Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 very high-yield · 📌 must memorize · ⚠️ exam trap · 🧠 memory trick · 🔬 lab connection · 💡 understand this.
1 · The three phases of testing — and where most errors happen
📚 Why this section: Calendar (Sample collection & errors) items: Preanalytical errors; Analytical errors; Postanalytical errors · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I F Pre-analytical variables and §I G Post-analytical interpretation (p.7); Ch.11 §III B Quality assurance plan (p.372) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Quality Assurance (p.227)
Every laboratory result passes through three phases. An error in ANY phase gives the doctor a wrong answer — even if the analyzer is perfect.
| Phase | Typical errors | What protects it |
|---|---|---|
| Pre-analytical | Wrong patient / unlabeled tube · not fasting · tourniquet too long · fist pumping · wrong tube or wrong order of draw · drawn above an IV · hemolysis · clotted or underfilled tube · delayed separation · wrong temperature or light exposure | Two patient identifiers, labeling at the bedside, written collection procedures, specimen rejection criteria, training of phlebotomists |
| Analytical | Failed calibration · QC out of control · reagent problems · interference (hemolysis, lipemia, icterus) · above linearity (hook effect, need dilution) | Internal QC (Levey-Jennings, Westgard), proficiency testing / EQA, maintenance |
| Post-analytical | Transcription error · result sent to the wrong chart · critical value not called · wrong units or reference interval · long turnaround time (TAT) | LIS interfacing, auto-verification, delta checks, critical-value policy, TAT monitoring |
📚 Book check — added from the two review books
- Inappropriate specimen type and mislabeled specimens are the MOST COMMON pre-analytical problems the laboratory sees. (Hubbard Ch.1, p.7)
- Quality Assurance (QA) covers ALL three phases: pre-analytical (specimen collection, transport, preservatives), analytical (test processing — QC data are used here) and post-analytical (how results are reported and acted upon). QC is only the analytical part of QA. (6.0 p.227; Hubbard Ch.11, p.372)
- The written procedure manual must give, for each test: criteria for specimen collection, processing and rejection; interfering substances and common sources of error; and result reporting, including panic (critical) values. (Hubbard Ch.11, pp.375–376)
📚 How the sections were chosen
Two sources: the fixed calendar list for this day and the chapter headings of A Concise Review of Clinical Laboratory Science (Hubbard). Check book: Clinical Laboratory Science Review, 5th ed. (6.0).
| # | Study section | Calendar items it covers | Book heading (Hubbard, page) |
|---|---|---|---|
| 1 | The 3 phases & where errors happen | Preanalytical / analytical / postanalytical errors | Ch.1 §I F–G (p.7); Ch.11 §III B (p.372) |
| 2 | Patient preparation & biological variables | Patient preparation; Fasting; Timing | Ch.11 §III B (p.372) |
| 3 | Patient identification & labeling | Preanalytical errors (identification) | Ch.1 §I F (p.7) |
| 4 | Venipuncture technique | Tourniquet effects; Contamination | Ch.1 §I F 1 (p.7) |
| 5 | Tubes, additives & order of draw | Order of draw; Tube additives; Serum vs plasma; Wrong anticoagulant | Ch.1 §I F 1a (p.7) |
| 6 | Capillary, arterial & special specimens | Tube additives; Storage; Stability | Ch.1 §IV C 5a (p.19); §IV I 3 (p.27) |
| 7 | Hemolysis, lipemia & icterus (HIL) | Hemolysis; Lipemia; Icterus; Interferences | Ch.1 §IV A 9 (p.16); §V (pp.32–34) |
| 8 | Processing, storage & stability | Sample storage; Stability; Delayed separation | Ch.1 §I F 2 (p.7) |
| 9 | Hematology & coagulation specimen problems | Improper fill; Wrong anticoagulant | Ch.3 §I A (p.87) |
| 10 | Recognizing & handling specimen errors | Analytical / postanalytical errors; Troubleshooting | Ch.11 §IV A (pp.375–376) |
| 11 | Top exam traps | Common examination troubleshooting | Ch.1 §I F (p.7) |
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.5 Preanalytical Variation and Pre-Examination Processes (pp.80–82); SUCCESS! in Clinical Laboratory Science, Ch.13, Q25, Q38 (pp.1011, 1013) and Ch.11, Q14 (p.937); Medical Laboratory Science Review (Harr), Ch.5, p.236, Q52.
2 · Patient preparation and biological (physiological) variables
📚 Why this section: Calendar (Sample collection & errors) items: Patient preparation; Fasting; Timing · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.11 §III B Quality assurance plan — patient preparation follows written guidelines (p.372) · Check book: Clinical Laboratory Science Review, 5th ed., Chemistry — Lipid specimen collection and handling (p.105); Cortisol diurnal variation (p.130)
Some changes are REAL changes inside the patient's body (not lab mistakes). We control them by standardizing how the patient is prepared, so the result can be compared with the reference interval.
Fasting
| Test | Preparation | Why |
|---|---|---|
| Fasting plasma glucose | No food or drink except water for at least 8 h | Food raises glucose |
| Triglycerides / calculated LDL | Traditionally 12 h fast (10–14 h) | Chylomicrons after a meal raise TG for hours; Friedewald LDL needs fasting TG |
| Total cholesterol & HDL | Fasting not essential | Least affected by a recent meal (Harr, p.259, Q57) |
| Oral Glucose Tolerance Test (OGTT) | Normal diet with ≥150 g carbohydrate/day for 3 days, then 8–14 h overnight fast; no smoking, coffee, or exercise during the test; seated | Carbohydrate restriction before the test gives a FALSELY ABNORMAL (diabetic-looking) curve |
| Ammonia | Fasting; no smoking for 8 h | Tobacco smoke can double ammonia |
Time of day (diurnal / circadian variation)
Posture
Total calcium rises about 5–10% from supine to upright; ionized calcium does not change. Aldosterone, renin and plasma metanephrines need the patient supine for at least 30 minutes before collection.
Other patient factors
- Exercise / muscle activity → ↑ CK, AST, LD, myoglobin, K⁺, lactate. Even fist clenching during venipuncture raises K⁺.
