1 · Solutions, concentration units and unit conversions
The simple idea: almost every number in the laboratory is a concentration = amount of solute (the thing dissolved) in an amount of solution (solute + solvent). If you know what the unit means, you can solve any calculation question in 3 steps: What am I given? → What do I want? → Which conversion connects them?
📌 The five concentration units you must know
| Unit | Means | Formula | Quick example |
|---|---|---|---|
| % w/v | grams of solute per 100 mL of solution | g ÷ mL × 100 | 5 g in 50 mL = 10 g/100 mL = 10% |
| % v/v | mL of solute per 100 mL of solution | mL ÷ mL × 100 | 200 mL acetic acid in 2 L = 10% |
| Molarity (M) | moles of solute per liter of solution | (g ÷ MW) ÷ L | 5.85 g NaCl (MW 58.5) in 0.5 L = 0.1 mol ÷ 0.5 L = 0.2 M |
| Molality (m) | moles of solute per kg of solvent | mol ÷ kg | Used for osmolality (mOsm/kg) |
| Normality (N) | gram-equivalents per liter | N = M × valence; GEW = MW ÷ valence | 1 GEW of HCl (36.5 g) in 1 L = 1 N (= 1 M); 1 M H₂SO₄ = 2 N |
International System of Units (SI)
- SI units have three classes: base, derived and supplemental. Base units cover the fundamental quantities: meter (length), kilogram (mass), second (time), ampere (current), kelvin (temperature), mole (amount of substance), candela (light) — Tietz's table also lists the katal (catalytic activity) for laboratory use.
- A derived unit is built from base units (e.g., mol/m³, pascal, joule). Recommended unit for electrolytes in plasma: mmol/L (mEq is the old unit).
- Prefixes: milli (10⁻³), micro (10⁻⁶), nano (10⁻⁹), pico (10⁻¹²), femto (10⁻¹⁵). So 1000 pg = 1 ng, and 0.1 ng is smaller than 1000 pg. One femtoliter (fL) = 10⁻¹⁵ L = 1 µm³ (MCV unit).
📌 Other conversions that appear in exams
| Convert | Formula | Example |
|---|---|---|
| °C → °F | °F = (°C × 1.8) + 32 | 100 °C = 212 °F; 37 °C = 98.6 °F; 30 °C = 86 °F |
| °F → °C | °C = (°F − 32) ÷ 1.8 | 70 °F = 21 °C |
| pounds → kg | lb ÷ 2.2 | 175.5 lb = 79.8 kg |
| pH ↔ [H⁺] | pH = −log[H⁺] | pH 7.0 = 10⁻⁷ mol/L = 100 nmol/L; [H⁺] 0.025 M → pH 1.6 |
| Hydrate salts | g hydrate = (MW hydrate ÷ MW anhydrous) × g anhydrous | 9.6 g CuSO₄ = 15.0 g CuSO₄·5H₂O |
| Concentrated acid (purity, specific gravity) | g/mL = SG × purity; then volume = grams needed ÷ g/mL | HNO₃ 68%, SG 1.42 → 0.9656 g/mL |
How this topic is organised
| # | Study section | Calendar items | Guide headings (Hubbard / 6.0) | Textbook chapter |
|---|---|---|---|---|
| 1 | Solutions, concentration & units | Concentrations · Percent solutions · Unit conversions | Hubbard Ch.1 §I A (pp.1–2) · 6.0 pp.231–233 | Tietz Ch.15 pp.199.e1–199.e2 |
| 2 | Dilutions & standard curves | Dilutions · Concentrations | Hubbard Ch.1 §I A 1d (p.2) · 6.0 pp.231–234 | Tietz Ch.16 pp.202–203, 208 |
| 3 | Clinical calculations | Important chemistry and hematology/coagulation calculations | 6.0 pp.233, 244 | Tietz Ch.34, 36, 50, 81; Turgeon Ch.5, 23, 26 |
| 4 | Reagents, water, glassware & basic equipment | Chemistry instrumentation (basic) · Quality | 6.0 pp.227–228 | Tietz Ch.15 pp.199.e3–199.e13; Strasinger Ch.7 p.131 |
| 5 | Quality assurance & quality management | Quality Assurance · Basic regulatory/quality principles | Hubbard Ch.11 §III–IV (pp.371–372) · 6.0 pp.227, 256 | Tietz Ch.3 pp.60–61, 60.e4–60.e14; Strasinger Ch.7 p.127 |
| 6 | Statistics, accuracy, precision & errors | Precision · Accuracy · Random/systematic errors | Hubbard Ch.1 §I B (pp.2–3) · 6.0 pp.228–230 | Tietz Ch.2 pp.15–16; Ch.6 p.132; Ch.9 pp.180, 186 |
| 7 | Quality control: Levey-Jennings & Westgard | Quality Control | Hubbard Ch.1 §I B · 6.0 p.230 | Tietz Ch.6 pp.129–150; Turgeon Ch.1 p.17 |
| 8 | Sensitivity, specificity & predictive values | Sensitivity · Specificity · Predictive values | 6.0 p.229 | Tietz Ch.2 pp.16, 45, 47, 51 |
| 9 | Chemistry instrumentation, automation & POCT | Chemistry instrumentation | Hubbard Ch.1 §I C–E, §II (pp.3–10) · 6.0 pp.234–238 | Tietz Ch.16, 17, 29, 30 |
| 10 | Hematology, coagulation, urinalysis & microbiology instruments; flow cytometry | Hematology · Coagulation · Microbiology · Urinalysis instrumentation · Flow cytometry | 6.0 pp.239–245 | Turgeon Ch.27; Tietz Ch.29, 84, 86; Jawetz Ch.47; Strasinger Ch.6 |
| 11 | Safety I: biological hazards, PPE & exposure | Biological hazards · PPE · Exposure response | Hubbard Ch.7 §XXIII · 6.0 pp.251–253 | Tietz Ch.15 pp.199.e17–199.e21, 199.e25; Strasinger Ch.1 pp.2–5; Turgeon Ch.1 pp.3–9 |
| 12 | Safety II: chemical, fire, electrical, radiation, waste & symbols | Chemical hazards · Fire · Electrical safety · Waste disposal · Safety symbols | 6.0 pp.253–255 | Tietz Ch.15 pp.199.e16–199.e24; Strasinger Ch.1 pp.6–8 |
| 13 | Laboratory operations | Documentation · Result reporting · Critical values · LIS · Regulation · Management | Hubbard Ch.11 §IV–VI (pp.372–382) · 6.0 p.256 | Tietz Ch.3, 4, 5, 29; Strasinger Ch.7; Turgeon Ch.1 |
| 14 | Comparison tables & exam traps | All items (review) | Hubbard Ch.1, 11 · 6.0 pp.227–256 | Tietz Ch.2–6, 15, 16 |
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.15, p.199.e1 (molarity, molality, gram-equivalent weight, normality; mmol/L recommended for electrolytes; base, derived and supplemental SI units), p.199.e2 (Table 15.2 derived units); Turgeon, Clinical Hematology 5th ed., Ch.26, p.502 (fL = 10⁻¹⁵ L) · Guide: Hubbard Ch.1 §I A (pp.1–2); 6.0 Lab Operations pp.231–233
2 · Dilutions, serial dilutions and standard curves
The simple idea: a dilution tells you how many parts of sample are in the TOTAL volume. "1:10" means 1 part sample in 10 parts total (1 part + 9 parts diluent).
🔥 Worked patterns (memorise the logic, not the numbers)
- 1 part bleach + 9 parts water = 1:10 → 1/10 = 10% bleach. If the bleach is 5% sodium hypochlorite, the final is 5% ÷ 10 = 0.5%.
- 1 part CSF + 3 parts saline = 1:4 → result × 4.
- 0.1 mL serum + 5 mL reagent + 4.9 mL water = 0.1 in 10 mL = 1:100.
- 1 mL serum + 9 mL water, analyzer reads 350 U/L → 350 × 10 = 3,500 U/L.
