatstone bridge; detects alcohols. Vapor pressure: thermocouple; misses ethanol/methanol.
- Calc osm = 2Na + Glu/18 + BUN/2.8. Gap = measured − calc (<10).
3. Electrophoresis
- pH 8.6 barbital: proteins negative → anode. pH below pI → positive → cathode. At pI → no movement.
- Anode → albumin, α1, α2, β, γ → cathode. Transferrin & C3 = β. Haptoglobin = α2.
- Electroendosmosis pushes γ toward cathode. Capillary electrophoresis: no support, EOF, high voltage.
- Stains: Ponceau S, amido black, Coomassie (protein); Oil red O (lipoprotein); ethidium bromide (DNA). Agarose = large DNA.
- Hollow band = too concentrated → dilute. Distorted = overload. No migration = lost buffer contact. Isoelectric focusing = ampholytes.
4. Chromatography & MS
- Principle: differential solubility/partition between stationary and mobile phases.
- TLC: non-volatile; Rf = spot ÷ solvent front. GC: volatile, elution by boiling point, FID detector. HPLC: pumped liquid, UV detector; reverse phase = polar mobile, nonpolar C18; isocratic vs gradient.
- Ion-exchange = charge. Gel filtration = size. Internal standard corrects extraction/injection.
- MS = mass-to-charge (m/z); vacuum. GC-MS = confirmation of drugs. LC-MS = electrospray. Tandem MS = newborn screening.
5. Immunoassays
- Chemiluminescence = most common automated & most sensitive (troponin). ELISA beats RIA: cost, no radiation.
- Competitive: signal ∝ 1/[analyte] (small molecules). Sandwich: signal ∝ [analyte]; specific Ab on solid phase; hook effect → dilute.
- Homogeneous (no separation): EMIT (enzyme on drug, G6PD, 340 nm), FPIA, CEDIA. Heterogeneous: ELISA, RIA.
- Indirect ELISA = antibody detection. Immunoprecipitation is NOT a labeled immunoassay.
- Poor washing/long color time → falsely high. Above top standard → dilute. Heterophile/HAMA → usually falsely high.
6. Flow cytometry
- Laser. FSC = size. SSC = granularity. Fluorescent CD antibodies = phenotype. Best for CD4 counts. Gating selects cells.
7. QC statistics
- CV% = SD/mean × 100 (compare precision at different means). SD = √[Σ(x−x̄)²/(n−1)]; variance = SD².
- ±1 SD 68% · ±2 SD 95% · ±3 SD 99.7%. Outside ±2 SD by chance 1/20; above +2 SD 1/40; two in a row 1/1600.
- Precision = reproducibility (random error). Accuracy = trueness (systematic error, bias). Reliability = accuracy + precision over time.
- Regression: intercept = constant error; slope = proportional error. r = association only.
8. Levey-Jennings & Westgard
- Set limits from ~20 days of data. Shift = sudden, ≥6 on one side (new lot, bad calibrator). Trend = gradual (deterioration).
- 1₂s warning · 1₃s random · 2₂s systematic · R₄s random (within run) · 4₁s systematic · 10x̄ systematic.
- Cusum = earliest trend detection. Youden = interlab. Delta check = specimen mix-up.
9. QC action & QA
- Out of control → don't report → follow procedure → find/fix cause → rerun controls → rerun patients → document.
- Calibrator ≠ control. 2 levels / 24 h minimum. QC = analytical phase. PT/EQA = accuracy vs other labs.
- Reference interval: 120 subjects (central 95%); verify with 20 (≤2 outside). Sensitivity TP/(TP+FN); specificity TN/(TN+FP); inversely related.
v1.1 · Extra high-yield points
- Potentiometry = volts · amperometry = current (PO₂, glucose, HPLC-ECD glassy carbon) · coulometry = charge (Cl⁻) · ASV = current (lead).
- Ag/AgCl reference potential set by KCl (Cl⁻). Clark: O₂ reduced at Pt cathode.
- Chemiluminescence: HRP + luminol; ALP + dioxetane; acridinium direct; ruthenium = ECL; singlet oxygen = LOCI. RIA labels antigen, IRMA labels antibody.
- HPLC detector = UV. Base peak = most abundant. ICP-MS = metals. m/z 16-17-18 = oxygen leak.
- Index of precision = CV. Different means → compare CV.
- L-J = all errors; Cusum = earliest trend; Youden = between labs; regression = method comparison.
- PV− = TN/(TN+FN). PV+ falls when prevalence is low. Six Sigma = error rate (3.4 per million).
- Centrifuge: tachometer, every 3 months. AMR = no dilution. Controls ≠ calibrators. PT samples = unknowns.
📚 Book check (Hubbard + 6.0)
- Flow absolute count = % positive × % lymphs × WBC (5,500 × 0.20 × 0.25 = 275/µL). AIDS = CD4 < 200/µL. T-gating: CD3 + CD4 (monocytes are CD4⁺ CD3⁻). (6.0 pp.244–245, 249)
- FITC = green · PE = yellow · isotypic control = background staining. CD45 all WBC · CD3 T · CD19 B · CD16/56 NK. (6.0 p.243)
- Electrophoresis errors: crescent bands = bent applicator/overload · split bands = applicator pressure · no bands = voltage too high · prealbumin fastest. (6.0 pp.237, 248)
- HPLC: noisy baseline = bubbles/particles · drifting baseline = overloaded column or dirty solvent. GC: retention time identifies; best for blood alcohol. Mixed ion-exchange resins = deionized water. (Hubbard Ch.1, p.9; 6.0 p.236)
- Standard reading too low → slope ↓ → results falsely high (and the reverse). (6.0 p.249)
- Primary standard = pure, weighed · secondary = value by comparison. Assayed control = manufacturer values · unassayed = lab sets them. CV = relative SD. (6.0 pp.227–228)
- CLIA: QC records 2 years · ≥2 levels per day · PT 3×/year, pass 80% (blood bank 100%). (Hubbard Ch.11, p.375)
- Freezing point ≫ vapor pressure osmolality → volatile alcohol. Osmometer errors: bath temperature, particles stop supercooling. (6.0 pp.235–236)