- Stress, anxiety → ↑ cortisol, catecholamines, glucose, WBC. Catecholamines: patient resting supine 30 min; medications stopped if possible.
- Smoking → ↑ carboxyhemoglobin, ↑ cortisol, ↑ glucose, ↑ ammonia, ↑ Hb/Hct.
- Alcohol → ↑ GGT, ↑ triglycerides, ↑ uric acid, ↑ MCV.
- Cooked meat → ↑ creatinine up to about 20%.
- Drugs → in vivo effects (e.g., diuretics lower K⁺) and in vitro interference (e.g., ascorbic acid lowers peroxidase-based glucose).
- Age, sex, pregnancy → unmodifiable; they are handled by correct reference intervals (e.g., ALP high in growing children and in pregnancy — placental ALP).
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4 (p.62) and Ch.5 (pp.81–88); Medical Laboratory Science Review (Harr), Ch.5, p.206 Q7, p.244 Q31, p.259 Q57, p.290 Q36, p.323 Q46; DoH Past Paper Exam — Chemistry, p.10, Q72.
3 · Patient identification and specimen labeling
📚 Why this section: Calendar (Sample collection & errors) items: Preanalytical errors (patient identification, labeling) · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I F Pre-analytical variables — two identifiers, mislabeled specimens (p.7) · Check book: Clinical Laboratory Science Review, 5th ed., Immunohematology — Pretransfusion testing, sample requirements (p.20)
A perfect result on the wrong patient is the most dangerous laboratory error — especially in blood bank (wrong blood in tube → ABO-incompatible transfusion).
| Situation | Correct action |
|---|---|
| Unlabeled or mislabeled specimen | Reject and request a new specimen. The lab never labels a tube for someone else from "information provided". |
| Unlabeled blood bank (pre-transfusion) sample brought by a doctor | Discard and request a new, properly labeled sample (SUCCESS!, p.565, Q142) |
| Irreplaceable specimen (CSF, tissue) mislabeled | Follow policy: the collector comes to the lab, identifies and signs; result is reported with a comment |
| Patient refuses venipuncture | Do NOT draw. Respect the refusal, notify the nurse/physician/supervisor and document (Harr, p.535, Q63) |
| Patient asleep / unconscious | Wake gently; identify by wristband; never draw without identification |
📚 Book check — added from the two review books
- Blood bank sample: the patient is identified with 2 unique identifiers, and the lab must be able to identify the phlebotomist and the date of collection. For the 3-day rule, the day of draw = day 0. Keep the patient sample and a donor RBC segment for 7 days after transfusion. (6.0 p.20)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4 (pp.61, 67, 76); SUCCESS! in Clinical Laboratory Science, Ch.5, Q142–Q144 (p.565); Medical Laboratory Science Review (Harr), Ch.9, p.532 Q46 and p.535 Q63.
4 · Venipuncture technique — step by step
📚 Why this section: Calendar (Sample collection & errors) items: Tourniquet effects; Contamination · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I F 1 Specimen collection (p.7) · Check book: Clinical Laboratory Science Review, 5th ed., Chemistry — Potassium specimen collection and handling (p.114)
Tourniquet and fist pumping
Sites to avoid
- Arm with an IV infusion — use the other arm. If both arms have IVs: draw BELOW the IV after the infusion has been stopped for at least 2 minutes (Tietz: ≥3 min, with clinician consent); discard the first 5 mL; document "drawn below IV".
- Same side as a mastectomy (lymphostasis) · arteriovenous fistula/graft · hematoma · scarred, burned or edematous areas.
- Drawing from a line (catheter): flush, then withdraw a discard volume (≈ 2× dead space, adults about 5–10 mL) before the sample; coagulation tests from a heparinized line may show heparin contamination.
Complications and fixes
| Problem | Cause / fix |
|---|---|
| Hematoma | Needle through both walls, partial insertion, or not enough pressure after removal |
| Hemolysis | Small needle (>23 G), wet alcohol, forcing blood through a syringe needle, vigorous shaking, traumatic draw |
| No blood flow | Bevel against vein wall → rotate/adjust needle slightly; lost vacuum → new tube |
| Fainting (syncope) | Stop, remove tourniquet and needle, lower the head, call for help |
| Accidental arterial puncture | Bright red pulsing blood → remove, firm pressure ≥5 min |
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4 (pp.62–67, 71) and Ch.5 (pp.87–88, MCQ 8 p.128.e16); Clinical Hematology: Theory and Procedures, 5th ed. (Turgeon), Ch.2 Principles of Blood Collection (review questions p.47); Medical Laboratory Science Review (Harr), Ch.5, p.244 Q31; DoH Past Paper Exam — Chemistry, p.7, Q2.
5 · Tubes, additives, serum vs plasma, and order of draw
📚 Why this section: Calendar (Sample collection & errors) items: Order of draw concepts; Tube additives; Serum vs plasma; Wrong anticoagulant · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I F 1a Evacuated blood tubes — order of draw (p.7) · Check book: Clinical Laboratory Science Review, 5th ed., Chemistry — Sample collection and handling (p.101); Coagulation — Specimen collection/handling (p.74); Hematology — Anticoagulants (p.55)
Serum = liquid left AFTER the blood clots (no fibrinogen, no clotting factors). Plasma = liquid from blood that did NOT clot because an anticoagulant was added (contains fibrinogen).