- 5 mL of a 1:25 dilution needs 5 ÷ 25 = 0.2 mL of reagent.
- Serial: 5 N diluted 1:4 = 1.25 N; then diluted 4:15 → 1.25 × 4/15 = 0.33 N.
- Titres: if 0.5 mL of a 1:300 dilution has 1 unit, 2 units are in 0.5 mL of a 1:150 (twice as concentrated).
📌 C₁V₁ = C₂V₂ (the "dilute a stock" formula)
| Question | Set-up | Answer |
|---|---|---|
| Glucose stock 2,000 mg/dL → 100 mL of 150 mg/dL | 2000 × V₁ = 150 × 100 | V₁ = 7.5 mL |
| 0.25 N NaOH → 100 mL of 0.05 N | 0.25 × V₁ = 0.05 × 100 | 20 mL |
| 95% alcohol → 5 L of 70% | 95 × V₁ = 70 × 5 | 3.7 L |
| 1.025 N NaOH → 1 L of 1.0 N | 1.025 × V₁ = 1.0 × 1000 | 975.6 mL |
Standard curves and Beer's law calculation
A standard (calibrator) has an exactly known concentration and is used to calibrate the method. A control is used to check that the calibrated method is still working. Because absorbance is directly proportional to concentration (Beer's law, A = abc), an unknown can be calculated from one standard:
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16, p.202 (absorbance = −log T, directly proportional to concentration), p.203 (Beer's law A = abc), p.208 (unknown compared directly with a calibrator; verification of linearity); Ch.15, p.199.e4 (primary and secondary reference materials); Turgeon, Clinical Hematology 5th ed., Ch.1, p.9 (Table 1.1: 1 mL bleach + 9 mL water = 1:10 = 0.5% sodium hypochlorite) · Guide: Hubbard Ch.1 §I A 1d (p.2); 6.0 pp.231–234
3 · Important chemistry, hematology and coagulation calculations
The simple idea: the exam does not ask for new formulas — it asks the classic ones again and again. Learn each formula with one example.
🔥 Chemistry formulas
| Test | Formula | Key point |
|---|---|---|
| Creatinine clearance | (U × V ÷ P) × (1.73 ÷ body surface area) | U and P in the same unit; V in mL/min (24-h volume ÷ 1440). Needs a timed urine. Result in mL/min/1.73 m². |
| Anion gap | Na⁺ − (Cl⁻ + HCO₃⁻) | High in lactic acidosis, ketoacidosis, renal failure, methanol, ethylene glycol, salicylate. |
| Calculated osmolality | 2 Na + glucose/18 + BUN/2.8 (mg/dL) | Osmolal gap = measured − calculated; a gap > about 10–15 suggests toxic alcohols. |
| LDL-C (Friedewald) | TC − HDL-C − TG/5 (mg/dL) | Not valid when TG > 400 mg/dL (4.5 mmol/L) or nonfasting. |
| Enzyme activity (IU/L) | (ΔA/min × total volume × 1000) ÷ (6.22 × light path × sample volume) | 6.22 = millimolar absorptivity of NADH at 340 nm. |
🔥 Hematology formulas
| Index | Formula | Example / normal |
|---|---|---|
| MCV (fL) | Hct (%) × 10 ÷ RBC (×10¹²/L) | 35% ÷ 4.0 × 10 = 87.5 fL · normal 80–100 fL |
| MCH (pg) | Hb (g/dL) × 10 ÷ RBC | not affected by an error in Hct |
| MCHC (g/dL) | Hb ÷ Hct × 100 | 15 ÷ 47 × 100 = 31.9 · normal 32–36 |
| Corrected reticulocyte | retic % × (patient Hct ÷ normal Hct 45) | 3% × 30/45 = 2.0% |
| Reticulocyte production index (RPI) | corrected retic ÷ maturation time (days) | RPI ≈ 1 normal; ≥3 in hemolysis with good marrow |
| Manual (hemocytometer) count | cells counted × dilution factor × 10 (depth) ÷ area counted (mm²) | 36 cells in 9 mm², 1:10 → 36 × 10 × 10 ÷ 9 = 400/µL = 0.4 × 10⁹/L |
| WBC corrected for nucleated RBCs | uncorrected WBC × 100 ÷ (100 + NRBC per 100 WBC) | NRBCs are counted as WBCs by many counters |
| Absolute count | % × total count | CD4: 5,500 × 0.20 lymph × 0.25 = 275/µL |
⚑ Not verified in the textbook — the hemocytometer formula and the NRBC correction formula are taken from Medical Laboratory Science Review (Harr), Ch.1, p.4 and p.6–7 (Turgeon Ch.25, p.469 gives only the counting rules). The exact calculated-osmolality formula and the anion-gap formula were not re-opened in Tietz for this page (they are taught with their textbook pages in MLS 03 and MLS 04).
🔥 Coagulation numbers
- INR = (patient PT ÷ mean normal PT)ISI. The ISI (International Sensitivity Index) is the exponent: WHO reference thromboplastin ISI = 1.0; a more sensitive reagent has ISI < 1.0. INR is used to monitor warfarin.
- Citrate tube ratio = 9 parts blood : 1 part 3.2% sodium citrate. When Hct > 55%, plasma volume is low, so use less citrate: C = (1.85 × 10⁻³) × (100 − Hct) × V blood.
- Platelet-poor plasma (platelets < 10 × 10⁹/L) for PT/aPTT: centrifuge about 1500 × g for at least 15 minutes. Low speed (about 150–200 × g) gives platelet-rich plasma for aggregation studies.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.34, p.352.e28 (concept of clearance), p.352.e33 (creatinine clearance, timed urine), p.352.e39 (adjustment to 1.73 m²); Ch.36, Box 36.4 (Friedewald equation); Ch.50, pp.694–695 (anion gap causes; osmolal gap and toxic alcohols); Ch.81, pp.1109.e7–1109.e8 (Box 81.1 citrate adjustment for Hct > 55%; PPP < 10 × 10⁹/L at 1500 × g ≥ 15 min); Ch.78, p.1096 (PRP for aggregometry); Turgeon, Clinical Hematology 5th ed., Ch.5, p.97 (corrected reticulocyte count and RPI); Ch.23, p.417 (ISI and INR formula); Ch.26, pp.501–502 (MCV, MCH, MCHC) · Guide: 6.0 pp.233, 244
4 · Reagents, reagent water, glassware, pipettes, centrifuges and balances
The simple idea: a good method gives a wrong result if the water is impure, the pipette is wrong or the centrifuge is too slow. These "basic" items are tested often.
📌 Reagent-grade water
- Made by distillation (vaporize and condense — oldest method; carries over volatile impurities), ion exchange / deionization (resins replace cations with H⁺ and anions with OH⁻, which form water), reverse osmosis (water forced through a semipermeable membrane), and UV oxidation. Water is filtered first.
- Clinical laboratory reagent water (CLRW) (formerly type I/II) is used for most testing: resistivity ≥ 10 MΩ·cm at 25 °C, < 10 CFU/mL, passed through a 0.22 µm filter, low organic content. Autoclave/wash water (formerly type III) is for washing glassware.
- More ions → lower resistivity and higher conductivity. Mixed-bed deionizers in series give > 10 MΩ·cm.
Chemicals and reference materials
Analytical reagent (AR/ACS) grade chemicals are used for reagents. A primary reference material (primary standard) is highly purified, stable, can be weighed directly and is used to prepare calibrators; a secondary standard is assayed against a primary one. NIST supplies Standard Reference Materials (SRMs).
🔥 Glassware and pipettes
| Item | Use / key fact |
|---|---|
| Class A glassware | Certified accuracy (NIST) — required for accurate work |
| Volumetric (transfer) pipette + volumetric flask | Most accurate single volume — use for standards and working solutions |
| Ostwald-Folin pipette | For viscous fluids (blood, serum); etched ring = blow-out |
| Serologic / Mohr (measuring) pipettes | Graduated; serologic is calibrated to the tip and blown out (etched ring) |
| Micropipettes | Usually to contain (TC) — rinse out; most automatic pipettes are air-displacement |
| Beakers, Erlenmeyer flasks, graduated cylinders | Not accurate enough for standards |
| Desiccator | Keeps chemicals dry; cobalt-treated desiccant is blue when dry and turns pink when it must be changed |
🔥 Centrifuges
- Report speed as RCF (× g), not rpm, because rotors have different radii.
