MLS STUDY SERIES · GENERAL LABORATORY · DAY 41

General Laboratory Principles

Solutions & units · Dilutions · Clinical calculations · Reagents, water & equipment · Quality assurance · Statistics, accuracy & precision · Quality control & Westgard rules · Sensitivity, specificity & predictive values · Chemistry instrumentation & POCT · Hematology, coagulation, urinalysis & microbiology instruments · Biological safety · Chemical, fire & electrical safety · Laboratory operations & LIS · Exam traps

Simple English · DoH Past Paper Exam → BOC6e → Harr → SUCCESS! → Other References → Additional · Works offline · questions · Version 1.0

📘 Study progress

Press ✓ Mark section complete at the end of each section after you study it. Opening the page does not mark anything. Tap a chip to jump. Labels: 🔥 very high-yield · 📌 must memorize · ⚠️ exam trap · 🧠 memory trick · 🔬 lab connection · 💡 understand this.

1 · Solutions, concentration units and unit conversions

The simple idea: almost every number in the laboratory is a concentration = amount of solute (the thing dissolved) in an amount of solution (solute + solvent). If you know what the unit means, you can solve any calculation question in 3 steps: What am I given? → What do I want? → Which conversion connects them?

📌 The five concentration units you must know

UnitMeansFormulaQuick example
% w/vgrams of solute per 100 mL of solutiong ÷ mL × 1005 g in 50 mL = 10 g/100 mL = 10%
% v/vmL of solute per 100 mL of solutionmL ÷ mL × 100200 mL acetic acid in 2 L = 10%
Molarity (M)moles of solute per liter of solution(g ÷ MW) ÷ L5.85 g NaCl (MW 58.5) in 0.5 L = 0.1 mol ÷ 0.5 L = 0.2 M
Molality (m)moles of solute per kg of solventmol ÷ kgUsed for osmolality (mOsm/kg)
Normality (N)gram-equivalents per literN = M × valence; GEW = MW ÷ valence1 GEW of HCl (36.5 g) in 1 L = 1 N (= 1 M); 1 M H₂SO₄ = 2 N
Mole = grams ÷ gram molecular weight (MW) Equivalent weight (GEW) = MW ÷ valence (number of H⁺, OH⁻ or charges) mEq/L = mmol/L × valence · mmol/L = (mg/dL × 10) ÷ atomic (molecular) weight
🔬 Electrolyte conversions (SI units): calcium 10 mg/dL × 10 = 100 mg/L ÷ 40 = 2.5 mmol/L (= 5.0 mEq/L, because Ca²⁺ is divalent). Chloride 369 mg/dL → 3690 ÷ 35.5 = 104 mmol/L. Potassium 19.5 mg/dL → 195 ÷ 39 = 5.0 mEq/L. Magnesium 2.0 mEq/L ÷ 2 = 1.0 mmol/L × 24.3 = 24.3 mg/L = 2.4 mg/dL.
⚠️ EXAM TRAP: "1/10" is 0.1 as a fraction = 10%, not 0.1%. And % w/v needs grams: 5 g in 50 mL is 10%, but 5 mg in 50 mL is only 0.01%.

International System of Units (SI)

  • SI units have three classes: base, derived and supplemental. Base units cover the fundamental quantities: meter (length), kilogram (mass), second (time), ampere (current), kelvin (temperature), mole (amount of substance), candela (light) — Tietz's table also lists the katal (catalytic activity) for laboratory use.
  • A derived unit is built from base units (e.g., mol/m³, pascal, joule). Recommended unit for electrolytes in plasma: mmol/L (mEq is the old unit).
  • Prefixes: milli (10⁻³), micro (10⁻⁶), nano (10⁻⁹), pico (10⁻¹²), femto (10⁻¹⁵). So 1000 pg = 1 ng, and 0.1 ng is smaller than 1000 pg. One femtoliter (fL) = 10⁻¹⁵ L = 1 µm³ (MCV unit).

📌 Other conversions that appear in exams

ConvertFormulaExample
°C → °F°F = (°C × 1.8) + 32100 °C = 212 °F; 37 °C = 98.6 °F; 30 °C = 86 °F
°F → °C°C = (°F − 32) ÷ 1.870 °F = 21 °C
pounds → kglb ÷ 2.2175.5 lb = 79.8 kg
pH ↔ [H⁺]pH = −log[H⁺]pH 7.0 = 10⁻⁷ mol/L = 100 nmol/L; [H⁺] 0.025 M → pH 1.6
Hydrate saltsg hydrate = (MW hydrate ÷ MW anhydrous) × g anhydrous9.6 g CuSO₄ = 15.0 g CuSO₄·5H₂O
Concentrated acid (purity, specific gravity)g/mL = SG × purity; then volume = grams needed ÷ g/mLHNO₃ 68%, SG 1.42 → 0.9656 g/mL
🧠 Memory: "Per cent = per hundred" (g/100 mL). "M = moles per Liter, m = moles per kilogram." "N = M × how many H⁺."

How this topic is organised

#Study sectionCalendar itemsGuide headings (Hubbard / 6.0)Textbook chapter
1Solutions, concentration & unitsConcentrations · Percent solutions · Unit conversionsHubbard Ch.1 §I A (pp.1–2) · 6.0 pp.231–233Tietz Ch.15 pp.199.e1–199.e2
2Dilutions & standard curvesDilutions · ConcentrationsHubbard Ch.1 §I A 1d (p.2) · 6.0 pp.231–234Tietz Ch.16 pp.202–203, 208
3Clinical calculationsImportant chemistry and hematology/coagulation calculations6.0 pp.233, 244Tietz Ch.34, 36, 50, 81; Turgeon Ch.5, 23, 26
4Reagents, water, glassware & basic equipmentChemistry instrumentation (basic) · Quality6.0 pp.227–228Tietz Ch.15 pp.199.e3–199.e13; Strasinger Ch.7 p.131
5Quality assurance & quality managementQuality Assurance · Basic regulatory/quality principlesHubbard Ch.11 §III–IV (pp.371–372) · 6.0 pp.227, 256Tietz Ch.3 pp.60–61, 60.e4–60.e14; Strasinger Ch.7 p.127
6Statistics, accuracy, precision & errorsPrecision · Accuracy · Random/systematic errorsHubbard Ch.1 §I B (pp.2–3) · 6.0 pp.228–230Tietz Ch.2 pp.15–16; Ch.6 p.132; Ch.9 pp.180, 186
7Quality control: Levey-Jennings & WestgardQuality ControlHubbard Ch.1 §I B · 6.0 p.230Tietz Ch.6 pp.129–150; Turgeon Ch.1 p.17
8Sensitivity, specificity & predictive valuesSensitivity · Specificity · Predictive values6.0 p.229Tietz Ch.2 pp.16, 45, 47, 51
9Chemistry instrumentation, automation & POCTChemistry instrumentationHubbard Ch.1 §I C–E, §II (pp.3–10) · 6.0 pp.234–238Tietz Ch.16, 17, 29, 30
10Hematology, coagulation, urinalysis & microbiology instruments; flow cytometryHematology · Coagulation · Microbiology · Urinalysis instrumentation · Flow cytometry6.0 pp.239–245Turgeon Ch.27; Tietz Ch.29, 84, 86; Jawetz Ch.47; Strasinger Ch.6
11Safety I: biological hazards, PPE & exposureBiological hazards · PPE · Exposure responseHubbard Ch.7 §XXIII · 6.0 pp.251–253Tietz Ch.15 pp.199.e17–199.e21, 199.e25; Strasinger Ch.1 pp.2–5; Turgeon Ch.1 pp.3–9
12Safety II: chemical, fire, electrical, radiation, waste & symbolsChemical hazards · Fire · Electrical safety · Waste disposal · Safety symbols6.0 pp.253–255Tietz Ch.15 pp.199.e16–199.e24; Strasinger Ch.1 pp.6–8
13Laboratory operationsDocumentation · Result reporting · Critical values · LIS · Regulation · ManagementHubbard Ch.11 §IV–VI (pp.372–382) · 6.0 p.256Tietz Ch.3, 4, 5, 29; Strasinger Ch.7; Turgeon Ch.1
14Comparison tables & exam trapsAll items (review)Hubbard Ch.1, 11 · 6.0 pp.227–256Tietz Ch.2–6, 15, 16

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.15, p.199.e1 (molarity, molality, gram-equivalent weight, normality; mmol/L recommended for electrolytes; base, derived and supplemental SI units), p.199.e2 (Table 15.2 derived units); Turgeon, Clinical Hematology 5th ed., Ch.26, p.502 (fL = 10⁻¹⁵ L) · Guide: Hubbard Ch.1 §I A (pp.1–2); 6.0 Lab Operations pp.231–233

2 · Dilutions, serial dilutions and standard curves

The simple idea: a dilution tells you how many parts of sample are in the TOTAL volume. "1:10" means 1 part sample in 10 parts total (1 part + 9 parts diluent).