| Stopper | Additive | How it works | Main use | Do NOT use for |
|---|---|---|---|---|
| Yellow / culture bottle | SPS (sodium polyanethol sulfonate) or ACD | SPS: anticoagulant + inhibits complement and phagocytosis | Blood culture (SPS); ACD for HLA/DNA | Chemistry |
| Light blue | 3.2% sodium citrate, ratio 9 blood : 1 citrate | Binds calcium (reversible — calcium added back in the test) | PT, aPTT, fibrinogen, D-dimer, factor assays | Chemistry (dilutes, binds Ca) |
| Black | Sodium citrate 1:4 | Binds calcium | Westergren ESR | — |
| Red / gold (SST) / tiger | None or clot activator ± gel | Clots → serum; gel separates serum from cells | Chemistry, serology, immunology, blood bank (red) | Tests needing whole blood/plasma |
| Green | Lithium or sodium heparin ± gel (PST) | Activates antithrombin → inhibits thrombin (and Xa) | STAT chemistry plasma, ammonia, blood gases (syringe), ionized Ca, K⁺ confirmation | Coagulation; CBC/smears (blue background, clumping); lithium levels (Li-heparin); PCR (inhibits) |
| Lavender / purple (pink = blood bank) | K₂EDTA (or K₃EDTA) | Chelates calcium irreversibly | CBC, blood smear, HbA1c, ESR (ICSH method), blood bank (pink), cyclosporine/tacrolimus, molecular | Calcium, K⁺, ALP, Mg, coagulation, iron |
| Gray | Sodium fluoride + potassium oxalate | Fluoride inhibits enolase (glycolysis); oxalate precipitates Ca | Glucose, lactate, blood alcohol | BUN by urease (fluoride inhibits urease), many enzymes, electrolytes (Na/K salts) |
| Royal (navy) blue | None or EDTA; trace-element free | Certified free of metals | Trace metals (zinc, copper, selenium), toxicology | — |
| Tan | K₂EDTA, lead free | Certified <0.25 µg/dL lead | Blood lead (whole blood) | — |
Order of draw (venipuncture, CLSI)
Why the order matters — additive carry-over
| Wrong order / contamination | Result |
|---|---|
| K-EDTA into a serum/heparin tube | K⁺ falsely HIGH; Ca²⁺, Mg²⁺ falsely LOW; ALP falsely LOW (Mg/Zn cofactors chelated). Classic pattern: very high K⁺ + very low Ca²⁺ in a well patient. |
| Clot activator / serum tube BEFORE citrate | Activates coagulation → shortened PT/aPTT |
| Heparin into citrate tube | Prolonged aPTT, thrombin time |
| Fluoride/oxalate into other tubes | ↑ Na⁺/K⁺, ↓ Ca²⁺, inhibited enzymes (ALP, amylase, urease) |
| Citrate into chemistry tube | ↑ Na⁺, ↓ Ca²⁺, dilution |
📚 Book check — added from the two review books
- Serum (red top) is used for most chemistry tests. When plasma is needed, heparin is better than oxalate, EDTA or citrate (these add Na⁺/K⁺ or bind calcium). (6.0 p.101)
- Plasma sent by mistake for serum protein electrophoresis → an extra fibrinogen band between β and γ (near the application point) that can look like a monoclonal band → recollect serum and repeat. (6.0 p.107; Hubbard Ch.1, p.16)
- With a straight needle and evacuated tubes, the light-blue (citrate) tube may be drawn FIRST — old fears of tissue-thromboplastin contamination are unfounded. (A discard tube is needed only with a butterfly set — see Section 4.) (6.0 p.74)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4, Table 4.2 (p.65), pp.69–71 and MCQ 3, 5 (p.79.e2); Clinical Hematology: Theory and Procedures, 5th ed. (Turgeon), Ch.2 (p.47); SUCCESS! in Clinical Laboratory Science, Ch.13, Q4, Q14, Q15 (pp.1009–1010) and Ch.1, Q154 (p.111); Medical Laboratory Science Review (Harr), Ch.2, p.44 Q3–Q4; Ch.5, p.303 Q38, p.313 Q16; DoH Past Paper Exam — Chemistry, p.12 Q94, p.23 Q97; DoH Past Paper Exam — Hematology/Blood Bank, p.5 Q64, p.14 Q88.
6 · Skin puncture, arterial blood gases and special specimens
📚 Why this section: Calendar (Sample collection & errors) items: Tube additives; Sample storage; Sample stability (special specimens) · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §IV C 5a Specimen requirements for blood gas analysis (p.19) and §IV I 3 Porphyrin specimens (p.27) · Check book: Clinical Laboratory Science Review, 5th ed., Chemistry — Glucose specimens (p.101); ACP specimen handling (p.111); Ammonia (pp.127, 135)
Skin (capillary) puncture
- Infants (<1 year): heel — medial or lateral plantar surface. Never the posterior curve, the arch, or an old puncture site. Depth <2.0 mm (avoid bone → osteomyelitis).
- Older children and adults: finger — palmar surface of the distal segment of the middle or ring finger, across the fingerprint.
- Warm the site (≈42 °C, 3–5 min) — especially for capillary blood gases ("arterialized" blood).
- Alcohol must dry · wipe away the FIRST drop (tissue fluid, alcohol) · do NOT squeeze hard ("milking" adds tissue fluid and hemolysis) · avoid air bubbles.
Arterial blood gases (ABG)
| Error | pH | PCO₂ | PO₂ | Why |
|---|---|---|---|---|
| Air bubble / exposure to air | ↑ | ↓ | ↑ (toward ≈150 mm Hg) | Room air has almost no CO₂ and more O₂ |
| Delay at room temperature | ↓ | ↑ | ↓ | Living cells (WBCs, platelets, reticulocytes) consume O₂, make CO₂ and lactic acid |
| Excess liquid heparin | ↓ (slightly) | ↓ (dilution) | shifts toward ≈150 (a low PO₂ rises) | Dilution with heparin that is in equilibrium with room air |
Before radial artery puncture, perform the modified Allen test to check ulnar collateral circulation.