- Balance tubes of equal weight opposite each other (difference ≤ 1% or as the manufacturer says), keep the lid closed, cap tubes, and never open until the rotor has stopped — this prevents aerosols.
- Check speed with a tachometer; Strasinger: centrifuge calibration about every 3 months (other books give 6 months or yearly — follow the manufacturer and the accreditation checklist). Disinfect centrifuges weekly.
- Other daily checks: refrigerator and water-bath temperatures are recorded daily; microscopes get an annual professional cleaning.
Balances
An analytical balance compares the unknown mass with known weights; modern electronic balances use an electromagnetic force to counterbalance the load. Calibration is checked with certified weights (Class S / ASTM class 1–type weights).
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.15, p.199.e3 (distillation, ion exchange, reverse osmosis, UV oxidation; Table 15.4 CLRW specifications, resistivity), p.199.e4 (CLRW vs autoclave/wash water; reagent grades; primary reference materials), p.199.e6 (class A glassware; pipettes to contain/to deliver), p.199.e7 (Ostwald-Folin and serologic blow-out pipettes; micropipettes calibrated to contain), p.199.e10 (RCF formula, radius to tube bottom, balancing ≤ 1%, centrifuge before unstoppering), p.199.e13 (balances, weights); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.6, p.83 (RCF from tachometer rpm), Ch.7, p.131 (refrigerator temperatures daily; centrifuge calibration every 3 months; weekly disinfection) · ⚑ Not verified in the textbook — cobalt desiccant colour change: source BOC Study Guide 6th ed., Ch.8, p.495, Q229 · Guide: 6.0 pp.227–228
5 · Quality assurance (QA) and quality management
The simple idea: Quality assurance (QA) = the whole system that makes sure the patient gets the right result, on the right patient, at the right time. Quality control (QC) is only one part of QA — it checks the analytical step.
📌 Correct order of the testing process
Receive the sample → run QC (confirm the analyzer is in control) → analyze → verify the result → print/release. A result is always verified before it is released.
Internal QC vs external quality assessment (EQA) / proficiency testing (PT)
| Internal QC | EQA / Proficiency testing | |
|---|---|---|
| Who sends samples? | The lab runs its own control material | An independent external organization sends unknown samples |
| Checks | Day-to-day precision and stability of each run | Accuracy compared with other laboratories (peer group) / target value |
| Rules | Westgard rules, Levey-Jennings | Test PT samples exactly like patient samples: no extra repeats, no averaging between staff, no discussion with other labs, no sending to another lab before submitting |
- CLIA (USA) PT: at least 5 challenges per analyte, 3 events per year; ABO, Rh and compatibility need 100%; other analytes need ≥ 80%. Even ungraded PT challenges must be reviewed.
- Comparing the lab mean with the peer-group mean uses the standard deviation index (SDI) = (lab mean − peer mean) ÷ peer SD; > 2 is significant.
📌 Quality management system (QMS) and accreditation
- ISO 15189 = the international QMS standard for medical laboratories (quality and competence). ISO 15190 = safety. CAP and The Joint Commission are accrediting bodies; accreditation = an agency evaluates the lab against predetermined standards.
- ISO 15189 clauses include document control, identification and control of nonconformities, corrective action, preventive action, continual improvement, internal audits and management review.
- Correction = fix the error itself. Corrective action = remove the cause of an error that already happened so it does not happen again. Preventive action = remove the cause of a potential problem before it occurs.
- Root cause analysis looks for the underlying cause and needs evidence that the fix works. Tools: PDCA (Plan-Do-Check-Act), Six Sigma (DMAIC: Define, Measure, Analyze, Improve, Control) — six sigma ≈ 3.4 defects per million; Lean removes waste; Pareto charts rank causes; fishbone diagrams list causes; Gantt charts follow project progress.
- Problem solving starts by gathering data (define the problem with facts) — e.g., for a turnaround-time (TAT) complaint, first collect current TAT data by shift.
- Training is required for new, transferred or promoted staff, new instruments and repeated performance problems — not because the medical director changes. Competency is assessed by direct observation, problem-solving exercises, blind samples and quizzes; signing that training is complete is not an assessment.
- IQCP (individualized quality control plan) = risk assessment → QC plan → quality assessment.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.3, p.60 (ISO 15189 as the QMS standard), p.60.e4 (risk ranking; root cause analysis), p.60.e8 (ISO 15189 clauses 4.9 nonconformities, 4.10 corrective action, 4.11 preventive action, 4.14 audits), pp.60.e13–60.e14 (root cause, PDCA, Six Sigma DMAIC), p.61 (EQA/PT benchmarking); Ch.6, pp.129–130 (internal QC vs EQA/PT from an independent organization with unknown values); Ch.5, p.127 (quality indicators: misidentified and unlabeled samples); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.7, p.127 (QA = pre-analytical, analytical and post-analytical factors; internal and external QC; PT) · ⚑ Not verified in the textbook — CLIA PT numbers (5 challenges, 3 events, 100%/80%) and SDI > 2: source Medical Laboratory Science Review (Harr), Ch.9, pp.528–529 and Ch.5, p.236 · Guide: Hubbard Ch.11 §III (p.371); 6.0 pp.227, 256
6 · Statistics, accuracy, precision, errors and reference intervals
The simple idea: repeat the same sample many times — the results spread around a centre. The centre tells you about accuracy; the spread tells you about precision.
📌 Basic statistics
| Term | Meaning | Formula / note |
|---|---|---|
| Mean | average | Σx ÷ n (e.g., 110,120,150,140,130 → 130) |
| Median | middle value (50th percentile) | not affected by outliers |
| Mode | most frequent value | |
| Range | highest − lowest | simplest measure of spread |
| Standard deviation (SD) | spread of results around the mean | √[Σ(x − x̄)² ÷ (n − 1)] = √variance |
| Variance | SD² | Σ(x − x̄)² ÷ (n − 1) |
| Coefficient of variation (CV) | relative SD — the index of precision | CV% = SD ÷ mean × 100; use CV to compare precision of tests with different means |
🔥 Accuracy vs precision
| Term | Definition | Affected by |
|---|---|---|
| Accuracy | closeness of a result to the true value | bias + imprecision |
| Trueness / bias | difference between the mean of many results and the true (target) value | systematic error |
| Precision (reproducibility) | closeness of repeated results to each other | random error; validated by running the same sample many times |
| Reliability ⚑ | ability to keep both accuracy and precision over time | |
| Analytical sensitivity | smallest amount that can be detected (detection limit) | |
| Analytical specificity | measures only the analyte it claims to measure (no interference) |
Widely scattered results whose average is near the target = accurate but imprecise (random error). Tight results far from the target = precise but inaccurate (systematic error).
Types of error
- Random error — unpredictable scatter (pipetting variation, bubbles, electronic noise) → poor precision. Detected by 1-3s and R-4s.
- Systematic error (bias) — all results shifted in one direction (calibrator, reagent lot, wavelength, temperature) → poor accuracy. Constant error = same amount at all levels (regression y-intercept); proportional error = grows with concentration (regression slope ≠ 1).
Method comparison and evaluation
- Linear regression (y = a + bx): intercept = constant error, slope = proportional error. Example ŷ = 2.10 + 1.01x → constant bias.
- Correlation coefficient (r) shows association only (r = 0 = no correlation) — do not use r alone to accept a method.
- Student's t-test compares means (bias); F-test compares variances (precision). Bias = mean y − mean x (246 − 235 = 11 mg/dL).
- Recovery experiments test proportional error; interference experiments test constant error; linearity verifies the reportable range (blank + at least 3 known levels).
- A reagent lot change needs patient-sample comparison (not just QC) because QC material can show matrix effects.