Dilution = volume of sample ÷ (volume of sample + volume of diluent) Dilution factor (DF) = the reciprocal → multiply the result by the DF Serial dilutions → multiply each step: 1:5 then 1:5 = 1:25

🔥 Worked patterns (memorise the logic, not the numbers)

  • 1 part bleach + 9 parts water = 1:10 → 1/10 = 10% bleach. If the bleach is 5% sodium hypochlorite, the final is 5% ÷ 10 = 0.5%.
  • 1 part CSF + 3 parts saline = 1:4 → result × 4.
  • 0.1 mL serum + 5 mL reagent + 4.9 mL water = 0.1 in 10 mL = 1:100.
  • 1 mL serum + 9 mL water, analyzer reads 350 U/L → 350 × 10 = 3,500 U/L.
  • 5 mL of a 1:25 dilution needs 5 ÷ 25 = 0.2 mL of reagent.
  • Serial: 5 N diluted 1:4 = 1.25 N; then diluted 4:15 → 1.25 × 4/15 = 0.33 N.
  • Titres: if 0.5 mL of a 1:300 dilution has 1 unit, 2 units are in 0.5 mL of a 1:150 (twice as concentrated).

📌 C₁V₁ = C₂V₂ (the "dilute a stock" formula)

QuestionSet-upAnswer
Glucose stock 2,000 mg/dL → 100 mL of 150 mg/dL2000 × V₁ = 150 × 100V₁ = 7.5 mL
0.25 N NaOH → 100 mL of 0.05 N0.25 × V₁ = 0.05 × 10020 mL
95% alcohol → 5 L of 70%95 × V₁ = 70 × 53.7 L
1.025 N NaOH → 1 L of 1.0 N1.025 × V₁ = 1.0 × 1000975.6 mL
⚠️ EXAM TRAP: C₁V₁ = C₂V₂ works only when both concentrations are in the same unit (both N, both %, both mg/dL). Convert first.

Standard curves and Beer's law calculation

A standard (calibrator) has an exactly known concentration and is used to calibrate the method. A control is used to check that the calibrated method is still working. Because absorbance is directly proportional to concentration (Beer's law, A = abc), an unknown can be calculated from one standard:

C unknown = (A unknown ÷ A standard) × C standard Standard 15 mg/dL, A = 0.50; patient A = 0.62 → (0.62/0.50) × 15 = 18.6 mg/dL Twice the concentration (same cuvette) → twice the absorbance (0.085 → 0.170)
🔬 Blanks: subtract the blank absorbance first. Triglyceride: (0.750 − 0.100 serum blank) ÷ 0.620 × 125 mg/dL = 131 mg/dL. When a result is above the linear (reportable) range, dilute the sample, re-run and multiply by the dilution factor — never report the "over-range" number.
🧠 Memory: "Dilution = part ÷ whole." "Serial = multiply." "Report = reading × DF."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16, p.202 (absorbance = −log T, directly proportional to concentration), p.203 (Beer's law A = abc), p.208 (unknown compared directly with a calibrator; verification of linearity); Ch.15, p.199.e4 (primary and secondary reference materials); Turgeon, Clinical Hematology 5th ed., Ch.1, p.9 (Table 1.1: 1 mL bleach + 9 mL water = 1:10 = 0.5% sodium hypochlorite) · Guide: Hubbard Ch.1 §I A 1d (p.2); 6.0 pp.231–234

3 · Important chemistry, hematology and coagulation calculations

The simple idea: the exam does not ask for new formulas — it asks the classic ones again and again. Learn each formula with one example.

🔥 Chemistry formulas

TestFormulaKey point
Creatinine clearance(U × V ÷ P) × (1.73 ÷ body surface area)U and P in the same unit; V in mL/min (24-h volume ÷ 1440). Needs a timed urine. Result in mL/min/1.73 m².
Anion gapNa⁺ − (Cl⁻ + HCO₃⁻)High in lactic acidosis, ketoacidosis, renal failure, methanol, ethylene glycol, salicylate.
Calculated osmolality2 Na + glucose/18 + BUN/2.8 (mg/dL)Osmolal gap = measured − calculated; a gap > about 10–15 suggests toxic alcohols.
LDL-C (Friedewald)TC − HDL-C − TG/5 (mg/dL)Not valid when TG > 400 mg/dL (4.5 mmol/L) or nonfasting.
Enzyme activity (IU/L)(ΔA/min × total volume × 1000) ÷ (6.22 × light path × sample volume)6.22 = millimolar absorptivity of NADH at 340 nm.

🔥 Hematology formulas

IndexFormulaExample / normal
MCV (fL)Hct (%) × 10 ÷ RBC (×10¹²/L)35% ÷ 4.0 × 10 = 87.5 fL · normal 80–100 fL
MCH (pg)Hb (g/dL) × 10 ÷ RBCnot affected by an error in Hct
MCHC (g/dL)Hb ÷ Hct × 10015 ÷ 47 × 100 = 31.9 · normal 32–36
Corrected reticulocyteretic % × (patient Hct ÷ normal Hct 45)3% × 30/45 = 2.0%
Reticulocyte production index (RPI)corrected retic ÷ maturation time (days)RPI ≈ 1 normal; ≥3 in hemolysis with good marrow
Manual (hemocytometer) countcells counted × dilution factor × 10 (depth) ÷ area counted (mm²)36 cells in 9 mm², 1:10 → 36 × 10 × 10 ÷ 9 = 400/µL = 0.4 × 10⁹/L
WBC corrected for nucleated RBCsuncorrected WBC × 100 ÷ (100 + NRBC per 100 WBC)NRBCs are counted as WBCs by many counters
Absolute count% × total countCD4: 5,500 × 0.20 lymph × 0.25 = 275/µL

⚑ Not verified in the textbook — the hemocytometer formula and the NRBC correction formula are taken from Medical Laboratory Science Review (Harr), Ch.1, p.4 and p.6–7 (Turgeon Ch.25, p.469 gives only the counting rules). The exact calculated-osmolality formula and the anion-gap formula were not re-opened in Tietz for this page (they are taught with their textbook pages in MLS 03 and MLS 04).

🔥 Coagulation numbers

  • INR = (patient PT ÷ mean normal PT)ISI. The ISI (International Sensitivity Index) is the exponent: WHO reference thromboplastin ISI = 1.0; a more sensitive reagent has ISI < 1.0. INR is used to monitor warfarin.
  • Citrate tube ratio = 9 parts blood : 1 part 3.2% sodium citrate. When Hct > 55%, plasma volume is low, so use less citrate: C = (1.85 × 10⁻³) × (100 − Hct) × V blood.
  • Platelet-poor plasma (platelets < 10 × 10⁹/L) for PT/aPTT: centrifuge about 1500 × g for at least 15 minutes. Low speed (about 150–200 × g) gives platelet-rich plasma for aggregation studies.
⚠️ EXAM TRAP: a falsely high Hct changes MCV and MCHC but not MCH (MCH uses Hb and RBC only). A high Hct with a normal citrate volume gives falsely long PT/aPTT (too much citrate for the small plasma volume).
🧠 Memory: "V-H-R": MCV uses Hct, MCH uses Hb, both divided by RBC; MCHC = Hb/Hct. "Correct the retic to 45."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.34, p.352.e28 (concept of clearance), p.352.e33 (creatinine clearance, timed urine), p.352.e39 (adjustment to 1.73 m²); Ch.36, Box 36.4 (Friedewald equation); Ch.50, pp.694–695 (anion gap causes; osmolal gap and toxic alcohols); Ch.81, pp.1109.e7–1109.e8 (Box 81.1 citrate adjustment for Hct > 55%; PPP < 10 × 10⁹/L at 1500 × g ≥ 15 min); Ch.78, p.1096 (PRP for aggregometry); Turgeon, Clinical Hematology 5th ed., Ch.5, p.97 (corrected reticulocyte count and RPI); Ch.23, p.417 (ISI and INR formula); Ch.26, pp.501–502 (MCV, MCH, MCHC) · Guide: 6.0 pp.233, 244

4 · Reagents, reagent water, glassware, pipettes, centrifuges and balances

The simple idea: a good method gives a wrong result if the water is impure, the pipette is wrong or the centrifuge is too slow. These "basic" items are tested often.