Specimens that need special handling
| Analyte | Special handling | What goes wrong |
|---|---|---|
| Ammonia | Heparin/EDTA plasma, full tube, on ice, separate and test within 30 min; no smoking; minimal stasis | Delay, stasis → ammonia ↑ (deamination of amino acids) |
| Lactate, pyruvate | Fluoride/oxalate (gray) on ice; no fist clenching; tourniquet released early | Glycolysis → lactate ↑ |
| Ionized calcium | Anaerobic (no air), heparinized whole blood (balanced/low heparin), on ice, test within 2 h; serum only if kept capped | Loss of CO₂ → pH ↑ → iCa ↓. EDTA, citrate, oxalate chelate Ca. |
| Bilirubin, carotene, porphyrins, vitamin A, B₆, methotrexate | Protect from LIGHT (amber tube or foil) | Photodegradation → falsely LOW bilirubin (up to ~50% per hour in direct light) |
| ACTH, PTH, gastrin, catecholamines | EDTA (lavender) on ice, rapid separation (refrigerated centrifuge) | Protease degradation → falsely low |
| Renin (exception!) | EDTA at ROOM temperature, centrifuge at room temperature, then freeze plasma | Chilling causes cryoactivation of prorenin → falsely HIGH renin |
| Cold agglutinins, cryoglobulins | Keep at 37 °C until serum is separated | Antibody binds to cells in the cold → falsely low |
| Lead | Whole blood (tan/royal blue EDTA); venous preferred | Lead is in RBCs; finger skin can contaminate |
| Carbon monoxide (COHb) | Whole blood (heparin/EDTA) | CO is bound to hemoglobin in RBCs |
| Trace metals | Royal (navy) blue tube | Ordinary stoppers contain zinc etc. |
| Blood culture | Disinfect skin (chlorhexidine/iodine), clean bottle tops, drawn FIRST; 2 sets | Skin contaminants (CoNS) |
📚 Book check — added from the two review books
- Acid phosphatase (ACP) loses activity at room temperature → separate serum from cells quickly and stabilize it (add disodium citrate monohydrate, or lower the pH to about 5.4 with acetic acid). Hemolysis gives falsely ↑ ACP (RBCs contain ACP); fluoride, oxalate and heparin give falsely ↓ ACP. (6.0 p.111; Hubbard Ch.1, p.32)
- Ammonia: a clotted (serum) sample is the wrong specimen — use heparin plasma on ice, centrifuge and test at once. Delays longer than 15 minutes can raise ammonia even on ice (amino acids/nitrogen compounds break down to ammonia). (6.0 pp.127, 135)
- Arterial and capillary glucose are about 2–3 mg/dL higher than venous glucose. (6.0 p.101)
- Urine porphyrins: protect from light — levels can fall by up to 50% in 24 h in light. Best specimen: fresh early-morning urine (10–20 mL) with no preservative. (Hubbard Ch.1, p.27)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4 (pp.67–71, 76–77) and MCQ 7, 9 (p.79.e2); Medical Laboratory Science Review (Harr), Ch.5, p.193 Q17, p.200 Q53, p.244 Q31, p.303 Q38, p.316 Q28; SUCCESS! in Clinical Laboratory Science, Ch.1, Q333–Q335 (p.133), Q403, Q405 (p.141); Clinical Hematology: Theory and Procedures, 5th ed. (Turgeon), Ch.2 (p.47); DoH Past Paper Exam — Chemistry, p.10 Q69, p.15 Q37.
7 · Hemolysis, lipemia and icterus — the "HIL" interferences
📚 Why this section: Calendar (Sample collection & errors) items: Hemolysis; Lipemia; Icterus; Interferences · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §IV A 9 Protein analysis — artifacts (p.16) and §V Enzymology — specimen errors (pp.32–34) · Check book: Clinical Laboratory Science Review, 5th ed., Chemistry — Hemolysis 'KLAMP' (p.101); Potassium (p.114); Bilirubin specimen handling (p.123)
Hemolysis (pink/red serum)
Hemolysis = red cells break and release their contents. In vitro hemolysis (during collection/handling) is far more common than true in vivo hemolysis.
| Hemolysis effect | Analytes |
|---|---|
| Greatly ↑ (do NOT report) | K⁺ (≈25× higher in RBCs), LD (≈150×), AST (≈40×), acid phosphatase, phosphate, Mg²⁺, iron, free Hb |
| Slight change | ALT, CK (adenylate kinase interference only when marked), total protein (↑) |
| Little / no effect | Na⁺, calcium (least affected electrolyte), glucose, urea, creatinine (method-dependent) |
| Falsely ↓ | Bilirubin (diazo inhibition), insulin (degraded by RBC proteases), troponin T in some assays |
Lipemia (milky serum)
- Pseudohyponatremia: indirect ISE (diluted) reads low; direct ISE (undiluted, blood gas analyzer) gives the true Na⁺ → report the direct ISE result.
- Remove lipids by ultracentrifugation (preferred) or high-speed centrifugation/lipid-clearing agents — but measure triglycerides and lipid-soluble drugs on the NATIVE sample BEFORE delipidation.
- CBC: correct the Hb with a plasma blank or saline replacement: Corrected Hb = Whole-blood Hb − [Plasma Hb × (1 − Hct/100)].
- Optical coagulation analyzers cannot detect the clot in grossly lipemic plasma → use an electromechanical (mechanical) clot detection method.
Icterus (dark yellow serum)
High bilirubin absorbs strongly at 400–540 nm. It causes spectral interference and chemical interference with peroxidase (Trinder) reactions (bilirubin reacts with H₂O₂) → falsely LOW cholesterol, triglycerides, uric acid, creatinine (enzymatic) in some methods.
📚 Book check — added from the two review books
- 6.0 memory aid: hemolysis — "help (c)KLAMP the leak": K⁺, LD, Aldolase and ACP, Mg²⁺, PO₄ all falsely ↑. Free hemoglobin also raises iron; bilirubin may be ↓ or ↑ depending on the hemoglobin and bilirubin levels (and the method). (6.0 pp.101, 135)
- Hemolysis artifacts on electrophoresis: SPE → an extra heavy β band (free Hb); LD isoenzymes → LD1 > LD2 "flipped" pattern that mimics myocardial infarction. (6.0 p.101; Hubbard Ch.1, pp.16, 34)
- Pseudohyperkalemia numbers: WBC ≥ 500,000/µL or platelets ≥ 700,000/µL; also excessive heel or finger squeezing in skin punctures. Normally plasma (and whole-blood) K⁺ is 0.1–0.7 mmol/L LOWER than serum K⁺, because platelets release K⁺ during clotting. (6.0 p.114)
- Lipemia gives falsely ↑ bilirubin. A non-fasting sample left in the refrigerator shows a creamy chylomicron layer on top of turbid serum (abnormal in a fasting sample). (6.0 pp.105, 123)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4 (pp.71–72) and Ch.5 (pp.89–100; MCQ 2–6, p.128.e16); Medical Laboratory Science Review (Harr), Ch.1 p.37 Q12; Ch.5 p.176 Q18, p.201 Q56, p.212 Q27, p.220 Q64, p.265 Q17, p.310 Q3, p.316 Q29; SUCCESS! in Clinical Laboratory Science, Ch.2, Q242, Q259–Q260 (pp.313–315); BOC Study Guide, 6th ed., Ch.4, Q278–Q279, Q296 (pp.264–267); DoH Past Paper Exam — Chemistry, p.1 Q10, p.11 Q76, p.12 Q98; DoH Past Paper Exam — Hematology/Blood Bank, p.9 Q27.