📌 Reference intervals
- Usually the central 95% of a healthy reference population (2.5th–97.5th percentiles), so about 5% of healthy people fall outside — an abnormal result is not always disease.
- To establish a new interval nonparametrically: at least 120 reference individuals (per partition).
- To verify (transfer) a manufacturer's interval: test 20 healthy people; accept if no more than 2 are outside.
- Reference intervals are the most important tool for interpreting a result; they are not critical values.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.2, p.15 (median, 2.5th and 97.5th percentiles, nonparametric 95% interval); Ch.6, p.132 (SD describes imprecision; ±1 SD 68%, ±2 SD 95%, 2.5% each tail; bias; accuracy combines bias and imprecision; trueness), pp.146–147 (reagent lot change verified with patient samples); Ch.9, p.180 (minimum 120 values for central 95 percentiles), p.186 (verification with 20 values, no more than two outside); Turgeon, Clinical Hematology 5th ed., Ch.1, p.17 (dispersion, trends, shifts) · ⚑ Not verified in the textbook — "reliability" definition: source SUCCESS! in Clinical Laboratory Science, Ch.13, p.1013, Q35; t-test/F-test, regression intercept/slope and linearity samples: source Medical Laboratory Science Review (Harr), Ch.5, pp.231–233 and Ch.9, p.533 · Guide: Hubbard Ch.1 §I B (pp.2–3); 6.0 pp.228–230
7 · Quality control: control material, Levey-Jennings charts and Westgard rules
The simple idea: a control is a sample with a known expected range. If the control is "in", we trust the patient results of that run. If it is "out", we do not report patients until the problem is fixed.
Control material
- Stable, matrix similar to patient specimens, concentrations at clinically important levels (normal and abnormal). The lab sets its own target mean and SD from its own runs — at least 20 results on different days to start; a cumulative SD over 6–12 months is more realistic.
- Do not use the manufacturer's (package-insert) ranges as your limits — they are too wide.
- When calculating next month's limits, include results between 2 and 3 SD (excluding them makes limits too narrow).
- Minimum frequency for most chemistry tests (CLIA): two levels every 24 hours (more for blood gases, hematology, POCT glucose).
🔥 Levey-Jennings chart
Control result (y-axis) plotted against day/run (x-axis) with lines at mean, ±1, ±2 and ±3 SD.
| Pattern | What you see | Typical cause | Error type |
|---|---|---|---|
| Shift | sudden jump; values stay on one side of the mean (≥ 6 consecutive) | new reagent or calibrator lot, recalibration, changed lamp/probe, wrong calibrator concentration | systematic |
| Trend | values gradually move up or down over ≥ 6 runs | deteriorating reagent/calibrator, aging lamp, evaporation, slow temperature drift | systematic |
| Dispersion / random | wide scatter both sides | pipetting, bubbles, poor mixing, power fluctuation | random |
📌 Westgard multirules
| Rule | Meaning | Detects | Action |
|---|---|---|---|
| 1-2s | one control outside ± 2 SD | warning (5% false alerts — 1 in 20 by chance) | inspect with other rules |
| 1-3s | one control outside ± 3 SD | random (or large systematic) | reject run |
| 2-2s | two consecutive controls (same level across runs, or two levels in one run) beyond the same ± 2 SD | systematic | reject |
| R-4s | within one run, one control above +2 SD and the other below −2 SD (difference > 4 SD) | random | reject |
| 4-1s | four consecutive controls beyond the same ± 1 SD | systematic | reject |
| 10x̄ | ten consecutive controls on the same side of the mean | systematic (high false-alert rate) | reject / investigate |
- 2-2s broken → systematic error: recalibrate and repeat controls and selected patients (repeating only the control does not fix it).
- Probability of two consecutive results beyond −2 SD by chance = 1/40 × 1/40 = 1/1,600.
- Other plots: Cusum — earliest sign of shift/trend; Tonks-Youden — compares labs (two control levels plotted against each other).
Patient-based checks
Delta check = compares a patient's result with the previous result (e.g., within 24–48 h). A big unexplained change suggests specimen misidentification, contamination (IV fluid) or error. Best for analytes with low biological variation (e.g., MCV, MCHC). Example: bilirubin 2.0 → 6.2 mg/dL next day with QC in range → perform a delta check and investigate.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.6, p.132 (Levey-Jennings plot), p.139 (minimum 20 results on different days; cumulative SD 6–12 months), p.140 (lab should assign its own target and SD; product-insert ranges too wide; Table 6.3 rules 1-3s, 2-2s, R-4s, 10x̄, 8-1s, CUSUM and the error they detect), p.141 (1-2s false-alert rate), pp.146–147 (reagent lot change), p.150 (Table 6.7 delta-check criteria for mislabeled or contaminated samples); Ch.5, p.123 (delta checks best for low-variation analytes such as MCV, MCHC); Turgeon, Clinical Hematology 5th ed., Ch.1, p.15 (delta check), p.17 (trends and shifts) · ⚑ Not verified in the textbook — "include 2–3 SD values when calculating new limits" and the CLIA frequency (2 levels/24 h): source Medical Laboratory Science Review (Harr), Ch.5, pp.228, 231 · Guide: Hubbard Ch.1 §I B; 6.0 p.230
8 · Diagnostic sensitivity, specificity and predictive values
The simple idea: draw a 2 × 2 table. Every diagnostic-test question becomes easy.
| Disease present | Disease absent | |
|---|---|---|
| Test positive | TP (true positive) | FP (false positive) |
| Test negative | FN (false negative) | TN (true negative) |
| Term | Question it answers | Formula |
|---|---|---|
| Sensitivity | Of people WITH disease, how many test positive? | TP ÷ (TP + FN) × 100 |
| Specificity | Of people WITHOUT disease, how many test negative? | TN ÷ (TN + FP) × 100 |
| Positive predictive value (PPV) | Of people who test positive, how many really have the disease? | TP ÷ (TP + FP) × 100 |
| Negative predictive value (NPV) | Of people who test negative, how many are really free of disease? | TN ÷ (TN + FN) × 100 |
| Efficiency (accuracy) | How many are classified correctly? | (TP + TN) ÷ total × 100 |
🔥 Prevalence changes predictive values — not sensitivity or specificity
- Sensitivity and specificity belong to the test (at a chosen cut-off). PPV, NPV and efficiency depend on prevalence.
- Higher prevalence → PPV rises, NPV falls. Rare disease → low PPV.
- Moving the cut-off: lower cut-off → higher sensitivity, lower specificity (more FP); higher cut-off → the opposite.
- ROC curve (receiver operating characteristic) = sensitivity plotted against (1 − specificity) for every cut-off; the point nearest the upper-left corner gives the best sensitivity with fewest false positives. Area under the curve compares tests.
- A screening test needs high sensitivity (few false negatives, high NPV); a confirmatory test needs high specificity.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.2, p.16 (diagnostic sensitivity = proportion of diseased with a positive result), p.45 (ROC curves: sensitivity/specificity pairs for all cut-offs), p.47 (predictive values depend highly on prevalence; Bayes formulas), p.51 (definitions of sensitivity, specificity, PPV, NPV) · Guide: 6.0 p.229
9 · Chemistry instrumentation, automation and point-of-care testing
The simple idea: every analyzer measures a signal (light absorbed, light emitted, light scattered, voltage, current) and turns it into a concentration by comparing it with calibrators.
🔥 Spectrophotometer — parts in order
- Tungsten (incandescent) lamp = visible light; deuterium/hydrogen (and xenon/mercury arc) = ultraviolet (UV, < 320–380 nm). 340 nm (NADH) is UV.
- Monochromator isolates a narrow band of wavelengths. Spectral bandwidth (width at half the peak transmittance) describes spectral purity.
- Double-beam instruments use a reference beam to correct for changes in light-source intensity.
- Verify performance: wavelength accuracy (didymium or holmium oxide filters), bandwidth, stray light, linearity and photometric accuracy. The first step for a new assay is to find the wavelength of maximum absorbance.