📌 Reagent-grade water

  • Made by distillation (vaporize and condense — oldest method; carries over volatile impurities), ion exchange / deionization (resins replace cations with H⁺ and anions with OH⁻, which form water), reverse osmosis (water forced through a semipermeable membrane), and UV oxidation. Water is filtered first.
  • Clinical laboratory reagent water (CLRW) (formerly type I/II) is used for most testing: resistivity ≥ 10 MΩ·cm at 25 °C, < 10 CFU/mL, passed through a 0.22 µm filter, low organic content. Autoclave/wash water (formerly type III) is for washing glassware.
  • More ions → lower resistivity and higher conductivity. Mixed-bed deionizers in series give > 10 MΩ·cm.

Chemicals and reference materials

Analytical reagent (AR/ACS) grade chemicals are used for reagents. A primary reference material (primary standard) is highly purified, stable, can be weighed directly and is used to prepare calibrators; a secondary standard is assayed against a primary one. NIST supplies Standard Reference Materials (SRMs).

🔥 Glassware and pipettes

ItemUse / key fact
Class A glasswareCertified accuracy (NIST) — required for accurate work
Volumetric (transfer) pipette + volumetric flaskMost accurate single volume — use for standards and working solutions
Ostwald-Folin pipetteFor viscous fluids (blood, serum); etched ring = blow-out
Serologic / Mohr (measuring) pipettesGraduated; serologic is calibrated to the tip and blown out (etched ring)
MicropipettesUsually to contain (TC) — rinse out; most automatic pipettes are air-displacement
Beakers, Erlenmeyer flasks, graduated cylindersNot accurate enough for standards
DesiccatorKeeps chemicals dry; cobalt-treated desiccant is blue when dry and turns pink when it must be changed

🔥 Centrifuges

RCF (× g) = 1.118 × 10⁻⁵ × r × (rpm)² r = radius in cm from the centre of rotation to the bottom of the tube Example: r = 10 cm, 2500 rpm → about 699 × g · head 60 cm diameter (r = 30 cm), 3000 rpm → about 2,700 × g
  • Report speed as RCF (× g), not rpm, because rotors have different radii.
  • Balance tubes of equal weight opposite each other (difference ≤ 1% or as the manufacturer says), keep the lid closed, cap tubes, and never open until the rotor has stopped — this prevents aerosols.
  • Check speed with a tachometer; Strasinger: centrifuge calibration about every 3 months (other books give 6 months or yearly — follow the manufacturer and the accreditation checklist). Disinfect centrifuges weekly.
  • Other daily checks: refrigerator and water-bath temperatures are recorded daily; microscopes get an annual professional cleaning.

Balances

An analytical balance compares the unknown mass with known weights; modern electronic balances use an electromagnetic force to counterbalance the load. Calibration is checked with certified weights (Class S / ASTM class 1–type weights).

⚠️ EXAM TRAP: deionization removes ions, not bacteria or organics; a 0.22 µm filter removes particles/bacteria; boiling and condensing is distillation.
🧠 Memory: "Volumetric for Very accurate." "Radius × rpm² — rpm counts twice."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.15, p.199.e3 (distillation, ion exchange, reverse osmosis, UV oxidation; Table 15.4 CLRW specifications, resistivity), p.199.e4 (CLRW vs autoclave/wash water; reagent grades; primary reference materials), p.199.e6 (class A glassware; pipettes to contain/to deliver), p.199.e7 (Ostwald-Folin and serologic blow-out pipettes; micropipettes calibrated to contain), p.199.e10 (RCF formula, radius to tube bottom, balancing ≤ 1%, centrifuge before unstoppering), p.199.e13 (balances, weights); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.6, p.83 (RCF from tachometer rpm), Ch.7, p.131 (refrigerator temperatures daily; centrifuge calibration every 3 months; weekly disinfection) · ⚑ Not verified in the textbook — cobalt desiccant colour change: source BOC Study Guide 6th ed., Ch.8, p.495, Q229 · Guide: 6.0 pp.227–228

5 · Quality assurance (QA) and quality management

The simple idea: Quality assurance (QA) = the whole system that makes sure the patient gets the right result, on the right patient, at the right time. Quality control (QC) is only one part of QA — it checks the analytical step.

Pre-analytical (pre-examination): test request → patient ID → patient preparation → collection → labeling → transport → processing ↓ Analytical (examination): calibration → QC → analysis → instrument maintenance → competency ↓ Post-analytical (post-examination): result verification → reporting → critical-value call → interpretation → turnaround time
🔥 Most laboratory errors are pre-analytical (wrong patient, unlabeled or mislabeled tube, wrong tube, hemolysis, delay). Post-analytical errors include transcription errors, failure to call a critical value and delayed reports. Labeling a specimen at the nurses' station is a pre-analytical (not post-analytical) error.

📌 Correct order of the testing process

Receive the sample → run QC (confirm the analyzer is in control) → analyze → verify the result → print/release. A result is always verified before it is released.

Internal QC vs external quality assessment (EQA) / proficiency testing (PT)

Internal QCEQA / Proficiency testing
Who sends samples?The lab runs its own control materialAn independent external organization sends unknown samples
ChecksDay-to-day precision and stability of each runAccuracy compared with other laboratories (peer group) / target value
RulesWestgard rules, Levey-JenningsTest PT samples exactly like patient samples: no extra repeats, no averaging between staff, no discussion with other labs, no sending to another lab before submitting
  • CLIA (USA) PT: at least 5 challenges per analyte, 3 events per year; ABO, Rh and compatibility need 100%; other analytes need ≥ 80%. Even ungraded PT challenges must be reviewed.
  • Comparing the lab mean with the peer-group mean uses the standard deviation index (SDI) = (lab mean − peer mean) ÷ peer SD; > 2 is significant.

📌 Quality management system (QMS) and accreditation

  • ISO 15189 = the international QMS standard for medical laboratories (quality and competence). ISO 15190 = safety. CAP and The Joint Commission are accrediting bodies; accreditation = an agency evaluates the lab against predetermined standards.
  • ISO 15189 clauses include document control, identification and control of nonconformities, corrective action, preventive action, continual improvement, internal audits and management review.
  • Correction = fix the error itself. Corrective action = remove the cause of an error that already happened so it does not happen again. Preventive action = remove the cause of a potential problem before it occurs.
  • Root cause analysis looks for the underlying cause and needs evidence that the fix works. Tools: PDCA (Plan-Do-Check-Act), Six Sigma (DMAIC: Define, Measure, Analyze, Improve, Control) — six sigma ≈ 3.4 defects per million; Lean removes waste; Pareto charts rank causes; fishbone diagrams list causes; Gantt charts follow project progress.
  • Problem solving starts by gathering data (define the problem with facts) — e.g., for a turnaround-time (TAT) complaint, first collect current TAT data by shift.
  • Training is required for new, transferred or promoted staff, new instruments and repeated performance problems — not because the medical director changes. Competency is assessed by direct observation, problem-solving exercises, blind samples and quizzes; signing that training is complete is not an assessment.
  • IQCP (individualized quality control plan) = risk assessment → QC plan → quality assessment.
⚠️ EXAM TRAP: "Assess the accuracy of a lab by comparing results on unknown samples with other labs" = proficiency testing, not internal QC, method validation or delta check.
🧠 Memory: "QC is inside, PT is outside." "Corrective = after, Preventive = before."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.3, p.60 (ISO 15189 as the QMS standard), p.60.e4 (risk ranking; root cause analysis), p.60.e8 (ISO 15189 clauses 4.9 nonconformities, 4.10 corrective action, 4.11 preventive action, 4.14 audits), pp.60.e13–60.e14 (root cause, PDCA, Six Sigma DMAIC), p.61 (EQA/PT benchmarking); Ch.6, pp.129–130 (internal QC vs EQA/PT from an independent organization with unknown values); Ch.5, p.127 (quality indicators: misidentified and unlabeled samples); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.7, p.127 (QA = pre-analytical, analytical and post-analytical factors; internal and external QC; PT) · ⚑ Not verified in the textbook — CLIA PT numbers (5 challenges, 3 events, 100%/80%) and SDI > 2: source Medical Laboratory Science Review (Harr), Ch.9, pp.528–529 and Ch.5, p.236 · Guide: Hubbard Ch.11 §III (p.371); 6.0 pp.227, 256

6 · Statistics, accuracy, precision, errors and reference intervals

The simple idea: repeat the same sample many times — the results spread around a centre. The centre tells you about accuracy; the spread tells you about precision.