8 · Specimen processing, transport, storage and stability
📚 Why this section: Calendar (Sample collection & errors) items: Sample storage; Sample stability; Delayed separation · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I F 2 Specimen transport and storage (p.7) · Check book: Clinical Laboratory Science Review, 5th ed., Chemistry — Glucose specimen collection and handling (p.101); Potassium (p.114)
Delayed separation of serum/plasma from cells
Refrigerating WHOLE blood (unseparated)
Coagulation specimen stability
| Test | Stability (capped, uncentrifuged or centrifuged) | What goes wrong |
|---|---|---|
| PT | Up to 24 h at room temperature | Refrigeration → cold activation of factor VII → PT falsely SHORT |
| aPTT (no heparin) | 4 h (room temperature or 2–4 °C) | Labile factors V and VIII decay → aPTT falsely LONG → request a new sample |
| aPTT/anti-Xa on heparin | Centrifuge within 1 h | Platelets release PF4 → neutralizes heparin → aPTT falsely short |
| Platelet-poor plasma | Platelets <10 × 10⁹/L (double centrifugation for lupus anticoagulant) | Residual platelets neutralize phospholipid-dependent tests |
Hematology specimen stability (EDTA)
- CBC best within 6 h; most parameters stable 24 h at room temperature.
- Standing at room temperature 6–24 h → RBCs swell: MCV ↑, Hct ↑, MCHC ↓; ESR ↓; WBC morphology changes (vacuoles, crenated cells). Make smears as soon as possible — ideally within 1 h (Turgeon) and no later than about 5 h (SUCCESS!), before neutrophils vacuolate and cells crenate.
- Platelet count stable about 24 h; ESR should be set up within 4 h (24 h at 4 °C).
Other stability facts
- CSF: deliver and test immediately (glucose falls; cells lyse; bacteria die). Do NOT refrigerate the microbiology tube.
- Light-sensitive: bilirubin, porphyrins, vitamin A, β-carotene, B₆ → amber tube/foil.
- Never freeze whole blood (hemolysis). Avoid repeated freeze–thaw; freezing destroys LD-5 (LD-4/5 cold labile).
- Serum for CK: stable ≈4 h at room temperature and a week at 4 °C; CK is NOT in RBCs, so slight hemolysis does not interfere. Calcium chelators (citrate, EDTA) remove Mg²⁺ needed by CK → do not use.
- Add-on tests: allowed only if the original specimen is the correct type, has enough volume, was stored correctly and is within the analyte's stability time.
- Transport: upright, capped (prevents evaporation, CO₂ loss, aerosols); pneumatic tubes with sudden stops can cause hemolysis.
- Centrifuge: balanced, lid closed until the rotor stops; speed checked with a tachometer; serum tubes clotted ≥30 min first.
📚 Book check — added from the two review books
- Glucose falls about 10 mg/dL per hour (≈7%) in whole blood, but refrigerated, separated serum is fairly stable for glucose. (6.0 p.101)
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4 (pp.70, 76–78; MCQ 10, p.79.e2) and Ch.5 (p.81); Medical Laboratory Science Review (Harr), Ch.2 p.46 Q20, p.68 Q9, p.72 Q16; Ch.5 p.265 Q17, p.267 Q22, p.316 Q28; SUCCESS! in Clinical Laboratory Science, Ch.1, Q149, Q175 (pp.111–113), Ch.3, Q59 (p.403), Ch.17, Q9 (p.1105); BOC Study Guide, 6th ed., Ch.4, Q290 (p.266); DoH Past Paper Exam — Chemistry, p.3 Q44, p.6 Q86, p.20 Q80.
9 · Hematology and coagulation specimen problems
📚 Why this section: Calendar (Sample collection & errors) items: Improper fill; Wrong anticoagulant (hematology and coagulation) · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.3 §I A Electronic cell counting — common errors (p.87) · Check book: Clinical Laboratory Science Review, 5th ed., Hematology — Hemoglobin, hematocrit and ESR sources of error (pp.55, 60, 70); Coagulation — Specimen collection/handling (p.74); Laboratory Operations — Erroneous CBC results (p.240)
Citrate tube: the 9 : 1 rule
C (mL citrate) = 0.00185 × V (mL blood) × (100 − Hct)
Example: Hct 70%, 4.5 mL blood → C = 0.00185 × 4.5 × 30 ≈ 0.25 mL citrate (instead of 0.5 mL).
- Clotted citrate sample → factors consumed → prolonged PT/aPTT (or shortened if partially activated) → check every coagulation tube for clots; reject.
- Hemolyzed coagulation sample → reject (activation of factors and platelets; optical interference).
- Heparin contamination (drawn from a heparinized line) → aPTT and thrombin time very long, PT nearly normal → reptilase time normal; perform heparin neutralization or recollect by venipuncture.
- Coagulation tests need platelet-poor plasma: centrifuge ≈1500 g for 15 min.