- Bichromatic analysis reads at two wavelengths to correct for background interference (lipemia, hemolysis).
📌 Other optical and electrochemical methods
| Method | Measures | Key facts / uses |
|---|---|---|
| Fluorometry | light emitted at a longer wavelength after excitation | Detector usually at 90° (right angle); xenon or mercury lamp (or laser); more sensitive than absorbance |
| Chemiluminescence | light from a chemical reaction (no excitation lamp) | Very sensitive immunoassays; electrochemiluminescence |
| Nephelometry | light scattered by particles (often at an angle, e.g., 90°) | Immunoglobulins, specific proteins |
| Turbidimetry | decrease in transmitted light due to particles | Measured like absorbance |
| Flame emission photometry | light emitted by atoms excited in a flame | Na, K, Li (older method) |
| Atomic absorption (AAS) | light absorbed by ground-state atoms | Hollow cathode lamp made of the element measured; Ca, Mg, lead, trace metals |
| Reflectance photometry | diffuse reflected light from a pad | Dry-slide chemistry, urine strip readers, glucose meters |
| Ion-selective electrode (ISE) | potential (voltage) related to ion activity | The membrane gives selectivity (valinomycin for K⁺, glass for Na⁺/pH); direct vs indirect ISE |
| Osmometry (freezing point) | freezing-point depression | Osmolality |
| Electrophoresis | migration of charged molecules in an electric field | Albumin fastest to the anode; crescent bands = overloading |
| Chromatography | separation by differential solubility/partition | TLC Rf = distance of spot ÷ distance of solvent front; GC needs volatile compounds; gel filtration separates by size |
| Mass spectrometry | mass-to-charge ratio (m/z) | Drug confirmation, newborn screening; MALDI-TOF for microbes |
Automation, middleware and point-of-care testing (POCT)
- Total laboratory automation links pre-analytical (barcoded labels, sorting, centrifuging, decapping), analytical and post-analytical (storage, retrieval) steps.
- Autoverification = rules-based release of results by the LIS or middleware (e.g., hold results if QC failed, flag critical values or delta failures for review).
- POCT = testing near the patient (bedside, ED, ICU, operating room). Main advantage: faster turnaround time; also small sample volume. It still needs QC, training and documentation.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16, p.202 (A = −log T), p.203 (Beer's law A = abc), p.204 (components of single- and double-beam spectrophotometers; tungsten for visible, deuterium/hydrogen, mercury and xenon arc for UV), p.205 (spectral bandwidth = spectral purity; hollow cathode lamp; stray light), p.208 (holmium oxide filter; wavelength accuracy, bandwidth, stray light, linearity, photometric accuracy), pp.209–210 (reflectance photometry; flame emission; AAS components), p.214 (right-angle fluorometer detection), p.216 (xenon source), p.220 (nephelometry and turbidimetry measure scattered light); Ch.17, p.224.e3 (ISE membrane controls selectivity); Ch.29, p.324 (total laboratory automation), p.324.e5 and p.324.e16 (autoverification, middleware, rules); Ch.30, p.326 (Box 30.2: POCT advantage = reduced turnaround time) · Guide: Hubbard Ch.1 §I C–E, §II (pp.3–10); 6.0 pp.234–238
10 · Hematology, coagulation, urinalysis and microbiology instruments; flow cytometry
🔥 Hematology analyzers — two counting principles
| Principle | How it works | Key points |
|---|---|---|
| Electrical impedance (Coulter principle) | Cells are non-conductive in a conductive diluent; each cell passing through a small aperture increases resistance → a voltage pulse | Number of pulses = count; pulse height = cell volume (MCV, RDW from the histogram) |
| Optical (light scatter) with hydrodynamic focusing | Cells flow one by one through a laser beam | Forward scatter ≈ size; side (90°) scatter ≈ granularity/internal complexity → 5-part differential |
- Hemoglobin is measured photometrically after lysis (cyanmethemoglobin or SLS methods).
- Interferences: marked leukocytosis falsely raises Hb (turbidity), RBC count and MCV; lipemia falsely raises Hb; giant platelets or clumps lower the platelet count; nucleated RBCs are counted as WBCs (correct the WBC); cold agglutinins falsely raise MCV and lower RBC.
- A right shift of the RBC histogram = larger cells (macrocytosis, e.g., pernicious anemia); left shift = microcytes.
📌 Coagulation analyzers — clot detection
- Electromechanical methods detect the fibrin clot by a change in conductivity or by a moving steel ball/probe that stops when the clot forms. Not affected by lipemia or icterus; possible error: sample carryover.
- Photo-optical methods detect the clot as a change in light transmission (optical density). Lipemia, icterus and hemolysis can interfere.
- Platelet aggregation uses platelet-rich plasma and light-transmission aggregometry.
Urinalysis instruments
- Reagent strip readers use reflectance photometry; they standardize colour reading and timing.
- Specific gravity: refractometer (refractive index), urinometer, harmonic oscillation densitometry.
- Automated sediment analyzers use flow cytometry or digital imaging of particles; workcells combine chemistry and particle modules.
- Routine sediment and Gram stains are examined with the bright-field microscope; phase-contrast helps pale casts; polarizing for crystals/lipids; fluorescence for special stains; dark-field for Treponema.
📌 Microbiology instruments
- Continuously monitored blood culture systems detect growth by the production/consumption of gas (usually CO₂) with a sensor; standard 5-day protocol.
- Mycobacterial liquid culture (MGIT) detects growth by fluorescence as oxygen is used.
- Automated identification and susceptibility systems (e.g., VITEK 2, Phoenix, MicroScan) read biochemical and growth changes in cards/panels.
- MALDI-TOF mass spectrometry: a colony smear + acid matrix on a metal target is hit by a laser; ionized proteins (mostly ribosomal proteins) fly through a vacuum tube; the time of flight gives an m/z spectral fingerprint compared with a database → rapid identification of bacteria and yeasts.
🔥 Flow cytometry
- Single cells in a fluid stream pass a laser (intense, stable, monochromatic) → forward scatter (size), side scatter (granularity) and fluorescence from fluorochrome-labelled monoclonal antibodies (immunophenotyping, CD markers).
- Steps: sample preparation → Fc-receptor blocking → antibody staining → acquisition and analysis (gating, e.g., CD45 vs side scatter). ⚑
- Uses: leukemia/lymphoma phenotyping, CD4 counts (absolute CD4 = WBC × lymph % × CD4 %), PNH, stem-cell (CD34) counts, DNA content.
Sources: Turgeon, Clinical Hematology 5th ed., Ch.27, p.527 (electrical impedance/Coulter principle through an aperture; optical detection with hydrodynamic focusing and laser), p.529 (laser properties; principles of flow cytometry), p.530 (forward scatter ≈ size, side scatter ≈ granularity), p.555 (electromechanical vs photo-optical clot detection); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.29, p.324.e15 (reflectance photometry), p.324.e18 (automated urinalysis: reflectometry strips and flow-cytometric or microscopic particle analysis); Ch.84, p.1201.e30 (continuously monitored blood culture systems detecting CO₂/O₂); Ch.86, p.1203.e12 (MGIT continuously monitors fluorescence); Ch.85 (automated AST platforms VITEK 2, Phoenix, MicroScan); Ch.78, p.1096 (light-transmission aggregometry with PRP); Jawetz, Melnick & Adelberg's Medical Microbiology 26th ed., Ch.47, p.765 (MALDI-TOF principle, ribosomal proteins, m/z 1000–20,000), p.769 (automated urinalysis instruments); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.5, p.55 (automated strip readers), Ch.6, p.87 (Table 6–5 microscopy types) · ⚑ Not verified in the textbook — Fc-receptor blocking step order: source DoH Past Paper Exam — General Lab, p.5, Q20; analyzer interferences: source Clinical Laboratory Science Review 5th ed. (6.0), pp.240, 248–249 · Guide: 6.0 pp.239–245
11 · Safety I — biological hazards, standard precautions, PPE and exposure response
The simple idea: infection needs a chain: source → way of transmission → susceptible host. Every safety rule breaks one link. Hand hygiene is the single most important way to break the chain.