📌 Basic statistics

TermMeaningFormula / note
MeanaverageΣx ÷ n (e.g., 110,120,150,140,130 → 130)
Medianmiddle value (50th percentile)not affected by outliers
Modemost frequent value
Rangehighest − lowestsimplest measure of spread
Standard deviation (SD)spread of results around the mean√[Σ(x − x̄)² ÷ (n − 1)] = √variance
VarianceSD²Σ(x − x̄)² ÷ (n − 1)
Coefficient of variation (CV)relative SD — the index of precisionCV% = SD ÷ mean × 100; use CV to compare precision of tests with different means
Gaussian (normal) distribution: mean ± 1 SD = 68% · ± 2 SD = 95% (95.5%) · ± 3 SD = 99.7% So 5 of 100 control results fall outside ± 2 SD by chance (1 in 20); 2.5% on each side (1 in 40)
🔬 Worked: mean 15.2 g/dL, SD 0.20 → ± 2 SD = 14.8–15.6 g/dL. Mean 100, SD 1.8 → 95% limits 96.4–103.6. Cholesterol control mean 137, 2 SD = 6 → SD 3 → CV = 3/137 = 2.19%. Ca mean 2.5 mmol/L, SD 0.3 → CV 12% (poorest precision even though its SD is smallest).

🔥 Accuracy vs precision

TermDefinitionAffected by
Accuracycloseness of a result to the true valuebias + imprecision
Trueness / biasdifference between the mean of many results and the true (target) valuesystematic error
Precision (reproducibility)closeness of repeated results to each otherrandom error; validated by running the same sample many times
Reliability ⚑ability to keep both accuracy and precision over time
Analytical sensitivitysmallest amount that can be detected (detection limit)
Analytical specificitymeasures only the analyte it claims to measure (no interference)

Widely scattered results whose average is near the target = accurate but imprecise (random error). Tight results far from the target = precise but inaccurate (systematic error).

Types of error

  • Random error — unpredictable scatter (pipetting variation, bubbles, electronic noise) → poor precision. Detected by 1-3s and R-4s.
  • Systematic error (bias) — all results shifted in one direction (calibrator, reagent lot, wavelength, temperature) → poor accuracy. Constant error = same amount at all levels (regression y-intercept); proportional error = grows with concentration (regression slope ≠ 1).

Method comparison and evaluation

  • Linear regression (y = a + bx): intercept = constant error, slope = proportional error. Example ŷ = 2.10 + 1.01x → constant bias.
  • Correlation coefficient (r) shows association only (r = 0 = no correlation) — do not use r alone to accept a method.
  • Student's t-test compares means (bias); F-test compares variances (precision). Bias = mean y − mean x (246 − 235 = 11 mg/dL).
  • Recovery experiments test proportional error; interference experiments test constant error; linearity verifies the reportable range (blank + at least 3 known levels).
  • A reagent lot change needs patient-sample comparison (not just QC) because QC material can show matrix effects.

📌 Reference intervals

  • Usually the central 95% of a healthy reference population (2.5th–97.5th percentiles), so about 5% of healthy people fall outside — an abnormal result is not always disease.
  • To establish a new interval nonparametrically: at least 120 reference individuals (per partition).
  • To verify (transfer) a manufacturer's interval: test 20 healthy people; accept if no more than 2 are outside.
  • Reference intervals are the most important tool for interpreting a result; they are not critical values.
⚠️ EXAM TRAP: to compare the precision of two tests with different means, use the CV, not the SD.
🧠 Memory: "Accuracy = Aim at the true value; Precision = Packed together." "Intercept = constant, slope = proportional."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.2, p.15 (median, 2.5th and 97.5th percentiles, nonparametric 95% interval); Ch.6, p.132 (SD describes imprecision; ±1 SD 68%, ±2 SD 95%, 2.5% each tail; bias; accuracy combines bias and imprecision; trueness), pp.146–147 (reagent lot change verified with patient samples); Ch.9, p.180 (minimum 120 values for central 95 percentiles), p.186 (verification with 20 values, no more than two outside); Turgeon, Clinical Hematology 5th ed., Ch.1, p.17 (dispersion, trends, shifts) · ⚑ Not verified in the textbook — "reliability" definition: source SUCCESS! in Clinical Laboratory Science, Ch.13, p.1013, Q35; t-test/F-test, regression intercept/slope and linearity samples: source Medical Laboratory Science Review (Harr), Ch.5, pp.231–233 and Ch.9, p.533 · Guide: Hubbard Ch.1 §I B (pp.2–3); 6.0 pp.228–230

7 · Quality control: control material, Levey-Jennings charts and Westgard rules

The simple idea: a control is a sample with a known expected range. If the control is "in", we trust the patient results of that run. If it is "out", we do not report patients until the problem is fixed.

Control material

  • Stable, matrix similar to patient specimens, concentrations at clinically important levels (normal and abnormal). The lab sets its own target mean and SD from its own runs — at least 20 results on different days to start; a cumulative SD over 6–12 months is more realistic.
  • Do not use the manufacturer's (package-insert) ranges as your limits — they are too wide.
  • When calculating next month's limits, include results between 2 and 3 SD (excluding them makes limits too narrow).
  • Minimum frequency for most chemistry tests (CLIA): two levels every 24 hours (more for blood gases, hematology, POCT glucose).

🔥 Levey-Jennings chart

Control result (y-axis) plotted against day/run (x-axis) with lines at mean, ±1, ±2 and ±3 SD.

PatternWhat you seeTypical causeError type
Shiftsudden jump; values stay on one side of the mean (≥ 6 consecutive)new reagent or calibrator lot, recalibration, changed lamp/probe, wrong calibrator concentrationsystematic
Trendvalues gradually move up or down over ≥ 6 runsdeteriorating reagent/calibrator, aging lamp, evaporation, slow temperature driftsystematic
Dispersion / randomwide scatter both sidespipetting, bubbles, poor mixing, power fluctuationrandom
⚠️ EXAM TRAP: a calibrator prepared at a lower concentration than labelled makes patient and control results read higher → upward shift. (Some books, e.g., Turgeon, define shift/trend over 3 days; Tietz and Harr use 6 or more points.)

📌 Westgard multirules

RuleMeaningDetectsAction
1-2sone control outside ± 2 SDwarning (5% false alerts — 1 in 20 by chance)inspect with other rules
1-3sone control outside ± 3 SDrandom (or large systematic)reject run
2-2stwo consecutive controls (same level across runs, or two levels in one run) beyond the same ± 2 SDsystematicreject
R-4swithin one run, one control above +2 SD and the other below −2 SD (difference > 4 SD)randomreject
4-1sfour consecutive controls beyond the same ± 1 SDsystematicreject
10x̄ten consecutive controls on the same side of the meansystematic (high false-alert rate)reject / investigate
Control out → do not report patients → follow the written out-of-control procedure → inspect (reagent, calibrator, instrument, sample) → correct (e.g., recalibrate) → repeat controls → repeat patient samples since the last good QC → document everything
  • 2-2s broken → systematic error: recalibrate and repeat controls and selected patients (repeating only the control does not fix it).
  • Probability of two consecutive results beyond −2 SD by chance = 1/40 × 1/40 = 1/1,600.
  • Other plots: Cusum — earliest sign of shift/trend; Tonks-Youden — compares labs (two control levels plotted against each other).