EDTA problems in the CBC
| Problem | Effect | Fix |
|---|---|---|
| Underfilled EDTA tube (excess EDTA) | RBCs shrink → Hct ↓, MCV ↓ (MCHC ↑); counts and Hb unchanged | Recollect with correct fill |
| EDTA-dependent platelet clumping / platelet satellitism | Platelet count falsely LOW; WBC may be falsely high; clumps or platelets ringing neutrophils on the smear | Recollect in sodium citrate; multiply the platelet count by 1.1 (dilution correction) |
| Cold agglutinin | RBC count ↓, MCV ↑, MCHC ↑ (>37), agglutinates on smear | Warm sample at 37 °C for 15–30 min and rerun |
| Lipemia (or very high WBC) | Hb falsely ↑ → MCH ↑, MCHC ↑; RBC and Hct correct | Plasma blank / saline replacement |
| Hemolyzed sample | RBC count and Hct falsely low; RBC fragments counted as platelets | Only Hb is valid → recollect |
| Clotted EDTA sample | Platelets and WBC falsely ↓ (consumed in clot) | Reject |
| Nucleated RBCs | Counted as WBCs | Corrected WBC = WBC × 100 ÷ (100 + nRBC) |
- EDTA is the anticoagulant of choice for CBC and smears (ICSH: K₂EDTA). Heparin causes platelet clumping and a blue smear background — use it only for osmotic fragility and RBC enzyme studies.
- Blood bank: EDTA (pink) plasma is preferred; EDTA prevents in-vitro complement binding, so a positive DAT with EDTA reflects in-vivo coating. Pre-transfusion samples must be no older than 3 days if the patient was transfused or pregnant in the last 3 months.
📚 Book check — added from the two review books
- CBC instrument errors (book table): lyse-resistant RBCs (Hb S, Hb C) → falsely ↑ WBC and Hb · giant platelets / platelet clumps → falsely ↑ WBC (giant platelets also ↑ RBC) and falsely ↓ platelets · schistocytes, microcytes → falsely ↑ platelets, falsely ↓ RBC · fragile WBCs → falsely ↓ WBC. (6.0 p.240)
- Hemoglobin (cyanmethemoglobin) errors: falsely ↑ with lipemia, icterus, very high WBC, abnormal globulins (paraproteins) and lysis-resistant Hb S/C; falsely ↓ with sulfhemoglobin (the only Hb form the method does not measure). (6.0 pp.55, 240)
- Glucose > 400 mg/dL → RBCs swell in the diluent → falsely ↑ MCV and Hct, ↓ MCHC. WBC > 50,000/µL → turbidity → falsely ↑ Hb (and MCV, Hct). (Hubbard Ch.3, p.87)
- Carryover: a LOW WBC count run right after a very HIGH count can be falsely high (2–3%; up to 5% when successive counts differ by > 3:1) → repeat the low count. (Hubbard Ch.3, p.87)
- ESR errors: tilted tube or high room temperature → falsely ↑ ESR; low temperature or an underfilled EDTA tube (excess EDTA shrinks RBCs, so they settle slowly) → falsely ↓ ESR. (6.0 pp.60, 70)
- Coagulation tubes: use plastic or siliconized glass — glass activates factor XII and platelets stick to it. Hemolyzed samples must NOT be used for platelet aggregation (RBCs contain ADP). (6.0 p.74)
Sources: Medical Laboratory Science Review (Harr), Ch.1 p.4 Q9, p.36 Q11, p.37 Q12; Ch.2 p.44 Q3–Q5, p.46 Q19, p.68 Q8; SUCCESS! in Clinical Laboratory Science, Ch.2, Q252–Q253, Q259–Q261 (pp.313–315), Ch.3, Q58–Q60 (pp.401–403); BOC Study Guide, 6th ed., Ch.4, Q278–Q296 (pp.264–267), Q425 (p.286); Tietz, 7th ed., Ch.5 MCQ 7–8 (p.128.e16); DoH Past Paper Exam — Hematology/Blood Bank, p.2 Q9, Q17, Q20, p.11 Q64.
10 · Recognizing specimen errors, and analytical and post-analytical errors
📚 Why this section: Calendar (Sample collection & errors) items: Analytical errors; Postanalytical errors; Common examination troubleshooting · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.11 §IV A CLIA 88 — procedure manual: rejection criteria, sources of error, panic values (pp.375–376) · Check book: Clinical Laboratory Science Review, 5th ed., Laboratory Operations — Delta check, sources of error (p.228); Chemistry — Q17 and rationale (pp.133, 136)
Recognize the pattern → name the cause
| Result pattern | Most likely cause | Action |
|---|---|---|
| Glucose very high (e.g., 1100 mg/dL), Na⁺ and Cl⁻ high/normal, K⁺ and HCO₃⁻ very low, other analytes diluted | Drawn ABOVE an IV of dextrose-saline (IV contamination) | Recollect from the other arm |
| K⁺ very high + Ca²⁺ very low (+ ALP, Mg low) in a stable patient | K-EDTA contamination (wrong order of draw or tube transfer) | Recollect |
| Lithium extremely high (e.g., 14 mmol/L) in a well patient | Collected in lithium heparin | Recollect in plain/red tube |
| K⁺, LD, AST, phosphate high; pink serum | In vitro hemolysis | Recollect |
| Low Na⁺ with very high TG or protein (indirect ISE); normal osmolality | Pseudohyponatremia | Direct ISE |
| Delta-check failure on many unrelated analytes | Wrong patient (misidentification) | Investigate, recollect |
| Delta flag on glucose, TG, phosphate, ALP only | Patient not fasting | Comment / recollect fasting |
| Albumin higher than total protein, or a result physiologically impossible | Random analytical error | Repeat the test |
| Anion gap <4 mmol/L, Na⁺ inappropriately low | Lab error (or rare cation excess) | Check Ca, albumin; repeat Na⁺ |
Delta check
A delta check compares a patient's present result with the previous result (usually within 24–72 h). A big difference that is not explained by the clinical situation suggests specimen misidentification, contamination or analytical error. It is the best tool to detect a mislabeled sample.
Analytical errors (review)
- QC out of control → do not report → follow the out-of-control procedure → correct → rerun controls and patients.
- Result above linearity → dilute and repeat (multiply by the dilution factor). Very high antigen in a sandwich immunoassay → hook effect (falsely LOW) → serial dilution.
- Interference from drugs, heterophile antibodies, HIL.
Post-analytical errors
- Transcription error, wrong units, wrong reference interval, report sent to wrong location.