📌 Standard Precautions
- 1996 CDC: Universal Precautions + body substance isolation = Standard Precautions — all blood and all body fluids, secretions and excretions except sweat are treated as potentially infectious, whatever the patient's "risk".
- Transmission-based precautions (contact, droplet, airborne) are added for specific infections.
- No special warning labels on patient specimens — all are treated as hazardous.
🔥 Personal protective equipment (PPE) and work practices
- PPE = gloves, fluid-resistant lab coats/gowns, face shields, masks, goggles, (N95 respirator for TB). A biological safety cabinet, eyewash or safety shower is equipment, not PPE.
- Gloves: disposable, single use; change between patients and when contaminated; wash hands after removing gloves; remove gloves before touching clean items (phone in clean area) or leaving the work area. Latex allergy → nitrile/vinyl/neoprene.
- Lab coats buttoned, removed before leaving the lab. Order at end of shift: disinfect area → remove gloves → remove lab coat → wash hands.
- Never mouth pipette; no eating, drinking, smoking, cosmetics, lip balm or contact-lens handling in the work area.
- Sharps: never recap, bend or cut needles; use safety-engineered devices; drop directly into a puncture-resistant sharps container. Most needlesticks happen during recapping.
- Centrifuge only capped tubes with the lid closed (aerosols).
- Specimens for M. tuberculosis (and other aerosol-transmitted agents) are processed in a biological safety cabinet (BSC) with HEPA filters; Class II BSCs are used in clinical labs.
Biosafety levels (BSL)
| Level | Agents |
|---|---|
| BSL-1 | not known to cause disease in healthy adults |
| BSL-2 | moderate risk; disease rarely serious, treatment available (most clinical labs, e.g., HBV, HIV specimens, Salmonella) |
| BSL-3 | serious/lethal disease by aerosol, treatment available (M. tuberculosis cultures, Brucella) |
| BSL-4 | life-threatening, no treatment (Ebola) |
🔥 Disinfection, sterilization and waste
- Bench tops and spills: freshly diluted 1:10 household bleach (about 0.5% sodium hypochlorite) — inactivates HIV and HBV. 1:10 bleach disinfects; it does not sterilize and it damages metal and lenses.
- Spill: put on PPE → absorb with disposable material → disinfect → discard in biohazard waste.
- Autoclave (moist heat): 121 °C at 15 psi for at least 15 minutes kills spores. Dry heat 160–170 °C; 56 °C only inactivates complement.
- Infectious waste (regulated medical waste) is separated from ordinary waste into labelled biohazard (red) bags/containers and incinerated, autoclaved or removed by a licensed company. Urine may be poured into the sink (then disinfect the sink).
- Prions resist routine decontamination.
📌 Bloodborne pathogens and exposure response
- Hepatitis B virus (HBV) is the main occupational infection risk (highest transmission after a needlestick, about 6–30% ⚑); HCV about 1.8% ⚑; HIV about 0.3% (3 per 1,000).
- Most common route of lab infection: parenteral inoculation (needlestick, cuts). Intact skin is a good barrier.
- Employer must offer HBV vaccine free of charge to at-risk staff and provide PPE (OSHA Bloodborne Pathogens Standard, exposure control plan).
- After a needlestick: wash the site with soap and water ⚑ (flush mucous membranes with water) → report immediately to the supervisor → medical evaluation, source-patient testing and post-exposure prophylaxis (PEP) as soon as possible; HBIG + vaccine if not immune.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.15, p.199.e17 (exposure control plan; TB control plan), p.199.e19 (Standard Precautions = all body fluids except sweat; exposure routes), p.199.e20 (transmission-based precautions; PPE list; latex alternatives; CLSI/OSHA rules: no mouth pipetting, wash hands, sharps containers, no warning labels, BSL-2 practices), p.199.e21 (HBV vaccine offered free), p.199.e24 (Table 15.14 ISO 15190), p.199.e25 (HEPA-filtered BSCs; safety elements); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.1, pp.2–3 (chain of infection; Standard Precautions), p.4 (OSHA bloodborne pathogens standard: no eating/cosmetics, free HBV vaccine, 1:10 bleach; report exposures immediately; PPE; centrifuge capped), p.5 (biological waste; urine into the sink; sharps); Turgeon, Clinical Hematology 5th ed., Ch.1, p.3 (needlesticks), p.4 (HIV risk ~3/1,000; sharps containers; recapping), p.5 (HBIG + vaccine, PEP), pp.8–9 (1:10 bleach = 0.5%); Jawetz, Medical Microbiology 26th ed., Ch.4, p.64 (121 °C at 15 psi for 15 min kills spores) · ⚑ Not verified in the textbook — HBV 6–30% and HCV 1.8% transmission risks and "wash with soap and water first": source DoH Past Paper Exam — General Lab, p.15, Q1; BSL definitions: source SUCCESS! in Clinical Laboratory Science, Ch.13, p.1016, Q58 · Guide: Hubbard Ch.7 §XXIII (p.287); 6.0 pp.251–253
12 · Safety II — chemical, fire, electrical, radiation, physical hazards, waste and symbols
📌 Chemical safety — who and what
- OSHA requires a written Chemical Hygiene Plan (CHP) with a designated chemical hygiene officer, work practices, SOPs, PPE, engineering controls (fume hoods, safety cabinets), training and medical consultation (paid by the employer, on work time).
- Hazard Communication Standard ("right to know"): labels and Safety Data Sheets (SDS, formerly MSDS) — a 16-section document (physical/chemical properties, fire, reactivity, health hazards, first aid, safe handling, storage, disposal). Must be available 24/7 and reviewed yearly. Consult the SDS before disposing of an old chemical or to learn a chemical's hazards.
- Labels on reagents: name, concentration, date prepared, initials, expiry, hazard. Secondary containers must be labelled.
🔥 Handling chemicals
- Add acid to water slowly (never water to acid). Carry bottles by the body, in rubber/plastic carriers.
- Skin or eye contact: flush with water at the safety shower or eyewash for at least 15 minutes, remove contaminated clothing; never neutralize on the skin; then medical attention.
- Large spill of a hazardous chemical (e.g., 10 gallons of formaldehyde) → evacuate first; small spills → spill kit.
- Incompatible: bleach (chlorine) + ammonia; perchloric acid + organics; azides build up explosive metal salts in pipes; diethyl ether forms explosive peroxides; picric acid becomes shock-sensitive when dry.
- Formaldehyde is a potential carcinogen; corrosives destroy tissue at the contact site; xylene, ethanol, methanol, acetone = flammable.
- Flammable solvents: OSHA-approved vented flammable safety cabinet; ≤ 5 gallons outside cabinets per room; safety cans (≤ 5 gallons); explosion-proof refrigerators. Xylene waste = EPA hazardous waste via a licensed hauler.
- Chemical gloves: nitrile or neoprene, chosen by the chemical and task.
- Compressed gas cylinders: secured upright (chained) at all times, valve cap on until use, moved on a hand truck.
🔥 Hazard identification systems
| NFPA 704 diamond | Colour | Hazard |
|---|---|---|
| Left | Blue | Health |
| Top | Red | Flammability (fire) |
| Right | Yellow | Reactivity / instability |
| Bottom | White | Special hazards (W̶ = reacts with water, OX = oxidizer, ACID, ALK, COR) |
Each quadrant is graded 0 (minimal) to 4 (severe). "4" in blue = deadly health hazard; "1" in blue = slightly hazardous. NFPA labels inform firefighters in an emergency. The GHS/OSHA system uses pictograms in a red diamond (flame, skull and crossbones, corrosion, health hazard, exclamation mark, gas cylinder, exploding bomb, flame over circle). DOT shipping labels show the UN hazard class 1–9.
| Sign | Meaning |
|---|---|
| Red/black/white DANGER | immediate hazard |
| Yellow/black CAUTION | potential hazard |
| Three interlocking circles | biohazard |
| Trefoil (three blades) | radiation |
📌 Fire safety
| Class | Burning material | Extinguisher |
|---|---|---|
| A | ordinary combustibles: wood, paper, cloth (a lab coat) | water, dry chemical |
| B | flammable liquids/gases, grease, solvents | dry chemical, CO₂, foam, halon — no water |
| C | energized electrical equipment | dry chemical, CO₂, halon — no water |
| D | combustible metals (magnesium, sodium) | special powder/sand; trained staff only |
| ABC | multipurpose | dry chemical — most common in labs |
Keep exits and corridors clear at all times; fire drills are not announced in advance; fire blankets smother burning clothing.