Patient-based checks

Delta check = compares a patient's result with the previous result (e.g., within 24–48 h). A big unexplained change suggests specimen misidentification, contamination (IV fluid) or error. Best for analytes with low biological variation (e.g., MCV, MCHC). Example: bilirubin 2.0 → 6.2 mg/dL next day with QC in range → perform a delta check and investigate.

🧠 Memory: "R for Random, 2-2 for sysTemaTic." "Shift = Sudden, Trend = Time."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.6, p.132 (Levey-Jennings plot), p.139 (minimum 20 results on different days; cumulative SD 6–12 months), p.140 (lab should assign its own target and SD; product-insert ranges too wide; Table 6.3 rules 1-3s, 2-2s, R-4s, 10x̄, 8-1s, CUSUM and the error they detect), p.141 (1-2s false-alert rate), pp.146–147 (reagent lot change), p.150 (Table 6.7 delta-check criteria for mislabeled or contaminated samples); Ch.5, p.123 (delta checks best for low-variation analytes such as MCV, MCHC); Turgeon, Clinical Hematology 5th ed., Ch.1, p.15 (delta check), p.17 (trends and shifts) · ⚑ Not verified in the textbook — "include 2–3 SD values when calculating new limits" and the CLIA frequency (2 levels/24 h): source Medical Laboratory Science Review (Harr), Ch.5, pp.228, 231 · Guide: Hubbard Ch.1 §I B; 6.0 p.230

8 · Diagnostic sensitivity, specificity and predictive values

The simple idea: draw a 2 × 2 table. Every diagnostic-test question becomes easy.

Disease presentDisease absent
Test positiveTP (true positive)FP (false positive)
Test negativeFN (false negative)TN (true negative)
TermQuestion it answersFormula
SensitivityOf people WITH disease, how many test positive?TP ÷ (TP + FN) × 100
SpecificityOf people WITHOUT disease, how many test negative?TN ÷ (TN + FP) × 100
Positive predictive value (PPV)Of people who test positive, how many really have the disease?TP ÷ (TP + FP) × 100
Negative predictive value (NPV)Of people who test negative, how many are really free of disease?TN ÷ (TN + FN) × 100
Efficiency (accuracy)How many are classified correctly?(TP + TN) ÷ total × 100
🔬 Worked: 21 of 24 cancer patients positive → sensitivity 21/24 = 87.5%. 45 of 50 benign subjects negative → specificity 90%. Sensitivity 80%, specificity 84%, prevalence 4% → PPV = (0.8 × 0.04) ÷ [(0.8 × 0.04) + (0.16 × 0.96)] = 17.2% — a good test still gives many false positives when the disease is rare.

🔥 Prevalence changes predictive values — not sensitivity or specificity

  • Sensitivity and specificity belong to the test (at a chosen cut-off). PPV, NPV and efficiency depend on prevalence.
  • Higher prevalence → PPV rises, NPV falls. Rare disease → low PPV.
  • Moving the cut-off: lower cut-off → higher sensitivity, lower specificity (more FP); higher cut-off → the opposite.
  • ROC curve (receiver operating characteristic) = sensitivity plotted against (1 − specificity) for every cut-off; the point nearest the upper-left corner gives the best sensitivity with fewest false positives. Area under the curve compares tests.
  • A screening test needs high sensitivity (few false negatives, high NPV); a confirmatory test needs high specificity.
⚠️ EXAM TRAP: diagnostic sensitivity (finds diseased people) ≠ analytical sensitivity (detects low amounts). Precision is not needed to calculate a predictive value.
🧠 Memory: "SeNsitivity rules OUT (SnOut), SPecificity rules IN (SpIn)." "Sensitivity = sick people; Specificity = sound people."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.2, p.16 (diagnostic sensitivity = proportion of diseased with a positive result), p.45 (ROC curves: sensitivity/specificity pairs for all cut-offs), p.47 (predictive values depend highly on prevalence; Bayes formulas), p.51 (definitions of sensitivity, specificity, PPV, NPV) · Guide: 6.0 p.229

9 · Chemistry instrumentation, automation and point-of-care testing

The simple idea: every analyzer measures a signal (light absorbed, light emitted, light scattered, voltage, current) and turns it into a concentration by comparing it with calibrators.

🔥 Spectrophotometer — parts in order

Light source → entrance slit → monochromator (filter, prism or diffraction grating) → exit slit → cuvette → detector (photodiode or photomultiplier tube, PMT) → readout
  • Tungsten (incandescent) lamp = visible light; deuterium/hydrogen (and xenon/mercury arc) = ultraviolet (UV, < 320–380 nm). 340 nm (NADH) is UV.
  • Monochromator isolates a narrow band of wavelengths. Spectral bandwidth (width at half the peak transmittance) describes spectral purity.
  • Double-beam instruments use a reference beam to correct for changes in light-source intensity.
  • Verify performance: wavelength accuracy (didymium or holmium oxide filters), bandwidth, stray light, linearity and photometric accuracy. The first step for a new assay is to find the wavelength of maximum absorbance.
  • Bichromatic analysis reads at two wavelengths to correct for background interference (lipemia, hemolysis).
🔥 Beer's law: A = a × b × c → absorbance is directly proportional to concentration; transmittance falls logarithmically. A = 2 − log %T (e.g., 68 %T → A = 0.168).

📌 Other optical and electrochemical methods

MethodMeasuresKey facts / uses
Fluorometrylight emitted at a longer wavelength after excitationDetector usually at 90° (right angle); xenon or mercury lamp (or laser); more sensitive than absorbance
Chemiluminescencelight from a chemical reaction (no excitation lamp)Very sensitive immunoassays; electrochemiluminescence
Nephelometrylight scattered by particles (often at an angle, e.g., 90°)Immunoglobulins, specific proteins
Turbidimetrydecrease in transmitted light due to particlesMeasured like absorbance
Flame emission photometrylight emitted by atoms excited in a flameNa, K, Li (older method)
Atomic absorption (AAS)light absorbed by ground-state atomsHollow cathode lamp made of the element measured; Ca, Mg, lead, trace metals
Reflectance photometrydiffuse reflected light from a padDry-slide chemistry, urine strip readers, glucose meters
Ion-selective electrode (ISE)potential (voltage) related to ion activityThe membrane gives selectivity (valinomycin for K⁺, glass for Na⁺/pH); direct vs indirect ISE
Osmometry (freezing point)freezing-point depressionOsmolality
Electrophoresismigration of charged molecules in an electric fieldAlbumin fastest to the anode; crescent bands = overloading
Chromatographyseparation by differential solubility/partitionTLC Rf = distance of spot ÷ distance of solvent front; GC needs volatile compounds; gel filtration separates by size
Mass spectrometrymass-to-charge ratio (m/z)Drug confirmation, newborn screening; MALDI-TOF for microbes

Automation, middleware and point-of-care testing (POCT)

  • Total laboratory automation links pre-analytical (barcoded labels, sorting, centrifuging, decapping), analytical and post-analytical (storage, retrieval) steps.
  • Autoverification = rules-based release of results by the LIS or middleware (e.g., hold results if QC failed, flag critical values or delta failures for review).
  • POCT = testing near the patient (bedside, ED, ICU, operating room). Main advantage: faster turnaround time; also small sample volume. It still needs QC, training and documentation.
⚠️ EXAM TRAP: the hollow cathode lamp belongs to atomic absorption, not to flow cytometry or fluorometry. A tungsten lamp is weak in UV, so it is not the light source of a fluorometer.
🧠 Memory: "Absorb–Emit–Scatter": spectrophotometry absorbs, fluorometry emits, nephelometry scatters.