- Critical (panic) values must be telephoned immediately to a responsible caregiver; the receiver reads back the result; document name, time, read-back. Examples: K⁺ <2.5 or >6.5 mmol/L, glucose <40–50 mg/dL.
- Long turnaround time (TAT) is a quality indicator; LEAN "value-stream mapping" studies each step from collection to result.
Specimen rejection criteria (summary)
📚 Book check — added from the two review books
- Anion gap as a QC check: several patients in a row with a high Na⁺ and a high anion gap (e.g., 20–22) → suspect the sodium measurement, not the patients — even when controls are in range. (6.0 pp.133, 136)
- Error type: specimen problems (wrong anticoagulant, drug interference, lipemia, hemolysis) and human errors are RANDOM errors — they affect single samples, not every result in the run. (6.0 p.228)
Sources: Medical Laboratory Science Review (Harr), Ch.5 p.236 Q52, p.313 Q13 & Q15–Q16, p.316 Q29 & Q31, p.323 Q46; Ch.9 p.535 Q65; SUCCESS! in Clinical Laboratory Science, Ch.13, Q38, Q44 (pp.1013–1014); Tietz, 7th ed., Ch.3 & Ch.4 (pp.76–78); DoH Past Paper Exam — Chemistry, p.1 Q10.
11 · Top exam traps
📚 Why this section: Calendar (Sample collection & errors) items: Common examination troubleshooting (review of traps) · Book framework: A Concise Review of Clinical Laboratory Science (Hubbard), Ch.1 §I F Pre-analytical variables (p.7) · Check book: Clinical Laboratory Science Review, 5th ed., Chemistry — Lipid specimen collection: turbid serum (p.105)
• Turbid (milky) serum = triglycerides (triglyceride-rich lipoproteins: chylomicrons and VLDL, the largest particles, scatter light). Cholesterol is carried mainly in small LDL/HDL particles that do not make serum turbid, and albumin is soluble. Turbidity usually appears when triglycerides are above ≈400 mg/dL.
• 24-hour urine: the collection STARTS by emptying the bladder and DISCARDING that urine (note the time); then collect all urine for 24 h, including the final void at the same time next day. Keeping the first urine gives more than 24 h of urine (over-collection). Measuring the total volume is the first step when the laboratory RECEIVES the container.
Top traps for this topic
- Most errors are PRE-analytical. QC checks only the analytical phase.
- Two identifiers; never room/bed number. Label at the bedside after collection. Unlabeled = reject.
- Tourniquet ≤1 min. Long tourniquet / fist pumping → ↑ K⁺, ↑ lactate, ↑ proteins and protein-bound substances.
- Order of draw: culture → blue → red/gold → green → lavender → gray. Skin puncture: EDTA first, serum last.
- EDTA contamination = high K⁺ + low Ca²⁺ (+ low ALP, Mg). EDTA: no calcium. Gray (fluoride): no BUN by urease. Lithium heparin: no lithium level.
- Citrate 3.2%, 9:1. Underfilled or Hct >55% → PT/aPTT falsely long. Reduce citrate: C = 0.00185 × V × (100 − Hct).
- Above an IV: very high glucose, low K⁺ and HCO₃⁻, diluted everything else.
- Hemolysis ↑ K⁺, LD, AST, ACP, phosphate, Mg; ↓ bilirubin (diazo). Na⁺ and calcium least affected.
- Lipemia: pseudohyponatremia (indirect ISE), falsely high Hb/MCHC. Measure TG before ultracentrifugation.
- Unseparated blood: glucose ↓ ~5–7%/h; K⁺ and phosphate ↑ (especially in the fridge); separate within 2 h.
- ABG: heparin, no air bubbles, analyze within 30 min. Air → PO₂ ↑, PCO₂ ↓, pH ↑. Delay → PO₂ ↓, PCO₂ ↑, pH ↓.
- Bilirubin: protect from light. Ammonia/lactate: on ice, test fast. Cold agglutinins: keep at 37 °C.
- PT: 24 h at room temperature (fridge → FVII activation → short PT). aPTT: 4 h (factors V/VIII decay).
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4 (p.72, timed urine) and Ch.5 (pp.91–92, lipemia); Medical Laboratory Science Review (Harr), Ch.5, p.316 Q29. Other traps: see Sections 1–10.
⚡ Last-Minute Review — Sample Collection, Handling & Laboratory Errors
1. Phases & errors
- Pre-analytical = most errors (≈60–70%). Analytical = fewest (QC protects it). Post-analytical = reporting, transcription, critical calls, TAT.
- QC in range ≠ good specimen. Labeling is PRE-analytical; phoning results is POST-analytical.
- Influence factor = real change in the patient (posture, food, exercise). Interference = false change in the test (hemolysis, lipemia, drugs).
2. Patient preparation
- Glucose fast ≥8 h · lipids/TG 12 h · TC & HDL OK nonfasting · OGTT: ≥150 g carb/day × 3 days (restriction → false diabetic curve).
- Cortisol/ACTH highest ~8 a.m. · K⁺ lower afternoon · iron higher morning.
- Supine → upright: plasma volume ↓ → proteins, enzymes, lipids, total Ca, cells ↑ (5–10%). Renin/aldosterone: supine ≥30 min.
- Exercise ↑ CK, AST, myoglobin, lactate, K⁺ · Cooked meat ↑ creatinine (≤20%) · Smoking ↑ COHb, ammonia · Meal ↑ glucose, TG, phosphate, ALP.
3. Identification
- Two identifiers (name + DOB/MRN); never room/bed. Open question: "Tell me your name."
- Label at the bedside, after collection: name, ID number, date/time, collector. Unlabeled/mislabeled → reject. Patient refuses → don't draw, notify, document.
- Blood bank: date of phlebotomy on tube; sample ≤3 days old if transfused/pregnant in last 3 months; two different ABO groups → new sample.
4. Venipuncture
- Median cubital vein · tourniquet ≤1 min, 7–10 cm above · 70% alcohol air-dry · bevel up, 15–30°.
- Long tourniquet / fist pumping → ↑ K⁺, ↑ lactate (DoH: lactic acid falsely ↑), ↑ proteins & total Ca, ↓ pH.