Electrical, radiation and physical hazards
- Equipment must be grounded (3-prong) or double insulated and UL-approved; no extension cords/multi-outlet adapters; never touch with wet hands; unplug wet equipment.
- Electric shock: remove the power source (circuit breaker, unplug, or move with a non-conducting object) — never pull the person with your hands.
- Radiation: exposure depends on time, distance and shielding; wear a dosimeter badge; pregnant staff avoid radioactive areas.
- Lifting: bend at the knees and hips, keep the load close, lift with the legs; store heavy containers on low shelves. Carpal tunnel prevention: keep wrists neutral. Ergonomics programs cover pipetting and workstations.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.15, p.199.e15 (chemical hygiene officer), p.199.e16 (CHP elements; SDS 16 sections available 24/7; eye washers, safety shields), p.199.e17 (SDS sections; GHS), p.199.e18 (hazard categories; UN 9 classes; DOT labels), p.199.e19 (GHS pictograms in red frame; NFPA 704 0–4, three coloured quadrants + special), p.199.e21 (acid to water; labelling reagents; perchloric acid; mercury; EPA hazardous waste), p.199.e22 (solvents in vented cabinets, 5 gal per room, safety cans; flash point; compressed gas rules; grounding, no extension cords; fire equipment, dry chemical extinguishers), p.199.e23 (Table 15.13 fire classes A, B, C; D and E fires; ISO 15190), p.199.e25 (radiation monitoring); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.1, p.6 (flush 15 minutes, do not neutralize; acid to water; CHP; NFPA 704 0–4), p.7 (MSDS contents; Fig.1–5 NFPA symbol; radiation time, distance, shielding; pregnancy; electrical safety), p.8 (Table 1–2 fire classes; PASS; RACE; physical hazards, lifting) · Guide: 6.0 pp.253–255
13 · Laboratory operations — specimens, documentation, reporting, critical values, LIS, regulation and management
📌 Specimen acceptance and tubes
- Patient identification with two identifiers (name + medical record number or date of birth), matched with the request at the bedside; label tubes at the bedside after collection.
- Unlabeled or mislabeled specimen → reject and request a new, properly labelled specimen. Do not relabel it in the lab, do not assume, and do not test it "while the ward finds out".
- Tube additives: light blue = sodium citrate (coagulation), lavender/purple = EDTA (CBC), green = heparin, grey = sodium fluoride/oxalate (glucose, lactate), red = no additive (serum), gold/red-grey = serum separator gel. Plasma contains fibrinogen; serum does not (used up in the clot).
- Order of draw (CLSI): blood culture → light blue (citrate) → red/gold (serum) → green (heparin) → lavender (EDTA) → grey (fluoride). Shaking tubes vigorously causes hemolysis (pre-analytical error).
🔥 Result reporting and critical values
- Critical (panic) value = a result that may be life-threatening (e.g., K⁺ 7.6 mmol/L) → notify the responsible clinician immediately, use read-back, and document who was told, when and by whom. Do not wait for a request.
- Communication of critical results is a patient-safety goal (The Joint Commission); failure to call critical values is a post-analytical error.
- Reports must identify the patient, test, result with units and reference interval, and the testing laboratory.
- Delta checks, autoverification rules and electronic reporting reduce post-analytical errors (no manual transcription).
- Patient results are confidential (HIPAA in the USA): released to the ordering physician and staff directly caring for the patient, not to employers or friends.
📌 Documents and records
- The procedure manual (SOPs): principle, specimen, reagents, calibration, QC, step-by-step procedure, calculations, reference and critical values, limitations, references, author, date and annual review. A package insert can be a reference within a procedure, not a replacement for it.
- QC records, maintenance logs, temperature logs, corrective actions, training and competency records. Improper documentation is the most common inspection deficiency.
- Electronic signatures and audit trails give traceability (who did what and when).
🔥 Laboratory information system (LIS)
- Functions: orders, barcode labels (accession numbers; symbologies such as Code 128, Code 39, ISBT 128), worklists, instrument interfaces (uni- or bidirectional; host query), QC management, result verification/autoverification, reports and billing.
- HL7 (Health Level Seven) = standard format for sending data between the LIS and the hospital/electronic medical record. LOINC and SNOMED CT are coding systems.
- LIS-calculated results (e.g., eGFR, LDL) must be validated when installed, after any change, and periodically — CAP: at least every 2 years ⚑.
- Changes are tested and documented (validation); data are protected by back-ups, firewalls and encryption.
Regulation, accreditation and management basics
- USA: CLIA '88 covers every lab testing human specimens; test complexity (waived, moderate, high) is categorized by the FDA; CMS enforces. OSHA = worker safety. Outside the USA, laboratories follow their national health authority's licensing rules, and many are accredited to ISO 15189 or by CAP.
- Credentialing/licensure: licensure = legal permission by a government body; certification = by a professional agency; accreditation = of a program or laboratory.
- Four management functions: planning, organizing, directing, controlling. Labor is the largest cost (about 60–70%). A full-time employee = 2,080 hours/year (40 h × 52). Percent over budget = (spent − budget) ÷ budget × 100.
- Discipline starts with verbal counseling; evaluations relate to the job description; professionalism = consistently following SOPs (quality before speed).
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4, p.65 (tube stopper colours and additives; order of draw), p.64 (uniform order of draw); Ch.5, p.123 (delta checks), p.127 (quality indicators: misidentified, unlabeled samples); Ch.3, p.60.e4 (critical results as a risk area); Ch.29, p.324.e2 (LIS order generates a barcoded label with a unique accession number), p.324.e16 (autoverification by LIS or middleware with critical values); Ch.15, p.199.e2 (HL7, LOINC, SNOMED CT); Turgeon, Clinical Hematology 5th ed., Ch.1, p.13 (errors by phase: failure to report critical values, LIS errors), p.15 (delta check); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.7, p.128 (procedure-manual contents; annual review), p.129 (pre-analytical factors; do not relabel; reject and recollect), p.134 (written procedures for critical values), p.135 (errors by phase) · ⚑ Not verified in the textbook — CAP 2-year validation of LIS calculations: source DoH Past Paper Exam — General Lab, p.1, Q3 and BOC Study Guide 6th ed., Ch.8, p.499, Q260; 2,080 hours, labor 60–70%, management functions: source Medical Laboratory Science Review (Harr), Ch.9, p.526 · Guide: Hubbard Ch.11 §IV–VI (pp.372–382); 6.0 p.256
14 · Comparison tables and exam traps
Look-alike words
| Word | Means | Do not confuse with |
|---|---|---|
| Accuracy | close to the true value | Precision — close to each other |
| Diagnostic sensitivity | positive in diseased people | Analytical sensitivity — lowest detectable amount |
| Specificity (diagnostic) | negative in healthy people | Analytical specificity — measures only the analyte |
| Standard / calibrator | known concentration used to calibrate | Control — checks the calibrated method |
| QC (internal) | daily run check, precision | PT/EQA — external unknowns, accuracy vs peers |
| Shift | sudden, stays on one side | Trend — gradual drift |
| R-4s / 1-3s | random error | 2-2s, 4-1s, 10x̄ — systematic error |
| Corrective action | after a problem happened | Preventive action — before a potential problem |
| Molarity | mol per L of solution | Molality — mol per kg of solvent; Normality — equivalents per L |
| Deionization | removes ions (resins) | Distillation — boil and condense; RO — membrane |
| Impedance (Coulter) | count + size by aperture pulses | Flow cytometry — laser scatter + fluorescence |
| Fluorometry | emitted light at 90° | Nephelometry — scattered light; AAS — absorbed (hollow cathode) |
| Biohazard | three interlocking circles | Radiation — trefoil |
| NFPA blue | health | Red — fire; yellow — reactivity; white — special |
⚠️ Top exam traps
- 1 + 9 parts = 1:10 (not 1:9) = 10% of the stock.