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.16, p.202 (A = −log T), p.203 (Beer's law A = abc), p.204 (components of single- and double-beam spectrophotometers; tungsten for visible, deuterium/hydrogen, mercury and xenon arc for UV), p.205 (spectral bandwidth = spectral purity; hollow cathode lamp; stray light), p.208 (holmium oxide filter; wavelength accuracy, bandwidth, stray light, linearity, photometric accuracy), pp.209–210 (reflectance photometry; flame emission; AAS components), p.214 (right-angle fluorometer detection), p.216 (xenon source), p.220 (nephelometry and turbidimetry measure scattered light); Ch.17, p.224.e3 (ISE membrane controls selectivity); Ch.29, p.324 (total laboratory automation), p.324.e5 and p.324.e16 (autoverification, middleware, rules); Ch.30, p.326 (Box 30.2: POCT advantage = reduced turnaround time) · Guide: Hubbard Ch.1 §I C–E, §II (pp.3–10); 6.0 pp.234–238

10 · Hematology, coagulation, urinalysis and microbiology instruments; flow cytometry

🔥 Hematology analyzers — two counting principles

PrincipleHow it worksKey points
Electrical impedance (Coulter principle)Cells are non-conductive in a conductive diluent; each cell passing through a small aperture increases resistance → a voltage pulseNumber of pulses = count; pulse height = cell volume (MCV, RDW from the histogram)
Optical (light scatter) with hydrodynamic focusingCells flow one by one through a laser beamForward scatter ≈ size; side (90°) scatter ≈ granularity/internal complexity → 5-part differential
  • Hemoglobin is measured photometrically after lysis (cyanmethemoglobin or SLS methods).
  • Interferences: marked leukocytosis falsely raises Hb (turbidity), RBC count and MCV; lipemia falsely raises Hb; giant platelets or clumps lower the platelet count; nucleated RBCs are counted as WBCs (correct the WBC); cold agglutinins falsely raise MCV and lower RBC.
  • A right shift of the RBC histogram = larger cells (macrocytosis, e.g., pernicious anemia); left shift = microcytes.

📌 Coagulation analyzers — clot detection

  • Electromechanical methods detect the fibrin clot by a change in conductivity or by a moving steel ball/probe that stops when the clot forms. Not affected by lipemia or icterus; possible error: sample carryover.
  • Photo-optical methods detect the clot as a change in light transmission (optical density). Lipemia, icterus and hemolysis can interfere.
  • Platelet aggregation uses platelet-rich plasma and light-transmission aggregometry.

Urinalysis instruments

  • Reagent strip readers use reflectance photometry; they standardize colour reading and timing.
  • Specific gravity: refractometer (refractive index), urinometer, harmonic oscillation densitometry.
  • Automated sediment analyzers use flow cytometry or digital imaging of particles; workcells combine chemistry and particle modules.
  • Routine sediment and Gram stains are examined with the bright-field microscope; phase-contrast helps pale casts; polarizing for crystals/lipids; fluorescence for special stains; dark-field for Treponema.

📌 Microbiology instruments

  • Continuously monitored blood culture systems detect growth by the production/consumption of gas (usually CO₂) with a sensor; standard 5-day protocol.
  • Mycobacterial liquid culture (MGIT) detects growth by fluorescence as oxygen is used.
  • Automated identification and susceptibility systems (e.g., VITEK 2, Phoenix, MicroScan) read biochemical and growth changes in cards/panels.
  • MALDI-TOF mass spectrometry: a colony smear + acid matrix on a metal target is hit by a laser; ionized proteins (mostly ribosomal proteins) fly through a vacuum tube; the time of flight gives an m/z spectral fingerprint compared with a database → rapid identification of bacteria and yeasts.

🔥 Flow cytometry

  • Single cells in a fluid stream pass a laser (intense, stable, monochromatic) → forward scatter (size), side scatter (granularity) and fluorescence from fluorochrome-labelled monoclonal antibodies (immunophenotyping, CD markers).
  • Steps: sample preparation → Fc-receptor blocking → antibody staining → acquisition and analysis (gating, e.g., CD45 vs side scatter). ⚑
  • Uses: leukemia/lymphoma phenotyping, CD4 counts (absolute CD4 = WBC × lymph % × CD4 %), PNH, stem-cell (CD34) counts, DNA content.
⚠️ EXAM TRAP: impedance = Coulter = size by pulse height. Light scatter + fluorescence = flow cytometry. Absorbance of stained cells is cytochemistry, not flow.
🧠 Memory: "Forward = Fat (size), Side = Speckles (granules)."

Sources: Turgeon, Clinical Hematology 5th ed., Ch.27, p.527 (electrical impedance/Coulter principle through an aperture; optical detection with hydrodynamic focusing and laser), p.529 (laser properties; principles of flow cytometry), p.530 (forward scatter ≈ size, side scatter ≈ granularity), p.555 (electromechanical vs photo-optical clot detection); Tietz Textbook of Laboratory Medicine, 7th ed., Ch.29, p.324.e15 (reflectance photometry), p.324.e18 (automated urinalysis: reflectometry strips and flow-cytometric or microscopic particle analysis); Ch.84, p.1201.e30 (continuously monitored blood culture systems detecting CO₂/O₂); Ch.86, p.1203.e12 (MGIT continuously monitors fluorescence); Ch.85 (automated AST platforms VITEK 2, Phoenix, MicroScan); Ch.78, p.1096 (light-transmission aggregometry with PRP); Jawetz, Melnick & Adelberg's Medical Microbiology 26th ed., Ch.47, p.765 (MALDI-TOF principle, ribosomal proteins, m/z 1000–20,000), p.769 (automated urinalysis instruments); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.5, p.55 (automated strip readers), Ch.6, p.87 (Table 6–5 microscopy types) · ⚑ Not verified in the textbook — Fc-receptor blocking step order: source DoH Past Paper Exam — General Lab, p.5, Q20; analyzer interferences: source Clinical Laboratory Science Review 5th ed. (6.0), pp.240, 248–249 · Guide: 6.0 pp.239–245

11 · Safety I — biological hazards, standard precautions, PPE and exposure response

The simple idea: infection needs a chain: source → way of transmission → susceptible host. Every safety rule breaks one link. Hand hygiene is the single most important way to break the chain.

📌 Standard Precautions

  • 1996 CDC: Universal Precautions + body substance isolation = Standard Precautions — all blood and all body fluids, secretions and excretions except sweat are treated as potentially infectious, whatever the patient's "risk".
  • Transmission-based precautions (contact, droplet, airborne) are added for specific infections.
  • No special warning labels on patient specimens — all are treated as hazardous.

🔥 Personal protective equipment (PPE) and work practices

  • PPE = gloves, fluid-resistant lab coats/gowns, face shields, masks, goggles, (N95 respirator for TB). A biological safety cabinet, eyewash or safety shower is equipment, not PPE.
  • Gloves: disposable, single use; change between patients and when contaminated; wash hands after removing gloves; remove gloves before touching clean items (phone in clean area) or leaving the work area. Latex allergy → nitrile/vinyl/neoprene.
  • Lab coats buttoned, removed before leaving the lab. Order at end of shift: disinfect area → remove gloves → remove lab coat → wash hands.
  • Never mouth pipette; no eating, drinking, smoking, cosmetics, lip balm or contact-lens handling in the work area.
  • Sharps: never recap, bend or cut needles; use safety-engineered devices; drop directly into a puncture-resistant sharps container. Most needlesticks happen during recapping.
  • Centrifuge only capped tubes with the lid closed (aerosols).
  • Specimens for M. tuberculosis (and other aerosol-transmitted agents) are processed in a biological safety cabinet (BSC) with HEPA filters; Class II BSCs are used in clinical labs.

Biosafety levels (BSL)

LevelAgents
BSL-1not known to cause disease in healthy adults
BSL-2moderate risk; disease rarely serious, treatment available (most clinical labs, e.g., HBV, HIV specimens, Salmonella)
BSL-3serious/lethal disease by aerosol, treatment available (M. tuberculosis cultures, Brucella)
BSL-4life-threatening, no treatment (Ebola)

🔥 Disinfection, sterilization and waste

  • Bench tops and spills: freshly diluted 1:10 household bleach (about 0.5% sodium hypochlorite) — inactivates HIV and HBV. 1:10 bleach disinfects; it does not sterilize and it damages metal and lenses.
  • Spill: put on PPE → absorb with disposable material → disinfect → discard in biohazard waste.
  • Autoclave (moist heat): 121 °C at 15 psi for at least 15 minutes kills spores. Dry heat 160–170 °C; 56 °C only inactivates complement.
  • Infectious waste (regulated medical waste) is separated from ordinary waste into labelled biohazard (red) bags/containers and incinerated, autoclaved or removed by a licensed company. Urine may be poured into the sink (then disinfect the sink).
  • Prions resist routine decontamination.