- Avoid: IV arm (both arms → below IV, stopped ≥2 min, discard 5 mL), mastectomy side, fistula, hematoma. Butterfly + citrate first → discard tube.
5. Tubes & order of draw
- Culture → light blue → red/gold → green → lavender → gray. Skin puncture: gases → EDTA first → other additives → serum last.
- Blue: 3.2% citrate, 9:1 (PT, aPTT) · Black: citrate 4:1 (ESR) · Green: heparin (STAT chem, ABG, iCa, ammonia) · Lavender: K₂EDTA (CBC, smear, HbA1c) · Gray: NaF + K-oxalate (glucose, lactate, alcohol) · Royal blue: trace metals · Tan: lead.
- EDTA: no Ca, K, ALP, Mg, coag. Gray: no BUN (urease), no K⁺. Li-heparin: no lithium. Heparin: no coag, no smears/platelets.
- EDTA contamination: K⁺ ↑↑ + Ca²⁺ ↓↓ (+ ALP, Mg ↓) in a well patient → recollect.
- Serum = plasma without fibrinogen; clot ≥30 min before spinning.
6. Special specimens
- ABG: heparin syringe, no air, no clots, ≤30 min (plastic, room temp). Air → PO₂ ↑, PCO₂ ↓, pH ↑. Delay → PO₂ ↓, PCO₂ ↑, pH ↓.
- On ice: ammonia, lactate, ACTH, PTH, (glass ABG) — but renin at room temperature (cold activates it). Warm 37 °C: cryoglobulins, cold agglutinins.
- Protect from light: bilirubin, porphyrins, carotene, vitamin A/B₆.
- Whole blood: lead, CO (COHb). Ionized Ca: anaerobic, balanced heparin, no citrate/oxalate/EDTA.
- Infant heel: medial/lateral plantar, <2 mm; wipe first drop; don't milk. Capillary glucose > venous.
7. Hemolysis, lipemia, icterus
- Hemolysis ↑ K⁺, LD, AST, ACP, phosphate, Mg, iron; ↓ bilirubin (diazo); least affected: Na⁺, Ca²⁺. High K⁺ in stable patient → check hemolysis → recollect.
- Pseudohyperkalemia: hemolysis, fist, delay, fridge, EDTA, thrombocytosis/leukocytosis → confirm with heparin plasma.
- Lipemia = triglycerides (not cholesterol): indirect ISE Na⁺ ↓ (use direct ISE); Hb, MCH, MCHC ↑ (plasma blank/saline replacement); optical coag fails (use mechanical). Measure TG before ultracentrifugation.
- Sample blank = sample + saline (no reagent); corrects color, not chemical interference.
8. Storage & stability
- Separate within 2 h. Unseparated: glucose ↓ 5–7%/h (~10 mg/dL/h); K⁺, phosphate, LD ↑; ammonia, lactate ↑.
- Fridge whole blood → K⁺ ↑ (pump stops). Overnight → K⁺ & phosphate ↑ (DoH).
- PT: 24 h room temp (fridge → FVII activation → short PT). aPTT: 4 h (V, VIII decay → long). Heparin aPTT: spin within 1 h.
- EDTA old: MCV ↑, Hct ↑, ESR ↓, poor morphology. ESR ≤4 h. CSF → immediately.
- Add-on: right tube + enough volume + within stability. Thawed sample: warm and mix.
9. Hematology/coag specimen errors
- Underfilled citrate or Hct >55% → PT/aPTT falsely long. C = 0.00185 × V × (100 − Hct).
- Clotted coag/CBC → reject. Line draw: long aPTT + long TT, normal PT → heparin contamination → neutralize/recollect.
- Underfilled EDTA → Hct ↓, MCV ↓. Platelet clumping/satellitism → citrate × 1.1. Cold agglutinin → warm 37 °C. Lipemia → plasma blank. MCHC > 37 = artefact (except spherocytes).
- EDTA for DAT: blocks in-vitro complement.
10. Recognize the pattern
- Glucose ↑↑, Na/Cl ↑, K⁺ & HCO₃⁻ ↓↓ → drawn above IV.
- Many analytes differ from yesterday → wrong patient (delta check). Glucose/TG/phosphate/ALP up only → not fasting.
- Lithium 14 mmol/L in well patient → Li-heparin tube. Albumin > total protein → random error → repeat.
- Critical value: verify specimen → call → read back → document.
📚 Book check (Hubbard + 6.0)
- Most common pre-analytical problems = wrong specimen type and mislabeled specimen. QA = all 3 phases; QC = analytical only. (Hubbard Ch.1, p.7; 6.0 p.227)
- Hemolysis "KLAMP": K⁺, LD, Aldolase/ACP, Mg²⁺, PO₄ ↑ (+ iron). SPE: extra β band. LD: flipped LD1 > LD2. (6.0 p.101; Hubbard Ch.1, pp.16, 34)
- Pseudohyperkalemia: WBC ≥ 500,000/µL, platelets ≥ 700,000/µL, squeezing a skin puncture. Plasma K⁺ is 0.1–0.7 mmol/L lower than serum K⁺. (6.0 p.114)
- Plasma by mistake on SPE → fibrinogen band between β and γ → recollect serum. (6.0 p.107)
- Ammonia: never serum (clotted); heparin plasma on ice, test within 15 min. ACP: separate fast and acidify (pH ≈5.4). (6.0 pp.111, 135)
- CBC artefacts: Hb S/C → ↑ WBC & Hb · giant platelets/clumps → ↓ PLT, ↑ WBC · schistocytes/microcytes → ↑ PLT · sulfhemoglobin → ↓ Hb · glucose > 400 → ↑ MCV · carryover → repeat low WBC after a high one. (6.0 p.240; Hubbard Ch.3, p.87)
- Coag: plastic/siliconized tubes; citrate may be first tube (straight needle); hemolyzed → no platelet aggregation (ADP). (6.0 p.74)
- Run of patients with high Na⁺ + high anion gap → sodium analyzer problem (even with controls in range). (6.0 p.136)