- % w/v = grams per 100 mL; 5 g/50 mL = 10%.
- 1 GEW HCl in 1 L = 1 N; 1 M H₂SO₄ = 2 N.
- About 5% of healthy people fall outside a reference interval; 1 control in 20 is outside ± 2 SD by chance.
- Prevalence changes PPV/NPV, not sensitivity/specificity.
- Out-of-control QC → do not report patients; fix, re-run controls and patients.
- Unlabeled tube → reject and recollect.
- Critical value → phone the clinician immediately with read-back.
- PT samples → test exactly like patients.
- Fire → Rescue first; extinguisher → Pull first; electrical fire = class C (no water).
- Chemical splash → 15 minutes of water, never neutralize.
- Needlestick → wash, report immediately, PEP; HBV is the highest risk.
- Bleach 1:10 for surfaces; autoclave 121 °C, 15 psi, ≥ 15 min.
- Platelet-poor plasma: about 1500 × g for 15 min; Hct > 55% → reduce citrate.
- RCF depends on radius and rpm²; report × g.
Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.2, pp.16, 47; Ch.3, p.60.e8; Ch.6, pp.132, 140; Ch.9, pp.180, 186; Ch.15, pp.199.e1–199.e25; Ch.16, pp.202–220; Ch.81, p.1109.e7; Turgeon, Clinical Hematology 5th ed., Ch.1 and Ch.27; Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.1 and Ch.7 · Guide: Hubbard Ch.1, Ch.11; 6.0 pp.227–256
⚡ Last-Minute Review — General Laboratory Principles
🔥 Concentrations & conversions
- % w/v = g/100 mL (5 g in 50 mL = 10%) · M = mol/L of solution · m = mol/kg solvent · N = equivalents/L = M × valence · GEW = MW ÷ valence.
- mmol/L = (mg/dL × 10) ÷ atomic weight · mEq/L = mmol/L × valence (Ca 10 mg/dL = 2.5 mmol/L = 5 mEq/L).
- °F = °C × 1.8 + 32 (100 °C = 212 °F) · °C = (°F − 32) ÷ 1.8 · pH 7 = 100 nmol/L H⁺ · lb ÷ 2.2 = kg.
🔥 Dilutions
- Dilution = part ÷ whole: 1 + 9 = 1:10 (10%; 5% bleach → 0.5%) · 1 + 3 = 1:4 (× 4) · 0.1 in 10 mL = 1:100.
- Serial = multiply (1:5 × 1:5 = 1:25) · C₁V₁ = C₂V₂ (same units!) · over-range → dilute, re-run, × DF.
- Beer's law: Cu = (Au/As) × Cs; A = 2 − log %T; double concentration = double absorbance.
🔥 Clinical formulas
- CrCl = U × V/P × 1.73/BSA · AG = Na − (Cl + HCO₃) · LDL = TC − HDL − TG/5 (not if TG > 400).
- MCV = Hct × 10/RBC · MCH = Hb × 10/RBC · MCHC = Hb/Hct × 100 · corrected retic = retic × Hct/45.
- Hemocytometer = cells × DF × 10 ÷ area · INR = (PT/MNPT)ISI · Hct > 55% → C = 1.85 × 10⁻³ × (100 − Hct) × V.
- PPP ≈ 1500 × g, 15 min (platelets < 10 × 10⁹/L) · RCF = 1.118 × 10⁻⁵ × r × rpm².
🔥 Statistics & QC
- CV = SD/mean × 100 (index of precision) · ±1 SD 68%, ±2 SD 95%, ±3 SD 99.7% · 1 in 20 outside ±2 SD by chance.
- Accuracy = true value · precision = reproducibility · random error → scatter (1-3s, R-4s) · systematic → bias (2-2s, 4-1s, 10x̄).
- Shift = sudden (new reagent/calibrator lot) · trend = gradual (deteriorating reagent, aging lamp) · low calibrator → upward shift.
- Out of control → stop reporting → fix → repeat controls and patients → document · new limits from ≥ 20 results.
- Regression: intercept = constant error, slope = proportional error · t-test = means · F-test = variances · r ≠ agreement.
- Reference interval = central 95% (5% of healthy outside) · establish with 120 · verify with 20 (≤ 2 outside).
🔥 Diagnostic performance
- Sens = TP/(TP+FN) · Spec = TN/(TN+FP) · PPV = TP/(TP+FP) · NPV = TN/(TN+FN) · efficiency = (TP+TN)/all.
- Prevalence changes PPV/NPV (rare disease → low PPV) · ROC upper-left = best cut-off · SnOut, SpIn.
🔥 Quality management
- QA = pre + analytical + post (most errors pre-analytical) · order: receive → QC → analyze → verify → release.
- PT/EQA = external unknowns for accuracy; test like a patient sample; no sharing, no averaging · SDI = (lab − peer)/peer SD.
- ISO 15189 = quality & competence · ISO 15190 = safety · corrective = after, preventive = before · Six Sigma DMAIC · IQCP = risk assessment → QC plan → QA.
🔥 Instruments
- Spectrophotometer: source → monochromator → cuvette → detector · tungsten = visible, deuterium = UV · bandwidth = spectral purity · didymium/holmium = wavelength check.
- Fluorometry = emitted light at 90° (xenon/mercury) · nephelometry = scatter · AAS = hollow cathode · ISE = membrane selectivity, activity · MS = m/z.
- Coulter = impedance (count + size) · optical: forward = size, side = granularity · flow cytometer light = laser; steps: prepare → Fc block → stain → acquire.
- Clot detection: electromechanical (carryover error) vs photo-optical (lipemia/icterus interfere) · urine strips = reflectance · blood culture = CO₂ sensor · MALDI-TOF = ribosomal-protein m/z fingerprint.
- Water: CLRW ≥ 10 MΩ·cm, 0.22 µm filter · deionization = H⁺/OH⁻ resins · Class A volumetric pipette + flask for standards · centrifuge tachometer check (Strasinger: every 3 months) · fridge temps daily · POCT = faster TAT.
🔥 Safety
- Standard Precautions = all body fluids except sweat · hand hygiene #1 · never recap · sharps container · BSC (HEPA) for TB · BSL-2 for routine clinical work.
- 1:10 bleach (0.5% NaOCl) for surfaces/spills · autoclave 121 °C / 15 psi / 15 min · separate infectious waste.
- HBV > HCV > HIV (≈ 30 / 1.8 / 0.3%) · free HBV vaccine · needlestick: wash, report, PEP.
- CHP + chemical hygiene officer (OSHA) · SDS 16 sections · acid to water · flush 15 min, never neutralize · evacuate for large spills · bleach + ammonia = toxic gas · ether peroxides · cylinders chained.
- NFPA: blue health, red fire, yellow reactivity, white special, 0–4 · RACE (Rescue first) · PASS (Pull first) · A ordinary, B liquids, C electrical (no water), D metals.
- Electrical: grounded, no extension cords; shock → cut the power, do not touch the person · radiation: time, distance, shielding.
🔥 Operations
- Two patient identifiers · unlabeled tube → reject & recollect · light blue citrate, lavender EDTA, green heparin, grey fluoride · plasma has fibrinogen, serum does not.
- Critical value (e.g., K⁺ 7.6) → call clinician now, read-back, document · delta check → misidentification · autoverification/electronic reports cut transcription errors.
- LIS: barcodes, HL7 interface, audit trail/e-signatures; validate calculations after changes (CAP: ≥ every 2 years) · CLIA: all labs; FDA classifies complexity · 1 FTE = 2,080 h/year · discipline starts with verbal counseling.