📌 Bloodborne pathogens and exposure response

  • Hepatitis B virus (HBV) is the main occupational infection risk (highest transmission after a needlestick, about 6–30% ⚑); HCV about 1.8% ⚑; HIV about 0.3% (3 per 1,000).
  • Most common route of lab infection: parenteral inoculation (needlestick, cuts). Intact skin is a good barrier.
  • Employer must offer HBV vaccine free of charge to at-risk staff and provide PPE (OSHA Bloodborne Pathogens Standard, exposure control plan).
  • After a needlestick: wash the site with soap and water ⚑ (flush mucous membranes with water) → report immediately to the supervisor → medical evaluation, source-patient testing and post-exposure prophylaxis (PEP) as soon as possible; HBIG + vaccine if not immune.
⚠️ EXAM TRAP: the three interlocking circles = biohazard. The trefoil with three blades = radiation.
🧠 Memory: "B is the Biggest" (HBV > HCV > HIV: 30 – 3 – 0.3). "121–15–15" for the autoclave.

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.15, p.199.e17 (exposure control plan; TB control plan), p.199.e19 (Standard Precautions = all body fluids except sweat; exposure routes), p.199.e20 (transmission-based precautions; PPE list; latex alternatives; CLSI/OSHA rules: no mouth pipetting, wash hands, sharps containers, no warning labels, BSL-2 practices), p.199.e21 (HBV vaccine offered free), p.199.e24 (Table 15.14 ISO 15190), p.199.e25 (HEPA-filtered BSCs; safety elements); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.1, pp.2–3 (chain of infection; Standard Precautions), p.4 (OSHA bloodborne pathogens standard: no eating/cosmetics, free HBV vaccine, 1:10 bleach; report exposures immediately; PPE; centrifuge capped), p.5 (biological waste; urine into the sink; sharps); Turgeon, Clinical Hematology 5th ed., Ch.1, p.3 (needlesticks), p.4 (HIV risk ~3/1,000; sharps containers; recapping), p.5 (HBIG + vaccine, PEP), pp.8–9 (1:10 bleach = 0.5%); Jawetz, Medical Microbiology 26th ed., Ch.4, p.64 (121 °C at 15 psi for 15 min kills spores) · ⚑ Not verified in the textbook — HBV 6–30% and HCV 1.8% transmission risks and "wash with soap and water first": source DoH Past Paper Exam — General Lab, p.15, Q1; BSL definitions: source SUCCESS! in Clinical Laboratory Science, Ch.13, p.1016, Q58 · Guide: Hubbard Ch.7 §XXIII (p.287); 6.0 pp.251–253

12 · Safety II — chemical, fire, electrical, radiation, physical hazards, waste and symbols

📌 Chemical safety — who and what

  • OSHA requires a written Chemical Hygiene Plan (CHP) with a designated chemical hygiene officer, work practices, SOPs, PPE, engineering controls (fume hoods, safety cabinets), training and medical consultation (paid by the employer, on work time).
  • Hazard Communication Standard ("right to know"): labels and Safety Data Sheets (SDS, formerly MSDS) — a 16-section document (physical/chemical properties, fire, reactivity, health hazards, first aid, safe handling, storage, disposal). Must be available 24/7 and reviewed yearly. Consult the SDS before disposing of an old chemical or to learn a chemical's hazards.
  • Labels on reagents: name, concentration, date prepared, initials, expiry, hazard. Secondary containers must be labelled.

🔥 Handling chemicals

  • Add acid to water slowly (never water to acid). Carry bottles by the body, in rubber/plastic carriers.
  • Skin or eye contact: flush with water at the safety shower or eyewash for at least 15 minutes, remove contaminated clothing; never neutralize on the skin; then medical attention.
  • Large spill of a hazardous chemical (e.g., 10 gallons of formaldehyde) → evacuate first; small spills → spill kit.
  • Incompatible: bleach (chlorine) + ammonia; perchloric acid + organics; azides build up explosive metal salts in pipes; diethyl ether forms explosive peroxides; picric acid becomes shock-sensitive when dry.
  • Formaldehyde is a potential carcinogen; corrosives destroy tissue at the contact site; xylene, ethanol, methanol, acetone = flammable.
  • Flammable solvents: OSHA-approved vented flammable safety cabinet; ≤ 5 gallons outside cabinets per room; safety cans (≤ 5 gallons); explosion-proof refrigerators. Xylene waste = EPA hazardous waste via a licensed hauler.
  • Chemical gloves: nitrile or neoprene, chosen by the chemical and task.
  • Compressed gas cylinders: secured upright (chained) at all times, valve cap on until use, moved on a hand truck.

🔥 Hazard identification systems

NFPA 704 diamondColourHazard
LeftBlueHealth
TopRedFlammability (fire)
RightYellowReactivity / instability
BottomWhiteSpecial hazards (W̶ = reacts with water, OX = oxidizer, ACID, ALK, COR)

Each quadrant is graded 0 (minimal) to 4 (severe). "4" in blue = deadly health hazard; "1" in blue = slightly hazardous. NFPA labels inform firefighters in an emergency. The GHS/OSHA system uses pictograms in a red diamond (flame, skull and crossbones, corrosion, health hazard, exclamation mark, gas cylinder, exploding bomb, flame over circle). DOT shipping labels show the UN hazard class 1–9.

SignMeaning
Red/black/white DANGERimmediate hazard
Yellow/black CAUTIONpotential hazard
Three interlocking circlesbiohazard
Trefoil (three blades)radiation

📌 Fire safety

RACE = Rescue people in danger → Alarm → Contain (close doors) → Extinguish or Evacuate PASS = Pull the pin → Aim at the base → Squeeze → Sweep side to side
ClassBurning materialExtinguisher
Aordinary combustibles: wood, paper, cloth (a lab coat)water, dry chemical
Bflammable liquids/gases, grease, solventsdry chemical, CO₂, foam, halon — no water
Cenergized electrical equipmentdry chemical, CO₂, halon — no water
Dcombustible metals (magnesium, sodium)special powder/sand; trained staff only
ABCmultipurposedry chemical — most common in labs

Keep exits and corridors clear at all times; fire drills are not announced in advance; fire blankets smother burning clothing.

Electrical, radiation and physical hazards

  • Equipment must be grounded (3-prong) or double insulated and UL-approved; no extension cords/multi-outlet adapters; never touch with wet hands; unplug wet equipment.
  • Electric shock: remove the power source (circuit breaker, unplug, or move with a non-conducting object) — never pull the person with your hands.
  • Radiation: exposure depends on time, distance and shielding; wear a dosimeter badge; pregnant staff avoid radioactive areas.
  • Lifting: bend at the knees and hips, keep the load close, lift with the legs; store heavy containers on low shelves. Carpal tunnel prevention: keep wrists neutral. Ergonomics programs cover pipetting and workstations.
⚠️ EXAM TRAP: the first action in a fire is Rescue, not "pull the alarm" — and the first step with an extinguisher is Pull the pin. Never use water on class B or C fires.
🧠 Memory: NFPA "Blue = Breathe (health), Red = Fire, Yellow = Yikes it reacts, White = Weird/special."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.15, p.199.e15 (chemical hygiene officer), p.199.e16 (CHP elements; SDS 16 sections available 24/7; eye washers, safety shields), p.199.e17 (SDS sections; GHS), p.199.e18 (hazard categories; UN 9 classes; DOT labels), p.199.e19 (GHS pictograms in red frame; NFPA 704 0–4, three coloured quadrants + special), p.199.e21 (acid to water; labelling reagents; perchloric acid; mercury; EPA hazardous waste), p.199.e22 (solvents in vented cabinets, 5 gal per room, safety cans; flash point; compressed gas rules; grounding, no extension cords; fire equipment, dry chemical extinguishers), p.199.e23 (Table 15.13 fire classes A, B, C; D and E fires; ISO 15190), p.199.e25 (radiation monitoring); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.1, p.6 (flush 15 minutes, do not neutralize; acid to water; CHP; NFPA 704 0–4), p.7 (MSDS contents; Fig.1–5 NFPA symbol; radiation time, distance, shielding; pregnancy; electrical safety), p.8 (Table 1–2 fire classes; PASS; RACE; physical hazards, lifting) · Guide: 6.0 pp.253–255

13 · Laboratory operations — specimens, documentation, reporting, critical values, LIS, regulation and management

📌 Specimen acceptance and tubes

  • Patient identification with two identifiers (name + medical record number or date of birth), matched with the request at the bedside; label tubes at the bedside after collection.
  • Unlabeled or mislabeled specimen → reject and request a new, properly labelled specimen. Do not relabel it in the lab, do not assume, and do not test it "while the ward finds out".
  • Tube additives: light blue = sodium citrate (coagulation), lavender/purple = EDTA (CBC), green = heparin, grey = sodium fluoride/oxalate (glucose, lactate), red = no additive (serum), gold/red-grey = serum separator gel. Plasma contains fibrinogen; serum does not (used up in the clot).
  • Order of draw (CLSI): blood culture → light blue (citrate) → red/gold (serum) → green (heparin) → lavender (EDTA) → grey (fluoride). Shaking tubes vigorously causes hemolysis (pre-analytical error).

🔥 Result reporting and critical values

  • Critical (panic) value = a result that may be life-threatening (e.g., K⁺ 7.6 mmol/L) → notify the responsible clinician immediately, use read-back, and document who was told, when and by whom. Do not wait for a request.
  • Communication of critical results is a patient-safety goal (The Joint Commission); failure to call critical values is a post-analytical error.
  • Reports must identify the patient, test, result with units and reference interval, and the testing laboratory.
  • Delta checks, autoverification rules and electronic reporting reduce post-analytical errors (no manual transcription).
  • Patient results are confidential (HIPAA in the USA): released to the ordering physician and staff directly caring for the patient, not to employers or friends.

📌 Documents and records

  • The procedure manual (SOPs): principle, specimen, reagents, calibration, QC, step-by-step procedure, calculations, reference and critical values, limitations, references, author, date and annual review. A package insert can be a reference within a procedure, not a replacement for it.
  • QC records, maintenance logs, temperature logs, corrective actions, training and competency records. Improper documentation is the most common inspection deficiency.
  • Electronic signatures and audit trails give traceability (who did what and when).

🔥 Laboratory information system (LIS)

  • Functions: orders, barcode labels (accession numbers; symbologies such as Code 128, Code 39, ISBT 128), worklists, instrument interfaces (uni- or bidirectional; host query), QC management, result verification/autoverification, reports and billing.
  • HL7 (Health Level Seven) = standard format for sending data between the LIS and the hospital/electronic medical record. LOINC and SNOMED CT are coding systems.
  • LIS-calculated results (e.g., eGFR, LDL) must be validated when installed, after any change, and periodically — CAP: at least every 2 years ⚑.
  • Changes are tested and documented (validation); data are protected by back-ups, firewalls and encryption.

Regulation, accreditation and management basics

  • USA: CLIA '88 covers every lab testing human specimens; test complexity (waived, moderate, high) is categorized by the FDA; CMS enforces. OSHA = worker safety. Outside the USA, laboratories follow their national health authority's licensing rules, and many are accredited to ISO 15189 or by CAP.
  • Credentialing/licensure: licensure = legal permission by a government body; certification = by a professional agency; accreditation = of a program or laboratory.
  • Four management functions: planning, organizing, directing, controlling. Labor is the largest cost (about 60–70%). A full-time employee = 2,080 hours/year (40 h × 52). Percent over budget = (spent − budget) ÷ budget × 100.
  • Discipline starts with verbal counseling; evaluations relate to the job description; professionalism = consistently following SOPs (quality before speed).
⚠️ EXAM TRAP: "verify then release": a result is verified before it is printed or released; QC is run before patient analysis.
🧠 Memory: "Critical? Call, Confirm (read-back), Chart."

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.4, p.65 (tube stopper colours and additives; order of draw), p.64 (uniform order of draw); Ch.5, p.123 (delta checks), p.127 (quality indicators: misidentified, unlabeled samples); Ch.3, p.60.e4 (critical results as a risk area); Ch.29, p.324.e2 (LIS order generates a barcoded label with a unique accession number), p.324.e16 (autoverification by LIS or middleware with critical values); Ch.15, p.199.e2 (HL7, LOINC, SNOMED CT); Turgeon, Clinical Hematology 5th ed., Ch.1, p.13 (errors by phase: failure to report critical values, LIS errors), p.15 (delta check); Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.7, p.128 (procedure-manual contents; annual review), p.129 (pre-analytical factors; do not relabel; reject and recollect), p.134 (written procedures for critical values), p.135 (errors by phase) · ⚑ Not verified in the textbook — CAP 2-year validation of LIS calculations: source DoH Past Paper Exam — General Lab, p.1, Q3 and BOC Study Guide 6th ed., Ch.8, p.499, Q260; 2,080 hours, labor 60–70%, management functions: source Medical Laboratory Science Review (Harr), Ch.9, p.526 · Guide: Hubbard Ch.11 §IV–VI (pp.372–382); 6.0 p.256

14 · Comparison tables and exam traps

Look-alike words

WordMeansDo not confuse with
Accuracyclose to the true valuePrecision — close to each other
Diagnostic sensitivitypositive in diseased peopleAnalytical sensitivity — lowest detectable amount
Specificity (diagnostic)negative in healthy peopleAnalytical specificity — measures only the analyte
Standard / calibratorknown concentration used to calibrateControl — checks the calibrated method
QC (internal)daily run check, precisionPT/EQA — external unknowns, accuracy vs peers
Shiftsudden, stays on one sideTrend — gradual drift
R-4s / 1-3srandom error2-2s, 4-1s, 10x̄ — systematic error
Corrective actionafter a problem happenedPreventive action — before a potential problem
Molaritymol per L of solutionMolality — mol per kg of solvent; Normality — equivalents per L
Deionizationremoves ions (resins)Distillation — boil and condense; RO — membrane
Impedance (Coulter)count + size by aperture pulsesFlow cytometry — laser scatter + fluorescence
Fluorometryemitted light at 90°Nephelometry — scattered light; AAS — absorbed (hollow cathode)
Biohazardthree interlocking circlesRadiation — trefoil
NFPA bluehealthRed — fire; yellow — reactivity; white — special

⚠️ Top exam traps

  • 1 + 9 parts = 1:10 (not 1:9) = 10% of the stock.
  • % w/v = grams per 100 mL; 5 g/50 mL = 10%.
  • 1 GEW HCl in 1 L = 1 N; 1 M H₂SO₄ = 2 N.
  • About 5% of healthy people fall outside a reference interval; 1 control in 20 is outside ± 2 SD by chance.
  • Prevalence changes PPV/NPV, not sensitivity/specificity.
  • Out-of-control QC → do not report patients; fix, re-run controls and patients.
  • Unlabeled tube → reject and recollect.
  • Critical value → phone the clinician immediately with read-back.
  • PT samples → test exactly like patients.
  • Fire → Rescue first; extinguisher → Pull first; electrical fire = class C (no water).
  • Chemical splash → 15 minutes of water, never neutralize.
  • Needlestick → wash, report immediately, PEP; HBV is the highest risk.
  • Bleach 1:10 for surfaces; autoclave 121 °C, 15 psi, ≥ 15 min.
  • Platelet-poor plasma: about 1500 × g for 15 min; Hct > 55% → reduce citrate.
  • RCF depends on radius and rpm²; report × g.
🧠 One-line summary: Right patient (2 IDs, labelled) → right sample (right tube, processed in time) → right method (calibrated, QC in, PT passed) → right result (verified, critical values called) → safe staff (PPE, SDS, RACE/PASS).

Sources: Tietz Textbook of Laboratory Medicine, 7th ed., Ch.2, pp.16, 47; Ch.3, p.60.e8; Ch.6, pp.132, 140; Ch.9, pp.180, 186; Ch.15, pp.199.e1–199.e25; Ch.16, pp.202–220; Ch.81, p.1109.e7; Turgeon, Clinical Hematology 5th ed., Ch.1 and Ch.27; Strasinger & Di Lorenzo, Urinalysis and Body Fluids 5th ed., Ch.1 and Ch.7 · Guide: Hubbard Ch.1, Ch.11; 6.0 pp.227–256